← Product Code [MNI](/productcode/MNI) · K984348

# TIMX LOW BACK SYSTEM (K984348)

_Depuy Motech Acromed · MNI · Jan 20, 1999 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K984348

## Device Facts

- **Applicant:** Depuy Motech Acromed
- **Product Code:** [MNI](/productcode/MNI.md)
- **Decision Date:** Jan 20, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3070
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The TiMX Low Back System when used with pedicle screws, in skeletally mature patients, is indicated for degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). Levels of fixation are for the thoracic, lumbar and sacral spine. The TiMX Low Back System is also indicated for pedicle screw fixation for severe spondylolisthesis (Grades 3 and 4) at L5-S1, in skeletally mature patients, when autogenous bone graft is used, when affixed to the posterior lumbosacral spine, and intended to be removed after solid fusion is attained. Levels of fixation are from L3-S1. The TiMX Low Back System, when not used with pedicle screws, is intended for posterior hook, wire, and/or sacral/iliac screw fixation from T1 to the ilium/sacrum. The non-pedicle screw indications are spondylolisthesis, degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), deformities (scoliosis, lordosis and kyphosis), tumor, fracture and previous failed fusion surgery.

## Device Story

TiMX Low Back System is a spinal fixation implant system; components include pedicle screws, hooks, and wires. Used by surgeons in clinical settings to stabilize the thoracic, lumbar, and sacral spine. Provides mechanical support for spinal fusion procedures; treats conditions like spondylolisthesis, scoliosis, and fractures. Device design remains unchanged; submission serves to add pedicle screw indications following a reclassification order.

## Clinical Evidence

No clinical data provided; substantial equivalence based on design and material consistency.

## Technological Characteristics

Manufactured from ASTM F-136 titanium alloy. System includes pedicle screws, hooks, wires, and sacral/iliac screws. Designed for posterior spinal fixation.

## Regulatory Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

## Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
 “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

## Reference Devices

- VSP System ([K984350](/device/K984350.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Jan. 11. 1999 6:01PM No. 3217 P. 7/13

K984348 1 of 2

こ ---------------------------------------------------------------------------------------------------------------------------------------------------------------------------

## ISOLA SPINAL SYSTEM Pedicle Screw Indications 510(k) SUMMARY

COMPANY:

DePuy AcroMed, Inc. 3303 Carnegie Avenue Cleveland, OH 44115

TRADENAME: TiMX Low Back System

Labeled for pedicle screw use: Class II CLASSIFICATION:

DESCRIPTION:

MATERIAL:

!!

INDICATIONS:

ri

All implant components are manufactured of ASTM F-136 titanium alloy. ·

The primary purpose of this premarket notification is to add

indications to the marketing clearance for the pedicle screws which

may be used as a spinal anchor in the TiMX Low Back System.

The TiMX Low Back System when used with pedicle screws, in_ is indicated for degenerative skeletally mature patients, objective evidence of neurologic spondylolisthesis with impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). Levels of fixation are for the thoracic, lumbar and sacral spine.

The TiMX Low Back System is also indicated for pedicle screw fixation for severe spondylolisthesis (Grades 3 and 4) at L5-S1, in skeletally mature patients, when autogenous bone graft is used, when affixed to the posterior lumbosacral spine, and intended to be removed after solid fusion is attained. Levels of fixation are from L3-81.

The TiMX Low Back System when not used with pedicle screws, is intended for posterior hook, wire, and/or sacral/iliac screw fixation from T1 to the ilium/sacrum. The non-pedicle screw indications are spondylolisthesis, degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), deformities (scoliosis, lordosis and kyphosis), turnor, fracture and previous failed fusion surgery.

9

JAN 20 1999

{1}------------------------------------------------

Jan. 11. 1999 6:01PM No. 3217 P. 8/13 K984348

## SUBSTANTIAL EQUIVALENCE:

ﺎ ﺃ

્વ

The TiMX Low Back System manufactured from titanium alloy is substantially equivalent, for purposes of this 510(k) adding indications, to itself. This 510(k) seeks to add labeled indications '; pursuant to a reclassification order. The device design remains the ાં વ same.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo is circular in shape, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. In the center of the circle is a stylized image of three human profiles facing to the right, with a wave-like design below them.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JAN 20 1999

Ms. Pam Corsillo Regulatory Submissions Associate DePuy AcroMed, Inc. 3303 Carneqie Avenue Cleveland, Ohio 44115

K984348 Re: TiMX Low Back System Trade Name: K984350 VSP System Trade Name: Requlatory Class: II Product Codes: MNI, MNH, and KWP December 3, 1998 Dated: Received: December 4, 1998

Dear Ms. Corsillo:

We have reviewed your Section 510(k) notifications of intent to market the devices referenced above and we have determined these devices are substantially equivalent (for the indications for use stated in the enclosures) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the devices, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your devices are classified (see above) into either class II (Special Controls) or class III (Premarket Approval), they may be subject to such additional controls. Existing major requlations affecting your devices can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your devices in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

Page 2 - Ms. Pam Corsillo

This letter will allow you to begin marketing your devices as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your devices to legally marketed predicate devices results in a classification for your devices and thus, permits your devices to proceed to the market.

If you desire specific advice for your devices on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diaqnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your devices, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Celia M. Witten, Ph.D., M.D.
Director

Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosures

{4}------------------------------------------------

, if

## K984348 510(k) Number (if known):

TiMX Low Back System (Titanium) Device Name:

Indications for Use: .

The TiMX Low Back System when used with pedicle screws, in skeletally mature patients, is indicated for degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). Levels of fixation are for the thoracic, lumbar and sacral spine.

The TiMX Low Back System is also indicated for pedicle screw fixation for,severe spondylolisthesis (Grades 3 and 4) at L5-S1, in skeletally mature patients, when autogenous bone graft is used, when affixed to the posterior lumbosacral spine, and intended to be removed after solid fusion is attained. Levels of fixation are from L3-S1.

The TiMX Low Back System, when not used with pedicle screws, is intended for posterior hook, wire, and/or sacral/iliac screw fixation from T1 to the ilium/sacrum. The non-pedicle screw indications are spondylolisthesis, degenerative disc disease (defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies), deformities (scoliosis, lordosis and kyphosis), tumor, fracture and previous failed fusion surgery. i . --------------------------------------------------------------------------------------------------------------------------------------------------------------------------

| (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF |
|------------------------------------------------------------------|
| NEEDED)                                                          |

|                                         | Concurrence of CDRH, Office of Device Evaluation (ODE) |                          |
|-----------------------------------------|--------------------------------------------------------|--------------------------|
|                                         |                                                        |                          |
| (Division Sign-Off)                     |                                                        |                          |
| Division of General Restorative Devices |                                                        |                          |
| 510(k) Number                           | K984348                                                |                          |
| Prescription Use (per 21 CRF 801.109)   | X                                                      |                          |
|                                         | OR                                                     | Over-The-Counter Use     |
|                                         |                                                        | (Optional Format 1-2-96) |

---

**Source:** [https://fda-staging.innolitics.com/device/K984348](https://fda-staging.innolitics.com/device/K984348)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).