Parker cuffed tubes are designed for oral intubation and are indicated for airway management.
Device Story
Parker Flex-Tip and FlexEMS endotracheal tubes are sterile, single-use cuffed tracheal tubes for airway management. Designed for oral intubation, these devices facilitate mechanical ventilation in anesthesia and emergent care. Key design modifications from standard tubes include a flexible, rounded, rear-beveled distal tip tapering toward the midline and two lateral Murphy eyes flanking the bevel. These features aim to reduce trauma to the nose, larynx, and trachea during insertion. The device is operated by clinicians in clinical settings. The tube's physical design allows for smoother navigation through the airway compared to traditional side-beveled tubes, potentially reducing mucosal injury. The device is intended for prescription use.
Clinical Evidence
Clinical evidence includes a cadaveric study using videoscopy to compare trauma during orotracheal and nasotracheal intubation, showing the Parker tube was less traumatic than standard tubes. A prospective comparative clinical study of 66 adult surgical patients (35 Parker tube, 31 standard tube) concluded the Parker tube is at least as safe and efficacious for orotracheal intubation as the predicate device.
Technological Characteristics
Cuffed tracheal tube; sizes 6.5, 7.0, 7.5, 8.0 mm ID. Features flexible, rounded, rear-beveled distal tip and dual lateral Murphy eyes. Complies with ASTM F1242-96 except for tip bevel and Murphy eye configuration. Sterile, single-use. Materials and biocompatibility identical to predicate.
Indications for Use
Indicated for airway management via oral intubation in patients requiring mechanical ventilation in anesthesia, emergent, and respiratory care settings.
Regulatory Classification
Identification
A tracheal tube is a device inserted into a patient's trachea via the nose or mouth and used to maintain an open airway.
{0}------------------------------------------------
JUL 15 1999
# K984528 Premarket Notification
# 510(k) Summary
Submitted by:
Parker Medical
2121 Herrick Ave. Cincinnati, Ohio 45208 Tel. 513-321-1097 Fax: 513-321-2097
109 Inverness Drive East, Suite J Englewood, CO 80112-5105 Tel. 303-799-1990 Fax: 303-799-1996
Contact person: Dr. Jeffrey D. Parker
Contact telephone: 513-321-1097
Date of Summary: March 18, 1999
Name of the Device
Trade names: Parker Flex-Tip Endotracheal Tube Parker FlexEMS Endotracheal Tube
Common name: Cuffed tracheal tube
Classification name: Tracheal tube
Classification:
Class II, 21 CFR 868.5730
Predicate device:
Kendall Sheridan CF Tracheal Tube (K822082)
{1}------------------------------------------------
510(k) Summary Page 2
### Product Description:
The Parker tracheal tubes are sterile, single use devices, for use in anesthesia and emergent and respiratory care where mechanical ventilation is required. The tube will be made in four sizes: 6.5, 7.0, 7.5, and 8.0 mm ID. It is identical to the predicate device in material composition, biocompatibility, packaging, sterilization, and in all structural respects, except for the following features:
- (1) Instead of a semi-rigid, side-beveled distal tip, the Parker tube has a flexible, slightly rounded, rear-beveled distal tip which tapers centrally toward the midline of the lesser curvature of the tube.
- (2) Instead of a single, lateral Murphy eye opposite the bevel, the Parker tube has two, facing Murphy eyes flanking the bevel.
- (3) Instead of a radiopaque stripe on the greater curvature of the tube and a cuff inflation circuit on the lesser curvature of the tube, the positions of these two features are reversed on the Parker tube.
#### Indications and Usage:
Parker tracheal tubes are designed for oral intubation and are indicated for airway management. Other, reinforced tracheal tubes should be used to reduce the potential for kinking whenever an unusual positioning of the head or neck is to be required following intubation.
#### Non-clinical Performance Data:
ASTM F1242-96 is widely regarded as a performance standard written to be applicable to "standard" fracheal tubes, such as the predicate device, but not necessarily to all aspects of specialized tubes, such as the Parker tube. Nevertheless, Parker tracheal tubes comply with ASTM F1242-96 in all respects except for the following:
- (1) Clause 6.4.1 provides for a tip which is beveled laterally. The predicate device meets this requirement. The Parker tube tip, by design, is beveled posteriorly, toward the greater curvature of the tube, rather than laterally.
{2}------------------------------------------------
510(k) Summary Page 3
- (2) Clause 6.5 provides for a Murphy eye on the side of the tube opposite the bevel. The predicate device meets this requirement. The Parker tube tip, by design, has two Murphy eyes, rather than one, and they are placed lateral to, rather than opposite, the bevel.
Clause 4.2.1 provides that the patient end of the tube shall be Note: "rounded without sharp points or rough edges." The predicate device meets this requirement. Although the Parker tube tip is tapered, its point is rounded, rather than sharp, and its edges are also rounded, rather than rough.
#### Clinical Performance Data:
To determine whether the unique tip design of the Parker tracheal tube would adversely affect its safety or efficacy, both orotracheal and nasotracheal intubations were carried out in adult human cadavers with standard tracheal tubes and with Parker tracheal tubes. These intubations were watched through a videoscope, and some were recorded. In this study, the tip of the Parker tracheal tube appeared to be clearly and consistently less traumatic to the nose, larynx, and trachea than standard tracheal tubes, including the predicate device.
A comparative clinical study of intubation was also carried out with IRB approval and informed consent on 66 adult patients orotracheally intubated for surgery. 31 of those patients were intubated with a standard tracheal tube and 35 were intubated with a Parker tracheal tube. The conclusion drawn from this study was that the Parker tracheal tube is at least as safe and efficacious for orotracheal intubation as standard tracheal tubes.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol that resembles an abstract representation of a human figure or a bird in flight, composed of three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 15 1999
Jeffrey D. Parker, M.D. Parker Medical 109 Inverness Drive East, Suite J Englewood, CO 80112-5105
Re: K984528 Parker Flex-Tip Tracheal Tube and Parker FlexEMS Tracheal Tube Requlatory Class: II (two) Product Code: 73 BTR Dated: April 22, 1999 April 22, 1999 Received:
Dear Dr. Parker:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 A substantially equivalent determination assumes compliance to 895. with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such Failure to comply with the GMP regulation may result in assumptions. regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2 - Jeffrey D. Parker, M.D.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Thomas J. Callehan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Page A4
#### 510(k) NUMBER: K984528
### DEVICE NAMES: PARKER FLEX-TIP TRACHEAL TUBE and PARKER FlexEMS TRACHEAL TUBE
### "INDICATIONS AND USAGE"
Parker cuffed tubes are designed for oral intubation and are indicated for airway management.
(PLEASE DO NOT WRITE BELOW THIS LINE. CONTINUE ON ANOTHER PAGE IF NEEDED.)
Office of Device Evaluation (ODE) Concurrence of CDRH,
Prescription Use (Per 21 CFR 801.109)
OR
Over-the-Counter-Use (Optional Format 1-2-96)
M. R
(Division Sign-Off)
Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number
TOTAL P.02
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.