PEDIATRIC CANNULATED OSTEOTOMY PLATES

K983967 · Biomet, Inc. · HRS · Dec 16, 1998 · Orthopedic

Device Facts

Record IDK983967
Device NamePEDIATRIC CANNULATED OSTEOTOMY PLATES
ApplicantBiomet, Inc.
Product CodeHRS · Orthopedic
Decision DateDec 16, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic, Pediatric

Indications for Use

Pediatric Osteotomy Plates are indicated for fixation of intertrochanteric osteotomies for treating congenital or traumatic deformities in children.

Device Story

Pediatric Osteotomy Plates are metallic bone fixation devices used in orthopedic surgery. Plates are surgically implanted to stabilize bone segments following intertrochanteric osteotomy procedures. Device is used in clinical settings by surgeons. Plates are affixed to bone using standard stainless steel bone screws. System includes specialized manual instrumentation: plate driver/extractors, chisels, goniometer guides, impactors, slide hammers, clamps, and guide wires. By providing rigid fixation, the plates facilitate bone healing and correction of congenital or traumatic deformities in pediatric patients.

Clinical Evidence

Bench testing only.

Technological Characteristics

Constructed from surgical implant grade 316LVM stainless steel conforming to ASTM F-138. Form factor consists of plates sized for infants, children, and adolescents. Fixation achieved via bone screws. Manual instrumentation provided for implantation.

Indications for Use

Indicated for fixation of intertrochanteric osteotomies to treat congenital or traumatic deformities in pediatric patients (infants, children, and adolescents).

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ DEC 16 1299 K983967 ## 510(K) SUMMARY Image /page/0/Picture/3 description: The image shows the logo for Biomet Inc. The logo is in black and white and features the company name in a bold, blocky font. The letters are all capitalized, and the "T" in Biomet is stylized with three horizontal lines. Date: October 29, 1998 CORPORATE HEA Company: Biomet, Inc. P.O. Box 587 Airport Industrial Park Warsaw, Indiana 46581-0587 Device Name: Pediatric Osteotomy Plates Common Name: Osteotomy Plates Classification: Plate, fixation, bone (21 CFR 888.3030) class II Device Description: Pediatric Osteotomy Plates are made from surgical implant grade 316LVM Stainless Steel conforming to the American Society for Testing and Materials ASTM) material standard ASTM F-138. The plates are affixed to bone utilizing currently marketed stainless steel bone screws. The plates are available in sizes suitable for use in infants, children, and adolescents. Special instrumentation is available including plate driver/extractors, chisels, goniometer guides, impactors, slide hammers, clamps, and guide wires. Intended Use: Pediatric Osteotomy Plates are indicated for fixation of intertrochanteric osteotomies for treating congenital or traumatic deformities in children. ## Substantially Equivalent Devices: Smith and Nephew Orthopaedics, Memphis, Tennessee Infant, Child, and Adolescent Osteotomy Blade Plates Zimmer, Inc., Warsaw, Indiana Intertrochanteric Osteotomy Blade-Plates for Children Synthes, Inc., Paoli, Pennsylvania Right Angled Plates for Intertrochanteric Femur Osteotomies in Infants, Children, and Adolescents The substantially equivalent devices are all similar in design and materials. The Biomet manufactured Pediatric Osteotomy plates do not raise any different questions regarding safety and effectiveness from the predicate devices. ## Potential Adverse Effects for Metallic Internal Fixation Devices: Nonunion or delayed healing of bone Implant breakage, bending, or migration Fracture of bone Infection Allergic reaction to foreign body MAILING ADDRESS P.O. Box 587 Warsaw, IN 46581-0587 1 SHIPPING ADDRESS Airport Industrial Park Warsaw, IN 46580 1 000025 OFFICE 219.267.6639 FAX 219.267.8137 1 E-MAIL biomet@biomet.com {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with three stripes representing the three levels of government: federal, state, and local. The eagle is surrounded by a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in capital letters. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DEC 1 6 1928 Mr. Lonnie Witham Senior Regulatory Affairs Specialist Biomet Inc. P.O. Box 587 46581-0587 Warsaw, Indiana K983967 Re: Pediatric Osteotomy Plates Trade Name: Regulatory Class: II Product Code: HRS Dated: October 28, 1998 November 6, 1998 Received: Dear Mr. Witham: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {2}------------------------------------------------ Page 2 - Mr. Lonnie Witham This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note one organism on compitied, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html". Sincerely yours, Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radioloqical Health Enclosure {3}------------------------------------------------ Page 1 of 1 510(k) Number: K983967 Device Name: Pediatric Osteotomy Plates Indications For Use: Pediatric Osteotomy Plates are indicated for fixation of intertrochanteric osteotomies for treating congenital or traumatic deformities in children. `; ## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.) Concurrence of CDRH, Office of Device Evaluation (ODE) × Prescription Use (Per 21 CFR 801.109) Over-The Counter Use (Optional Format 1-2-96) fcośeefo OR (Division Sign-Off) Division of General Restorative Devices 510(k) Number 4983967
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...