← Product Code [GEX](/productcode/GEX) · K984486

# PULSEMASTER DENTAL LASER SYSTEM (K984486)

_American Dental Technologies, Inc. · GEX · May 13, 1999 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K984486

## Device Facts

- **Applicant:** American Dental Technologies, Inc.
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** May 13, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The American Dental Technologies, Inc. PulseMaster Dental Laser Systems are indicated for use for selective ablation of enamel (first degree) dental caries.

## Device Story

PulseMaster Dental Laser System is an Nd:YAG pulsed laser (1064 nm) for selective ablation of first-degree enamel caries. Device comprises a cabinet housing laser head, power supply, cooling system, and microprocessor control panel, connected to a fiber optic delivery system. Operated by dental professionals in clinical settings. Laser energy is delivered to the tooth surface to ablate carious enamel. Output is controlled via the microprocessor interface. Clinical benefit includes safe and effective removal of enamel caries compared to traditional high-speed handpieces.

## Clinical Evidence

Evidence includes bench testing comparing temporal pulse structures and effects on enamel caries against a pulsed Er:YAG laser. Clinical evidence includes an extracted tooth study evaluating histological effects, and a multi-center, blinded, randomized clinical study comparing the PulseMaster to high-speed handpieces, establishing safety and effectiveness.

## Technological Characteristics

Nd:YAG laser system; 1064 nm wavelength; pulsed emission. Components: cabinet (laser head, power supply, cooling system, microprocessor control panel) and fiber optic delivery system. Class II device (21 CFR 878.48).

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- Centauri Dental Laser ([K932683](/device/K932683.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K9844486

## Attachment I 510(K) Summary American Dental Technologies PulseMaster Dental Laser Systems

This 510(K) Summary of safety and effectiveness for the American Dental Technologies Philo of of(x) Dammary of Submitted in accordance with the requirements of the SMDA 1990 and following guidance concerning the organization and content of a 510(K) summary.

| Applicant:                                                                              | American Dental Technologies, Inc.                                                                                                                                                                                                                                                                                                                                                                                        |
|-----------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address:                                                                                | 18860 West Ten Mile Road<br>Southfield, MI 48075-2657                                                                                                                                                                                                                                                                                                                                                                     |
| Contact Person:                                                                         | John Sulewski                                                                                                                                                                                                                                                                                                                                                                                                             |
| Telephone:                                                                              | (248) 395-3900 - Phone<br>(248) 395-3801 – Fax                                                                                                                                                                                                                                                                                                                                                                            |
| Preparation Date:                                                                       | 12-16-98                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Device Trade Name:                                                                      | PulseMaster Dental Laser Systems                                                                                                                                                                                                                                                                                                                                                                                          |
| Common Name:                                                                            | Nd:YAG Pulsed Laser                                                                                                                                                                                                                                                                                                                                                                                                       |
| Classification Name:                                                                    | Instrument, Surgical, Powered, Laser<br>79-GEX<br>21 CFR 878-48                                                                                                                                                                                                                                                                                                                                                           |
| Legally Marketed Predicate Device:                                                      | The Centauri Dental Laser manufactured by Premier<br>Laser Systems, #3 Morgan Street, Irvine, CA ,<br>previously cleared for market under 510(k) number<br>K932683.                                                                                                                                                                                                                                                       |
| Description of the American Dental<br>Technologies PulseMaster Dental<br>Laser Systems  | The American Dental Technologies PulseMaster Dental<br>Laser Systems are Nd:YAG lasers producing laser<br>emission at 1064 nm.  The lasers consist of two<br>interconnected sections: The cabinet which houses<br>the laser head, the power supply, the cooling system<br>and the microprocessor with control panel; and the fiber<br>optic delivery system.                                                              |
| Intended use of the American Dental<br>Technologies PulseMaster Dental<br>Laser Systems | The American Dental Technologies, Inc. PulseMaster<br>Dental Laser Systems are indicated for use for selective<br>ablation of first degree enamel caries.                                                                                                                                                                                                                                                                 |
| Nonclinical Performance Data:                                                           | Bench test data were collected under controlled<br>scientific conditions to establish the substantial<br>equivalence of the PulseMaster Dental Laser Systems<br>to a pulsed Er:YAG Dental Laser. We measured the<br>temporal pulse structures and effects on enamel caries<br>and found the two lasers to be equivalent.                                                                                                  |
| Clinical Performance Data:                                                              | An extracted tooth study was conducted to evaluate<br>the histological effects of the laser treatment.  The<br>histology indicated that the device is safe for the<br>intended use.  Clinical data from a multi-center, blinded,<br>randomized study comparing the PulseMaster to the<br>highspeed handpiece established the safety and<br>effectiveness of the PulseMaster Dental Laser Systems<br>for the Intended use. |
| Conclusion:                                                                             | The American Dental Technologies PulseMaster Dental<br>Laser Systems are substantially equivalent to another<br>existing surgical laser system in commercial distribution<br>for selective ablation of enamel caries and to<br>commercially available High Speed Dental Handpieces.                                                                                                                                       |
| Additional Information:                                                                 | None requested at this time                                                                                                                                                                                                                                                                                                                                                                                               |

{1}------------------------------------------------

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three stripes extending from its head, representing the department's mission to protect the health of all Americans and provide essential human services.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 1 3 1999

Mr. John G. Sulewski Product Manager American Dental Technologies, Inc. 18860 West Ten Mile Road Southfield. Michigan 48075-2657

Re: K984486 Trade Name: PulseMaster Dental Laser Systems Regulatory Class: II Product Code: GEX Dated: March 15, 1999 Received: March 16, 1999

Dear Mr. Sulewski:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (OS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

## Page 2 - Mr. John G. Sulewski

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Celia M. Witten, Ph.D., M.D.
Director

Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## INDICATION FOR USE STATEMENT

510(k) Number. K 984486 New Submission

Device Name: PulseMaster Dental Laser Systems

Indications for Use:

The American Dental Technologies, Inc. PulseMaster Dental Laser Systems are indicated for use for selective ablation of enamel (first degree) dental caries.

(Please do not write below this line - Continue on another page if needed)

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off)
Division of General Restorative Devices K984486
510(k) Number

Prescription Use
(per 21 CFR 801.109)

OR

Over-the-Counter Use__________________________________________________________________________________________________________________________________________________________

---

**Source:** [https://fda-staging.innolitics.com/device/K984486](https://fda-staging.innolitics.com/device/K984486)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).