← Product Code [KGO](/productcode/KGO) · K984395

# SURGEON'S GLOVES (K984395)

_Sri Anusham Rubber Industries Pvt., Ltd. · KGO · May 21, 1999 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K984395

## Device Facts

- **Applicant:** Sri Anusham Rubber Industries Pvt., Ltd.
- **Product Code:** [KGO](/productcode/KGO.md)
- **Decision Date:** May 21, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4460
- **Device Class:** Class 1
- **Review Panel:** General, Plastic Surgery

## Indications for Use

Surgeon's glove is a medical device intended to be worn by operating room personel to protect a surgical wound from contamination.

## Device Story

Surgeon's gloves manufactured from natural rubber latex; intended for use by operating room personnel to provide a barrier against contamination of surgical wounds. Device is a pre-powdered, single-use sterile glove. Performance relies on physical barrier properties, including tensile strength, elongation, and freedom from holes. Used in clinical surgical settings to protect both patient and clinician. Effectiveness is verified through bench testing against ASTM D3577-95 standards for dimensions, physical properties, and sterility.

## Clinical Evidence

Bench testing only. No clinical data provided. Testing included physical property evaluation (tensile strength, elongation at break before/after aging), dimension measurements, powder content, protein content, moisture content, and biocompatibility assessment. All results met or exceeded ASTM D3577-95 specifications.

## Technological Characteristics

Natural rubber latex surgeon's gloves. Dimensions, tensile strength, and elongation comply with ASTM D3577-95. Thickness: 0.12mm (cuff), 0.16mm (palm), 0.19mm (finger). Sterilization per IP/USP standards. Biocompatible material.

## Regulatory Identification

A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

## Predicate Devices

- Class 1- Surgeon's Gloves (Pre-Powdered) 79KGO (meets ASTM D 3577-91)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

|  | SRI ANUSHAM RUBBER INDUSTRIES<br>PRIVATE LIMITED<br>100% EXPORT ORIENTED UNIT | TEL: 91 4652 33091<br>FAX: 91 4652 32871<br>TLX: 436 209 SIVA IN | Pioneer Manikandan Buildings<br>Vadasery, Nagercoil-629001.<br>Tamil Nadu, S India. | 6 |
|--|-------------------------------------------------------------------------------|------------------------------------------------------------------|-------------------------------------------------------------------------------------|---|
|--|-------------------------------------------------------------------------------|------------------------------------------------------------------|-------------------------------------------------------------------------------------|---|

MAY 21 1999

K984395

| 1.0               | 510K SUMMARY as required by : 807.92(c)                                                                                                  |
|-------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| 2.0               | APPLICANT.                                                                                                                               |
| NAME              | : SRI ANUSHAM RUBBER INDUSTRIES (P) LTD<br>PIONEER MANIKANDAN BUILDINGS<br>VADASERRY, NAGERCOIL-629001<br>TAMIL NADU , S.INDIA           |
| PHONE             | : + 91 4652 33091                                                                                                                        |
| FAX               | : + 91 4652 32871                                                                                                                        |
| CONTACT PERSON    | : MR. N.PARAMASIVAN<br>MANAGING DIRECTOR                                                                                                 |
| DEVICE TRADE NAME | : NIL                                                                                                                                    |
| 3.                | COMMON NAME : Surgeon's Glove                                                                                                            |
| 4.                | LEGALLY MARKETED DEVICE TO WHICH<br>COMPANAY CLAIMING EQUIVALENCE :                                                                      |
|                   | Class 1- Surgeon's Gloves (Pre-Powdered) 79KGO<br>that meets all the requirements of<br>ASTM D 3577 - 91.                                |
|                   | Type 1 - Gloves compounded primarily from Natural Rubber<br>Latex.                                                                       |
| 5.                | DESCRIPTION OF THE DEVICE:                                                                                                               |
|                   | Class 1- Surgeon's Gloves (Pre-Powdered) 79KGO<br>that meets all the requirements of<br>ASTM D 3577 - 91.                                |
| 6                 | INTENDED USE OF THE DEVICE:                                                                                                              |
|                   | Surgeon's glove is a medical device intended to be worn by<br>operating room personel to protect a surgical wound from<br>contamination. |

こ。
 『

{1}------------------------------------------------

#### TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO PREDICATE DEVICE. om and end and and ear and ein him me min me me me me une me and min me une me uit me was was was was was was was was was was was was was was was was was was was was was was

| Measured Parameters of Latex<br>Surgeon's Gloves Manufactured<br>by SRI ANUSHAM RUBBER INDUS-<br>TRIES PRIVATE LIMITED |      |           | ASTM D 3578<br>Requirement for<br>Latex Surgeon's<br>Glove |
|------------------------------------------------------------------------------------------------------------------------|------|-----------|------------------------------------------------------------|
| Characteristics                                                                                                        | Size | Value     | Prescribed                                                 |
| 1. Length                                                                                                              | 5.5  | 270-272mm | 245 mm minimum                                             |
|                                                                                                                        | 6    | 270-272mm | 265 mm minimum                                             |
|                                                                                                                        | 6.5  | 270-272mm | 265 mm minimum                                             |
|                                                                                                                        | 7    | 270-272mm | 265 mm minimum                                             |
|                                                                                                                        | 7.5  | 270-272mm | 265 mm minimum                                             |
|                                                                                                                        | 8    | 270-272mm | 265 mm minimum                                             |
|                                                                                                                        | 8.5  | 270-272mm | 265 mm minimum                                             |
|                                                                                                                        | 9    | 270-272mm | 265 mm minimum                                             |
| 2. Width                                                                                                               | 5.5  | 68mm      | 70 +/- 6 mm                                                |
|                                                                                                                        | 6    | 73mm      | 76 +/- 6 mm                                                |
|                                                                                                                        | 6.5  | 79mm      | 83 +/- 6 mm                                                |
|                                                                                                                        | 7    | 87mm      | 89 +/- 6 mm                                                |
|                                                                                                                        | 7.5  | 92mm      | 95 +/- 6 mm                                                |
|                                                                                                                        | 8    | 103mm     | 105 +/- 6 mm                                               |
|                                                                                                                        | 8.5  | 108mm     | 108 +/- 6 mm                                               |
|                                                                                                                        | 9    | 112mm     | 114 +/- 6 mm                                               |

- ri Thickness at Cuff,Palm and Finger. tip of all the sizes is 0.12mm,0.16mm and 0.19mm respectively. As per ASTM D3577-95 the minimum thickness specfied at cough,palm and finger is 0.1mm.
7 .. 0

.

62

{2}------------------------------------------------

# PHYSICAL PROPERTIES

|                               | BEFORE AGEING |                         | AFTER AGEING |                         |
|-------------------------------|---------------|-------------------------|--------------|-------------------------|
| CHARACTERISTICS               | *SARI VALUE   | ASTM D 3577 REQUIREMENT | SARI VALUE   | ASTM D 3577 REQUIREMENT |
| Tensile Strength              | 31 mpa        | 24 mpa min.             | 22mpa        | 18 mpa min.             |
| Elongation at<br>Break %      | 950-975%      | 750% min.               | 850-900%     | 560% min.               |
| Modulus at 500%<br>Elongation | 3.4 mpa       | 5.5 mpa<br>(max)        | -            | -                       |

Carrier

*SARI - SRI ANUSHAM RUBBER INDUSTRIES FRIVATE LIMITED

PERFORMANCE REQUIREMENT :
----------------------------------------------------------------------------------------------------------------------------------------------------

| Characteristics    | Related Defects                                                      | Level followed By |                   | AQL followed by SARI | AQL Value as per ASTM D3577 |
|--------------------|----------------------------------------------------------------------|-------------------|-------------------|----------------------|-----------------------------|
|                    |                                                                      | SARI              | As per ASTM D3577 |                      |                             |
| Sterility          | Fails Sterility                                                      | As per IP         | As per USP        | NA                   | NA                          |
| Freedom from Holes | Holes                                                                | S4                |                   | 1.5                  | 1.5                         |
| Dimensions         | Width<br>Length<br>Thickness                                         | S2                | S4                | 4                    | 4                           |
| Physical Property  | Tensile Strength<br>Elongation at Break<br>before & after<br>ageing. | S2                | S2                | 4                    | 4                           |

63

් පිහිටි පිහිටා හා දෙකින් පිහිට

{3}------------------------------------------------

# POWDER CONTENT :

100000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000000

|               | FOR ANY FOR CONTENT OF THE FOR CONSTITUTION OF THE FOR COLLECTION OF CREATING OF CREATING OF CREATING OF CREAT OF CREAT OF CREAT OF CREAT OF CREAT OF CREAT OF CREAT OF CREASE      |
|---------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| SARI VALUE    | ASTM REQUIREMEN                                                                                                                                                                     |
| 120 +/-<br>20 | 144 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 100 10<br>. |

# PROTEIN CONTENT:

| SARI VALUE                                                                                                                                                                     | NE LEA LEA LES PRODUCE CALL CAR CAR CAR CAR CAR CAR CAR CAR CAR COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION CARE CREATING CARE CREATING CARA CON CREAT C                    |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| The for the first for the part and one for the first and the first and the mean and me me more and can and me me more see many ment me more see more see many ment me more see | FDA REQUIREMENT                                                                                                                                                                                   |
| 80+/-20ppm                                                                                                                                                                     | Value not fixed<br>(2) [2] [2] [2] [2] [2] [2] [2] [2] [2] [2] [2] [2] [2] [2] [2] [2] [2] [2] [2] [2] [2] [2] [2] [2] [2] [2] [2] [2] [2] [2] [2] [2] [2] [2] [2] [2] [2] [2] [2] [2] [2] [2] [2 |

### MOISTURE CONTENT :

| SARI VALUE | suit free new and the first least and the new see and the new seen price for the many and for the first for the first for the first for the first for the first for the first<br>FDA REQUIREMENT<br>" " the new reas not the case and not new see and may not me and me not me see and may not see and may not the may not not not not not not not not not not not not not not not |
|------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 0.8%       | Value not fixed<br>E MICH EN ENTERES OF FEE SERVET FOR COLLECTION CONTRIVER OF CHE THE THE THE THE FORE THE FINE FOR THE FINE FOR THE FINE FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FINE FOR THE F                                                                                                                                                                                  |

## BIO-COMPATABILITY

| SARI GLOVE                                                                                                                                                                     | FIFT THE CHIE PER FINE FREE FOR THE FOR THE FOR THE FOR THE FOR THE FOR THE FOR THE FOR THE FOR THE FOR THE FOR THE FOR THE FOR THE FOR THE FOR THE FOR THE FOR THE FOR THE FO<br>FDA REQUIREMENT |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| The me nee and hear for any for the long may been and on the new the new see and one and one and one and one and one meet the will be fill and the look has not one the line t | Biologically Compatible                                                                                                                                                                           |
| Biologically Compatible                                                                                                                                                        | THE CHILDER OF THE COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION COLLECTION C                    |

t

{4}------------------------------------------------

#### 日 . മ PERFORMANCE DATA

:

:

:

 : 上 The Performance Test Data of the Surgeon's Gloves manufactured by SRI ANUSHAM RUBBER INDUSTRIES PVT. LIMITED is given here-under:

| Measured Parameters of Latex Surgon's Gloves<br>manufactured by SRI ANUSHAM RUBBER INDUSTRIES<br>PRIVATE LIMITED |      |              |
|------------------------------------------------------------------------------------------------------------------|------|--------------|
| Characteristics                                                                                                  | Size | Value        |
| 1. Length                                                                                                        | 5.5  | 270 mm       |
|                                                                                                                  | 6    | 270 - 272 mm |
|                                                                                                                  | 6.5  | 270 - 272 mm |
|                                                                                                                  | 7    | 270 - 272 mm |
|                                                                                                                  | 7.5  | 270 - 272 mm |
|                                                                                                                  | 8    | 270 - 272 mm |
|                                                                                                                  | 8.5  | 270 - 272 mm |
|                                                                                                                  | 9    | 270 - 272 mm |
| 2. Width                                                                                                         | 5.5  | 68mm         |
|                                                                                                                  | 6    | 73mm         |
|                                                                                                                  | 6.5  | 79mm         |
|                                                                                                                  | 7    | 87mm         |
|                                                                                                                  | 7.5  | 92mm         |
|                                                                                                                  | 8    | 103mm        |
|                                                                                                                  | 8.5  | 106mm        |
|                                                                                                                  | 9    | 112mm        |

ાં છ Thickness at cough ,palm and finger of all the sizes is 0.12, 0.16 and 0.19mm.

-

65

{5}------------------------------------------------

#### FHYSICAL PROPERTIES

| CHARACTERISTICS       | Before ageing | After Ageing |
|-----------------------|---------------|--------------|
| Tensile Strength      | 29 mpa        | 22 mpa       |
| Elongation at break % | 850%          | 800%         |

#### INSPECTION LEVEL OF AQL

| CHARACTERISTICS     | RELATED DEFECTS                                                 | LEVEL        | AQL |
|---------------------|-----------------------------------------------------------------|--------------|-----|
| Sterility           | Fails Sterility                                                 | As per<br>IP |     |
| Freedom from holes  | Holes                                                           | S4           | 1.5 |
| Dimensions          | Length,Width &<br>Thickness                                     | S2           | 4   |
| Physical Properties | Tensile Strength<br>Elongation at Break<br>before & after aging | S2           | 4   |

| POWDER CONTENT    | 120 +/- 20mg per glove  |
|-------------------|-------------------------|
| PROTEIN CONTENT   | 80 +/- 20 ppm           |
| MOISTURE CONTENT  | 0.7% maximum            |
| BIO-COMPATABILITY | Biologically Compatible |
| 9. CLINICAL DATA  | Not Applicable          |

CONCLUSION OF PERFORMANCE OF TEST DATA:

The Surgeon's Gloves manufactured by SRI ANUSHAM RUBBER INDUSTRIES PRIVATE LIMITED;

* Meet or exceed the ASTM D3577-95 Specifications

* Meet FDA Pin Hole Requirements

* Meet Labelling Claim as shown in data under S.No.6

ANY OTHER INFORMATION:

Any other information required by FDA in respect of PRODUCT SAFETY AND EFFECTIVENESS shall be provided on demand.

66

{6}------------------------------------------------

Image /page/6/Picture/2 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, consisting of a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 21 1999

SRI Anusham Rubber Industries Pvt. Ltd. C/O Ms. Latha Kumaraswamy Authorised/Designated Agent U.S.A. P.O. Box 5206 Pleasanton, California 94566 U.S.A.

K984395 Re : Surgeon's Gloves Trade Name: Requlatory Class: I Product Code: KGO March 17, 1999 Dated: Received: March 19, 1999

Dear Ms. Kumaraswamy:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existinq major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) requlation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{7}------------------------------------------------

### Page 2 - Ms. Kumaraswamy

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note ene regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Suser Pungs

Timothy A. Ulatowski Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

510(k) Number (if known): Device Name: 2800 Indications For Use:

| INDICATIONS FOR USE                                                                                                                                |                                         |
|----------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------|
| APPLICANT                                                                                                                                          | : SRI ANUSHAM RUBBER INDUSTRIES PVT.LTD |
| 510(K) NO                                                                                                                                          | :                                       |
| DEVICE NAME                                                                                                                                        | : SURGEON'S GLOVES                      |
| INDICATIONS FOR USE:                                                                                                                               |                                         |
| Surgeon's Glove is a medical device intended to be worn by operating room personnel during surgery to protect a surgical wound from contamination. |                                         |

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Olin S. him

(Division and I 510(k) Nu

Prescription Use (Per 21 CFR 801.109)

OR

Over-The-Counter Use X

(Optional Format 1-2-96)

---

**Source:** [https://fda-staging.innolitics.com/device/K984395](https://fda-staging.innolitics.com/device/K984395)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).