Intravascular diagnostic catheters are used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels.
Device Story
The AVE 4F SiteSeer (II) is an intravascular diagnostic catheter used in the cardiovascular system. It functions by being inserted into the heart and vessels to record intracardiac pressures, sample blood, and deliver substances such as radiopaque contrast media. The device is available in various curve styles to facilitate navigation. It is intended for use by clinicians in a hospital or clinical setting. The catheter provides a pathway for diagnostic procedures, allowing healthcare providers to obtain physiological data and administer contrast agents to visualize the vascular system, thereby aiding in cardiovascular diagnosis and decision-making.
Clinical Evidence
Bench testing only. Performance data included tensile strength, bending stiffness, straight-line torque, high-pressure static leak, flow rate, and natural frequency testing. Biocompatibility testing conducted in compliance with ISO 10993 included acute intracutaneous reactivity, acute systemic toxicity, cytotoxicity, hemolysis, material-mediated pyrogen, and sensitization tests. Shelf-life testing confirmed performance specifications are met through the expiration date.
Technological Characteristics
Intravascular diagnostic catheter; 4F diameter; available in various curve styles. Materials, packaging, and sterilization methods are consistent with predicate devices. Biocompatibility verified per ISO 10993 standards.
Indications for Use
Indicated for use in the cardiovascular system to record intracardiac pressures, sample blood, and introduce substances (including radiopaque contrast media) into the heart and vessels. No contraindications specified.
Regulatory Classification
Identification
An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.
Predicate Devices
AVE SiteSeer™ Catheter
AVE Envision™ Catheter
Cordis Infiniti™ Angiographic Catheter
Cordis Vista™ Bright tip Catheter
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image contains a sequence of characters and numbers. The sequence starts with the letter 'K', followed by the number '9', then '84', '62', and ends with the number '7'. The characters and numbers are written in a handwritten style.
... ...
## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION VI.
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.
| <b>A. Submitter Information:</b> | |
|-----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <b>Name:</b> | Arterial Vascular Engineering, Massachusetts |
| <b>Address:</b> | 129 Concord Road, Billerica, MA 01821 |
| <b>Phone:</b> | (978) 739-3116 |
| <b>Fax:</b> | (978) 777-0390 |
| <b>Contact Person:</b> | Fred Boucher |
| | Regulatory Affairs Manager |
| <b>Date of Preparation:</b> | 6/28/99 |
| <b>B. Device Name</b> | |
| <b>Trade Name:</b> | AVE 4F SiteSeer (II) Cardiovascular<br>Angiographic Catheter |
| <b>Common Name:</b> | Cardiovascular Angiographic Catheter |
| <b>Classification Name:</b> | Diagnostic Intravascular Catheter<br>/Percutaneous Catheter |
| <b>C. Predicate Device Name(s):</b> | 1. AVE SiteSeer™ Catheter<br>2. AVE Envision™ Catheter<br>3. Cordis Infiniti™ Angiographic Catheter<br>4. Cordis Vista™ Bright tip Catheter |
| <b>D. Device Description</b> | The 4F SiteSeer (II) catheter is an<br>intravascular diagnostic catheter for use in the<br>cardiovascular system. |
| <b>E. Intended Use</b> | Intravascular diagnostic catheters are used to<br>record intracardiac pressures, to sample blood,<br>and to introduce substances into the heart and<br>vessels. |
| <b>F. Technological Characteristics</b> | The AVE 4F SiteSeer II catheter<br>is similar to the AVE 4F SiteSeer and<br>Cordis Infiniti and Vista Bright Tip<br>catheters regarding materials of<br>construction, packaging, and<br>sterilization. |
| | The indications for use are similar to<br>both the AVE Angiographic catheters<br>and the Cordis catheters. They are all |
{1}------------------------------------------------
indicated to deliver media or substances into the vascular system. The actual indications are listed below:
The AVE 4F SiteSeer II is indicated for use in the cardiovascular system. Their primary function is to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels.
Cordis Infiniti indications are The designed to deliver radiopaque contrast media to selected sites in the vascular system. Similar to the AVE and Cordis catheters, the AVE SiteSeer II catheter will be available in various curve styles.
- AVEM conducted tensile strength, bending G. Performance Date stiffness, straight-line torque, high-pressure static leak, flow rate and natural frequency testing. The test results demonstrate that the SiteSeer II catheter is substantially 4F equivalent to the AVE SiteSeer™ catheter, the AVE Envision™ Catheter, or the Cordis Vista™ Bright tip Catheter.
also conducted the following AVEM biocompatibility tests in compliance with ISO 10993: acute intracutaneous reactivity, acute systemic toxicity, cytotoxicity, hemolysis. material-mediated pyrogen and sensitization test. All tests demonstrated that the 4F SiteSeer II catheter is biocompatible.
Shelf-life testing on the AVE 4F SiteSeer II catheter demonstrates that the device meets all performance specifications within its labeled life expiration date.
Based on the results of the above testing. AVEM concludes that the AVE 4F SiteSeer catheter II is substantially equivalent to the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes representing the department's mission. The eagle is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA". The text is arranged in a circular fashion around the eagle.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 27 1999
Mr. Fred Boucher Regulatory Affairs Manager AVE Massachusetts, Inc. 129 Concord Road Billerica, MA 01821-0566
Re : K984627 4F SiteSeer™ II Cardiovascular Angiographic Catheter Regulatory Class: II (two) Product Code: 74 DQO Dated: June 18, 1999 Received: June 21, 1999
Dear Mr. Boucher:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic You may, therefore, market the device, subject to the Act (Act). general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, FDA will verify such assumptions. Failure to comply with
{3}------------------------------------------------
Page 2 - Mr. Fred Boucher
the GMP requlation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal Laws or Regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on the labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requlation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Christopher, Abon for.
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## INDICATIONS FOR USE D.
AVE 4F SiteSeer™ Cardiovascular Angiographic Catheter Device Name:
- An intravascular diagnostic catheter is a device used to Indications for Use: record intracardiac pressure, to sample blood, and to introduce substances into the heart and vessels.
None Contraindications:
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use X (Per 21 CFR 801.109)
OR
Over-the-Counter Use
Chr. M. Attridge for Callahan
(Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices K984627 510(k) Number _
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.