K984141 · Sienco, Inc. · JBP · May 28, 1999 · Hematology
Device Facts
Record ID
K984141
Device Name
GBACT + KIT
Applicant
Sienco, Inc.
Product Code
JBP · Hematology
Decision Date
May 28, 1999
Decision
SESE
Submission Type
Special
Regulation
21 CFR 864.7140
Device Class
Class 2
Indications for Use
The gbACT+ Kit is an in-vitro diagnostic product designed for use with the Sonoclot Coagulation & Platelet Function Analyzer System. The gbACT+ test is an activated whole blood clotting time test. It may also be used with citrated whole blood or plasma. The gbACT+ Kit is intended for general purpose global hemostasis monitoring including: clot detection, fibrin formation, platelet function, and hyperfibrinolysis. The monitoring information is typically used for anticoagulant management at low to moderate heparin levels (0 to 2 units per ml), hypercoagulable and/or hypocoagulable screening, platelet function assessment, and hyperfibrinolysis screening.
Device Story
The gbACT+ Kit is an in-vitro diagnostic reagent system used with the Sonoclot Coagulation & Platelet Function Analyzer. The kit consists of pink plastic lidded cuvettes containing glass bead contact activators and a magnetic stir bar. The device measures the mechanical impedance of a blood sample as it clots; the analyzer tracks the movement of the stir bar within the sample. As fibrin forms and platelets interact, the viscosity changes, altering the stir bar's motion. The system produces quantitative and qualitative outputs, including Activated Clotting Time (ACT/Onset), Clot RATE, and the 'Sonoclot Signature' graph. These outputs are used by clinicians to manage anticoagulation, assess platelet function, and screen for hyperfibrinolysis. The device is intended for professional use in clinical settings to guide therapeutic decisions regarding hemostasis.
Clinical Evidence
Bench testing only. Comparison of gbACT+ and standard SonACT performance using native whole blood from a normal population. Provided typical reference ranges for ACT/Onset (125-220s for gbACT+ vs 85-145s for SonACT), Clot RATE, and Time to Peak. Representative 'Sonoclot Signature' graphs provided to demonstrate functional equivalence in clot detection and rate measurement.
Technological Characteristics
IVD reagent kit; pink plastic lidded cuvettes containing glass bead contact activator and magnetic stir bar. Operates via mechanical impedance sensing within the Sonoclot Coagulation & Platelet Function Analyzer. Quantitative and qualitative outputs derived from stir bar motion analysis. Quality control via Reference Plasma Quality Control Kit.
Indications for Use
Indicated for global hemostasis monitoring in patients requiring assessment of clot detection, fibrin formation, platelet function, and hyperfibrinolysis. Used for anticoagulant management at low to moderate heparin levels (0-2 units/ml) and screening for hypercoagulable or hypocoagulable states. Suitable for use with whole blood or citrated plasma.
Regulatory Classification
Identification
An activated whole blood clotting time tests is a device, used to monitor heparin therapy for the treatment of venous thrombosis or pulmonary embolism by measuring the coagulation time of whole blood.
{0}------------------------------------------------
#### MAY 28 1000 Sienco, Inc. 4892 Van Gordon St. - Unit 203 Wheat Ridge, CO USA 80033 303/420-2204(Fax) 303/420-1148 800/432-1624
# 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: _K 984 141
- (1) Contact Person: Barbara DeBiase, Director, Sienco Inc. Summary Preparation Date: 1 November 1998
- (2) Device Trade Name: gbACT+ Kit Common Name: Activated Clotting Time (ACT) Test Classification Name: Activate Whole Blood Clotting Time Test Classification: Class II, 21 CFR 864.7140
- (3) Identification of predicate device to which substantial equivalence is being claimed: SonACT Kit, 510(k) # K952560
- (4) and (5) Device Description and Intended Use: The gbACT+ Kit is an in-vitro diagnostic product designed for use with the Sonoclot Coagulation & Platelet Function Analyzer System. Each kit contains glass bead cuvettes, probes and instructions for use. Current package sizes of 100 (#800-0412) and 24 (#800-0411) tests are available. The lidded pink cuvettes contain a controlled amount of the glass bead activator and a magnetic stir bar. A mason box is used for the kit package.
The gbACT+ test is an activated whole blood clotting time test. It may also be used with citrated whole blood or plasma. The gbACT+ Kit is intended for general purpose global hemostasis monitoring including: clot detection, fibrin formation, platelet function, and hyperfibrinolysis. The monitoring information is typically used for anticoagulant management at low to moderate heparin levels (1 to 2 units per ml), hypercoagulable and/or hypocoagulable screening, platelet function assessment, and hyperfibrinolysis screening.
. "
{1}------------------------------------------------
#### (6) Technological Characteristic Comparison of the gbACT+ Kit to the standard SonACT Kit:
| ITEM | gbACT+ | STANDARD SONACT |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Classification Name | Activated Whole Blood Clotting<br>Time Test | Activated Whole Blood Clotting<br>Time Test |
| IVD Reagent for<br>Use With | Sonoclot Coagulation & Platelet<br>Function Analyzer | Sonoclot Coagulation & Platelet<br>Function Analyzer |
| Intended Use | General purpose global hemostasis<br>monitoring, hypercoagulable and<br>hyperfibrinolysis screening,<br>platelet function assessment,<br>anticoagulation management (low<br>to moderate heparin levels) | General purpose global hemostasis<br>monitoring, hypercoagulable and<br>hyperfibrinolysis screening,<br>platelet function assessment,<br>anticoagulation management (low<br>to high heparin levels) |
| Results Provided | Quantitative results for Activated<br>Clotting Time (Onset /ACT Time).<br>Qualitative and quantitative fibrin<br>formation information (Clot RATE,<br>Sonoclot Signature). Qualitative<br>and quantitative platelet function<br>information from Sonoclot<br>Signature. Qualitative and<br>quantitative hyperfibrinolysis<br>information from Sonoclot<br>Signature. | Quantitative results for Activated<br>Clotting Time (Onset/ACT Time).<br>Qualitative and quantitative fibrin<br>formation information (Clot RATE<br>Sonoclot Signature). Qualitative<br>and quantitative platelet function<br>information from Sonoclot<br>Signature. Qualitative and<br>quantitative hyperfibrinolysis<br>information from Sonoclot<br>Signature. |
| Test Design | Pink plastic lidded cuvette<br>containing contact activator (glass<br>beads ) and magnetic stir bar | Colorless plastic lidded cuvette<br>containing contact activator (celite)<br>and magnetic stir bar |
| Kit Contents | Cuvettes, probes and instructions<br>for use | Cuvettes, probes and instructions<br>for use |
| Quality Control | Reference Plasma Quality Control<br>Kit, Sienco part #900-1318 | Reference Plasma Quality Control<br>Kit, Sienco part #900-1318 |
510(k) Summary - Page 2 of 3
:
.
..
ー…。
『
!
{2}------------------------------------------------
- (7) Performance Comparison of the gbACT+ Kit to the standard SonACT Kitt Expected
Performance Comparison of the goACT The to the standard SonACT in the lables below.
values for the gbACT+ Kit and the standard SonACT to estandard values for the gDACT+ Tric and the standard Somro-Print as pro-SonACT test are also included.
#### gbACT+ Typical Results and Examples
| gbACT+ Test (#800-0411, 800-0412)<br>Native Whole Blood - Normal Population, No Heparin | |
|-----------------------------------------------------------------------------------------|----------------------------------|
| Result | Normal Range |
| ACT/Onset | 125 - 220 seconds |
| Clot RATE | 10-45 Clot Signal Units / minute |
| Time to Peak | <20 minutes |
### Normal A: Native Whole Blood: gbACT+
Image /page/2/Figure/4 description: The image shows three lines of text related to medical measurements. The first line reads "Onset/ACT: 167", indicating a value of 167 for Onset/ACT. The second line reads "ClotRATE: 15", indicating a ClotRATE value of 15. The third line reads "Time to Peak: 8", indicating a Time to Peak value of 8.
Image /page/2/Figure/5 description: The image shows a graph of a clot signal over time. The y-axis represents the clot signal, and the x-axis represents time. The graph shows an initial increase in the clot signal, followed by a peak, and then a gradual decrease. The onset is labeled as 169, and the rate is 15.
# Normal B: Native Whole Blood: gbACT+
Onset/ACT: 149 ClotRATE: 9 Time to Peak: 12
Image /page/2/Figure/20 description: The image shows a graph of a clot signal. The x-axis is labeled "SIEN.D. INC." and ranges from 0 to 20. The y-axis is labeled "Clot Signal" and ranges from 0 to 120. The graph shows two peaks, one at approximately x=5 and y=40, and another at approximately x=12 and y=80. The text "Onset=149" and "帆-1.3" are also visible on the graph.
#### SonACT Typical Results and Examples for Comparison
| Standard SonACT Test (#800-0432, #800-0431)<br>Native Whole Blood - Normal Population, No Heparin | |
|---------------------------------------------------------------------------------------------------|----------------------------------|
| Result | Normal Range |
| ACT/Onset | 85-145 seconds |
| Clot RATE | 15-45 Clot Signal Units / minute |
| Time to Peak | <30 minutes |
# Normal A: Native Whole Blood: SonACT
Image /page/2/Figure/12 description: The image shows a graph of clot signal over time. The graph shows an initial increase in clot signal, followed by a peak and then a gradual decrease. The image also includes some text, which indicates that the onset/ACT is 89, the clot rate is 22, and the time to peak is 17. The y axis is labeled clot signal and the x axis is not labeled.
### Normal B: Native Whole Blood: SonACT
Image /page/2/Figure/14 description: The image shows a graph of a clot signal over time. The graph shows the clot signal increasing to a peak, then decreasing. The image also includes the following values: Onset/ACT: 91, ClotRATE: 10, and Time to Peak: 20.
Barbara DeBiase, Director
rbara DeBiase, Director
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the top half of the circle. In the center of the seal is a stylized image of an eagle with three lines representing its wings. The eagle is facing to the right.
MAY 2 8 1939
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Barbara Debiase SIENCO, INC. 4892 Van Gordon Street Unit 203 80033 Wheatridge, CO
- Re: K984141 Trade Name: gbACT + Kit Requlatory Class: II Product Code: JBP Dated: May 11, 1999 Received: May 12, 1999
Dear Ms. Debiase:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the qeneral controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
#### Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours.
Steven Sutman
Steven I. Gutman, M.D, M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page_ 1_of_1
510(k) Number (if known): \$944 / 41
Device Name:
### Indications For Use:
The gbACT+ Kit is an in-vitro diagnostic product designed for use with the Sonoclot Coagulation & Platelet Function Analyzer System. The gbACT+ test is an activated whole blood clotting time test. It may also be used with citrated whole blood or plasma. The gbACT+ Kit is intended for general purpose global hemostasis monitoring including: clot detection, fibrin formation, platelet function, and hyperfibrinolysis. The monitoring information is typically used for anticoagulant management at low to moderate heparin levels (0 to 2 units per ml), hypercoagulable and/or hypocoagulable screening, platelet function assessment, and hyperfibrinolysis screening.
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| | <img alt="Signature" src="signature.png"/> |
|-----------------------------------------|--------------------------------------------|
| (Division Sign-Off) | |
| Division of Clinical Laboratory Devices | |
| 510(k) Number | K984141 |
| Prescription Use | <div style="text-align:center;">✓</div> |
|----------------------|-----------------------------------------|
| (Per 21 CFR 801.109) | |
OR
| Over-The-Counter Use | _________________ |
|--------------------------|-------------------|
| (Optional Format 1-2-96) | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.