The Tigran TM Hip Prosthesis for total hip arthoplasty consists of three basic parts: femoral heads, femoral stems, and the instrumentation necessary for proper implantation of these components. The Tigran TM Hip Prosthesis is designed for implantation into the human femur in total or hip replacement and is indicated for the following: 1) Patients suffering from severe hip pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head, and nonunion of previous fractures of the femur 2) Patients with congenital hip dysplasia, protrusion acetabuli, or slipped capital femoral epiphysis; 3) Patients suffering from disability due to previous fusion; 4) Patients with previously failed endoprostheses and/or total hip components in the affected extremity: 5) Patients with acute femoral neck fractures.
Device Story
Tigran Hip Prosthesis is a modular, cementless femoral hip system for total hip arthroplasty. Components include a Titanium Ti-6Al-4V alloy femoral stem with a rectangular cross-section and surface finish for rotational stability/fixation, and a Cobalt-Chromium-Molybdenum alloy femoral head (28mm diameter). The system utilizes a Morse-type tapered neck for modular assembly. Instrumentation is provided for implantation. Used by orthopedic surgeons in clinical settings to replace diseased or damaged hip joints. The device provides a stable articulating surface when paired with a compatible UHMWP acetabular component. Benefits include pain relief and restoration of hip function for patients with degenerative or traumatic hip conditions.
Clinical Evidence
Long-term clinical testing was performed to demonstrate safety and performance. No specific metrics or study results were provided in the summary document.
Technological Characteristics
Materials: Titanium Ti-6Al-4V alloy (femoral stem), Cobalt-Chromium-Molybdenum alloy (femoral head). Design: Modular, cementless, straight femoral stem with rectangular cross-section, Morse-type tapered neck. Surface: Prepared finish for fixation. Compatibility: Designed for use with UHMWP acetabular components. Sterilization: Not specified.
Indications for Use
Indicated for patients with severe hip pain/disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis, nonunion of femoral fractures, congenital hip dysplasia, protrusion acetabuli, slipped capital femoral epiphysis, disability from previous fusion, failed endoprostheses/total hip components, or acute femoral neck fractures.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
{0}------------------------------------------------
-Ponmed Limited
Confidential
Tigran 510(k)
P981117
Date of current draft 226/99
Section 16.0 510(k) Summary
K984492
## 510K(k) SUMMARY
SUBMITTER:
Ponmed Limited
DATE PREPARED: November 17th, 1998 Tigran™ Hip Prosthesis DEVICE NAME: CLASSIFICATION NAME: Femoral Hip Prosthesis
PREDICATE DEVICES:
The Tigran TM Hip Prosthesis is substantially equivalent to the Zimmer Versys Hip System (510(k) Number K 964769) and the DePuy Stability™ Hip System (K 915594).
## Device Description:
The Tigran™ Femoral Prosthesis for hip arthroplasty consists of three basic parts: femoral heads, femoral stems, and the instrumentation necessary for proper implantation of these components. The femoral stem is made from Titanium Ti-6A1-4V Alloy, a high fatigue strength material with a long history of clinical use and exceptional The surface of the femoral stem is prepared with a surface finish and biocompatibility. pattern to provide optimal initial and lasting fixation in bone. The femoral stem has a collar and is modular in design, incorporating a Morse-type tapered neck and utilizing modular femoral heads. The Femoral stem is straight and with a rectangular crosssectional shape to provide optimal, rotational stability. The Tigran™ Prosthesis Femoral Stem is available in four neck sizes (short, medium, long and extra long).
The femoral head is fabricated from Cobalt-Chromium-Molybdenum Alloy and comes in three neck configurations (short, medium and long) and in one head diameter (28mm). The various neck lengths provided allow this prosthesis to fit different variations found in the human femora.
The titanium and chrome-cobalt alloy components can be used together, although they are made from dissimilar alloys. In vitro electrochemical corrosion studies have shown that no detrimental changes occur when the two alloys are coupled.
Special instruments are available to assist in mating, disassembling, and the accurate
# 000003
{1}------------------------------------------------
· Ponned Limited
| Confidential | Filename<br>Tigran 510(k) | File No.<br>P981112 | Date of current draft<br>2/26/99 |
|--------------|---------------------------|---------------------|----------------------------------|
|--------------|---------------------------|---------------------|----------------------------------|
installation of the implant components.
The Tigran™ Femoral Prosthesis (femoral stem and head) may be used with any acetabular prosthesis with an articulating surface made of ultra high molecular weight polyethylene (UHMWP) and designed to be used with a 28 mm CoCrMo Alloy femoral head.
## Intended Use:
The Tigran TM Hip Prosthesis for total hip arthoplasty consists of three basic parts: femoral heads, femoral stems, and the instrumentation necessary for proper implantation of these components. The Tigran TM Hip Prosthesis is designed for implantation into the human femur in total or hip replacement and is indicated for the following:
- 1) Patients suffering from severe hip pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head, and nonunion of previous fractures of the femur
- 2) Patients with congenital hip dysplasia, protrusion acetabuli, or slipped capital femoral epiphysis;
- 3) Patients suffering from disability due to previous fusion;
- 4) Patients with previously failed endoprostheses and/or total hip components in the affected extremity:
- 5) Patients with acute femoral neck fractures.
This indication statement is essentially the same as the indication statement for the predicate device.
#### Comparison to Predicate Devices:
All hip systems listed in the previous table are substantially equivalent to each other and to the Tigran TM Hip Prosthesis which are intended for cementless fixation into the intramedullary canal for pathological or degenerative conditions involving the femur and / or acetabulum. All predicate devices feature a porous or roughened surface and are modular in design. All incorporate pure titanium, a Titanium alloy, or hydroxy apetitie as a surface texture coating or interface for the femoral stem. Each has Morse-type proximal neck taper that mates with a femoral head which, in turn articulates upon the acetabular component. All predicate devices and the Tigran TM Hip Prosthesis are manufactured from metal alloys that have a history of successful clinical use in orthopedic applications.
# 000004
{2}------------------------------------------------
- Ponned Limited
| Confidential | Filename | File No. | Date of current draft |
|--------------|---------------|----------|-----------------------|
| | Tigran 510(k) | P981112 | 2/26/99 |
The Tigran ™ Hip Prosthesis is substantially equivalent to the Zimmer Versys Hip System (510(k) Number K 964769) and the DePuy Stability™ Hip System (K 915594).
Examination of the information pertaining to the Tigran ™ Hip Prosthesis demonstrates that this device is substantially equivalent in composition, intended use, packaging and labeling to other hip prostheses currently approved for commercial distribution in the United States by the FDA. There are no significant differences between these marketed products and our proposed device.
## Summary of Non-Clinical Tests:
Non-clinical mechanical testing was performed to demonstrate that the femoral stem met or exceed the appropriate testing standards / requirements
## Clinical Test Results:
Long-term clinical testing was performed and reported on to demonstrate the safety and performance of the Tigran™ Hip Prosthesis.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. The bottom half of the circle contains a stylized image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 1 5 1999
Jeffrey R. Shideman, Ph.D. Ponmed Limited 7307 Glouchester Drive Edina, Minnesota 55435
Re: K984492 Tigran™ Hip Prosthesis Trade Name: Regulatory Class: II Product Codes: LPH and JDI Dated: November 17, 1998 December 17, 1998 Received:
#### Dear Dr. Shideman:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
{4}------------------------------------------------
Page 2 - Jeffrey R. Shideman, Ph.D.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Witten, Ph.D., M.D.
Director
Division of General and
Restorative Devices
Office of Device Evaluation
Center for Devices and
Radiological Health
Enclosure
{5}------------------------------------------------
**Page** / of /
#### 510 (k) NUMBER (IF KNOWN): K984492
DEVICE NAME: Tigran Hip Prosthesis
INDICATIONS FOR USE:
Indications for Use Statement
- 1) Patients suffering from severe hip pain and disability due to rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, collagen disorders, avascular necrosis of the femoral head, and nonunion of previous fractures of the femur
- 2). Patients with congenital hip dysplasia, protrusion acetabuli, or slipped capital femoral epiphysis;
- 3) Patients suffering from disability due to previous fusion;
- 4) Patients with previously failed endoprostheses and/or total hip components in the affected extremity;
- 5) Patients with acute femoral neck fractures.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use 11/1 (Per 21 CFR 801.109)
OR
Over-The-Counter-Use
(Optional Format 1-2-96)
to coely
(Division Sign-Off)
Division of General Restorative Devices K984492
510(k) Number ***_*****_**
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.