K984167 · GE Medical Systems · LNI · Jan 12, 1999 · Radiology
Device Facts
Record ID
K984167
Device Name
SAGE 7
Applicant
GE Medical Systems
Product Code
LNI · Radiology
Decision Date
Jan 12, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
SAGE 7 is a package of software tools for MR spectroscopic data processing and display. SAGE 7 is indicated for use for a wide variety of MR spectroscopic data management tasks, including file handling, display, processing/modeling, analysis, storage and hard copy output. The resultant spectroscopic presentations, when interpreted by a trained physician, can provide physiological / chemical information that can be useful in determining a diagnosis.
Device Story
SAGE 7 is a software toolkit for MR spectroscopy data management, operating on GE Signa Horizon Systems and Advantage Windows workstations. It processes raw spectroscopy data to reconstruct, analyze, and display spectra and spectroscopic images. The system enables trained clinical spectroscopists to perform tasks such as file handling, modeling, and numerical analysis (e.g., peak positions, widths, heights, areas). Outputs are viewed by physicians to derive physiological or chemical information for diagnostic support. The device enhances the previously marketed SA/GE Analysis option by improving the user interface and leveraging modern workstation computational power to perform sophisticated curve-fitting algorithms that were previously impractical.
Clinical Evidence
Bench testing only. The device was evaluated for compliance with international medical equipment safety standard IEC 601-2-33 (Particular requirements for the safety of magnetic resonance equipment for medical diagnosis).
Technological Characteristics
Software-based MR spectroscopy analysis toolkit. Operates on GE Signa Horizon Systems and Advantage Windows workstations. Implements standard spectroscopic processing algorithms including Fourier transform, phasing, peak-picking, integration, and curve-fitting. Complies with IEC 601-2-33 safety standards.
Indications for Use
Indicated for MR spectroscopic data management tasks, including file handling, display, processing, modeling, analysis, storage, and output, to provide physiological/chemical information for diagnostic support by a trained physician.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
SA/GE Analysis option
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized script. The letters are enclosed within a circular border, which has decorative swirls or flourishes around the perimeter. The logo is presented in black and white.
page 1 of 2
K984167
P.O. Box 414. W-709 Milwaukee. WI 53201 USA
JAN 1 2 1999
## SUMMARY OF SAFETY AND EFFECTIVENESS
- o This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.87(h).
- o Identification of Submitter Larry A. Kroger, Ph.D., 414-544-3894, November 19, 1998
- o Identification of the Product SAGE 7 Spectroscopy Analysis Option
| Manufactured by: | GE Medical Systems<br>3200 N. Grandview Blvd.<br>Waukesha, WI 53188 |
|------------------|---------------------------------------------------------------------|
|------------------|---------------------------------------------------------------------|
- 0 Marketed Devices
The SAGE 7 Spectroscopy Analysis option is an enhancement to the currently marketed SA/GE Analysis option in distribution since the reclassification of Magnetic Resoncance Diagnostic devices (ref. 53 FR 5078, Feb. 1, 1989).
- Device Description
0
The GE SAGE 7 Spectroscopy Analysis Software is designed to operate on GE Signa Horizon Systems and GE Advantage Windows workstations. It provides a toolkit of software applications to handle a wide variety of tasks associated with spectroscopy data management. The SAGE (Spectroscopy Analysis GE) package provides the capability for the trained clinical spectroscopist to reconstruct, analyze, and display spectra and spectroscopic images to provide information to support a diagnosis. SAGE 7 provides a tool for the spectroscopist to process raw spectroscopy data into spectra or spectroscopic images, view them, and optionally perform numerical analyses to obtain spectroscopic parameters (ex. peak positions, widths, heights, areas, etc.).
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined and enclosed within a circle. The letters and the circle are in black, creating a strong contrast. The logo is a well-known symbol associated with the General Electric company.
page 2 of 2
P.O. Box 414, W-709 Milwaukee, WI 53201 USA
# SUMMARY OF SAFETY AND EFFECTIVENESS
#### o Indications for Use
SAGE 7 is a package of software tools for MR spectroscopic data processing and display. SAGE 7 is indicated for use for a wide variety of MR spectroscopic data management tasks, including file handling, display, processing/modeling, analysis, storage and hard copy output. The resultant spectroscopic presentations, when interpreted by a trained physician, can provide physiological / chemical information that can be useful in determining a diagnosis.
#### 0 Comparison with Predicate
The SAGE 7 Spectroscopy Analysis option is an enhancement to the SA/GE Analysis option. The basic operations (ex. Fourier transform, phasing, peak-picking, integration, fitting, etc.) have not significantly changed, and are provided in both SAGE 7 and the SA/GE Analysis option. What has changed is the sophistication of the user interface. In addition, the increased computational capability of modern workstations have made possible as routine operations some sophisticated curve-fitting algorithms that were impractical on older computer hardware.
### o Summary of Studies
The SAGE 7 Spectroscopy Analysis option was evaluated to International medical equipment safety standard IEC 601-2-33 titled Particular requirements for the safety of magnetic resonance equipment for medical diagnosis. The SAGE 7 Spectroscopy Analysis option is comparable to the SA/GE Spectroscopy option.
- 0 Conclusions
It is the opinion of GE that the SAGE 7 Spectroscopy Analysis option is substantially equivalent to the SA/GE Spectroscopy option. SAGE 7 does not include any new indications for use, nor does use of this device result in any new potential hazards.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES". The text is in all caps and is in a bold, sans-serif font. The text is likely part of a document or sign.
Rockville MD 20850
Image /page/2/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard
JAN 1 2 1999
Larry A. Kroger, Ph.D. Senior Regulatory Programs Manager GE Medical Systems P.O. Box 414 . W-709 Milwaukee, WI 53201
Re: K984167
SAGE 7 Spectroscopy Dated: November 19, 1998 Received: November 20, 1998 Regulatory class: II 21 CFR 892.1000/Procode: 90 LNI
Dear Dr. Kroger:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration,
If your device is classified (see above) into either class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for vour device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Capt. Daniel G. Schultz, M.D. Acting Director, Division of Reproductive. Abdominal. Ear. Nose and Throat. and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________
Device Name: SAGE 7 Spectroscopy
Indications For Use:
SAGE 7 is a package of software tools for MR spectroscopic data processing and display. SAGE 7 is indicated for use for a wide variety of MR spectroscopic data management tasks, including file handling, display, processing/modeling, analysis, storage and hard copy output. The resultant spectroscopic presentations, when interpreted by a trained physician, can provide physiological / chemical information that can be useful in determining a diagnosis.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
David A. Seymon
(Division Sign-Off) Division of Reproductive, Abdominal, E and Radiological I 510(k) Number
Prescription Use_ (Per 21 CFR 801.109)
OR
Over-The-Counter Use_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.