K984023 · Dynamic Healthcare Technologies, Inc. · LLZ · Dec 8, 1998 · Radiology
Device Facts
Record ID
K984023
Device Name
MAXVIEW DIAGNOSTIC WORKSTATION
Applicant
Dynamic Healthcare Technologies, Inc.
Product Code
LLZ · Radiology
Decision Date
Dec 8, 1998
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Dynamic PACSPlus is a powerful image management tool designed for use by technologists. radiologists and other healthcare professionals. Dynamic PACSPlus is fully compatible with the RADPlus and Maxifile RIS as well as other Hospital and Radiology Information Systems. The product provides client/server-based multimedia results viewing and distribution.
Device Story
PACSPlus v2.0 is an information-enabled PACS integrating RIS data with medical images. Input sources include imaging modalities via DICOM or selected non-DICOM interfaces. InfoBroker Servers capture images and reconcile them with RIS orders; Image and Information Servers store, index, and distribute data. Users access exams via patient identifiers, location, or physician through the web-based Dynamic WebSight viewer. The system displays full imaging exam history, including film and digital results. Used in clinical settings by radiologists, technologists, and healthcare professionals to review data and support clinical decision-making. The device does not contact patients; it serves as a tool for professional evaluation of radiological data.
Clinical Evidence
No clinical data provided. The submission relies on bench testing and technical comparison to the predicate device.
Technological Characteristics
Client/server-based PACS software. Integrates with RIS via DICOM and non-DICOM interfaces. Features web-based viewing (Dynamic WebSight). Includes wavelet lossy image compression via Pegasus PICtools. Operates on networked infrastructure for image storage, indexing, and distribution.
Indications for Use
Indicated for use by radiology professionals to capture, store, distribute, and display radiological images and information over local and wide area networks.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Pegasus PICtools Medical Compression toolkit (K982146)
Submission Summary (Full Text)
{0}------------------------------------------------
#### DEC 8 1998 SUMMARY OF SAFETY AND EFFECTIVENESS for SPECIAL 510(k) MODIFICATION SUBMISSION
DEVICE CLASSIFICATION NAME: PACS System (Picture Archiving and Communications System) 892.2050 REGULATION NUMBER: PACS System (Picture Archiving and Communications COMMON/USUAL NAME: System) PACSPlus, version 2.0 TRADE/PROPRIETARY NAME: PREVIOUS 510(k) DEVICE NAME: MaxiView Radiology Workstation PREVIOUS 510(k) DEVICE NUMBER: K930500 ESTABLISHMENT LICENSE NO .: 1933179
# REASON FOR SPECIAL 510(k) SUBMISSION:
The predicate device has been modified to incorporate changes in technology. A comparison chart is included as a part of the Special 510(k) submission.
# SYSTEM DESCRIPTION:
"Dynamic PACSPlus Version 2.0 is an information-enabled PACS (Picture Archiving and Communication System) that integrates all information from the RIS with images from PACS. Image capture is available from most imaging modalities using DICOM interfaces and selected non-DICOM interfaces. PACSPlus InfoBroker Servers are used as gateways that capture the DICOM images, reconcile the DICOM information with order information from the RIS, and provide access to additional exam and patient information including results. The PACSPlus Image and Information Server stores, indexes, and distributes images and information. Images can be viewed through the web-based Dynamic WebSight viewer. Exams can be accessed by patient identifiers, location, or by physician. The entire imaging exam history is shown (film and digital) to allow the viewer to select any results or available digital images. " (source: Product Fact Sheet, Dynamic Healthcare Technologies, Inc.)
### INTENDED USE:
Dynamic PACSPlus is a powerful image management tool designed for use by technologists. radiologists and other healthcare professionals. Dynamic PACSPlus is fully compatible with the RADPlus and Maxifile RIS as well as other Hospital and Radiology Information Systems. The product provides client/server-based multimedia results viewing and distribution.
# SAFETY STATEMENT:
Dynamic Healthcare Technologies, Inc.'s PACSPlus product, version 2.0, is a set of software applications and interfaces that do not directly come into contact with the patient. Further, as specified in our Users' Guide, this product is intended for use by trained professionals only, who utilize their education and experience to evaluate and review the data obtained and make a decision using that professional judgement. We believe this software product to be a safe and effective device to be on the market.
# CONTACT INFORMATION:
Dynamic Healthcare Technologies, Inc. 615 Crescent Executive Center, Fifth Floor Lake Mary, FL 32746-2109
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract image of what appears to be an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
10008 DEC ರ್ಧ
Karen White Senior Consultant, Regulatory Affairs Dynamic Healthcare Technologies, Inc. 615 Crescent Executive Court Fifth Floor Lake Mary, FL 32746
Re: K984023
Dynamic PACSPLUS™ Version 2.0 Dated: November 12, 1998 Received: November 12, 1998 Regulatory class: II 21 CFR 892.2050/Procode: 90 LLZ
Dear Ms. White:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (2) CFR Part 820) and that, through periodic OS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diggiostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrf/dsmaldsmamain.html".
Sincerely yours,
Kilian Yiu
Lillian Yin, Ph.D. Director, Division of Reproductiv Abdominal. Ear. Nose and Thro and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known): K984023
Dynamic PACSPlus, version 2.0 Device Name:
# Indications for Use Statement
Dynamic Healthcare Technologies' PACSPlus, version 2.0, is indicated for use by radiology professionals to capture, store, distribute and display radiological images and information over local and Wide Area Networks. The only editing of the images is via the wavelet lossy compression feature, using the Pegasus PICtools ™ Medical Compression toolkit, a 510(k)-cleared medical device product (K#982146).
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Vivid C. Segmen
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
510(k) Number K984023
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use
- .
(Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.