BDPROBETEC ET CHLAMYDIA TRACHOMATIS AND NEISSERIA GONORRHOEAE AMPLIFIED DNA ASSAY
K984631 · Becton Dickinson Microbiology Systems · LSL · Nov 4, 1999 · Microbiology
Device Facts
Record ID
K984631
Device Name
BDPROBETEC ET CHLAMYDIA TRACHOMATIS AND NEISSERIA GONORRHOEAE AMPLIFIED DNA ASSAY
Applicant
Becton Dickinson Microbiology Systems
Product Code
LSL · Microbiology
Decision Date
Nov 4, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3390
Device Class
Class 2
Indications for Use
The BDProbeTec™ ET Chlamydia trachomatis (CT) and Neisseria gonorrhoeae (GC) Amplified DNA Assays, when tested with the BDProbeTec ET System, use Strand Displacement Amplification (SDA) technology for the direct, qualitative detection of Chlamydia trachomatis and Neisseria gonorrhoeae DNA in endocervical swabs, male urethral swabs, and in female and male urine specimens as evidence of infection with C. trachomatis, N. gonorrhoeae, or of co-infection with both C. trachomatis and N. gonorrhoeae. Specimens may be from symptomatic or asymptomatic females for the BDProbeTec ET CT and GC Assays, from symptomatic or asymptomatic males for the BDProbeTec ET CT Assay, and from symptomatic males for the BDProbeTec ET GC Assay. A separate Amplification Control is an option for inhibition testing (BDProbeTec ET CT/GC/AC Reagent Pack).
Device Story
Device uses Strand Displacement Amplification (SDA) and fluorescent energy transfer (ET) to detect C. trachomatis and N. gonorrhoeae DNA in clinical specimens (swabs/urine). Processed samples are added to priming microwells, then transferred to sealed amplification microwells containing DNA polymerase and restriction endonuclease. A thermally controlled fluorescent reader monitors wells for amplified products. The system calculates MOTA (Method Other Than Acceleration) scores, comparing them to pre-determined cutoffs to report positive/negative results. An optional Amplification Control (AC) identifies reaction inhibition. Used in clinical laboratory settings by trained personnel. Output assists clinicians in diagnosing C. trachomatis and N. gonorrhoeae infections, enabling timely treatment and management of co-infections.
Clinical Evidence
Multicenter study of 4108 CT and 4105 GC specimens from 2109 patients. Compared BDProbeTec ET results to culture and patient infected status. For CT, overall sensitivity 92.0% and specificity 94.9% vs. culture. For GC, overall sensitivity 96.1% and specificity 98.2% vs. culture. Performance evaluated across symptomatic/asymptomatic populations and specimen types (swab/urine). Study included assessment of potential interfering substances (leukocytes, blood, etc.).
Technological Characteristics
Uses homogeneous SDA technology and fluorescent energy transfer (ET) detection. Reagents are dried in microwell strips. Employs DNA polymerase and restriction endonuclease for amplification. Closed system design for contamination control. Operates on a thermally controlled fluorescent reader. Qualitative assay format.
Indications for Use
Indicated for qualitative detection of C. trachomatis and N. gonorrhoeae DNA in endocervical swabs, male urethral swabs, and urine. Patient population: symptomatic or asymptomatic females (CT/GC), symptomatic or asymptomatic males (CT), and symptomatic males (GC).
Regulatory Classification
Identification
Neisseria spp. direct serological test reagents are devices that consist of antigens and antisera used in serological tests to identify Neisseria spp. from cultured isolates. Additionally, some of these reagents consist of Neisseria spp. antisera conjugated with a fluorescent dye (immunofluorescent reagents) which may be used to detect the presence of Neisseria spp. directly from clinical specimens. The identification aids in the diagnosis of disease caused by bacteria belonging to the genus Neisseria, such as epidemic cerebrospinal meningitis, meningococcal disease, and gonorrhea, and also provides epidemiological information on diseases caused by these microorganisms. The device does not include products for the detection of gonorrhea in humans by indirect methods, such as detection of antibodies or of oxidase produced by gonococcal organisms.
Predicate Devices
Chlamydia cell culture
Neisseria gonorrhoeae culture
Abbott LCx® Chlamydia trachomatis Assay
Abbott LCx® Neisseria gonorrhoeae Assay
Submission Summary (Full Text)
{0}------------------------------------------------
# <984631
NOV - 4 1999
#### 510(k) SUMMARY
BECTON DICKINSON MICROBIOLOGY SYSTEMS SUBMITTED BY: 7 LOVETON CIRCLE SPARKS, MD 21152
Colleen Rohrbeck, Regulatory Affairs Associate CONTACT:
- (410) 316-4988 TELEPHONE:
- October 29, 1999 PREPARED:
BDProbeTec™ ET Chlamydia trachomatis and Neisseria DEVICE NAME: gonorrhoeae Amplified DNA Assays
PREDICATE
#### Chlamydia cell culture DEVICES: Neisseria gonorrhoeae culture Abbott LCx® Chlamydia trachomatis Assay Abbott LCx® Neisseria gonorrhoeae Assay
The BDProbeTec™ ET Chlamydia trachomatis and INTENDED USE: Neisseria gonorrhoeae Amplified DNA Assays, when tested with the BDProbeTec™ ET System, use Strand Displacement Amplification (SDA) technology for the direct, qualitative detection of Chlamydia trachomatis and Neisseria gonorrhoeae DNA in endocervical swabs, male urethral swabs, and in female and male urine specimens as evidence of infection with C. trachomatis, N. gonorrhoeae, or of co-infection with C. trachomatis and N. gonorrhoeae. Specimens may be from symptomatic or asymptomatic females for the BDProbeTec ET CT and GC Assays, from symptomatic or asymptomatic males for the BDProbeTec ET CT Assay, and from symptomatic males for the BDProbeTec ET GC Assay. A separate Amplification Control is an option for inhibition testing (BDProbeTec ET CT/GC/AC Reagent Pack).
{1}------------------------------------------------
#### DEVICE DESCRIPTION:
The BDProbeTec™ ET Chlamydia trachomatis (CT) and Neisseria gonorrhoeae (GC) amplified DNA assays utilize homogeneous SDA technology as the amplification method and fluorescent energy transfer (ET) as the detection method to test for the presence of CT and GC in clinical specimens.
For each assay, the SDA reagents are dried in two separate microwell strips. First, the processed sample is added to the Priming Microwell which contains the amplification primers, fluorescent labeled detector probe, and other reagents necessary for amplification. However, because no enzymes are present in the priming microwell strips, no amplification occurs at this step. After incubation, the reaction mixture is transferred to the Amplification Microwell, which contains two enzymes (a DNA polymerase and a restriction endonuclease) necessary for SDA. It is in this latter microwell in which amplification and detection occurs. The Amplification Microwells are sealed to prevent contamination and then incubated in a thermally controlled fluorescent reader which monitors each test well for the generation of amplified products. The presence of CT and GC is determined by relating the BDProbeTec ET MOTA (Method Other Than Acceleration) scores for the sample to pre-determined cutoff values. The MOTA score is a metric used to assess the magnitude of signal generated as a result of the reaction.
If the CT/GC Reagent Pack is used, each sample and control are tested in two discrete microwells: C. trachomatis and N. gonorrhoeae. Results are reported through an algorithm as positive or negative. If the CT/GC/AC Reagent Pack is used, each sample and control are tested in three discrete microwells: C. trachomatis, N. gonorthoeae, and the Amplification Control. The purpose of the Amplification Control is to identify a sample that may inhibit the SDA reaction. Results are reported through an algorithm as positive, negative, or indeterminate.
### DEVICE COMPARISON:
Tables 1 and 2 summarize the similarities and differences between the BDProbeTec™ ET CT and GC Amplified DNA Assays and the predicate devices.
... ..
{2}------------------------------------------------
| Feature | BDProbeTec™ ET | Abbott LCx® | CT Cell Culture |
|-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | The BDProbeTec™ ET CT<br>Amplified DNA Assay, when<br>used with the BDProbeTec™<br>ET System, uses SDA<br>technology for the direct,<br>qualitative detection of CT<br>DNA in endocervical swabs,<br>male urethral swabs, and in<br>female and male urine<br>specimens as evidence of<br>infection with<br><i>C. trachomatis</i> .<br>Specimens may be from<br>symptomatic or asymptomatic<br>males and females. A<br>separate Amplification Controlis an option for inhibition<br>testing (BDProbeTec ET<br>CT/AC Reagent Pack). | The LCx CT Assay uses<br>LCR™/<br>Ligase Chain Reaction /<br>amplification technology in<br>the LCx Probe System for the<br>direct, qualitative detection of<br>plasmid DNA of CT in female<br>endocervical and male<br>urethral swab specimens or<br>in female and male urine<br>specimens from symptomatic<br>and asymptomatic males and<br>females. | CT cell culture uses<br>tissue cell culture in<br>microtiter plates or<br>dram vials for the<br>direct, qualitative and<br>quantitative detection<br>of CT in a variety of<br>direct specimens<br>such as female<br>endocervical and<br>male urethral swabs. |
| Type of Assay | Amplified DNA Probe | Amplified DNA Probe | Tissue culture |
| Technology | SDA | LCR | Culture |
| Amplification Target | Cryptic plasmid | Cryptic plasmid | Not applicable |
| Detection format | Simultaneous amplification &<br>detection | Amplification followed by<br>detection | Not applicable |
| Qualitative or<br>Quantitative | Qualitative | Qualitative | Qualitative and/or<br>quantitative |
| Assay Formats | CT and CT/GC | CT | CT |
| Amplification Control | Optional | No | No |
| Number of Controls /<br>Run | Negative Control (1)<br>Positive Control (1) | Negative Control (2X)<br>Calibrator (2X)<br>Positive Control | At least 1 positive<br>and 1 negative<br>control per batch |
| Contamination Control<br>Method | Closed System | Chelating metal complex &<br>oxidizing reagent (added by<br>LCx) | None |
| Dedicated Laboratory<br>Area | 1 Room | 2 Rooms (separate sample<br>processing and<br>amplification/detection) | Biological safety<br>hood |
| Dried Reagents | Yes | No | No |
| Samples to be Tested | Endocervical swabs<br>Urethral swabs<br>Urine (Male/Female) | Endocervical swabs<br>Urethral swabs<br>Urine (Male/Female) | Endocervical swabs<br>Urethral swabs |
| Swab Specimen<br>Transport | 2-27°C up to 4-6 days;<br>media-free transport | 2-30°C up to 4 days;<br>liquid transport media | 2-8°C up to 48 hrs. or<br>frozen culture<br>transport medium |
| Urine Specimen<br>Transport 2-8°C | 2-8°C up to 4-6 days | 2-8°C up to 4 days | Not available |
| Urine Specimen<br>Transport 15-27°C | 15-27°C up to 2 days | Not available | Not available |
| Feature | BDProbeTec™ ET | Abbott LCx® | GC Culture |
| Intended Use | The BDProbeTec™ ET CT/GC<br>Amplified DNA Assay, when used<br>with the BDProbeTec™ ET<br>System, uses SDA technology for<br>the direct, qualitative detection of<br>CT/GC DNA in endocervical<br>swabs, male urethral swabs, and<br>in female and male urine<br>specimens as evidence of infection<br>with CT, GC, or of co-infection with<br>both CT and GC. Specimens may<br>be from symptomatic or<br>asymptomatic females for the<br>BDProbeTec ET CT and GC<br>Assays, from symptomatic or<br>asymptomatic males for the<br>BDProbeTec ET CT Assay, and<br>from symptomatic males for the<br>BDProbeTec ET GC Assay. A<br>separate Amplification Control is<br>an option for inhibition testing<br>(BDProbeTec ET CT/GC/AC Reagent Pack). | The LCx GC Assay uses LCR™<br>(Ligase Chain Reaction)<br>amplification technology in the<br>LCx Probe System for the direct,<br>qualitative detection of a specific<br>target nucleic acid sequence in<br>the Opa gene of GC in female<br>endocervical and male urethral<br>swab specimens or in female and<br>male urine specimens from<br>symptomatic and asymptomatic<br>males and females. | GC culture may use a<br>variety of selective<br>culture media to grow<br>and isolate gram<br>negative diplococci.<br>Identification of N.<br>gonorrhoeae relies on<br>biochemical and/or other<br>identification methods. |
| Type of Assay | Amplified DNA Probe | Amplified DNA Probe | Growth & detection |
| Technology | SDA | LCR | Culture |
| Amplification Target | Pilin gene inverting protein<br>homolog | Opa gene | Not applicable |
| Detection Format | Simultaneous amplification &<br>detection | Amplification followed by<br>detection | Not applicable |
| Qualitative or Quantitative | Qualitative | Qualitative | Qualitative and<br>quantitative |
| Assay Formats | CT/GC (GC not available<br>separately) | GC | GC |
| Amplification Control | Optional | No | No |
| Number of Controls / Run | Negative Control (1)<br>Positive Control (1) | Negative Control (2X)<br>Calibrator (2X)<br>Positive Control | 1 positive and 1<br>negative control per<br>batch |
| Contamination Control<br>Method | Closed System | Chelating metal complex &<br>oxidizing reagent (added by LCx) | None |
| Dedicated Laboratory Area | 1 Room | 2 Rooms (separate sample<br>processing and<br>amplification/detection) | None |
| Dried Reagents | Yes | No | No |
| Samples to be Tested | Endocervical swabs<br>Urethral swabs<br>Urine (Male/Female) | Endocervical swabs<br>Urethral swabs<br>Urine (Male/Female) | Endocervical swabs<br>Urethral swabs<br>Urine (Male/Female) |
| Swab Specimen Transport | 2-27°C up to 4-6 days;<br>media-free transport | 2-30°C up to 4 days;<br>liquid transport media | Inoculate into selective<br>media. Incubate at 35-<br>39°C in a CO₂ enriched<br>atmosphere immediately<br>after incubation. |
| Urine Specimen Transport<br>2-8°C | 2-8°C up to 4-6 days | 2-8°C up to 4 days | |
| Urine Specimen Transport<br>15-27°C | 15-27°C up to 2 days | Not available | |
Table 1: Device Comparison - Chlamydia trachomatis (CT)
{3}------------------------------------------------
# Table 2: Device Comparison - Neisseria gonorrhoeae (GC)
{4}------------------------------------------------
#### SUMMARY OF PERFORMANCE DATA:
# ANALYTICAL STUDIES:
Precision of the BDProbeTec™ ET CT and GC Amplified DNA Assay was demonstrated by testing a five member panel consisting of four dilutions coinoculated with CT and GC in sample diluent and a negative (uninoculated sample diluent). The five member panel is made up of samples containing 0-100 C. trachomatis EBs/rxn and 0-100 N. gonorrhoeae cells/rxn. This precision panel was run at two clinical sites and internally. Six replicates of each panel were run twice a day for three days. No significant run-to-run or site-to-site variability was observed.
Reproducibility was determined by evaluating one panel consisting of seeded swab specimens, and another panel consisting of seeded buffer to simulate urine specimens. The 30 member swab panels contained 12 replicates of a level seeded with both 500 EBs/rxn (CT) and 500 cells/rxn (GC), 12 replicates of a level seeded with both 50 EBs/rxn (CT) and 30 cells/rxn (GC) and six unseeded samples. The 30 member urine panels contained 12 replicates of a level seeded with both 600 EBs/rxn (CT) and 500 cells/rxn (GC), 12 replicates of a level seed with both 115 EBs/rxn (CT) and 100 cells/rxn (GC) and six unseeded samples. Results were combined across 23 operators and across all sample levels (negative, low level, high level) to estimate reproducibility. Eighteen of 23 (78%) operators were at least 95% reproducible with CT swab specimens: 14/23 (61%) of the operators were at least 95% reproducible for CT buffer specimens. For GC specimens, 22/23 (96%) of the operators were at least 95 % reproducible with GC swab specimens and 20/23 (87%) achieved 95% reproducibility with GC buffer specimens.
The analytical sensitivity (limit of detection) of the BDProbeTec™ ET CT and GC Amplified DNA Assay was determined by diluting 15 C. trachomatis serovars and 39 N. gonorrhoeae strains in CT/GC Diluent. Quantitated CT cultures were diluted to 0,5,15,35,70 and 200 EBs per reaction for each serovar. Quantitated GC cultures were diluted to 0,5,10,15,and 25 cells per reaction for each strain. Samples were processed and assayed in triplicate. The analytical sensitivity of the CT serovars ranged from 5-200 EBs per reaction with a median of 35 EBs per reaction. The LOD of the 39 N. gonorrhoeae strains ranged from 5-25 cells per reaction with a median of 5 cells per reaction. These strains included 14 ATCC strains (including six different N. gonorrhoeae auxotypes) and 25 clinical isolates obtained from geographically diverse sites.
{5}------------------------------------------------
A total of 156 bacteria, viruses, and yeast were tested with the BDProbeTec™ ET CT and GC Amplified DNA Assay. Bacterial isolates were tested using at least 10° Colony Forming Units (CFU)/ml or equivalent copies of genomic DNA. Viruses were tested using at least 108 Plaque Forming Units (PFU)/ml or equivalent copies of genomic DNA. The organisms tested include those commonly found in the urogenital tract as well as others. For Chlamydia trachomatis, all results were negative as expected. Three N. cinerea strains were tested in the BDProbeTec ET GC assay. Of these, two were repeatedly positive. Sixteen N. subflava strains were tested in triplicate. Two strains were positive in one of three replicates. When the new strains were prepared and tested again, all results were negative. Eight N. lactamica strains were tested in triplicate. One strain was positive in one of the three replicates. When that strain was prepared and tested again, all results were negative.
Potential interfering substances which may be encountered in swab and/or urine specimens were tested with the BDProbeTec™ ET CT and GC Amplified DNA Assays. Potential interfering substances were evaluated in the absence of target or with 200 CT EBs per reaction and 200 GC cells per reaction. False negative results may be caused by leukocytes and blood > 5% in swabs and by leukocytes, blood, bilirubin, and phenazopyridine in urine. When using the AC, these substances may also cause indeterminate results.
## CLINICAL STUDIES:
Performance characteristics for the BDProbeTec™ ET CT and GC Amplified DNA Assays were established in a multicenter study at seven geographically diverse clinical sites. The final data analysis included 4108 CT and 4105 GC specimens collected from 2109 patients attending sexually transmitted disease clinics, OB/GYN clinics, family planning clinics, adolescent clinics, and emergency rooms. Paired specimens (swab and urine) were collected from 2020 of the 2109 patients. Four endocervical swabs and one urine specimen were collected from female patients. The swabs were tested by cell culture for CT, culture for GC, the BDProbeTec ET assay, and a commercially available amplification method (AMP1). The endocervical swab collection order was rotated throughout the study to minimize effects of collection order. For males, two urethral swabs and one urine specimen were collected. The first swab was used for GC culture and then the BDProbeTec ET assay. The second swab was used for CT cell culture. Male and female urine specimens were tested on both the BDProbeTec ET system and AMP1. If cell culture was negative, but either amplification assay was positive, a DFA test was performed from the cell culture transport medium. A different commercially available amplification assay (AMP2) was performed from culture transport medium for those male patients who had a positive urine AMP1 test and the corresponding swabs were culture negative.
{6}------------------------------------------------
Performance characteristics for CT and GC were calculated both with and without the amplification control (AC). Data are presented without the AC. Assay interpretation differences resulting from use of the AC are footnoted at the bottom of each table.
BDProbeTec ET C. trachomatis results were compared to culture and patient infected status. Performance estimates for each specimen type and symptomatic status are shown in Table 1. A patient was considered if (1) the culture was positive, or (2) positive results were obtained for both AMP1 (in either the swab or urine) and DFA, or (3) AMP1 was positive in both swab and urine paired specimens. Data on pregnant females are footnoted at the bottom of Table 1. Of the 1.419 female swab specimens tested in the clinical evaluations by the BDProbeTec ET CT assay, 101 (7.1%) were classified as grossly bloody and 242 (17.1%) as moderately bloody. Assay performance with moderately to grossly bloody swabs was not statistically different than assay performance with non-bloody or lightly bloody swabs.
BDProbeTec ET N. gonorrohoeae results were compared to culture and patient infected status. Performance estimates for each specimen type and symptomatic status are shown in Table 2. A patient was considered if (1) the culture was positive or (2) in females, if AMP1 was positive in both swab and urine (paired specimens). Data on pregnant females are footnoted at the bottom of Table 2. Of the 1.411 female swab specimens tested in the clinical evaluations by the BDProbeTec ET GC assay, 102 (7.2%) were classified as grossly bloody and 242 (17.2%) as moderately bloody. Assay performance with moderately to grossly bloody swabs was not statistically different than assay performance with non-bloody or lightly bloody swabs.
{7}------------------------------------------------
Overall performance of the BDProbeTec™ ET Chlamydia trachomatis and Neisseria gonorrhoeae Amplified DNA Assays on the BDProbeTec™ ET System, is substantially equivalent1 to CT cell culture and GC culture methods that were in use prior to May 28, 1976 and to the Abbott LCx ® Chlamydia trachomatis Assay and the Abbott LCx® Neisseria gonorrhoeae Assay.
1 The term "substantial equivalence" as used in this 510(k) notification is limited to the definition of substantial equivalence as found in the Federal Food, Drug and Cosmetic Act, as amended and as applied under 21 CFR 807, Subpart E under which a device can be marketed without pre-market approval or reclassification. A determination of substantial equivalency under this notification is not intended to have any bearing whatsoever on the resolution of patent infringement suits or any other patent matters. No statements related to, or in support of substantial equivalence herein shall be construed as an admission against interest under the US Patent Laws or their application by the courts.
{8}------------------------------------------------
| Specimen<br>Type | S/A | Performance Compared to Culture | | Performance Compared to Patient<br>Infected Status | | #Indeterminate<br>Initial/Final<br>(With AC) | DFA or AMP1 Positive in<br>either specimen type/<br>BDPT Positive- Patient<br>Infected Negative |
|------------------|--------|---------------------------------|--------------------------------|----------------------------------------------------|--------------------------------|----------------------------------------------|-------------------------------------------------------------------------------------------------|
| FS | S | 90.9% (50/55)<br>80.0-97.0 | 97.6% (531/544)<br>95.9-98.7 | 88.7% (55/62)<br>78.1-95.3 | 98.5% (529/537)<br>97.1-99.4 | 3/1 | 3/8 |
| | A | 100% (47/47)<br>92.5-100 | 96.1% (743/773)<br>94.5-97.4 | 96.8% (61/63)<br>89.0-99.6 | 97.9% (741/757)<br>96.6-98.8 | 6/0 | 8/16 |
| | Total | 95.1% (97/102)<br>88.9-98.4 | 96.7% (1274/1317)<br>95.6-97.6 | 92.8% (116/125)<br>86.8-96.7 | 98.1% (1270/1294)<br>97.3-98.8 | 9/1 | 11/24 |
| FU1 | S | 75.9% (41/54)2<br>62.4-86.5 | 97.3% (506/520)<br>95.5-98.5 | 77.0% (47/61)3<br>64.5-86.8 | 98.2% (505/513)<br>97.0-99.3 | 71/34 | 4/8 |
| | A | 91.3% (42/46)<br>79.2-97.6 | 96.9% (694/716)<br>95.4-98.1 | 83.9% (52/62)4<br>72.3-92.0 | 98.3% (688/700)<br>97.0-99.1 | 90/47 | 5/12 |
| | Total5 | 83.0% (83/100)<br>74.2-88.2 | 97.1% (1200/1236)<br>96.0-98.0 | 80.5% (99/123)<br>72.4-87.1 | 98.4% (1193/1213)<br>97.4-99.0 | 161/81 | 9/20 |
| MS | S | 95.8% (92/96)<br>89.7-98.3 | 89.9% (356/396)<br>86.5-92.7 | 95.5% (105/110)<br>89.7-98.5 | 92.9% (355/382)<br>89.9-95.3 | 1/0 | 16/27 |
| | A | 88.2% (15/17)<br>63.6-98.5 | 95.9% (162/169)<br>91.7-98.3 | 89.5% (17/19)<br>66.9-98.7 | 97.0% (162/167)<br>93.2-99.0 | 1/0 | 2/5 |
| | Total | 94.7% (107/113)<br>88.8-98.0 | 91.7% (518/565)<br>89.1-93.8 | 94.6% (122/129)<br>89.1-97.8 | 94.2% (517/549)<br>91.9-96.0 | 2/0 | 18/326 |
| MU1 | S | 95.8% (91/95)<br>89.6-98.8 | 86.5% (340/393)<br>82.7-89.7 | 95.4% (104/109)<br>89.6-98.5 | 89.4% (339/379)<br>85.9-92.4 | 20/10 | 28/40 |
| | A | 88.2% (15/17)<br>63.6-98.5 | 94.7% (161/170)<br>90.2-97.6 | 89.5% (17/19)<br>66.9-98.7 | 95.8% (161/168)<br>91.6-98.3 | 16/3 | 5/7 |
| | Total | 94.6% (106/112)<br>88.7-98.0 | 89.0% (501/563)<br>86.1-91.5 | 94.5% (121/128)<br>89.1-97.8 | 91.4% (500/547)<br>88.7-93.6 | 36/13 | 33/477 |
| Total8 | | 92.0% (393/427)<br>89.1-94.4 | 94.9% (3493/3681)<br>94.1-95.6 | 90.7% (458/505)<br>87.8-93.1 | 96.6% (3480/3603)<br>95.9-97.1 | 208/95 | 71/123 |
Tawe 1: BDProbeTec ET CT Results vs. Culture and Patient Infected Sieuds
1 Comparison cultures for female and male urine specifical and male urethral swab specimens, respectively,
² With A.C. two final indeetminates reported (instead of false negative), resulting in increase in sensitivity from 7.3 to 96.9%
* With AC, wo final indexeminates reported instead of false negative, resulting in increase in somitively from 77.0% v81.4% and decrease in specificity from 98.4% to 98.1%
* With AC, one final indeterminate reported (insealting in increase in sensitivity from 83.9% to 8.2% and decrease in specificity from 93.3% b98.2%.
5 With AC, female urine sensitivity for culture were 85.7% and 96.8%, respectively; and for patient infected status were 8.3% and 98.1%, respectively
6 13 of 16 of the AMP1 urine positives were confirmed by AMP2 testing.
7 14 of 30 of the AMP1 urine positives were confirmed by AMP2 testing.
8 With AC, total sensitivity and specificity for culture were 92.7% and for patient infected status were 91.4% and 96.5%, respectively
Note Separate performance characteristics were collected from pregant females. Sensibity and specificity compared o patient infected status for female swals and urines were 94.4% (17/18) , 98.4% (122/124) and 83.3% (15/18) , 100% (120/120), respectively.
{9}------------------------------------------------
| Specimen<br>Type | S/A | Compared to Culture | | Compared to Patient Infected Status | | # Indeterminate<br>Initial/Final | AMP1 Positive in either<br>specimen type /<br>BDPT Positive- Patient<br>Infected Negative |
|------------------|--------|------------------------------|--------------------------------|-------------------------------------|--------------------------------|----------------------------------|-------------------------------------------------------------------------------------------|
| | | Sensitivity<br>95% C.I. | Specificity<br>95% C.I. | Sensitivity<br>95% C.I. | Specificity<br>95% C.I. | | |
| FS | S | 95.8% (46/48)2<br>85.7-99.5 | 98.7% (545/552)<br>97.4-99.5 | 96.1% (49/51)3<br>86.5-99.5 | 99.3% (545/549)<br>98.1-99.8 | 3/1 | 1/4 |
| | A | 97.1% (34/35)<br>85.1-99.9 | 99.2% (770/776)<br>98.3-99.7 | 97.4% (37/38)<br>86.2-99.9 | 99.6% (770/773)<br>98.9-99.9 | 5/0 | 1/3 |
| | Total4 | 96.4% (80/83)<br>89.8-99.2 | 99.0% (1315/1328)<br>98.3-99.5 | 96.6% (86/89)<br>90.5-99.3 | 99.5% (1315/1322)<br>98.9-99.8 | 8/1 | 2/7 |
| FU1 | S | 84.8% (39/46)<br>71.1-93.7 | 99.2% (527/531)<br>98.1-99.8 | 83.7% (41/49)<br>70.3-92.7 | 99.6% (526/528)<br>98.6-100 | 79/38 | 0/2 |
| | A | 88.2% (30/34)<br>72.5-96.7 | 99.0% (713/720)<br>98.0-99.6 | 86.5% (32/37)<br>71.2-95.5 | 99.3% (712/717)<br>98.4-99.8 | 75/48 | 1/5 |
| | Total | 86.3% (69/80)<br>76.7-92.9 | 99.1% (1240/1251)<br>98.4-99.6 | 84.9% (73/86)<br>75.5-91.7 | 99.4% (1238/1245)<br>98.8-99.8 | 154/86 | 1/7 |
| MS5 | S | 98.4% (187/190)<br>95.5-99.7 | 94.8% (290/306)<br>91.6-97.0 | 98.4% (187/190)<br>95.5-99.7 | 94.8% (290/306)<br>91.6-97.0 | 1/0 | 16/16 |
| MU1,6 | S | 97.9% (185/189)<br>94.7-99.4 | 94.4% (286/303)<br>91.2-96.7 | 97.9% (185/189)<br>94.7-99.4 | 94.4% (286/303)<br>91.2-96.7 | 28/15 | 14/17 |
| Total | | 96.1% (521/542)<br>94.1-97.6 | 98.2% (3131/3188)<br>97.7-98.6 | 95.8% (531/554)<br>93.8-97.4 | 98.5% (3129/3176)<br>98.2-99.0 | 191/102 | 33/47 |
#### Table 2: BDProbeTec ET GC Results vs. Culture and Patient ...rected Status
1 Comparison cultures for female and male urine performed on endocervical and male urethral swab specimens, respectively.
2With AC one indeterminate reported (instead of false negative), resulting in increase in sensitivity from 95.8 to 97.9%
3With AC, one indeterminate reported (instead of false negative), resulting in increase in sensitivity from 96.1 to 98.0%
1 With AC, female swabs sensitivity and specificity for culture were 97.6% and 99.0%, respectively, and for patient infected status were 97.8% and 99.5%, respectively
• Data from swab specimens collected from 187 asymptomatic males were insufficient number (4) of infected patients to alequately determine performance characteristics.
· Data from wine speciment collected from 188 asymptomatic males were insufficient number (4) of infected patients to adequately determine performance characteristics.
Note: Separate characteritics vere calculated from pregant females. Sensibily and specificity and specificity composed o patient infected stans for finale swas and urines were 100% (2/2) , 98.6% (137/139) and 100% (2/2) , 98.5% (133/135), respectively.
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines representing hair or movement.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
NOV - 4 1999
Ms. Colleen Rohrbeck Regulatory Affairs Associate Becton Dickinson Microbiology Systems 7 Loveton Circle Sparks. Maryland 21152
K984631 Re:
> Trade Name: BDProbeTec™ ET Chlamydia trachomatis and Neisseria gonorrhoeae Amplified DNA Assay BDProbeTec™ ET Chlamydia trachomatis Amplified DNA Assay Regulatory Class: II, I Product Code: LSL, MKZ Dated: August 10, 1999 Received: August 11, 1999
Dear Ms. Rohrbeck:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{11}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"
Sincerely yours,
Steven Butman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{12}------------------------------------------------
Page 1 of 1
510(k) Number (if known): K984631
Device Name: BDProbeTec™ ET Chlamydia trachomatis and Neisseria gonorrhoeae Amplified DNA Assay BDProbeTec™ ET Chlamydia trachomatis Amplified DNA Assay
Indications For Use:
The BDProbeTec™ ET Chlamydia trachomatis (CT) and Neisseria gonorrhoeae (GC) Amplified DNA Assays, when tested with the BDProbeTec ET System, use Strand Displacement Amplification (SDA) technology for the direct, qualitative detection of Chlamydia trachomatis and Neisseria gonorrhoeae DNA in endocervical swabs, male urethral swabs, and in female and male urine specimens as evidence of infection with C. trachomatis. N. gonorrhoeae, or of co-infection with both C. trachomatis and N. gonorrhoeae. Specimens may be from symptomatic or asymptomatic females for the BDProbeTec ET CT and GC Assays, from symptomatic or asymptomatic males for the BDProbeTec ET CT Assay, and from symptomatic males for the BDProbeTec ET GC Assay. A separate Amplification Control is an option for inhibition testing (BDProbeTec ET CT/GC/AC Reagent Pack).
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Woody Dubose
finical Laboratory Devices 510(k) Number
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.