MEDTRONIC KHONSARI ANNULOPLASTY BAND, MODEL H609M

K984435 · Medtronic Vascular · KRH · Jan 7, 1999 · Cardiovascular

Device Facts

Record IDK984435
Device NameMEDTRONIC KHONSARI ANNULOPLASTY BAND, MODEL H609M
ApplicantMedtronic Vascular
Product CodeKRH · Cardiovascular
Decision DateJan 7, 1999
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.3800
Device ClassClass 2
AttributesTherapeutic

Indications for Use

For use in those patients undergoing surgery of diseased or damaged mitral valves in whom the surgeon determines that the valve can be preserved by employing the appropriate surgical repair. The annuloplasty bands provide support for and restrict dilatation of the posterior mitral valve annulus.

Device Story

Implantable, flexible, adjustable annuloplasty band (Model H609M) for mitral valve repair; reduces and stabilizes mitral valve annulus. Device consists of braided polyester body with four adjustable drawstrings allowing independent shortening of annular segments. Features flexible radiopaque marker for visualization and eight surgeon's flags for suture placement. Supplied on a holder with a malleable handle stem to facilitate seating; drawstrings release automatically upon holder removal. Used by cardiac surgeons in an operating room setting. Provides structural support to restrict posterior mitral valve annulus dilatation, facilitating valve repair and preservation.

Clinical Evidence

Bench testing only. Verification and validation studies included manufacturing process validation, physical performance testing to verify functionality post-sterilization and accelerated aging, and sterilization validation (SAL 10^-6 at 25kGy). No clinical data provided.

Technological Characteristics

Implantable, flexible, adjustable band. Materials: braided polyester fabric, non-metallic radiopaque marker. Configuration: open band (no stiffener). Connectivity: none. Sterilization: validated at 25kGy minimum dose. No software or electronic components.

Indications for Use

Indicated for patients undergoing surgical repair of diseased or damaged mitral valves where valve preservation is determined appropriate by the surgeon.

Regulatory Classification

Identification

An annuloplasty ring is a rigid or flexible ring implanted around the mitral or tricuspid heart valve for reconstructive treatment of valvular insufficiency.

Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Annuloplasty Rings 510(k) Submissions.”

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ KG84435 # SUMMARY OF SAFETY AND EFFECTIVENESS ## MEDTRONIC KHONSARI™ ANNULOPLASTY BAND This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA and CFR 807.92. #### SUBMITTER INFORMATION I | Company Name: | Medtronic Heart Valves (Medtronic) | |--------------------------------------|---------------------------------------------------| | Company Address: | 7000 Central Avenue N.E.<br>Minneapolis, MN 55432 | | Company Phone:<br>Company Facsimile: | (612) 514-2700<br>(612) 514-2890 | | Contact Person: | Julie Sherman<br>Product Regulation Manager | | Date Summary Prepared: | December 10, 1998 | #### II DEVICE IDENTIFICATION | Trade/Proprietary Name: | Medtronic KHONSARI™ Annuloplasty Band | |-------------------------|---------------------------------------| | | [Model H609M] | | 21 CFR Reference: | 870.3800 | | 21 CFR Common Name: | Ring, Annuloplasty | | Class: | Pre-amendment Class III | | Panel: | CV (74) KRH | #### III IDENTIFICATION OF PREDICATE DEVICE | Device | 510(k) # | |-------------------------------------------------------|----------| | Medtronic SCULPTOR® Annuloplasty Ring<br>[Model 605M] | K905175 | ## The 510(k) also includes references to the following marketed devices: | Medtronic Posterior Annuloplasty Ring [Model 607] | K960356 | |----------------------------------------------------|---------| | Cosgrove-Edwards™ Annuloplasty System [Model 4600] | K923367 | | Medtronic DURAN Annuloplasty Ring [Model H608H] | K980534 | {1}------------------------------------------------ #### IV DEVICE DESCRIPTION The Medtronic KHONSARI Annuloplasty Band [Model H609M] is an implantable, flexible, adjustable band. The band reduces and stabilizes the mitral valve annulus in patients undergoing mitral valve repair. The body of the band is made of braided polyester and has drawstrings that permit segments of the annulus to be independently shortened. The band contains a flexible (non-metallic) radiopaque marker that enables radiographic visualization of its entire length. Eight radially spaced surgeon's flags indicate recommended locations for implantation sutures. Each annuloplasty band is attached to a holder. A circumferential groove in the holder protects the drawstrings from entrapment by the implanting sutures. The stem of the handle is malleable to facilitate seating of the band during placement. The four adjustable drawstrings are coiled within cylindrical canisters on the holder. A retaining suture secures each end of the band to the holder. The drawstrings will automatically release from the holder as it is removed from the band. #### V DESCRIPTION OF INTENDED USE The Medtronic KHONSARI Annuloplasty Band is for use in those patients undergoing surgery of diseased or damaged mitral valves in whom the surgeon determines that the valve can be preserved by employing the appropriate surgical repair. The annuloplasty bands provide support for and restrict dilatation of the posterior mitral valve annulus. #### VI SUBSTANTIAL EQUIVALENCE The Medtronic KHONSARI Annuloplasty Band [Model H609M] is substantially equivalent to the Medtronic SCULPTOR Annuloplasty Ring [Model 605M]. The KHONSARI Annuloplasty Band is manufactured based on modifications to the manufacturing processes of the SCULPTOR Annuloplasty Band. Both devices contain equivalent raw materials and are produced using equivalent manufacturing, packaging and sterilization processes. Both devices are indicated for surgical repair of mitral valves. The KHONSARI Annuloplasty Band is also equivalent to the Posterior Annuloplasty Band [Model 607] manufactured by Medtronic and the Cosgrove-Edwards™ Annuloplasty System [Model 4600] by virtue of the flexible, open band configuration (no stiffener) of all three devices. {2}------------------------------------------------ #### TECHNOLOGICAL CHARACTERISTICS VII Both the KHONSARI Annuloplasty Band and the SCULPTOR Annuloplasty Ring are manufactured using the same white polyester braided fabric as is used in the predicate device. The radiopaque marker, identification tag and suture used for sewing of the polyester fabric and surgeon markers remains the same. The drawstrings are made from the same raw material. Both products are manufactured using equivalent manufacturing, packaging and sterilization processes. The KHONSARI Annuloplasty Band is manufactured as an open configuration with a completely flexible band, whereas the SCULPTOR Annuloplasty Ring is manufactured as a complete flexible ring with a a rigid anterior portion. #### PERFORMANCE DATA VIII The KHONSARI Annuloplasty Band was subjected to verification and validation studies to demonstrate that the modifications to the predicated device are appropriate and do not affect the intended use or performance of the device. Manufacturing process validation was performed. The modified manufacturing processes meet all procedure requirements. Physical performance studies were conducted to verify that the device performs as intended after routine sterilization and accelerated aging cycles. Sterilization validation of the device was also completed. Based on the results of the study, the sterility assurance level (SAL) of the sterlization process was qualified at 10 ° sterilization level at a minimum sterilization dose of 25kGy. The verification/validation studies demonstrate that the modifications to the manufacturing process are appropriate and do not affect the intended use of the product. No changes have been made to the manufacturing or sterilization of this device to warrant new or additional biocompatibility testing of the device components. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle or bird-like figure with its wings spread. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JAN 19 1999 Ms. Julie Sherman Product Requlation Manager Medtronic Cardiac Surgery 7000 Central Avenue, NE Minneapolis, MN 55432-3576 Re : K984435 Medtronic Khonsari™ Annuloplasty Band [Model H609M] Regulatory Class: III (Three) Product Code: KRH Dated: December 10, 1998 Received: December 11, 1998 Dear Ms. Sherman: This letter corrects our substantially equivalent letter of January 7, 1999, regarding the incorrect trade name and model number . We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent [(for the indications for use stated in the enclosure)] to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of The general controls provisions of the Act include the Act. requirements for annual reqistration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbrandinq and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good {4}------------------------------------------------ Page 2 - Ms. Julie Sherman Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General (QS) regulation (21 CFR Part 820) and that, through periodic QS inspections, FDA will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, the Food and Drug Administration (FDA) may publish further announcements concerning your device in the Federal Please note: this response to your premarket Register. notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to continue marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at their toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Thomas J. Callahan Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## Page 1 of 1 510(k) Number (if known): K98 Device Name: Medtronic KHONSARI ™ Annuloplasty Band [Model H609M] ## Indications for Use: For use in those patients undergoing surgery of diseased or damaged mitral valves in whom the surgeon determines that the valve can be preserved by employing the appropriate surgical repair. The annuloplasty bands provide support for and restrict dilatation of the posterior mitral valve annulus. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use V Per 21 CFR 801.109 OR Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ Bette K Campbell (Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices X 98.44 510(k) Number_
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...