K984502 · I-Flow Corp. · MEB · Mar 4, 1999 · General Hospital
Device Facts
Record ID
K984502
Device Name
NERVE BLOCK INFUSION KIT
Applicant
I-Flow Corp.
Product Code
MEB · General Hospital
Decision Date
Mar 4, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Nerve Block Infusion Kit is intended to provide continuous infusion of a local anesthetic near a nerve for regional anesthesia and pain management for preoperative, perioperative and postoperative general and orthopedic surgery. Additional routes of infusion include percutaneous, subcutaneous, epidural and into the intraoperative (soft tissue / body cavity) site and synovial cavity.
Device Story
Disposable elastomeric infusion pump; provides continuous, fixed-rate delivery of local anesthetics. Operates via strain energy of elastomeric membranes forcing fluid through fixed-diameter tubing or glass orifice; no external power required. Kit includes pump, catheter, needle, syringe, and accessories. Used in hospital, home, or alternative care sites; operated by clinicians or patients. Provides ambulatory regional anesthesia/pain management. Output is continuous infusion; healthcare providers monitor infusion progress visually via pump volume. Benefits include portable, non-electronic pain management without risk of fluid runaway.
Clinical Evidence
Bench testing only. Flow rate performance testing conducted at nominal fill volumes (50-500 ml) and standard operating conditions (31ºC skin temperature, 0.9% NaCl). All models demonstrated average flow rates within ±15% accuracy claim at 95% confidence interval. Biological testing performed in conformance with ISO 10993 Part 1 for all fluid path components.
Technological Characteristics
Elastomeric infusion pump; mechanical power via elastomeric membrane strain energy. Flow control via fixed-diameter tubing or glass orifice. Fill volumes 50-500 ml; flow rates 0.5-10 ml/hr. Fluid path materials identical to PainBuster pump. Sterilization via ETO or radiation. No external power source. No software or electronic components.
Indications for Use
Indicated for patients undergoing general or orthopedic surgery requiring regional anesthesia or pain management via continuous local anesthetic infusion. Routes include percutaneous, subcutaneous, epidural, and intraoperative soft tissue/body cavity or synovial sites.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
{0}------------------------------------------------
3/4/99
Premarket Notification - 510(k) Summary of Safety and Effectiveness Nerve Block Infusion Kit Page 2 of 4
# GENERAL INFORMATION 1.0
### Purpose of Submission 1.1
K984502
- This submission is intended to notify the Federal Food and Drug 1.1.1 Administration that I-Flow Corporation intends to market a new product line to be called the Nerve Block Infusion Kit.
- Trade Name: Nerve Block Infusion Kit 1.1.2
- Common Name: Elastomeric Infusion Pump Kit 1.1.3
- 1.1.4 Classification Name: Pump, Infusion, Elastomeric
- Classification Panel: General Hospital and Personal Use Device 1.1.5
#### 1.2 Statement of Equivalence
- The Nerve Block Infusion Kit is substantially equivalent in total or in part to the 1.2.1 following: the I-Flow PainBuster Infusion Kit (K980558, K982946), the Sqarlato Pain Control Infusion Pump (PCIP) (K896422), the I-Flow Homepump C-Series (K944692), the B. Braun Continuous Nerve Block Set (pre-amendment device) and the McKinley Outbound Syringe Infuser (K982256).
- The Nerve Block Infusion Kit includes components that are legally marketed 1.2.2 (either pre-amendment devices or devices that have been granted permission to market via premarket notification regulation).
- The pumps used in the Nerve Block Infusion Kit are identical in design and 1.2.3 materials to the pumps used in the PainBuster Infusion Kit.
- 1.3 Definitions (from Taber's Cyclopedic Medical Dictionary, 1997)
- Nerve Block: The induction of regional anesthesia by preventing sensory 1.3.1 nerve impulses from reaching the centers of consciousness. This is usually done on a temporary basis, by using chemical or electrical means. In the former case. it is accomplished by injecting an anesthetic solution, such as procaine, around the nerve but at some distance, or by anesthetizing nerve endings in the region itself (infiltration).
# PHYSICAL SPECIFICATIONS AND DESCRIPTIONS 2.0
## 2.1 Description of the Nerve Block Infusion Kit
- The Nerve Block Infusion Kit is nearly identical to the I-Flow PainBuster 2.1.1 Infusion Kit with the exception of a new intended use, the addition of three new pump models and the addition of three new kit components (insulated Tuohy needle, hemostasis valve assembly and hookup wire).
- 2.1.2 The kit is comprised of an elastomeric infusion pump (K944692) and various components such as catheter, needle, syringe, dressing, tape, gauze, hemostasis valve assembly, hookup wire and accessories such as carry case, power ring and clothing attachment clip (E-Clip).
- 2.1.2.1 The PainBuster kit contains all the above components except for the insulated Tuohy needle, hemostasis valve assembly and hookup wire.
{1}------------------------------------------------
- The Nerve Block Infusion Kit is intended to provide continuous infusion of a 2.1.3 local anesthetic near a nerve for regional anesthesia and pain management for pre-operative, perioperative and postoperative general and orthopedic surgery. Additional routes of administration include percutaneous, subcutaneous, epidural and into the intraoperative (soft tissue / body cavity) site and synovial cavity.
- The Nerve Block pump is a disposable device intended for single patient use. 2.1.4
- The Nerve Block pump is suitable for use as an ambulatory device and is 2.1.5 intended for use in the hospital, home environment or alternative care sites.
### 2.2 Product Configuration
- The Nerve Block kit models are available in fill volumes from 50 to 500 ml and 2.2.1 flow rates from 0.5 to 10 ml/hr.
- Each model consists of a kit with a Nerve Block pump and the following 2.2.2 optional components/accessories:
- Catheter, needle, hemostasis valve assemlby, hookup wire, 2.2.2.1 syringe, dressing, carry case, E-clip, antiseptic skin swabs, tape, gauze and power ring.
#### 2.3 Components and Materials
All fluid path components of the Nerve Block pump are identical to the fluid path components of the PainBuster pump.
#### 2.4 Pumping Mechanism
- The pressure that pumps the fluid comes from the strain energy of the 2.4.1 elastomeric membranes which are forced to expand when the pump is filled.
- The administration set consists of fixed diameter flow control tubing or glass 2.4.2 orifice.
#### 2.5 Power Requirements
- 2.5.1 The Nerve Block pump is a mechanical pump that utilizes elastomeric membranes for power. No additional external power source is required.
## OPERATIONAL SPECIFICATIONS AND DESCRIPTIONS 3.0
## Standard Operating Conditions: 3.1
| <= 10 | 500 ml volume |
|-----------|---------------------------------|
| <= 9 | 270 ml volume |
| <= 4 | 125 ml volume |
| <= 3 | 65 ml volume |
| | 31ºC skin temperature (90ºF) |
| 0.9% NaCl | |
| | 14 to 7 psi pressure source |
| 16" | |
| | ±15% at 95% confidence interval |
| | |
- 3.2 Flow Rate Performance Data: Testing occurred at standard operating conditions. Testing occured at nominal fill volumes for each model. All models produced an average flow rate within the ±15% accuracy claim.
{2}------------------------------------------------
## 3.3 Safety / Alarm Functions
- The Nerve Block pump provides a continuous fixed flow and as such is not 3.3.1 subject to fluid runaway conditions similar to that of some electronic pumps. This device contains no alarms or indicators.
# BIOLOGICAL SPECIFICATIONS 4.0
- Biological testing is in conformance with ISO 10993 Part 1 for all fluid path 4.1 components of the Nerve Block pump.
# 5.0 CHEMICAL AND DRUG SPECIFICATIONS
## 5.1 Compatibility
- There are no specific drugs referenced in the labeling for the Nerve Block 5.1.1 Infusion Kit.
- The Nerve Block Infusion Kit is intended for use with general local anesthetics 5.1.2 and epidural medications.
# 6.0 INTENDED USE
- The Nerve Block Infusion Kit is intended to provide continuous infusion of a local 6.1 anesthetic near a nerve for regional anesthesia and pain management for preoperative, perioperative and postoperative general and orthopedic surgery. Additional routes of infusion include percutaneous, subcutaneous, epidural and into the intraoperative (soft tissue / body cavity) site and synovial cavity.
- 6.2 The Nerve Block pump is single patient use only.
- No testing has been conducted to determine the efficacy of the Nerve Block pump for 6.3 the delivery of blood, blood products, lipids or fat emulsions. The Nerve Block pump is not intended for the delivery of blood, blood products, lipids or fat emulsions.
- 6.4 The Nerve Block pump is suitable for use as an ambulatory device and is intended for use in the hospital, home environment or alternative care sites.
# 7.0 STANDARDS
- 7.1 There are currently no standards established for elastomeric infusion pumps.
# 8.0 PACKAGING
- 8.1 Packaging is suitable for either radiation or ETO sterilization.
# 9.0 STERILIZATION INFORMATION
- 9.1 The methods of sterilization are ETO gas.
# 10.0 COMPARISON TO LEGALLY MARKETED DEVICES
- 10.1 The Nerve Block Infusion Kit has similar routes of administration and components as the following predicate devices: the B. Braun Continuous Nerve Block Set . PainBuster Infusion Kit, Sqarlato Pain Control Infusion Pump (PCIP), Homepump C-Series and McKinley Outbound Syringe Infuser.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with healthcare. The caduceus is depicted with three lines forming a shape resembling a human profile. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the caduceus.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR - 4 1999
Robert J. Bard, Esq., R.A.C. Vice President Regulatory and Legal Affairs I-Flow Corporation 20202 Window Drive Lake Forest, California 92630
Re : K984502 Nerve Block Infusion Kit Trade Name: Regulatory Class: II Product Code: MEB December 17, 1998 Dated: December 18, 1998 Received:
Dear Mr. Bard:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਚੋ substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) requlation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2 - Mr. Bard
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fdaggov/cdrh/dsmamain.html".
Sincerely yours,
Timothy A. Ulatowski
Direttor Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known): _
Device Name: _ Nerve Block Infusion Kit
# Indications for Use:
The Nerve Block Infusion Kit is intended to provide continuous infusion of a local anesthetic near a nerve for regional anesthesia and pain management for pre-operative, perioperative and nostoperative general and orthopedic surgery. Additional routes of infusion include percutaneous, subcutaneous, epidural and into intraoperative sites.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUED ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Palbociclib Ceearde
(Division Sign-Off) Division of Dental, Infection Control, and General Hospital Devices 510(k) Number _
V Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use_
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.