K983917 · Theratronics, Inc. · IWB · Dec 4, 1998 · Radiology
Device Facts
Record ID
K983917
Device Name
THERATRON ELITE
Applicant
Theratronics, Inc.
Product Code
IWB · Radiology
Decision Date
Dec 4, 1998
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.5750
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The unit is intended to be used in: - delivering the intended dose at a specified position; . - delivering the radiation in accordance with the selected relationship of the . radiation to the patient (fixed or moving beam therapy, beam modifying device, etc.); - . delivering the radiation without causing unnecessary risk to the patient, the operator, other persons, or the immediate environment.
Device Story
Theratron Elite is a cobalt teletherapy device comprising a source head, collimator, gantry, main frame, base, and counterweight. It functions by delivering radiation beams to patients for therapy. The primary evolution from predicate devices (Theratron 780E/1000E) is the replacement of push-button controls with a PC-based remote control console featuring a graphical user interface and 'Record and Verify' communication capability. Operated by clinical staff in a radiation therapy environment, the device allows for precise beam positioning and dose delivery. The PC interface facilitates treatment planning and verification, enhancing operational control and safety. The device benefits patients by providing controlled, accurate radiation therapy while minimizing risks to the patient, operator, and environment.
Clinical Evidence
Bench testing only. Validation demonstrated compliance with product development procedures and confirmed the device meets published specifications and performs equivalently to predicate devices.
Technological Characteristics
Radionuclide radiation therapy system. Components: source head, collimator, gantry, main frame, base, counterweight. Control system: PC-based remote console with graphical user interface and 'Record and Verify' communication. Energy source: Cobalt radiation source.
Indications for Use
Indicated for patients requiring radiation therapy via fixed or moving beam delivery. Used by clinicians for dose delivery at specified positions.
Regulatory Classification
Identification
A radionuclide radiation therapy system is a device intended to permit an operator to administer gamma radiation therapy, with the radiation source located at a distance from the patient's body. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts (including beam-limiting devices), and accessories.
{0}------------------------------------------------
Theratron Elite 510(k) Summary
| Submitter: | Theratronics International Limited<br>413 March Road, P.O. Box 13140<br>Kanata, Ontario<br>Canada K2K 2B7 |
|--------------------|-----------------------------------------------------------------------------------------------------------|
| Phone: | 613-591-2100 |
| Fax: | 613-592-3816 |
| Contact person: | E. S. Martell, Vice President Quality & Regulatory Affairs |
| Date: | Nov 16, 1998 |
| Trade name: | Theratron Elite |
| Common name: | Cobalt Teletherapy device |
| Classif. name: | Radionuclide Radiation Therapy System |
| Equivalent device: | Theratron 780E (K964606), Theratron 1000E (K964607) |
# Description of device:
The Theratron Elite consists of a source head, collimator, gantry, main frame, base, controls and a pendulum or beam-stopper style counterweight. The design of this device is similar to predicate devices Theratron 780E and 1000E. The key difference is that the 'push-button' style control console on the predicate devices has been replaced with a modern PC based remote control console with graphic user interface and 'Record and Verify' communication capability.
# Intended use of device:
The unit is intended to be used in:
- delivering the intended dose at a specified position; .
- delivering the radiation in accordance with the selected relationship of the . radiation to the patient (fixed or moving beam therapy, beam modifying device, etc.);
- . delivering the radiation without causing unnecessary risk to the patient, the operator, other persons, or the immediate environment.
# Summary of comparison to predicate device.
See Section 2.
# Summary of conclusions drawn from nonclinical tests
The Theratron Elite development and validation is in compliance with Theratronics product development procedures. The validation demonstrates that the device meets its published specifications, performs as well or better than the predicate device to which it is substantially equivalent, and is safe and effective for its intended use.
{1}------------------------------------------------
Image /page/1/Picture/16 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of a human figure with three heads, arranged in a vertical stack. The figure is positioned to the right of the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA", which is arranged in a circular pattern around the figure.
1000 4 DEC
E.S. Martell Vice President Quality and Regulatory Affairs Theratronics International, Ltd. 413 March Road Kanata, Ontario CANADA
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Re: K983917 Theratron Elite 80 and 100 Dated: October 29, 1998 Received: November 4, 1998 Regulatory class: II Procode: 90 IWB CFR 892.5750
Dear Mr. Martell:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4591 for Radiology devices, or 594-4613 for Ear, Nose and Throat devices. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or at (301) 443-6597.
Sincerely yours,
Willian Yin
Lillian Yin, Ph.D.
Director. Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{2}------------------------------------------------
## Intended use of device 3.
The device is intended to be used in:
- Delivering the intended dose of radiation at a specified position; .
- Delivering the radiation in accordance with the selected relationship of the . radiation to the patient (fixed or moving beam therapy, beam modifying device, etc);
- Delivering the radiation without causing unnecessary risk to the patient, the . operator, other persons or the immediate environment.
Vinid C. Flynn
sion Sign-Off Division of Reproductive, Abdominal, ENT, and Radiological Dev 510(k) Number
**Prescription Use**
(Per 21 CFR 801.109)***_***_ and the image.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.