K984313 · Hyox Systems , Ltd. · CBF · Oct 22, 1999 · Anesthesiology
Device Facts
Record ID
K984313
Device Name
HTU3
Applicant
Hyox Systems , Ltd.
Product Code
CBF · Anesthesiology
Decision Date
Oct 22, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5470
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
To promote the movement of oxygen from the environment to a patient's tissue by means of pressurization that is greater than atmospheric pressure, with non critical patients for the treatment of: 1. Air or Gas Embolism; 2. Carbon Monoxide Poisoning and Smoke Inhalation, Carbon Monoxide Poisoning Complicated by Cyanide Poisoning; 3. Clostridal Myonecrosis; 4. Crush Injury, Compartment Syndrome, Acute Traumatic Ischemias; 5. Decompression Sickness; 6. Enhancement of Healing in Selected Problem Wounds; 7. Exceptional Blood Loss; 8. Necrotizing Soft Tissue Infections; 9. Osteomyelitis (Refractory); 10. Radiation Tissue Damage; 11. Skin Grafts & Flaps (Compromised); 12. Thermal Burns; and 13. Adjunctive Hyperbaric Oxygen in Intracranial Abscess.
Device Story
HTU3 Hyperbaric Chamber provides hyperbaric oxygen therapy by pressurizing patient environment to three times atmospheric pressure. Device facilitates oxygen transport to tissues. Modifications from predecessor HTU2 include increased steel thickness, modified acrylic window shape/thickness, and transition to free-piston compressor mechanism to meet ASME/PVHO-1 safety standards. Compression and decompression times adjusted to accommodate higher pressure, though rates remain consistent. Used in clinical settings for treatment of various ischemic, infectious, and wound-related conditions.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and compliance with ASME/PVHO-1 safety standards.
Technological Characteristics
Hyperbaric chamber; steel construction; acrylic windows; free-piston compressor; operates at 3x atmospheric pressure; complies with ASME/PVHO-1 safety standard for pressure vessels for human occupancy.
Indications for Use
Indicated for non-critical patients requiring hyperbaric oxygen therapy for conditions including air/gas embolism, carbon monoxide poisoning, clostridial myonecrosis, crush injury/compartment syndrome, decompression sickness, problem wound healing, exceptional blood loss, necrotizing soft tissue infections, refractory osteomyelitis, radiation tissue damage, compromised skin grafts/flaps, thermal burns, and intracranial abscess.
Regulatory Classification
Identification
A hyperbaric chamber is a device that is intended to increase the environmental oxygen pressure to promote the movement of oxygen from the environment to a patient's tissue by means of pressurization that is greater than atmospheric pressure. This device does not include topical oxygen chambers for extremities (§ 878.5650).
{0}------------------------------------------------
# OCT 22 1999
## 510(k) SUMMARY
## Hyox Systems, Ltd.'s HTU3 Hyperbaric Chamber
# Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Hyox Systems, Ltd. Pressure Products House Westhill Industrial Estate, Westhill Aberdeen, AB32 6TQ Scotland, United Kingdom
Contact Person:
Ms. Kathleen Scanlan Quality Assurance Manager Hyox Systems, Ltd. Pressure Products House Westhill Industrial Estate, Westhill Aberdeen, AB32 6TQ Scotland, United Kingdom
Date Prepared: February 3, 1999
#### Name of Device
HTU3 Hyperbaric Chamber
### Common or Usual Name
Hyperbaric Chamber
# Classification Name
Hyperbaric Chamber
#### Predicate Devices
- Hyox Systems, Ltd.'s, HTU2 Hyperbaric Chamber (1)
- Sechrist Industries's Sechrist 2500E Hyperbaric Chamber (2)
- Sechrist Industries's Sechrist 3200 Hyperbaric Chamber (3)
{1}------------------------------------------------
## Substantial Equivalence
The HTU3 Hyperbaric Chamber and the predicate devices listed above have the same intended use and very similar principles of operation and technological characteristics. Specifically, the HTU3 and the predicate devices are intended to promote the movement of oxygen from the environment to a patient's tissue by means of pressurization that is greater than atmospheric pressure. The HTU3 is indicated for use in treating: (1) Air or Gas Embolism; (2) Carbon Monoxide Poisoning and Smoke Inhalation, Carbon Monoxide Poisoning Complicated by Cyanide Poisoning; (3) Clostridal Myonecrosis; (4) Crush Injury, Compartment Syndrome. Acute Traumatic Ischemias; (5) Decompression Sickness: (6) Enhancement of Healing in Selected Problem Wounds; (7) Exceptional Blood Loss: (8) Necrotizing Soft Tissue Infections: (9) Osteomvelitis (Refractory): (10) Radiation Tissue Damage; (11) Skin Grafts & Flaps (Compromised); (12) Thermal Burns; and (13) Adjunctive Hyperbaric Oxygen in Intracranial Abscess.
Moreover, the HTU3 is a modified version of the HTU2. To comply with the ASME/PVHO-1 safety standard for pressure vessels for human occupancy, Hyox increased the operating pressure of the Company's hyperbaric therapy unit from two times atmospheric pressure to three times atmospheric pressure, the operating pressure of the Sechrist devices. To facilitate this change, certain minor modifications to the HTU2 were necessary to configure the HTU3. including: (1) an increase in the compression time, as well as an increase in the slow, fast, and emergency decompression times (although the compression and the decompression rates remain unchanged); (2) an increase in the thickness of steel used to manufacture the hyperbaric chamber: (3) an increase in thickness of the device's acrylic windows; and (4) a change from a compressor that functions using a diaphragm unit to a compressor that works using a free piston mechanism. Finally, also to comply with the ASME/PVHO-1 safety standard for pressure vessels for human occupancy, Hyox has made two additional modifications. Hyox has modified the shape of the chamber's acrylic windows and the grade of steel from which the hyperbaric chamber is manufactured.
None of these differences raises new questions of safety or effectiveness. Thus, the HTU3 Hyperbaric Chamber is substantially equivalent to the HTU2, the Sechrist 2500E, and the Sechrist 3200 hyperbaric chambers,
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines representing its wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged around the symbol in a circular fashion.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 22 1999
Mr. Edward C. Wilson, Jr. Hyox Systems, Ltd. c/o Hogan & Hartson L.L.P. 555 Thirteenth Street, N.W. Washington, DC 20004-1109
K984313 Re: HTU3 Hyperbaric Therapy Unit Regulatory Class: II (two) Product Code: 73 CBF Dated: July 27, 1999 Received: July 27, 1999
Dear Mr. Wilson:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Druq, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP requlation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2 - Mr. Edward C. Wilson, Jr.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
teanne Bi Wentworth-
- fri, Wolf Sapirstein, M.D. Acting Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Number (if known):_ K984313
Device Name: Hyox Systems Ltd.'s HTU3 Hyperbaric Chamber
Indications for Use:
To promote the movement of oxygen from the environment to a patient's tissue by means of pressurization that is greater than atmospheric pressure, with non critical patients for the treatment of:
- 1. Air or Gas Embolism;
- Carbon Monoxide Poisoning and Smoke Inhalation, 2. Carbon Monoxide Poisoning Complicated by Cyanide Poisoning;
- Clostridal Myonecrosis; 3.
- Crush Injury, Compartment Syndrome, Acute Traumatic 4. Ischemias;
- Decompression Sickness; 5.
- Enhancement of Healing in Selected Problem Wounds; 6.
- Exceptional Blood Loss; 7.
- Necrotizing Soft Tissue Infections; 8.
- Osteomyelitis (Refractory); 9.
- 10. Radiation Tissue Damage;
- 11, Skin Grafts & Flaps (Compromised);
- 12. Thermal Burns; and
- 13. Adjunctive Hyperbaric Oxygen in Intracranial Abscess.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
to Hillelth
(Division Sign-Off) Division of Cardiovascular. Respiratory, and Neurological Devices
510(k) Number K984313
Prescription Use
(Per 21 C.F.R. 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.