K984567 · Ventana, Inc. · MYA · Aug 12, 1999 · Hematology
Device Facts
Record ID
K984567
Device Name
VENTANA ER PRIMARY ANTIBODY (CLONE 6F11)
Applicant
Ventana, Inc.
Product Code
MYA · Hematology
Decision Date
Aug 12, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 864.1860
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K984567 · Aug 12, 1999
VENTANA ER PRIMARY ANTIBODY (CLONE 6F11)
Ventana, Inc.
Archived clinical breast cancer tissue biopsies; Patient medical records containing concurrent ER SBA/DCC analysis results
A retrospective study was conducted to validate the device's performance by comparing immunohistochemical staining results against historical clinical data (SBA/DCC) obtained from archived patient samples.
Sensitivity, specificity, and concordance of ER IHC relative to SBA/DCC
Allred, et. al. study; Retrospective analysis of prognostic value; Follow-up/Duration: 5 years
820 patients (no adjuvant therapy) and 800 patients (endocrine therapy); Sample Size: 1620
Not applicable for this study
Disease free survival (DFS)
Indications for Use
Ventana ER primary antibody (clone 6F11) is intended for laboratory use for the qualitative detection of estrogen receptor (ER) antigen in sections of formalin fixed, paraffin embedded normal and neoplastic tissue on a Ventana automated slide staining immunohistochemistry device. It is indicated as an aid in the management, prognosis and prediction of therapy outcome of breast cancer within the context of the patient's clinical history and other diagnostic tests evaluated by a qualified pathologist.
Device Story
Monoclonal antibody (clone 6F11) binds to estrogen receptor (ER) antigen in nuclear regions of normal/neoplastic tissue; used on Ventana automated slide staining devices. Process: tissue pretreatment/antigen enhancement; inhibitor application to reduce endogenous peroxidase; primary antibody incubation; biotin-conjugated secondary antibody application; streptavidin-enzyme conjugate binding; chromogenic substrate application. Output: visualized precipitating enzyme product on microscope slides. Operated by laboratory personnel; results interpreted by qualified pathologist via light microscopy. Output aids in breast cancer prognosis and therapy prediction by identifying ER-positive status. Benefits: provides standardized immunohistochemical assessment of ER status to guide clinical decision-making.
Clinical Evidence
Retrospective study of 100 breast cancer cases comparing Ventana ER IHC (clone 6F11) to SBA/DCC cytosol receptor assay. Primary endpoints: sensitivity, specificity, and concordance. Results at ≥11% tumor cell nuclei cutoff: 98.6% sensitivity (95% CI: 92.5-100%), 78.6% specificity (95% CI: 59.1-91.7%), and 93% concordance (95% CI: 86-97%). Additional specificity testing performed on 76 normal and 19 neoplastic tissues.
Technological Characteristics
Mouse anti-human ER monoclonal antibody (clone 6F11, IgG kappa). Reagent concentration ~1 mg/ml. Utilizes indirect biotin-avidin-enzyme detection system. Requires antigen enhancement (microwave/citrate buffer). Compatible with Ventana automated slide staining devices. Visualization via chromogenic enzyme substrate (DAB, AEC, or Alkaline Phosphatase).
Indications for Use
Indicated for use in the management, prognosis, and prediction of therapy outcome of breast cancer in patients. Used for qualitative detection of ER antigen in formalin-fixed, paraffin-embedded tissue sections to aid in the diagnosis of ER-positive tumors.
Regulatory Classification
Identification
Immunohistochemistry test systems (IHC's) are in vitro diagnostic devices consisting of polyclonal or monoclonal antibodies labeled with directions for use and performance claims, which may be packaged with ancillary reagents in kits. Their intended use is to identify, by immunological techniques, antigens in tissues or cytologic specimens. Similar devices intended for use with flow cytometry devices are not considered IHC's.
Special Controls
(2) Class II (special control, guidance document: “FDA Guidance for Submission of Immunohistochemistry Applications to the FDA,” Center for Devices and Radiologic Health, 1998). These IHC's are intended for the detection and/or measurement of certain target analytes in order to provide prognostic or predictive data that are not directly confirmed by routine histopathologic internal and external control specimens. These IHC's provide the pathologist with information that is ordinarily reported as independent diagnostic information to the ordering clinician, and the claims associated with these data are widely accepted and supported by valid scientific evidence. Examples of class II IHC's are those intended for semiquantitative measurement of an analyte, such as hormone receptors in breast cancer.
Predicate Devices
Abbot Laboratories' ER-ICA Monoclonal, Immunocytochemical Assay for the Detection of Human Estrogen Receptor (P880026)
Submission Summary (Full Text)
{0}------------------------------------------------
# K984567
## 510(k) SUMMARY Ventana ER Primary Antibody (Clone 6F11)
#### Submitted By:
Ventana Medical Systems, Inc. 3865 N. Business Center Drive Tucson, Arizona 85705 Tel: 520-887-2155 Fax: 520-887-2558 Contact: Judith Frederick, MS, CCRA Date: August 4, 1999
FDA/CDRH/ODE/DHHS
6 AUG 99
13 52
RECEIVED
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is K984567.
#### Name of the Device:
| Trade Name: | Ventana ER Primary Antibody (Clone 6F11) |
|-----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Antibody for aid in the detection of estrogen receptor (ER) antigen in histological tissue sections |
| Classification: | To the best of our knowledge ER antibody for aid in the detection of ER antigen in histological tissue sections is classified as class II and codified at 21 CFR 864.1860; Medical Devices; Classification/Reclassification Immunohistochemistry Reagents and Kits, June 3, 1998. |
Predicate Device: Ventana ER primary antibody (clone 6F11) is substantially equivalent to commercially available Abbot Laboratories' ER-ICA Monoclonal, Immunocytochemical Assay for the Detection of Human Estrogen Receptor, approved by the FDA as PMA # P880026 and downclassified to class II by 21 CFR 864.1860 Immunohistochemistry Reagents and Kits on June 3, 1998.
Description: Ventana's ER Primary Antibody (clone 6F11) is a monoclonal antibody which specifically binds to estrogen receptor antigen located in the nuclear region of a variety of normal and neoplastic tissues. The dispenser contains approximately 5 ug (50 test) or 25 ug (250 test) of mouse anti-human ER monoclonal antibody. The total protein concentration of the reagent is approximately 1 mg/ml. Specific antibody concentration is approximately 1 ug/ml. Clone 6F11 is immunoglobulin class IgG, light chain kappa. There is no known irrelevant antibody in the preparation.
Intended Use: Ventana ER primary antibody (clone 6F11) is intended for laboratory use for the qualitative detection of estrogen receptor (ER) antigen in sections of formalin fixed, paraffin embedded normal and neoplastic tissue on a Ventana automated slide staining immunohistochemistry device. It is indicated as an aid in the management, prognosis and
Ventana Medical Systems, Inc.
{1}------------------------------------------------
prediction of therapy outcome of breast cancer within the context of the patient's clinical history and other diagnostic tests evaluated by a qualified pathologist.
Technology of the device and predicate device: The demonstration of antigens in tissue and cells by immunostaining is a process involving first the binding of an antibody to an antigen of interest and second, visualization of the bound primary antibody by an indirect biotin-avidin system coupled to an enzyme. Ventana ER primary antibody (clone 6F11) is located by a biotin conjugated secondary antibody formulation which recognizes rabbit or mouse immunoglobulins. This step is followed by the addition of a streptavidin-enzyme conjugate which binds to the biotin on the secondary antibody. The primary antibody-secondary antibody-avidin enzyme complex is visualized by using a precipitating enzyme generated product. Ventana ER Primary Antibody (clone 6F11) was developed for use on Ventana's automated slide staining devices. Briefly, the procedure is as follows:
- 1. Pretreatment of tissues or cells on a microscope slide with an antigen-enhancement procedure is required.
- 2. An inhibitor solution is added to the slide on which antigen-enhanced tissues or cells are affixed. The inhibitor reduces the endogenous peroxidase activity in the tissue. This step takes 4 minutes at 37° C. The slide is washed with buffer and covered with liquid coverslip.
- 3. Optimized ER Primary Antibody (clone 6F11) is applied, mixed and incubated for 4 to 32 minutes at 37° C. The slide is washed with buffer and covered with liquid coverslip.
- 4. Streptavidin-enzyme reagent is added to the slide, mixed and incubated at 37° C. (The time of incubation is indicated in each detection kit insert.) The slide is washed with buffer and covered with liquid coverslip.
- 5. The chromogenic enzyme substrate is applied to the slide and mixed. The solutions are incubated with mixing at 37° C. (The time of incubation is indicated in each detection kit insert.) The slide is washed and ready for counterstaining, copper enhancement, or can be prepared for microscopic reading immediately.
Abbott ER-ICA Monoclonal detects estrogen receptor in frozen tissue sections affixed to glass slides in a technically similar fashion to the immunostaining process used for staining paraffin embedded tissue. The frozen sections are fixed in formaldehyde, methanol and acetone, then treated with Blocking Reagent to prevent nonspecific binding. The primary antibody (the Ig fraction of a rat monoclonal antibody specific for ER receptor) is then added to the tissue. Visualization of the bound primary antibody is via an indirect peroxidase anti-peroxidase (PAP) system. ER-ICA Monoclonal is located by a goat anti-rat bridging antibody. Anti-Rat PAP complex which recognizes the bridging antibody is then added. This step is followed by the addition of a precipitating enzyme generated product (hydrogen peroxidase) and a chromogen substrate solution, containing DAB. The resulting insoluble reddish brown precipitation allows visualization of the bound primary antibody ?. The two devices are similar in that both systems specifically bind to ER proteins in breast carcinoma tissues, both employ similar detection chemistry principles for visualization of the product, and both aid in the diagnosis and prognosis of breast carcinoma.
{2}------------------------------------------------
#### Performance Data:
1. Tissue Specificity: The staining patterns of 76 normal tissues were examined. These included adrenal, bone marrow, breast, cerebrum, cervix, colon, endometrium, esophagus, heart, kidney, liver, lung, mesothelium, ovary, pancreas, peripheral nerve, prostate, salivary gland, skeletal muscle, skin, small intestine, spleen, stomach, testis, thyroid, tonsil, and thymus. Tissues not tested as recommended in "FDA Guidance for Submission of Immunohistochemistry Applications to the FDA", were cerebellum, pituitary, and parathyroid. Staining was considered to be appropriate, with the exception of the unexpected negative staining of one case of breast tissue, and one case of endometrium. The pathologist noted however, that in the case of the breast, the tissue was poorly preserved, so the negative staining may be due to fixation. Positive nuclear staining was found in the lobular and ductal cells of the breast (2 of 3 cases), epithelia and fibromuscular cells of the cervix (3 of 3 cases), the glandular epithelia, stromal and smooth muscle cells of the endometrium (2 of 3 cases), and the surface epithelium and stromal cells of the ovary (3 of 3 cases). All other tissues were negative.
The staining patterns of 19 neoplastic tissues were also evaluated by a qualified pathologist. Neoplastic tissues examined were lung, prostate, colon, lymphoma, bladder, stomach, cervix, and ovary. Positive nuclear staining occurred in the tumor tissue of both cases of ovarian cancer (100% of tumor cell nuclei stained) and in the surrounding normal tissue of two cases of cervical cancer. All other tissues were negative.
Prior to staining with Ventana ER primary antibody (Clone 6F11), the dried, deparaffinized sections used for the specificity studies were treated for antigen enhancement. Slides containing the tissue sections were placed in citrate buffer (pH 5.75) and heated under pressure in a microwave for approximately 30 minutes. The slides were allowed to stand for an additional 30 minutes (approximate), then transferred to fresh citrate buffer until ready to run on the Ventana automated slide stainer. Optimized ER Primary Antibody (clone 6F11) was applied, mixed and incubated 32 minutes at 37° C. Endogenous biotin was blocked using Ventana Medical Systems' Endogenous Biotin Blocking Kit (catalogue number 250-050). Detection used was Ventana's DAB Kit (catalogue number 250-001). DAB Detection includes: 1) An inhibitor solution, added to reduce the endogenous peroxidase activity in the tissue. 2) Universal biotinylated secondary antibody. 3) Avidin-HRPO. 4 and 5) Basic DAB solution and hydrogen peroxide. The two reagents are applied together and mixed on the specimen slide. And 6) Copper enhancement. The incubation time for each reagent was pre-programmed at Ventana.
2. Validation of Assay Performance by Comparison with an FDA approved chemical cytosol receptor assay for ER, serum binding assay/ dextran coated charcoal (SBA/DCC): The safety and effectiveness of Ventana's ER Primary Antibody (clone 6F11) have been examined through clinical testing by Ventana Medical Systems, Inc. Ventana verified that ER Primary Antibody (clone 6F11) detects ER in breast carcinoma tissue in an equivalent manner to SBA/DCC by examining 100 cases of breast cancer for expression of ER. The cases were randomly chosen from an archived collection of breast carcinoma biopsies (300 cases total) at one investigative site. The study was retrospective in design, using archived tissue samples embedded in paraffin
{3}------------------------------------------------
blocks that had concurrent ER SBA/DCC analyses in the patient file. For antigen enhancement, the investigator used a 10 minute autoclave steam heating procedure and the Biogenex Citra antigen retrieval solution. The staining protocol was identical to that used in the specificity studies.
Quality control procedures included a positive tissue control (uterus) placed on every slide run in addition to the case tissue. Each case had two slides stained with Ventana ER Primary Antibody (Clone 6F11) and one slide with Ventana Negative Control Ig (catalogue # 250-2014). For the test to be considered valid, the positive control tissue was expected to exhibit nuclear staining of the endometrial glands and stroma. These components were to be negative when stained with Negative Control Ig. In addition, a negative tissue control slide (ER-negative breast carcinoma) was included for every batch of samples processed and run on the Ventana automated slide staining device. This negative tissue control was stained with ER Primary Antibody (Clone 6F11) to ensure that the antigen enhancement and other treatment procedures did not create false positive staining. The study cases were evaluated within the context of the performance of the controls.
Staining of ER primary antibody (clone 6F11) was assessed both as a percentage of cells stained and by staining intensity. Scoring of tumor cell percentages was estimated and categorized into the following cell count ranges: 0, 1-10, 11-25, 26-75, and over 75 percent. If the scorer was uncertain of the category, 100 tumor cells in each of three fields at 40X magnification was counted and the values averaged for a true count percentage. The emphasis in data analysis was on percent positive cells rather than stain intensity. Scoring of the benign elements was also performed to aid in discrepancy resolution. The pathologist scoring the IHC was masked to the results of the ER SBA/DCC. Results were as follows:
| Cut off for % Positive Cells | TP | TN | FP | FN | TOTAL |
|------------------------------|----|----|----|----|-------|
| Raw ER ≥1% | 71 | 15 | 12 | 2 | 100 |
| Raw ER ≥ 11% | 71 | 21 | 6 | 2 | 100 |
| Resolved ER ≥1% | 71 | 16 | 12 | 1 | 100 |
| Resolved ER > 11% | 71 | 22 | 6 | 1 | 100 |
## Table 1: ER IHC vs. ER SBA/DCC
Scoring Criteria: ER IHC positivity was compared to ER SBA/DCC at different cut-off values to determine the cut-off value for ER Primary Antibody (Clone 6F11) that best agreed with the SBA/DCC results. Sensitivity, specificity, and concordance of the ER test, relative to that of the ER SBA/DCC test. improved using a cut-off value of > 11% compared to > 1%. Ventana therefore recommends evaluating slides with a cut-off for positivity of > 11% positive staining of tumor cell nuclei. Cell intensity was variable and was not used as a criteria for positivity. ER SBA/DCC was scored as positive at greater than or equal to 10 femtomoles (fm).
Discrepancy Resolution: Because SBA/DCC cannot distinguish between positivity of pathological and benign elements, the protocol allowed for the resolution of false negative ER
{4}------------------------------------------------
IHC to true negative if benign elements were positive and pathological elements did not stain. The results of these analyses were noted as "Resolved". A possible explanation for the negative SBA/DCC results of the 6 cases is estrogen receptor degradation due to poor tissue handling during sample collection'. The one false negative ER-IHC case may have been due to sample over-fixation.
Agreement: ER IHC with Ventana's ER Primary Antibody (Clone 6F11) showed the best agreement with ER SBA/DCC (x2 = 67.82, P < 0.0001) when ≥ 11% of tumor cell nuclei were scored as positive. 93% of the cases were concordant, with a 95% confidence interval of 86 -97%.
Sensitivity: Resolved sensitivity relative to ER SBA/DCC at a cut off of > 11% was 98.6% (71/72), with a 95% confidence interval of 92.5 - 100%.
Specificity: Resolved specificity relative to ER SBA/DCC at a cut off of > 11% was 78.6% (22/28), with a 95% confidence interval of 59.1 - 91.7%. One possible reason for the low specificity in relation to the other values, was the few number of SBA/DCC negative cases in the randomly selected pool (n=28).
3. Prognosis and prediction: Allred, et. al. also investigated the prognostic value of ER, using formalin fixed, paraffin embedded tissue and clone 6F11. In a group of 820 patients receiving no adjuvant therapy, he found the difference in disease free survival (DFS) at five years follow up to be ten per cent. In a subset of 800 patients receiving endocrine therapy, however, ER positivity by IHC was associated with a 25% improvement in DFS over 5 years' duration'.
Conclusions: Ventana claims that ER Primary Antibody (clone 6F11) is substantially equivalent to Abbott ER-ICA Monoclonal, subject of P880026, which was downclassified from class III to Class II as a result of 21 CFR 864.1860, Immunohistochemistry Reagents and Kits. Performance characteristics of the Ventana ER Primary Antibody were validated in a comparison with an FDA approved chemical cytosol receptor assay for ER, serum binding assay/dextran coated charcoal (SBA/DCC). Demonstrated performance characteristics were: 98.6% relative sensitivity, 78.6% relative specificity, 93% concordance. Ventana further claims that ER Primary Antibody (clone 6F11) is a useful aid in assessing prognosis and prediction of therapy outcome for the breast cancer patient.
## Bibliography:
- 1. Allred, D. C. et. al. Prognostic and predicitve facotrs in breast cancer by immunohistochemical analysis. Mod. Pathol., 11: 155-168, 1998.
- 2. Bevitt, D. J. et. al. New monoclonal antibodies to oestrogen and progesterone receptors effective for paraffin section immunohistochemistry. J. Pathol. 183: 228-232, 1997.
- 3. Abbott ER-ICA Monoclonal Package Insert. Abbott Laboratories Diagnostics Division. Customer Support Center (USA) 1-800-323-9100.
Ventana Medical Systems, Inc.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features an abstract image of an eagle with three stripes forming its body and head. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the top of the seal. The seal is black and white.
AUG 12 1999
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Ms. Judith Frederick, MS, CCRA Senior Clinical Coordinator Ventana Medical Systems, Inc. 3865 North Business Center Drive Tucson, Arizona 85705
Re: K984567
Trade Name: Ventana ER Primary Antibody (Clone 6F11) Regulatory Class: II Product Code: MYA Dated: July 22, 1999 Received: July 26, 1999
Dear Ms. Frederick:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{6}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours.
Steven Sutman
Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
510(k) Number (if known): K984567
Device Name: ER Primary Antibody (Clone 6F11)
# Indications For Use:
Ventana Medical Systems' ER Primary Antibody (Clone 6F11) may be used in immunohistochemical methods to aid in the identification of estrogen receptor (ER) antigen in normal and neoplastic cells as an aid in the diagnosis of ER positive tumors. The anti-ER (Clone 6F11) is intended for laboratory use to qualitatively stain ER antigen in sections of formalin fixed, paraffin embedded tissue on a Ventana automated slide staining device. It is further intended as an aid in the assessment of prognosis in the evaluation of breast cancer. Detection chemistries used by Ventana to detect ER are DAB, AEC, Alkaline phosphatase red and blue. Light microscopy is used to detect the staining of cell components. It is indicated for use in the management, prognosis and prediction of therapy outcome of breast cancer.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division Sign-Up
Division of Clinical Laboratory Devices
510(k) Number K984567
**Prescription Use**
(Per 21 CFR 801.109)
OR
**Over-The-Counter Use**
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.