IMMUNO-TROL CONTROL CELLS

K984216 · Coulter Corp. · JPK · Mar 16, 1999 · Hematology

Device Facts

Record IDK984216
Device NameIMMUNO-TROL CONTROL CELLS
ApplicantCoulter Corp.
Product CodeJPK · Hematology
Decision DateMar 16, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 864.8625
Device ClassClass 2

Indications for Use

IMMUNO-TROL™ Control Cells (Immuno-Trol) is an for an assayed, lysable whole blood quality control product for immunophenotyping analysis using monoclonal antibody reagents and flow cytomerry. It provides a positive cell control that is processed in the same manner as a whole blood sample. This allows verification of reagent performance and the methods used for staining targeted cells, lysing erythrocytes, and analyzing samples by flow cytometry. CD populations intended "For In Vitro Diagnostic Use" and those intended "For Research Use In Diagnostic Procedures." are given in separate tables in the TABLE OF EXPECTED RESULTS section of the Immuno-Trol package insert.

Device Story

Assayed, lysable whole blood quality control product; provides positive cell control for immunophenotyping. Input: whole blood sample processing workflow (staining with monoclonal antibodies, erythrocyte lysis). Operation: processed identically to patient whole blood samples; used in flow cytometry analysis. Output: verification of reagent performance, staining methods, and analysis accuracy. Used in clinical laboratories; operated by laboratory technicians/technologists. Healthcare providers use output to validate flow cytometry results; ensures reliability of patient immunophenotyping data. Benefits: provides standardized quality control for clinical diagnostic and research immunophenotyping procedures.

Clinical Evidence

Bench testing only. Studies evaluated vial-to-vial reproducibility, lot-to-lot reproducibility, stability (closed/open vial and stored prepared sample), absolute count method verification, and assay precision. Results demonstrated Immuno-Trol meets all performance specifications and provides percent positive and absolute count results comparable to NWB and CYTO-TROL predicates. All values met pre-defined acceptance criteria.

Technological Characteristics

Assayed, lysable whole blood control containing leukocytes with surface antigens. Compatible with standard monoclonal antibody reagents and flow cytometry instrumentation. Supports both direct (Flow-Count) and indirect absolute count methods. Stability: 90 days closed vial (2-8°C), 30 days open vial (2-8°C).

Indications for Use

Indicated for use as a positive cell control in immunophenotyping analysis of whole blood samples via flow cytometry. Used to verify reagent performance, sample preparation, staining, and erythrocyte lysis procedures. Applicable to CD populations for both in vitro diagnostic and research use.

Regulatory Classification

Identification

A hematology quality control mixture is a device used to ascertain the accuracy and precision of manual, semiautomated, and automated determinations of cell parameters such as white cell count (WBC), red cell count (RBC), platelet count (PLT), hemoglobin, hematocrit (HCT), mean corpuscular volume (MCV), mean corpuscular hemoglobin (MCH), and mean corpuscular hemoglobin concentration (MCHC).

Special Controls

*Classification.* Class II (special controls). Except when intended for use in blood components, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ | COULTER CORPORATION<br>PO. BOX 169015<br>Miami, Florida 33116-9015 USA | Date: January 22, 1999 | |------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Customer Service: (800) 526-7694 | Title: Summary of Safety and Effectiveness Information For 510(k) Premarket Notification | | Product Information: (800) 526-6932<br>(800) 327-6531 (305) 327-6531 | Product: IMMUNO-TROL <sup>TM</sup> Control Cells | | www.coulter.com | | | Company: | Coulter Corporation<br>11800 SW 147 Avenue<br>Miami, FL 33196-2500 | | Contact: | Dr. Marion S. Gaide (M/C: 31-B06)<br>Senior Regulatory Affairs Specialist<br>Premarket Regulatory Affairs<br>(305) 380-2594 | | Telephone: | (305) 380-2594 | | Common or Usual or Classification Name: | Whole Blood Control for Imunophenotyping | | Product Classification: | Product Code: 81 JPK; C.F.R. Section: 864.8625; Classification<br>Panel: Hematology and Pathology Devices; Device Class: II | | Intended Use: | IMMUNO-TROL <sup>TM</sup> Control Cells (Immuno-Trol) is an assayed, lysable whole blood quality control<br>product for immunophenotyping analysis using monoclonal antibody reagents and flow cytometry. It<br>provides a positive cell control that is processed in the same manner as a whole blood sample. This<br>allows verification of reagent performance, and the methods used for staining of targeted cells, lysing<br>erythrocytes, and analyzing samples by flow cytometry. CD populations intended "For In Vitro<br>Diagnostic Use" and those intended "For Research Use Only. Not For Use In Diagnostic Procedures."<br>are given in separate tables in the TABLE OF EXPECTED RESULTS section of the Immuno-Trol<br>package insert. | | | Substantial Equivalence: 510(k) Premarket Notification: K894651 | | | COULTER <sup>TM</sup> CYTO-TROL <sup>TM</sup> Control Cells Kit (CYTO-TROL) | | Product Comparison: | Immuno-Trol is essentially identical to the combined capabilities and intended uses of the two Predicate<br>System components, Normal Whole Blood (NWB) and CYTO-TROL. It also achieves the same<br>purpose as two other Hematology Quality Control Materials for immunophenotyping by flow<br>cytometry currently distributed "For In Vitro Diagnostic Use." Similar to CYTO-TROL, Immuno-Trol<br>provides an assayed positive cell control material consisting of leukocytes with surface antigens<br>detectable by monoclonal antibodies using, for example, CYTO-STAT <sup>®</sup> , CYTO-STAT <sup>®</sup> /COULTER<br>CLONE <sup>®</sup> and IOTest <sup>®</sup> Reagents. However, Immuno-Trol expands the capabilities of CYTO-TROL by<br>featuring more surface antigens. The NWB component of the Predicate System also consists of<br>leukocytes with surface antigens detectable by monoclonal antibodies and may be used in lieu of an<br>assayed cell control product. However, NWB is unassayed material and the specific surface antigens<br>present in a sample are unknown prior to immunophenotyping analysis using one or more monoclonal<br>antibody reagents and flow cytometry. Further, when using NWB to QC immunophenotyping<br>parameters and procedures, the percent positive and absolute count results obtained for targeted cellular<br>populations must be compared to results representative of the Site's normal human donor pool. Similar<br>to NWB, the erythrocyte component of Immuno-Trol can be lysed using a whole blood lysing method<br>such as COULTER <sup>®</sup> ImmunoPrep <sup>TM</sup> Reagent System and Workstation (ImmunoPrep). NWB and<br>CYTO-TROL are the same in that the Standard (Indirect) Method which combines hematology and<br>flow | surface antigens detected are defined by the reagent(s) being assessed for monoclonal antibody activity. Science Serving Humanity flow cytometry results can be used to calculate absolute count results. Immuno-Trol and NWB are the same in that both can be used with the absolute count reagent, COULTER® Flow-Count™ Fluorospheres (Flow-Count) for the direct determination of absolute count by flow cytometry. Finally, Immuno-Trol, NWB and CYTO-TROL are the same in that the gating population, gating method and . 4232216-F R 6-97 {1}------------------------------------------------ Product testing to assess the performance of Immuno-Trol is described below. Studies were designed Product Testing: in line with instructions for use in the product package insert and performance specifications. NWB and CYTO-TROL samples were also assayed for comparison purposes. The results of product testing demonstrated that Immuno-Trol meets all performance specifications and provides Percent Positive (%) and Absolute Count Results comparable to those of NWB and CYTO-TROL and also within the Acceptance Criteria for each leukocyte surface antigen given in the Immuno-Trol package insert. 1. Vial-to-Vial Reproducibility: Reproducibility (that is, Homogeneity) was assessed by measuring the Red Blood Cell (RBC) and White Blood Cell (WBC) concentrations of filled vials of Immuno-Trol on the COULTER® STKS™ Analyzer (STKS). The filled vials were sampled from Immuno-Trol lots at regular intervals across the fill. The study results were expressed in terms of cells/mL and all values met the Acceptance Criteria. These data demonstrate the consistency and reproducibility of Immuno-Trol filled vials. #### 2. Lot-to-Lot Reproducibility: Reproducibility was assessed by testing Immuno-Trol on different instrument-operator combinations using one-color, two-color three-color and four-color monoclonal antibody reagents. Testing produced overall percent positive (%) and absolute count (as applicable) results for the leukocyte surface antigens given in the Immuno-Trol Package Insert. The study results were expressed in terms of percent positive (%) and absolute count (cells/uL) and all values met the Acceptance Criteria. These data demonstrate the consistency and reproducibility of Immuno-Trol between lots. #### 3. Stability Studies: Closed Vial, Open Vial and Stored Prepared Sample stability characteristics and claims for Immuno-Trol were assessed with two study protocols: "Closed Vial: 90 days at 2-8℃; Open Vial: 30 days at 2-8℃" and "Stored Prepared Sample: 2 hours at 20-25°C; 24 hours at 2-8°C." Immuno-Trol was tested using one-color, threecolor and four-color monoclonal antibody reagents under the different storage and temperature conditions over test periods extending up to and beyond the respective 90-day and 30-day dains. The study results were expressed in terms of percent positive (%) and absolute count (cells/uL) and all values met the Acceptance Criteria. These data demonstrate Immuno-Trol meets both reagent and sample stability characteristics and claims under the storage and temperature conditions studied. - 4. Absolute Count Method Verification: Absolute count method was assessed by testing Immuno-Trol using three-color and four-color monoclonal antibody reagents and the Flow-Count (Direct) Method and the Standard (Indirect) Method for absolute count determination. NWB was also tested under the same conditions. The study results were expressed in terms of percent positive (%) and absolute count (cells/uL) and all values met the Acceptance Criteria. In addition, both the Flow-Count (Direct) Method and the Standard (Indirect) Method provided essentially identical absolute count results for Immuno-Trol and NWB. These data clearly demonstrate the suitability of using the Flow-Count (Direct) Method for absolute count determination of Immuno-Trol leukocyte surface antigens. - 5. Assay Precision: Assay precision (that is, reproducibility) was assessed by testing Immuno-Trol on different instrument-operator combinations and over multiple test days using one-color and four-color and four-color monoclonal antibody reagents. NWB and CYTO-TROL were also tested under the same conditions. The study results were expressed in terms of percent positive (%) and absolute count (cells/uL) and analyzed in terms individual and combined assay precision for the various study parameters. All values met the Acceptance Criteria. In addition, Immuno-Trol, NWB and CYTO-TROL varied little as evidenced by the means, standard deviations and coefficients of variation for the different sets of measurements. These data demonstrate Immuno-Trol performs in a manner comparable to the Predicate System components, NWB and CYTO-TROL. Marion S. Gaide, Ph.D. Senior Regulatory Affairs Specialist Corporate Regulatory Affairs January 22, 1999 Date trolse {2}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/2 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle. ## MAR 1 6 1999 Marion S. Gaide, Ph.D. Senior Regulatory Affairs Specialist Premarket Regulatory Affairs Coulter Corporation 11800 SW 147 Avenue Miami, Florida 33196-2500 Re: K984216 Trade Name: IMMUNO-TROL™ Control Cells Regulatory Class: II Product Code: JPK Dated: February 22, 1999 Received: February 23, 1999 ### Dear Dr. Gaide: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 {3}------------------------------------------------ ## Page 2 Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Steven Sutman Steven I. Gutman, M.D. M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Page 1 of 1 # INDICATIONS FOR USE 510(k) Number (if known): K984216 IMMUNO-TROL™ Control Cells Device Name: #### Indications For Use: IMMUNO-TROL™ Control Cells (Immuno-Trol) is an for an assayed, lysable whole blood quality control product for immunophenotyping analysis using monoclonal antibody reagents and flow cytomerry. It provides a positive cell control that is processed in the same manner as a whole blood sample. This allows verification of reagent performance and the methods used for staining targeted cells, lysing erythrocytes, and analyzing samples by flow cytometry. CD populations intended "For In Vitro Diagnostic Use" and those intended "For Research Use In Diagnostic Procedures." are given in separate tables in the TABLE OF EXPECTED RESULTS section of the Immuno-Trol package insert. Immunophenotyping analysis by flow cytometry involves the identification and enumeration of targeted cells in whole blood samples. Whole blood samples are stained with monoclonal antibodies and erythrocytes are lysed prior to flow cytometric analysis. A positive cell control is required to verify reagent performance, sample performance, sample preparation methods, and staining procedures. A positive cell control should mimic a representative whole blood sample in terms of monoclonal antibody performance, erythrocyte lysing, and flow cytometric analysis. Steve G. Maher (Division Sign-Om Division of Clinical Laboratory Devices 510(k) Number L984216 (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) V Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use 510i4us2
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...