IMMULITE® Myoglobin is a solid-phase, two-site chemiluminescent enzyme immunometric assay for use with the IMMULITE Automated Analyzer and designed for the quantitative measurement of myoglobin in serum. It is intended strictly for in vitro use as an aid in the diagnosis of acute myocardial infarction (AMI).
Device Story
IMMULITE® Myoglobin is an in vitro diagnostic assay used with the IMMULITE Automated Analyzer to measure myoglobin in patient serum. The device utilizes a solid-phase, two-site chemiluminescent enzyme immunometric assay. A polystyrene bead coated with monoclonal anti-myoglobin antibody captures myoglobin from the sample, forming an antibody-sandwich complex with an alkaline phosphatase-conjugated polyclonal antibody. Following incubation and washing, a phosphate ester substrate is added; hydrolysis by alkaline phosphatase produces an unstable intermediate that emits light. The luminometer measures photon output, which is directly proportional to myoglobin concentration. Results are used by clinicians to aid in the diagnosis of acute myocardial infarction (AMI), assess reperfusion after thrombolytic treatment, and monitor for reinfarction. The test provides early detection of AMI due to rapid myoglobin diffusion following cardiac cell membrane disruption.
Clinical Evidence
Method comparison study performed on 172 patient samples with myoglobin concentrations ranging from 13 to 831 ng/mL. Linear regression analysis comparing IMMULITE® Myoglobin to the ACS:180 predicate yielded a correlation coefficient (r) of 0.981. Mean values were 61 ng/mL for IMMULITE® and 76 ng/mL for the predicate.
Technological Characteristics
Solid-phase, two-site chemiluminescent enzyme immunometric assay. Solid phase: polystyrene bead coated with monoclonal anti-myoglobin antibody. Detection: alkaline phosphatase-conjugated polyclonal antibody; phosphate ester of adamantyl dioxetane substrate. Energy source: automated analyzer luminometer. Connectivity: integrated with IMMULITE Automated Analyzer. Software: automated assay processing and signal quantification.
Indications for Use
Indicated for the quantitative measurement of myoglobin in serum as an aid in the diagnosis of acute myocardial infarction (AMI).
Regulatory Classification
Identification
A myoglobin immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the myoglobin (an oxygen storage protein found in muscle) in serum and other body fluids. Measurement of myoglobin aids in the rapid diagnosis of heart or renal disease.
{0}------------------------------------------------
NOV 2 5 1998
K983970
# 510(k) Summary Safety and Effectiveness
This summary of safety and effectiveness information has been prepared in accordance with the requirements of SMDA 1990 and 21 CFR Part 807.92.
| Name: | Diagnostic Products Corporation |
|-----------------------------------------------|--------------------------------------------------------------------------------------------------------|
| Address: | 5700 West 96th Street<br>Los Angeles, California 90045-5597 |
| Telephone Number: | (310) 645-8200 |
| Facsimile Number: | (310) 645-9999 |
| Contact Person: | Edward M. Levine, Ph.D.<br>Director of Clinical Affairs |
| Date of Preparation: | November 5, 1998 |
| Device Name:<br>Trade: | IMMULITE® Myoglobin |
| Catalog Number: | LKMY1 (100 tests), LKMY5 (500 tests) |
| Common: | Reagent system for the determination of myoglobin in<br>serum. |
| Classification: | Class II device, 75-DDR (21CFR 866.5680) |
| Manufacturer: | Diagnostic Products Corporation<br>5700 West 96th Street<br>Los Angeles, California 90045-5597 |
| Sole U.S. Importer: | Diagnostic Products Corporation<br>5700 West 96th Street<br>Los Angeles, California 90045-5597 |
| Establishment Registration<br>Number | DPC's Registration Number is 2017183 |
| Substantially Equivalent<br>Predicate Device: | ACS:180 Myoglobin (K974325)<br>Manufactured by Chiron Diagnostics |
| Description of Device: | IMMULITE® Myoglobin is a clinical device for use with<br>the IMMULITE® Automated Immunoassay Analyzer. |
{1}------------------------------------------------
# Intended Use of the Device:
IMMULITE® Myoglobin is a solid-phase, two-site chemiluminescent enzyme immunometric assay for use with the IMMULITE Automated Analyzer and designed for the quantitative measurement of myoglobin in serum. It is intended strictly for in vitro use as an aid in the diagnosis of acute myocardial infarction (AMI).
#### Performance Equivalence:
Diagnostic Products Corporation (DPC) asserts that the IMMULITE® Myoglobin produces substantially equivalent results to other commercially marketed myoglobin assays, such as the ACS:180 Myoglobin. Each product is intended strictly for in vitro diagnostic use to aid in the clinical diagnosis of acute myocardial infarction.
#### Summary and Explanation of the Test:
Acute myocardial infarction (AMI) disrupts cardiac cell membranes, releasing intracellular cardiac proteins into the vascular system. Some of the proteins, including myoglobin, creatine kinase-MB (CK-MB), lactate dehydrogenase type 1 (LD1), and cardiac troponin subunits I and T, have proven useful in diagnosing AMI. The optimal clinical utility of each marker depends on specific protein characteristics. Myoglobin, the smallest of the markers, diffuses rapidly throughout the vascular system and provides the earliest indication of AMI. Myoglobin levels become elevated 2-3 hours after chest pain onset and peak within 6-9 hours. Other markers reach peak levels 12-19 hours after pain onset. The kidneys rapidly eliminate the 17.8 kDa protein from the system, restoring normal circulating concentrations within 24 hours. Since the protein rapidly clears from the system, myoglobin concentrations can reliably indicate reinfarction. Additionally, myoglobin measurements can preclude AMI: two consecutive low measurements, the first upon patient admission and the second 1-2 hours later, negatively predict AMI in nearly all cases. Myoglobin measurements provide early detection of reperfusion after thrombolytic treatment as well. Some unrelated patients conditions produce high myoglobin levels, decreasing test specificity. Myoglobin participates in aerobic metabolism in both skeletal and cardiac muscle cells, and high levels accompany various muscle traumas. Renal failure and other kidney problems produce high myoglobin levels as well. Most of the complications have distinct clinical symptoms that allow reliable differential diagnosis.
{2}------------------------------------------------
## Technology Comparison:
Provided for the reviewer is a comparison of DPC's IMMULITE® Myoglobin technology vs. the ACS: 180 Myoglobin technology: This section does not contain any new information for a reviewer who is familiar with the DPC IMMULITE® System based upon the review of previous IMMULITE® assay submissions.
IMMULITE® Myoglobin is a solid-phase, two-site chemiluminescent immunometric assay. The solid phase, a polystyrene bead enclosed within an IMMULITE Test Unit, is coated with a monoclonal antibody specific for myoglobin. While the patient serum sample and alkaline phosphatase-conjugated polyclonal antibody are incubated for approximately 30 minutes at 37 "C in the Test Unit with intermittent agitation, myoglobin in the sample is bound to form an antibody sandwich complex. Unbound conjugate is then removed by a centrifygal wash, after which substrate is added and the Test Unit is incubated for a further 10 minutes.
The chemiluminescent substrate, a phosphate ester of adamantyl dioxetane, undergoes hydrolysis in the presence of alkaline phosphatase to yield an unstable intermediate. The continuous production of this intermediate results in the sustained emission of light, thus improving precision by providing a window for multiple readings. The bound complex - and thus also the photon output, as measured by the luminometer - is proportional to the concentration of myoglobin in the sample.
The ACS:180 Myoglobin is a two-site sandwich immunoassay using direct chemiluminescent technology, which uses constant amounts of two antibodies. The first antibody, in the Lite Reagent, is a polyclonal goat anti-myoglobin antibody labeled with acridinium ester. The second antibody, in the Solid Phase, is a monoclonal mouse anti-myoglobin antibody covalently coupled to paramagnetic particles.
The system automatically performs the following steps:
- Dispenses 10 uL of sample into a cuvette .
- Dispenses 100 uL of Lite Reagent and incubates the reagents for 2.5 minutes at 37 °C .
- . Dispenses 200 uL of Solid Phase and incubates the reagents for 5.0 minutes at 37 °C
- Separates, aspirates, and washed the cuvettes with reagent water
- o Dispenses 300 uL each of Reagent 1 and Reagent 2 to initiate the chemiluminescent reaction
- . Reports results according to the selected option, as described in the system operating instructions or in the online help system
A direct relationship exists between the amount of myoglobin present in the patient sample and the amount of relative light units (RLUs) detected by the system.
{3}------------------------------------------------
## Method Comparison:
The IMMULITE® Myoglobin procedure was compared to a commercially available solid phase immunoassay for myoglobin (ACS:180) on 172 patient samples, with myoglobin concentrations ranging from approximately 13 to 831 ng/mL. Linear regression analysis yielded the following statistics:
(IMMULITE®) = 0.87 (ACS: 180) = 5.7 ng/mL r = 0.981
61 ng/mL (IMMULITE®) Means 76 ng/mL (ACS:180)
#### Conclusion:
The data presented in this summary of safety and effectiveness is the data that the Food and Drug Administration used in granting DPC substantial equivalence for IMMULITE® Myoglobin.
Edward Heenan
Edward M. Levine, Ph.D. Director of Clinical Affairs
11/6/98
**Date**
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized caduceus with three intertwined snakes and a bowl. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged around the seal's perimeter. The seal is black and white.
NOV 2 5 1998
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Edward M. Levine, Ph.D. Director of Clinical Affairs Diagnostic Products Corporation
5700 West 96th Street Los Angeles, CA 90045-5597
Re: K983970 Trade Name: IMMULITE® Myoglobin Regulatory Class: II Product Code: 82 DDR November 05, 1998 Dated: November 09, 1998 Received:
Dear Dr. Levine:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in In addition, FDA may publish further regulatory action. announcements concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}------------------------------------------------
Page 2
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Putman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radioloqical Health
Enclosure
{6}------------------------------------------------
## 49:39 510(k) Number (if known): Device Name: IMMULITE® Myoglobin
Indications For Use:
IMMULITE® Myoglobin is a solid-phase, two-site chemiluminescent enzyme immunometric assay for use with the IMMULITE Automated Analyzer and designed for the quantitative measurement of myoglobin in serum. It is intended strictly for in vitro use as an aid in the diagnosis of acute myocardial infarction (AMI).
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
for k Mon famang
(Division Sign-Off)
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number A983970
✓
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-
96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.