The Acticoat® Composite Wound Dressing is an effective barrier to bacterial penetration. The barrier function of the dressing may help reduce infection in partial and full thickness wounds including decubitus ulcers, venous stasis ulcers, diabetic ulcers, first and second degree burns, donor sites, and surgical wounds. Acticoat dressings may be used over debrided and grafted partial thickness wounds
Device Story
Acticoat Composite Wound Dressing is a 3-ply sterile dressing; absorbent rayon core; upper polyurethane film layer; lower silver-coated high-density polyethylene (HDPE) mesh. Silver coating consists of silver and oxygen alloy. Moisture activates dressing, releasing silver ions into/around wound site to provide antimicrobial barrier. Used in clinical settings for wound management; applied by healthcare professionals; secured with tape, gauze, or elastic bandages. Dressing acts as physical barrier to bacterial penetration; helps reduce infection risk. Benefits include antimicrobial protection, moisture management, and cushioning. Performance characteristics similar to untreated gauze regarding absorptivity, moisture content, abrasion resistance, adhesion, and tensile strength.
Clinical Evidence
Bench testing only. Performance data includes silver dissolution, absorptivity, moisture content, drop penetration, vapour transmission, tensile strength, biocompatibility (skin irritation, sensitization, cytotoxicity), silver ion exposure levels, and in vitro antimicrobial activity studies.
Technological Characteristics
3-ply dressing: absorbent rayon core, polyurethane film top layer, silver-coated high-density polyethylene (HDPE) mesh bottom layer. Silver coating is silver/oxygen alloy. Gas permeable, soft, pliable, cushioning. Sterilized via gamma irradiation. No software or electronic components.
Indications for Use
Indicated for patients with partial and full thickness wounds, including decubitus, venous stasis, and diabetic ulcers, first and second degree burns, donor sites, and surgical wounds. May be used over debrided and grafted partial thickness wounds. Contraindicated for third degree burns.
Regulatory Classification
Identification
A liquid bandage is a sterile device that is a liquid, semiliquid, or powder and liquid combination used to cover an opening in the skin or as a dressing for burns. The device is also used as a topical skin protectant.
{0}------------------------------------------------
K983833
# ATTACHMENT 8
# 510(K) SUMMARY FOR THE WESTAIM CORPORATION'S ACTICOAT® COMPOSITE WOUND DRESSING
# Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
Westaim Biomedical Division The Westaim Corporation 10102 - 114 Street Fort Saskatchewan, Alberta T8L 3W4 CANADA
Phone: (403) 992-5380 Facsimile: (403) 992-5325
Contact Person: Robert Burrell, Ph.D. Phone: (403) 992-5380 Facsimile: (403) 992-5325
> Jonathan S. Kahan, Esq. Regulatory Counsel for The Westaim Corporation Hogan & Hartson, L.L.P. 555 Thirteenth Street NW Washington, DC 20004-1109 Phone: (202) 637-5794 Facsimile: (202) 637-5910
Date Prepared: March 1, 1998
{1}------------------------------------------------
# Name and Device and Name/Address of Sponsor
Acticoat® Composite Wound Dressing
### Sponsor
Westaim Biomedical Division The Westaim Corporation 10102 - 114 Street Fort Saskatchewan, Alberta T8L 3W4 CANADA
Phone: (403) 992-5380 Facsimile: (403) 992-5325
Contact Person: Robert Burrell, Ph.D. Phone: (403) 992-5380 Facsimile: (403) 992-5325
# Common or Usual Name
Composite wound dressing
## Classification Name
Wound dressing
# Predicate Devices
The Company's Acticoat® Composite Wound Dressing, covered by this submission, is substantially equivalent to other legally marketed wound dressings. Specifically, Actional Composite Wound Dressing is substantially equivalent The Westaim Corporation's Acticoat® Antimicrobial Barrier Dressing (K955453) and Bioderm's Composite Wound Dressing. The wound contact layer is also substantially equivalent to the Conformant 28 Wound Dressing.
### WESTAIM
{2}------------------------------------------------
# Intended Use
The Acticoat® Composite Wound Dressing is an effective barrier to bacterial penetration. The barrier function of the dressing may help reduce infection in partial and full thickness wounds including decubitus ulcers, venous stasis ulcers, diabetic ulcers, first and second degree burns, donor sites, and surgical wounds. Acticoat dressings may be used over debrided and grafted partial thickness wounds
# Technological Characteristics and Substantial Equivalence
#### Description of the Acticoat® Composite Wound Dressing A.
Acticoat® Composite Wound Dressing is a 3-ply dressing consisting of an absorbent rayon core with an upper layer of polyurethane film and a lower layer of silvercoated high density polyethylene mesh designed to be a barrier against microbial infections of a wound. The silver in the coating is an alloy of silver and oxygen. The coating is highly porous, and the film has enhanced solubility in water-based fluids. Additionally, the performance characteristics of the Acticoat® Composite Wound Dressing are similar to those found in untreated gauze-based dressings in the areas of absorptivity and moisture content. abrasion resistance, adhesion, and tensile strength.
#### Substantial Equivalence of the Acticoat® Composite Wound Dressing B.
The predicate devices also provide the same or similar functions, characteristics, and accessories as described above for the Acticoat® Composite Wound Dressing. The dressings have the same intended use and are recommended for the same indications (e.g., ulcers, trauma wounds, surgical wounds, abrasion, lacerations, and donor sites).
Like the subject composite wound dressing the Acticoat " Antimicrobial Barrier Dressing consists of an absorbent material and the dressing releases silver ions into and around the wound site when activated by moisture. Both products are provided sterile to the user, and both have an antimicrobial effect. Additionally, both dressings are non-toxic, nonirritating, and non-sensitizing. Similar to the subject device, the silver coated dressing is gas permeable and has a soft, pliable, cushioning texture.
Both the subject dressing and its predicates are effective over similar periods of time (i.e., between dressing changes). Additionally, the dressings are secured to the wound site via traditional dressing methods such as tape, gauze, or elastic bandage. The predicate dressings, like the composite wound dressings, also have been shown to have low adhesivity to the wound site.
## AWESTAIM
{3}------------------------------------------------
Although there are some differences between the Acticoat® Composite Wound Dressing and its predicates, these differences are minor and raise no new questions of safety or effectiveness. First, the Bioderm device is described as a similar 3-ply gauze bandage without the silver coating on the lower layer of HDPE. Similarly, the Acticoat® Antimicrobial Barrier Dressing is described as a device similar to the subject dressing but lacks the occlusive polyurethane film and is not as absorptive. The labelling for all of the devices indicate that the frequency of dressing changes should be determined, in part, by the amount of wound exudate and general condition of the predicate silver coated dressing is also sterilized by gamma irradiation as is the subject composite wound dressing. These few differences in technological characteristics between the Acticoat® Composite Wound Dressing and its predicate devices are minor and do not present any new issues of safety or effectiveness.
# Performance Data
The Acticoat® Composite Coated Dressing was subjected to the following performance tests:
- Silver dissolution ●
- Absorptivity and moisture content .
- Drop penetration and vapour transmission .
- Tensile strength
- Biocompatibility studies (including skin irritation, sensitization and . cytotoxicity)
- Silver ion exposure levels
- In vitro studies of antimicrobial activity .
In all instances, the Acticoat Composite Wound Dressing is both effective for its intended use and functions in a substantially equivalent manner to the predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing the body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 1 0 1998
Westaim Biomedical, Inc. c/o Jonathan S. Kahan Regulatory Counsel for The Westaim Corporation Hogan and Hartson L.L.P. 555 Thirteenth Street NW Washington, DC 20004
Re: K983833
Trade Name: Acticoat® Composite Wound Dressing Regulatory Class: Unclassified Product Code: KMF Dated: October 30, 1998 Received: October 30, 1998
Dear Mr. Kahan:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act). You may, therefore, market your device subject to the general controls provisions of the Federal Food, Drug, and Cosmetic Act (Act) and the following limitations:
- This device may not be labeled for use on third degree burns. 1.
- This device may not be labeled as having any accelerating effect on the rate 2. of wound healing or epithelization.
- 3. This device may not be labeled as a long-term, permanent, or no-change dressing, or as an artificial (synthetic) skin.
- This device may not be labeled as a treatment or a cure for any type of 4. wound.
The labeling claims listed above would be considered a major modification in the intended use of the device and would require a premarket notification submission (21 CFR 807.81).The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practices, labeling, and prohibitions against misbranding and adulteration.
{5}------------------------------------------------
Page 2 - Mr. Jonathan S. Kahan
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval) it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations (CFR), Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practices (GMP) for Medical Devices: General GMP regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition. FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or 301-443-6597 or at its internet address http://www.fda.gov/cdrh/dsmamain.html.
Sincerely yours,
M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
iK983833
HOGAN & HARTSON LL.P.
510(k) Number (if known): __ K983833
Device Name: Acticoat@ Composite Wound Dressing
Indications For Use:
The Acticoat@ Composite Wound Dressing is an effective barrier to bacterial penetration. The barrier function of the dressing may help reduce infection in partial and full thickness wounds including decubitus ulcers, venous stasis ulcers, diabetic ulcers, first and second degree burns, donor sites, and surgical wounds. Acticoat® dressings may be used over debrided and grafted partial thickness wounds.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) Over-The-Counter OR Prescription Use Use (Per 21 CFR 801.109) (Optional Format 1-2-96)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.