← Product Code [DDG](/productcode/DDG) · K983880

# TRANSFERRIN SPQ III ANTIBODY REAGENT SET (K983880)

_DiaSorin, Inc. · DDG · Dec 18, 1998 · Immunology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K983880

## Device Facts

- **Applicant:** DiaSorin, Inc.
- **Product Code:** [DDG](/productcode/DDG.md)
- **Decision Date:** Dec 18, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 866.5880
- **Device Class:** Class 2
- **Review Panel:** Immunology

## Indications for Use

FOR IN VITRO DIAGNOSTIC USE. The Transferrin SPQ™ III ANTIBODY REAGENT SET is designed for the quantitative determination of transferrin in human serum by immunoprecipitin analysis.

## Device Story

The Transferrin SPQ™ III Antibody Reagent Set is an in vitro diagnostic kit used for the quantitative measurement of transferrin levels in human serum. The device utilizes immunoprecipitin analysis to detect and quantify transferrin. It is intended for use by laboratory professionals in a clinical setting. The modification from the predicate device involves the use of a higher titer primary antibody and the elimination of the patient sample diluent. The output is a quantitative concentration value of transferrin, which assists clinicians in assessing patient iron metabolism and nutritional status.

## Clinical Evidence

Clinical testing demonstrated that the performance characteristics of the modified device were substantially equivalent to those of the predicate device.

## Technological Characteristics

Immunoprecipitin analysis reagent set. Modification includes higher titer primary antibody and removal of patient sample diluent. In vitro diagnostic use.

## Regulatory Identification

A transferrin immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the transferrin (an iron-binding and transporting serum protein) in serum, plasma, and other body fluids. Measurement of transferrin levels aids in the diagnosis of malnutrition, acute inflammation, infection, and red blood cell disorders, such as iron deficiency anemia.

## Predicate Devices

- SPQ™ Antibody Reagent Set II for Transferrin

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## DEC 18 1998

# 11983880

### Special 510(k): Device Modification: Transferrin SPQ™ III Antibody Reagent Set

## 10 510(k) Summary

October 30, 1998

. .

| Submitted By:                  | Judith J. Smith<br>DiaSorin, Inc.<br>9175 Guilford Rd. Suite 100<br>Columbia, MD 21046                                                                                                                                                        |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name Of Device:<br>Trade Name: | Transferrin SPQ™ III Antibody Reagent Set                                                                                                                                                                                                     |
| Common Name:                   | Transferrin (TRF) Immunological Test Kit                                                                                                                                                                                                      |
| Classification Name:           | Transferrin, antigen, antiserum, Control (82 DDG)                                                                                                                                                                                             |
| Device Classification          | Class II                                                                                                                                                                                                                                      |
| Predicate Device:              | SPQ™ Antibody Reagent Set II for Transferrin                                                                                                                                                                                                  |
| Device Description:            | Transferrin (TRF) Immunological Test Kit                                                                                                                                                                                                      |
| Intended Use:                  | FOR IN VITRO DIAGNOSTIC USE.<br>The Transferrin SPQ™ III ANTIBODY REAGENT SET is designed<br>for the quantitative determination of transferrin in human serum by<br>immunoprecipitin analysis.                                                |
| Technological<br>Comparison:   | The modified device has the same technological basis as the<br>predicate device.                                                                                                                                                              |
| Labeling Comparison:           | The labeling of the modified assay is substantially equivalent to that<br>of the predicate device. Changes in labeling directly reflect the<br>device modification (higher titer primary antibody, elimination of<br>patient sample diluent). |
| Nonclinical Testing:           | Nonclinical testing demonstrated that the performance<br>characteristics of the modified device were substantially equivalent<br>to those of the predicate device.                                                                            |
| Clinical Testing:              | Clinical testing demonstrated that the performance characteristics<br>of the modified device were substantially equivalent to those of the<br>predicate device.                                                                               |
| Conclusions from<br>Testing:   | The modified device is substantially equivalent to the predicate<br>device.                                                                                                                                                                   |

{1}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Image /page/1/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of three human profiles facing right, stacked on top of each other.

DEC 1 8 1998 Ms. Judith J. Smith Corporate Director Worldwide Requlatory Affairs And Ouality Systems DiaSorin, Inc. 9175 Guilford Road, Suite 100 Columbia, Maryland 21046

Re: K983880 Trade Name: Transferrin SPQ™ III Antibody Reagent Set Requlatory Class: II Product Code: DDG Dated: October 30, 1998 November 2, 1998 Received:

Dear Ms. Smith:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 A substantially equivalent determination assumes compliance to 895. with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

Food and Drug Administration 2098 Gaither Road Rockville MD 20850

{2}------------------------------------------------

#### Page 2

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diaqnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.qov/cdrh/dsma/dsmamain.html".

Sincerely yours,

Steven Putman

Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

#### Indications for Use 3

510(k) Number: _ K 983880

Device Name: Transferrin SPQ™ III Antibody Reagent Set

Indications For Use:

### FOR IN VITRO DIAGNOSTIC USE.

The Transferrin SPQTM III ANTIBODY REAGENT SET is designed for the quantitative determination of transferrin in human serum by immunoprecipitin analysis.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

| (Division Sign-Off)                     |         |
|-----------------------------------------|---------|
| Division of Clinical Laboratory Devices |         |
| 510(k) Number                           | f983880 |

| Prescription Use<br>(Per 21 CFR 801.109) | OR | Over-The-Counter Use |
|------------------------------------------|----|----------------------|
|------------------------------------------|----|----------------------|

(Optional Format 1-2-96)

---

**Source:** [https://fda-staging.innolitics.com/device/K983880](https://fda-staging.innolitics.com/device/K983880)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
