← Product Code [JOJ](/productcode/JOJ) · K984613

# BROWNE SPORICIDIN PHENOL INDICATOR FOR SPORICIDIN STERILIZING AND DISINFECTING SOLUTION (K984613)

_Albert Browne , Ltd. · JOJ · Feb 4, 1999 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K984613

## Device Facts

- **Applicant:** Albert Browne , Ltd.
- **Product Code:** [JOJ](/productcode/JOJ.md)
- **Decision Date:** Feb 4, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.2800
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

The Browne Sporicidin Phenol Indicator for Sporicidin Sterilizing and Disinfecting Solution (Browne Sporicidin Phenol Indicator) is a phenol concentration monitor for use in phenol-containing germicides with a minimum effective concentration (MEC) of 1.3% phenol. The Browne Sporicidin Phenol Indicator is dedicated for use with Sporicidin Sterilizing and Disinfecting Solution.

## Device Story

Device is a chemical indicator strip for monitoring phenol concentration in Sporicidin Sterilizing and Disinfecting Solution. Input: immersion of indicator pad into germicide solution. Operation: chemical reaction between phenol in solution and active ingredient in pad forms diazo addition product; subsequent pH increase triggers color change in pH-sensitive dye. Output: visual color change (yellow to green) indicating phenol concentration relative to 1.3% MEC. Used in clinical settings; operated by healthcare staff. Result read between 1-2 minutes post-immersion. Output informs user if germicide solution remains effective for disinfection; false negatives may lead to unnecessary solution disposal, but disinfection procedure is not compromised.

## Clinical Evidence

Bench testing only. Indicators tested in Sporicidin solution at 1.6% and 1.3% phenol concentrations. Results showed 100% failure rate (detection of concentration below MEC) at 1.3% concentration. No false positives observed. Some false negatives occurred, particularly with high detergent/salt contaminants or inexperienced users, leading to conservative solution disposal without compromising disinfection safety.

## Technological Characteristics

Non-sterile, disposable polypropylene strip with indicator pad. Sensing principle: chemical reaction (diazo addition) causing pH-dependent color change. No energy source. Shelf life: 2 years unopened; 90-day in-use life.

## Regulatory Identification

Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.

## Predicate Devices

- Browne GA Indicator ([K922481](/device/K922481.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

FEB 4 1995

K984613

## 510(k) SUMMARY BROWNE SPORICIDIN PHENOL INDICATOR FOR SPORICIDIN STERILIZING AND DISINFECTING SOLUTION

#### SUBMITTER NAME AND ADDRESS 1.

Mr. Alan Charlton Albert Browne Ltd. Chancery House 190 Waterside Road Hamilton Industrial Park Leicester LE5 1QZ United Kingdom

- 2. DEVICE NAME

| Proprietary Name:    | Browne Sporicidin Phenol Indicator for Sporicidin<br>Sterilizing and Disinfecting Solution |
|----------------------|--------------------------------------------------------------------------------------------|
| Common/Usual Name:   | Browne Sporicidin Phenol Indicator                                                         |
| Classification Name: | Physical/Chemical Sterilization Process Indicator                                          |

#### PREDICATE DEVICES 3.

Browne GA Indicator (K922481)

#### INTENDED USE 4.

The Browne Sporicidin Phenol Indicator for Sporicidin Sterilizing and Disinfecting Solution (Browne Sporicidin Phenol Indicator) is a phenol concentration monitor for use in phenol-containing germicides with a minimum effective concentration (MEC) of 1.3% phenol.

The Browne Sporicidin Phenol Indicator is dedicated for use with Sporicidin Sterilizing and Disinfecting Solution.

Browne Sporicidin Phenol Indicator Additional Information - K984613

1/25/99

Page A-1

{1}------------------------------------------------

K984613

#### 5. DEVICE DESCRIPTION

The Browne Sporicidin Phenol Indicator and the substantially equivalent device are chemical indicator strips intended to monitor the concentration of the active ingredient of a liquid chemical germicide. The devices indicate, via a color change, if the concentration of the active ingredient exceeds the MEC for the solution.

The Browne Sporicidin Phenol Indicator consists of a polypropylene strip with an indicator pad on one end. The indicator pad is impregnated with an indicator solution which changes color from yellow to green in liquid chemical germicides with a phenol concentration above the MEC of 1.3%.

The Browne Sporicidin Phenol Indicator and the predicate device have a 2-year shelf life from the date of manufacture in the unopened bottle. Containers of both devices have a 90 day in-use life.

#### TECHNOLOGICAL CHARACTERISTICS ર્ણ

The technological characteristics of the Browne GA Indicator are similar to that of the Browne Sporicidin Phenol Indicator described in this submission. Both devices are non-sterile, disposable strips containing an indicator pad impregnated with an indicator solution which changes color in a germicide solution at the appropriate concentration of the active ingredient of the liquid chemical germicide,

The mechanism of action for inducing a color change is identical for the Browne Sporicidin Phenol Indicator and the Browne GA Indicator. The active ingredient of the indicator solution in both devices interacts with the active ingredient in the germicide solution to produce a hydroxyl ion which ionizes a pH-sensitive dye.

In the proposed device, the phenol component of the Sporicidin Sterilizing and Disinfecting Solution reacts with the active ingredient in the indicator solution of the test strip to form a diazo addition product (STEP 1). If sufficient phenol is present, the increase in pH causes a color change in the pH indicator dye (STEP 2).

Browne Sporicidin Phenol Indicator Additional Information - K984613

1125/99

{2}------------------------------------------------

$$\forall \mathcal{A} \forall \emptyset \forall \mathcal{P}$$

| STEP 1 |                                                                                                                                                                                |                                         |   |                           |                 |        |
|--------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------|---|---------------------------|-----------------|--------|
| ArOH   | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | Ar'N,                                   | t |                           | ArN=NAr'        | + NaOH |
| phenol |                                                                                                                                                                                | active ingredient<br>indicator solution |   | diazo addition<br>product |                 |        |
| STEP 2 |                                                                                                                                                                                |                                         |   |                           |                 |        |
| NaOH   | +                                                                                                                                                                              | pH sensitive dye                        |   |                           | green color dye |        |

The indicator color is dependent on the phenol concentration of the germicide solution, and the time after exposure when the results of the test are read, as described in the Table A-1 below.

Table A-1. Color Development for the Sporicidin Phenol Indicator Strip

| Time<br>(minutes) | Phenol Concentration (%) |                                |       |
|-------------------|--------------------------|--------------------------------|-------|
|                   | ≤1.3                     | 1.4-1.5                        | ≥1.6  |
| <1                |                          | yellow, green/yellow           |       |
| 1-2               | yellow                   | yellow, green/yellow,<br>green | green |
| >2                |                          | yellow, green/yellow           |       |

During the first 60 seconds after the test strip has been dipped into the Sporicidin Sterilizing and Disinfecting Solution, the yellow test strip will begin to develop a green color.

At 60 seconds (1 minute), the strip will exhibit a uniform green color (except for the top 2 mm of the strip) if the concentration of phenol in the Sporicidin Sterilizing and Disinfecting Solution is ≥1.6%. The strip will appear parchy green/yellow or yellow if the solution contains 51.3% phenol. In the concentration range of 1.4-1.5% phenol, the strip may appear yellow, green/yellow or green.

Browne Sporicidin Phenol Indicator Additional Information - K984613

1/25/99

. the . word wise the child the sell to

{3}------------------------------------------------

K984613

From 60 to 120 Seconds (1 to 2 minutes) the color of the strip is stable. A color reading must be taken during this time period for the results to accurately reflect the phenol concentration of the Sporicidin Sterilizing and Disinfecting Solution.

After 120 seconds (2 minutes) the color of the strip regresses toward the original yellow color. The rate of regression is dependent on the phenol concentration of the Sporicidin Sterilizing and Disinfecting Solution.

The indicator solution of the proposed device changes color from yellow to green to indicate a PASS, while that of the Browne GA Indicator changes color from yellow to purple to indicate a PASS. The difference in colors chosen for these indicators does not represent a difference in the mechanism of action for the production of a color change.

#### PERFORMANCE TESTING 7.

The performance characteristics of the Browne Sporicidin Phenol Indicator were established by testing indicators in Sporicidin Sterilizing and Disinfecting Solution containing 1.6% and 1.3% phenol. The test indicators showed a 100% failure rate in the 1.3% solution when the testing was performed according to the Instructions for Use. Some false negatives were observed. The frequency of false negatives increased when the indicators were tested in the presence of high levels of detergent and salt contaminants, or when testing was performed by inexperienced users.

No false positives were recorded in any of the testing conducted. The false negatives reported would have caused the user to discard the solution unnecessarily. No disinfection procedure would have been compromised.

The data demonstrates that the Browne Sporicidin Phenol Indicator is an effective monitor for the phenol component of the Sporicidin Sterilizing and Disinfecting Solution with a phenol MEC of 1.3%.

Browne Sporicidin Phenol Indicator Additional Information - K984613

1/25/99

Page A-4

{4}------------------------------------------------

Image /page/4/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines representing hair or clothing.

FEB 4 1999 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Albert Browne Ltd. C/O Cynthia J.M. Nolte, Ph.D. Associate Regulatory Consultant Medical Device Consultants, Incorporated 49 Plain Street 02760 North Attleboro, Massachusetts

Re : K984613 Browne Sporicidin Phenol Indicator Trade Name: for Sporicidin Sterilizing and Disinfecting Solution Requlatory Class: II Product Code: JOJ December 28, 1998 Dated: December 29, 1998 Received:

Dear Dr. Nolte:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. ਉ substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531

{5}------------------------------------------------

Page 2 - Dr. Nolte

through 542 of the Act for devices under the Electronic chrough Jr. Or on Control provisions, or other Federal laws or requlations.

This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA debelibed is possible equivalence of your device to a legally marketed predicate device results in a classification for your marketed predicate acits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on compreadios as (advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address

"http://www.fda.gov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Timothy A. Ulatowski

Direc Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

# 510(k) Number (if known): _ 1984613

### Browne Sporicidin Phenol Indicator For Sporicidin Sterilizing and Device Name: Disintecting Solution

Indications For Use:

The Browne Sporicidin Phenol Indicator For Sporicidin Sterilizing and Disinfecting Solution (Browne Sporicidin Phenol Indicator) is a phenol concentration monitor for use in phenol-containing germicide solutions with a minimum effective concentration of 1.3% phenol.

The Browne Sporicidin Phenol Indicator is dedicated for use with Sporicidin Sterilizing and Disinfecting Solution.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Qian S. Lin

(Division Sign-Off) Division of Dental, Infection Control, and General Hospital Devices C 98 Ub 510(k) Number

Prescription Use (Per 21 CFR 801.109)

OR Over-The-Counter Use X

---

**Source:** [https://fda-staging.innolitics.com/device/K984613](https://fda-staging.innolitics.com/device/K984613)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).