The PVS BASICS ™ (paflufocon E) rigid gas permeable contact lenses are indicated for daily wear as recommended by the eye care practitioner. The PVS BASICS ™ (paflufocon E) rigid gas permeable spherical, aspheric and bifocal contact lenses are indicated for the correction of visual acuity in not-aphakic persons with non-diseased eyes who are nearsighted (myopic), farsighted (hyperopic) and who may exhibit corneal astigmatism up to 4.00 diopters or less that does not interfere with visual acuity. PVS BASICS ™ (paflufocon E) toric contact lenses are indicated to correct astigmatism of up to 6.00 diopters. PVS BASICS ™ Bifocal lenses are indicated for presbyopic persons (far or near sighted) including astigmatic corrections up to +4.00 D requiring add power of up to + 4.00 D.
Device Story
PVS BASICS are rigid gas permeable (RGP) contact lenses manufactured from paflufocon E polymer. Lenses are prescribed by eye care practitioners for daily wear to correct refractive errors including myopia, hyperopia, astigmatism, and presbyopia. The device functions as a corrective optical lens placed directly on the cornea. Clinical benefit is achieved through the physical correction of visual acuity. The device is available in spherical, aspheric, toric, and bifocal configurations, with optional violet tinting (D&C Violet #2 and D&C Red #17).
Clinical Evidence
Randomized, double-masked clinical trial comparing paflufocon E (PVS Basics) to predicate paflufocon A (FluoroPerm 92). Study enrolled 70 eyes for PVS Basics and 22 for control. Primary endpoints included visual acuity, comfort, lens-eye relationship, movement, slit lamp findings, and functional wetting. Results showed no adverse reactions, corneal ulcers, iritis, or significant staining/edema/neovascularization in either group. Biocompatibility testing included ISO 10993 ocular irritation (rabbits), cytotoxicity (L-929 mouse fibroblasts), and delayed dermal sensitization (guinea pigs), all showing no evidence of toxicity or irritation.
Technological Characteristics
Material: Paflufocon E polymer. Water content: <1%. Dk: 67. Refractive index: 1.454. Specific gravity: 1.10. Hardness: 81.7. Dimensions: 7.0-10.5 mm chord diameter, 0.05-0.70 mm center thickness. Powers: -20.00 to +12.00 Diopters. Bifocal add: +0.25 to +4.00 Diopters. Tint: Violet (D&C Violet #2, D&C Red #17).
Indications for Use
Indicated for daily wear in non-aphakic, non-diseased patients (myopic, hyperopic, or presbyopic) to correct visual acuity. Corrects corneal astigmatism up to 4.00D (spherical/aspheric/bifocal) or up to 6.00D (toric). Bifocal lenses indicated for presbyopia with add power up to +4.00D.
Regulatory Classification
Identification
A rigid gas permeable contact lens is a device intended to be worn directly against the cornea of the eye to correct vision conditions. The device is made of various materials, such as cellulose acetate butyrate, polyacrylate-silicone, or silicone elastomers, whose main polymer molecules generally do not absorb or attract water.
{0}------------------------------------------------
4/21/99
K 9 8 4 4 3 6
# APPENDIX H 510(k) Summary
Company Name: Paragon Vision Sciences Submitter. Address: 945 East Impala Ave. Phone: 602-892-7602 Fax: 602-892-3226 Registration: Owner Operator # 9024618
Manufacturer:
Company Name: Paragon Vision Sciences Address: 945 East Impala Ave. Mesa AZ 85204 Phone: 602-892-7602 Fax: 602-892-3226: Registration. Site Registration #2020433
Official Correspondent: William E. Meyers, Ph.D. % Paragon Vision Sciences Address: 945 East Impala Ave. Mesa AZ 85204 Phone: 602-507-7606 Fax: 602-892-3226 Reason for 510(k) Submission: Material change
Date of submission 12/10/98
Device Identification:
Trade Name: PVS BASICS ™ Common Name: contact lens Classification Name: rigid gas permeable contact lens for daily wear Reference: 21 CFR 886.5916;rigid gas permeable contact lens, Class II- daily wear
Indications For Use:
The PVS BASICS ™ (paflufocon E) rigid gas permeable contact lenses are indicated for daily wear as recommended by the eye care practitioner. The PVS BASICS ™ (paflufocon E) rigid gas perical, aspheric and bifocal contact lenses are indicated for the correction of visual acuity in not-aphakic persons with non-diseased eyes who are nearsighted (myopic), farsighted (hyperopic) and who may exhibit corneal astigmatism up to 4.00 diopters or less that does not interfere with visual acuity. PVS BASICS ™ (paflufocon E) toric contact lenses are indicated to correct astigmatism of up to 6.00 diopters. PVS BASICS ™ Bifocal lenses are indicated for presbyopic persons (far or near sighted) including astigmatic corrections up to +4.00 D requiring add power of up to + 4.00 D.
{1}------------------------------------------------
The lenses have the following dimensions and characteristics:
| Parameters | PVS BASICS TM |
|--------------------------------|---------------|
| Material | Paflufocon E |
| Indication | Daily wear |
| Water content | < 1% |
| Dk (35)° | 67* |
| Light transmittance Violet**** | 92% |
| Wetting Angle*** | 16 |
| Hardness | 81.7 |
| refractive index | 1.454 |
| Specific gravity | 1.10 |
| color | violet |
*Oxygen Permeability 67 x10-11Dk** at 35'C
** (cm2/sec)(mL 02/mL x mm Hg) Method of Irving Fatt, Ph.D.
*** after soaking in conditioning solution
*** +12 D, 7mm OZ, 7.8 BC, .45mm CT, harmonic mean thickness over 7mm = 0.329 mm.
#### Lens Parameters:
| Chord Diameter | 7.0 to 10.5 mm |
|------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Center Thickness | 0.05 to 0.70 mm |
| Base Curve | 6.50 to 9.00 mm |
| Powers | 20.00 to +12.00 Diopters |
| | 11:0 -- 1 4 1 1 D ------------------------------------------------------------------------------------------------------------------------------------------------------------ |
Bifocal Add Powers ... +0.25to +4.00 Diopters Concentric Bifocal Add Diameter ................................................................................................................................................................. Monocentric Bifocal Add Diameter ................................................................................................................................................................. Monocentric Bifocal Prism ..................... 1.0 to 2.5 Diopters
The PVS BASICS 714 (paflufocon E) rigid gas permeable contact lenses are available in violet color. The violet tinted lenses contain D & C Violet # 2 and D & C Red #17.
The PVS BASICS 714 parturocon E) Dk 65 contact lens is substantially equivalent to the FluoroPerin 92 Dk 92 Rigid Gas Permeable Contact Lens marketed by Paragon Vision Sciences which is presently approved for daily wear under PMA (P820063). The physical, optical and chemical properties of the PVS BASICS ™ (paffufocon E) contact lens are substantially equivalent to the FluoroPerm 92 (paflufocon A).
In addition to the physicochemical comparison, toxicity and human clinical studies were conducted.
The test article, PVS Basics (AKA patlufocon E), Batch 234880-6, was extracted in 0.9% sodium chloride USP solution (SC) and cottonseed oil, NF (CSO). These extracts were evaluated for ocular irritation base on the International Organization for Standardization 10993: Biological Evaluation of Medical Devices, Part 10: Tests for Irritation and Sensitization.
A 0.2 ml dose of the appropriate test article extract was instilled in the right eye of three test rabbits. Similarly, the corresponding reagent control was instilled into the left eye of each rabbit to serve as the control conditions were evaluated at 1, 24, 48, and 72 hours after the single exposure.
Under the conditions of this study, there was no evidence of irritation in the test eye or any rabbit.
{2}------------------------------------------------
The SC and CSO test article extracts would not be considered irritants to the ocular tissue of the rabbit.
An in vitro biocompatibility test, based on the International Organization (ISO 10993-5) guidelines, was conducted on the test article. PVS Basics (AKA patlufocon E). Lot HS401710-8, in order to determine the potential for in vitro cytotoxicity. A single lens was placed on triplicate agarose surfaces directly overlaying confluent monolayers of L-929 mouse fibroblast cells. Similarly, triplicate negative and positive control wells were presared. After incubating at 37°C in 5% or for 24-26 hours, the cell cultures were examined macroscopically for cell decolorization around the test article and controls to determine the zone of cell lysis (if any). The cultures were then examined microscopically (100X) to verify any decolorized zones and to determine cell morphology in proximity to and beneath the test article.
The negative controls and the positive controls performed. Under the conditions of this study, the test article showed no evidence of causing cell lysis or toxicity. The test article was not cytotoxic and passed this ISO study.
A guinea pig maximization test of PVS Basics (AKA paffuroon E), Batch 234880-6, was conducted to evaluate the potential for delaved dermal contact sensitization. This study was conducted based on the international Organization for Standardization 10993; Biological Evaluation of Medical Devices, Part 10: Tests for Irritation and Sensitization.
The test article was extracted in 0.9% sodium chloride USP (SC) and contonsed oil. NF (CSO). Each extract was intradermally injected and occlusively patched to test guinea pies (per extract) in an attempt to induce sensitization. The vehicle was similarly injected and occlusively patched to five control guinea pigs (per vehicle). Following a recovery period, the test and control animals received a challenge patch of the appopriate test and the reagent control. All sites were scored at 24, 48 and 72 hours after patch removal.
Under the conditions of this study, the SC and CSO test article extracts showed no evidence of causing delayed demall contact sensitization in the guinea pig.
A clinical evaluation of the safety and effectiveness of the paflufocon E (PVS Basic™) polymer lens was reefformed. This randomized double-masked controlled trial evaluated the clinical performance of lenses produced using the new pathutocon E material in comparison to control lenses made with FDA approved prodicate material, paflufoon A (FluoroPerm 92TN).
Decision criteria were established for the equivalence of the materials these variables were: Visual Acuity, Comfort, Lens-Eye Relationship, Movement, Slit lamp Findings, Functional Wetting,
This clinical trial established Patlufocon E as safe and effective when compared to the currently marketed predicate control material, Paflufocon A (FluoroPerm 92™). The tables below summarize the results of the study.
{3}------------------------------------------------
#### EYES ENROLLED IN THE STUDY AND DISTRIBUTION AND STATUS
| Status | | Number of Eyes | |
|------------------------|-----------------|------------------------|----------------------|
| | | Paflufocon A<br>TOTALS | PVS Basics<br>TOTALS |
| Enrolled Dispensed | | | |
| Completed | | 22 | 68 |
| Active | | | |
| | Dispensing | 22 | 70 |
| | First Follow Up | 22 | 68 |
| | Final Visit | 22 | 68 |
| Total Active | | 22 | 68 |
| Discontinued | | 0 | 2 |
| Incomplete | | 0 | 0 |
| Total Dispensed | | 22 | 70 |
| Enrolled Not Dispensed | | 0 | 0 |
| Total Enrolled | | 22 | 70 |
## SUMMARY OF FINDINGS RELATED TO SAFETY
| | Paflufocon A<br>TOTALS | PVS Basics<br>TOTALS | |
|----------------------------|------------------------|----------------------|-----|
| NUMBER OF EYES | 22 | 68 | |
| DISCONTINUED EYES | 0 | 2 | |
| AVERAGE WEARING TIME | $14.09$ | $13.58$ | TMC |
| ALL ADVERSE REACTIONS | 0 | 0 | |
| ALL CORNEAL ULCERS | 0 | 0 | |
| ALL IRITIS | 0 | 0 | |
| STAINING REPORTS > GRADE 2 | 0 | 0 | |
| EDEMA REPORTS >GRADE 2 | 0 | 0 | |
| INJECTION REPORTS >GRADE 2 | 0 | 0 | |
| NEOVASC. REPORTS >GRADE 2 | 0 | 0 | |
| VISITS | 44 | 136 | |
| MISSED VISITS | 0 | 0 | |
{4}------------------------------------------------
#### SUMMARY OF SYMPTOMS PROBLEMS AND COMPLAINTS
#### Completed Control Eyes
| | Initial Dispensing<br>Visit | | Intermediate Visits | | | |
|------------------------|-----------------------------|---------|---------------------|---------|--------|---------|
| | 1 | | 2 | | | |
| Total Eyes at Visit | 22 | | 22 | | 22 | |
| | Number | Percent | Number | Percent | Number | Percent |
| None | 19 | 86.4 | 16 | 72.7 | 14 | 63.6 |
| Discomfort | 3 | 13.6 | 4 | 18.2 | 6 | 27.3 |
| Excess Tearing * | 0 | 0 | 0 | 0 | 0 | 0 |
| Photophobia | 0 | 0 | 0 | 0 | 0 | 0 |
| Halos | 0 | 0 | 0 | 0 | 0 | 0 |
| Itching/Burning | 0 | 0 | 1 | 4.5 | 4 | 18.2 |
| Spectacle Blur* | 0 | 0 | 0 | 0 | 0 | 0 |
| Variable Vision | 0 | 0 | 0 | 0 | 4 | 18.2 |
| Blurred Vision | 2 | 9.1 | 3 | 13.6 | 8 | 36.4 |
| Lens Needs Cleaning | 2 | 9.1 | 3 | 13.6 | 2 | 9.1 |
| Other | | | | | | |
| Dryness | 2 | 9.1 | 3 | 13.6 | 6 | 27.3 |
| Total Positive Reports | 9 | | 14 | | 24 | |
### Completed Trial Eyes
| | Initial Dispensing<br>Visit | | Intermediate Visits | | | |
|------------------------|-----------------------------|---------|---------------------|---------|--------|---------|
| | | | 1 | | 2 | |
| Total Eyes at Visit | 70 | | 68 | | 68 | |
| | Number | Percent | Number | Percent | Number | Percent |
| None | 55 | 78.6 | 52 | 76.5 | 45 | 66.2 |
| Discomfort | 6 | 8.6 | 4 | 5.9 | 8 | 11.8 |
| Excess Tearing | 0 | 0 | 0 | 0 | 0 | 0 |
| Photophobia | 0 | 0 | 0 | 0 | 0 | 0 |
| Halos | 2 | 2.9 | 0 | 0 | 2 | 2.9 |
| Itching/Burning | 1 | 1.4 | 2 | 2.9 | 2 | 2.9 |
| Spectacle Blur | 0 | 0 | 0 | 0 | 0 | 0 |
| Variable Vision | 0 | 0 | 6 | 8.8 | 5 | 7.4 |
| Blurred Vision | 4 | 5.7 | 4 | 5.9 | 2 | 2.9 |
| Lens Needs Cleaning | 0 | 0 | 6 | 8.8 | 4 | 5.9 |
| Other | Dryness | | | | | |
| Dryness | 5 | 7.1 | 5 | 7.4 | 6 | 8.8 |
| Total Positive Reports | 18 | | 27 | | 29 | |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 2 1 1993
William E. Meyers, Ph.D. Vice President, Science & Technology PARAGON VISION SCIENCES 947 East Impala Mesa, AZ 85204
Re: K984436 Trade Name: PVS BASICS ™ (paflufocon E) Rigid Gas Permeable Contact Lenses For Daily Wear (Spherical, Aspheric, Toric and Bifocal) Clear and Violet Tinted (with D&C Violet # 2 and D & C Red # 17) Regulatory Class: II Product Code: 86 HQD Dated: March 2, 1999 Received: March 3, 1999
Dear Dr. Meyers:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect
{6}------------------------------------------------
Page 2 - William E. Meyers, Ph.D.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
A Ralph Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### Indications Statement
510(k) Number (if known):
Device Name: PVS BASICS TM (paflufocon E)Contact Lens
Indications For Use:
The PVS BASICS "M (paflufocon E) rigid gas permeable contact lenses are indicated for daily wear as recommended by the eye care practitioner. The PVS BASICS ™ (paflufocon E) rigid gas permeable spherical, aspheric and bifocal contact lenses are indicated for the correction of visual acuity in not-aphakic persons with non-diseased eyes who are nearsighted (myopic), farsighted (hyperopic) and who may exhibit corneal astigmatism up to 4.00 diopters or less that does not interfere with visual acuity. PVS BASICS ™ (paflufocon E) toric contact lenses are indicated to correct astigmatism of up to 6.00 diopters. PVS BASICS ™ Bifocal lenses are indicated for presbyopic persons (far or near sighted) including astigmatic corrections up to +4.00 D requiring add power of up to + 4.00 D.
> PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use OR
Over-The Counter Use
(Optional Format 1-2-96)
Erica
(Division Sign-Off)
Division of Ophthalmic Devices
510(k) Number K 984436
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.