# DRIED GRAM-NEGATIVE AND GRAM-POSITIVE MIC/COMBO PANELS (K983922)

_Dade Microscan, Inc. · JWY · Jun 17, 1999 · Microbiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K983922

## Device Facts

- **Applicant:** Dade Microscan, Inc.
- **Product Code:** [JWY](/productcode/JWY.md)
- **Decision Date:** Jun 17, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.1640
- **Device Class:** Class 2
- **Review Panel:** Microbiology

## Indications for Use

To determine gram-negative and gram-positive bacterial susceptibility against the antimicrobial agent Cefdinir.

## Device Story

MicroScan® Dried MIC/Combo Panels determine antimicrobial susceptibility of bacteria to Cefdinir. Panels contain dried antimicrobial agent; inoculated with bacterial samples using Turbidity or Prompt methods. Panels processed via autoScan-4® or WalkAway® instruments. System measures bacterial growth inhibition to determine Minimum Inhibitory Concentration (MIC). Used in clinical laboratories by technicians/microbiologists to guide antibiotic therapy decisions. Provides quantitative susceptibility data to assist clinicians in selecting appropriate antimicrobial treatment for patients with bacterial infections.

## Clinical Evidence

Bench testing only. External evaluations compared dried Cefdinir panels against NCCLS frozen reference panels using fresh/stock efficacy isolates and stock challenge strains. Gram-negative Essential Agreement: 97.9%. Gram-positive Essential Agreement: 94.0%. Reproducibility and precision testing performed using Turbidity and Prompt inoculum methods on autoScan-4® and WalkAway® instruments. Quality control performance acceptable.

## Technological Characteristics

Dried microdilution MIC/Combo panels containing Cefdinir (0.015 - 8 mcg/ml). Compatible with autoScan-4® and WalkAway® automated instruments. Supports Turbidity and Prompt inoculum methods.

## Regulatory Identification

An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

## Predicate Devices

- NCCLS Frozen Cefdinir Reference Panels

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

JUN 17 1999

### Attachment G

## 510(k) Summary

#### 510(k) Submission Information:

| Device Manufacturer: | Dade MicroScan Inc.                                               |
|----------------------|-------------------------------------------------------------------|
| Contact name:        | Cynthia Van Duker, Sr. Regulatory Affairs Specialist              |
| Fax:                 | 916-374-3144                                                      |
| Date prepared:       | November 3, 1998                                                  |
| Product Name:        | Microdilution Minimum Inhibitory Concentration (MIC) Panels       |
| Trade Name:          | MicroScan® Dried Gram-Negative and Gram-Positive MIC/Combo Panels |
| Intended Use:        | To determine antimicrobial agent susceptibility                   |
| 510(k) Notification: | New antimicrobial - Cefdinir                                      |
| Predicate device:    | NCCLS Frozen Cefdinir Reference Panels                            |

#### 510(k) Summarv:

The proposed MicroScan® Dried Gram-Negative and Gram-Positive MIC/Combo Panel with Cefdinir demonstrated substantially equivalent performance when compared with an NCCLS frozen Cefdinir Reference Panel, as defined in the FDA DRAFT document "Review Criteria for Assessment of Antimicrobial Susceptibility Devices" (dated May 31, 1991).

The Premarket Notification (510[k]) presents data in support of the new antimicrobial, Cefdinir, for the MicroScan® Dried Gram-Negative and Gram-Positive MIC/Combo Panels.

Both the gram-negative and gram-positive external evaluations were conducted with fresh and stock Efficacy isolates and stock Challenge strains. The external evaluations were designed to confirm the acceptability of the proposed Dried Cefdinir panels by comparing their performance with an NCCLS frozen Cefdinir Reference panel.

The Dried gram-negative Cefdinir panel demonstrated acceptable performance with an overall Essential Agreement of 97.9% when compared with the frozen Cefdinir Reference panel. The Dried gram-positive Cefdinir panel demonstrated acceptable performance with an overall Essential Agreement of 94.0% when compared with the frozen Cefdinir Reference panel.

Inoculum and instrument reproducibility testing was conducted; both the gram-negative and the gram-positive Dried Cefdinir panels demonstrated acceptable reproducibility and precision, regardless of which inoculum method (i.e., Turbidity and Prompt), or instrument (autoScan-4® and WalkAway®) was used.

Quality Control performance was acceptable for both the gram-negative and the grampositive Dried Cefdinir panels.

The following limitations will be added to the labeling for Cefdinir:

. .

{1}------------------------------------------------

"Do Not Report" with S. pneumoniae and viridans streptococci (MIC and Breakpoint dilutions).

"Do Not Report" with methicillin-resistant, coagulase-negative staphylococci (MIC and Breakpoint dilutions).

"Use Alternate Method" with Providencia spp. (MIC and Breakpoint dilutions).

"The performance of Cefdinir has not been established with Stationary and Log Inoculum methods."

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract image of an eagle or bird with three stylized lines representing its wings or feathers.

Public Health Service

JUN 17 1999

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Sharon K. Cullen Director RA/Compliance Dade MicroScan. Inc. 1584 Enterprise Boulevard West Sacramento. CA 95691

Re: K983922

> Trade Name: MicroScan® Dried Gram-Negative and Gram-Positive MIC/Combo Panels/Cefdinir Regulatory Class: II Product Code: JWY Dated: May 21, 1999 Received: May 24, 1999

Dear Ms. Cullen:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or

{3}------------------------------------------------

### Page 2

Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770)488-7655.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll free number (800) 638-2041 or at (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html"

Sincerely yours,

Steven Sutman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Attachment F

## Indications for Use Statement

| 510(k) No.:          | K983922                                                                                                                                                                                                                                                                                                                                                                                                        |                                                                                                               |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| Device Name:         | MicroScan® Dried Gram-Negative and Gram-Positive<br>MIC/Combo Panels with Cefdinir (0.015 - 8 mcg/ml on the<br>gram-negative and gram-positive panels)                                                                                                                                                                                                                                                         |                                                                                                               |
| Indications for Use: | To determine gram-negative and gram-positive bacterial susceptibility against the antimicrobial agent Cefdinir.                                                                                                                                                                                                                                                                                                |                                                                                                               |
|                      | Organisms with indications for testing* include:                                                                                                                                                                                                                                                                                                                                                               |                                                                                                               |
|                      | Gram-Negative Bacteria                                                                                                                                                                                                                                                                                                                                                                                         | Gram-Positive Bacteria                                                                                        |
|                      | None**                                                                                                                                                                                                                                                                                                                                                                                                         | <i>Staphylococcus aureus</i> ***<br>(Including methicillin-<br>susceptible, β-lactamase<br>producing strains) |
|                      |                                                                                                                                                                                                                                                                                                                                                                                                                | <i>Streptococcus pyogenes</i>                                                                                 |
|                      | *As taken from the indications and Usage section of the<br>manufacturer's package insert (Issued: March 1998).                                                                                                                                                                                                                                                                                                 |                                                                                                               |
|                      | **No gram-negative organisms for which Microscan is seeking<br>clearance are indicated for testing in the Indications and Usage<br>section of the manufacturer's package insert.                                                                                                                                                                                                                               |                                                                                                               |
|                      | ***Cefdinir is inactive against methicillin-resistant <i>Staphylococcus</i>                                                                                                                                                                                                                                                                                                                                    |                                                                                                               |
|                      | The following secondary organisms included for the testing of<br>MicroScan® panels with Cefdinir have in vitro data but the safety<br>and effectiveness in treating clinical infections have not been<br>established: <i>Citrobacter (diversus) koseri, Escherichia coli,<br/>Klebsiella pneumoniae, Proteus mirabilis, Staphylococcus<br/>epidermidis (methicillin-susceptible), Streptococcus agalactiae</i> |                                                                                                               |

:

{5}------------------------------------------------

The MicroScan® Dried Gram-Negative and Gram-Positive The MicroScall - Dried Crain 1105 --------------------------------------------------------------------------------------------------------------------------------------------MIC/Combo I and viridans streptococci.

Woody Dubois

Division of Clinical Laboratory Devices 510(k) Number_K 98 3922

PRESCRIPTION USE X

---

**Source:** [https://fda-staging.innolitics.com/device/K983922](https://fda-staging.innolitics.com/device/K983922)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com