These devices are used to introduce or remove gametres, zygotes(s), preembryo)s) and/or embryo(s) into or from the body.
Device Story
Prodimed Microinjection Pipettes are specialized glass tools used in assisted reproduction. The device consists of holding and injection pipettes manufactured from borosilicate glass capillary tubes. The injection pipette features a beveled tip, optionally with a sharp spike, to facilitate non-traumatic penetration of the oocyte during ICSI. The holding pipette is designed to secure the oocyte during manipulation. These tools are used in a clinical laboratory setting by embryologists or clinicians under a microscope. The operator manually manipulates the pipettes using an instrument stand to control gamete movement. The device enables precise handling of gametes, zygotes, and embryos, directly supporting the fertilization process in assisted reproductive technology.
Clinical Evidence
No clinical data provided. Bench testing only, including verification of physical dimensions, material composition, and manufacturing standards (ISO 9001, EN 46001).
Technological Characteristics
Materials: Borosilicate glass (Boron Oxide, Sodium Oxide). Dimensions: 78mm length, 0.69mm ID/0.97mm OD capillary tubes; finished tips 50µm/20µm (holding) and 7µm/5µm (injection). Principle: Manual mechanical manipulation of gametes via micro-injection/holding. Sterilization: Not specified. Connectivity: None. Software: None.
Indications for Use
Indicated for I.C.S.I. (Intra cytoplasmic sperm injection) procedures.
Regulatory Classification
Identification
Assisted reproduction microtools are pipettes or other devices used in the laboratory to denude, micromanipulate, hold, or transfer human gametes or embryos for assisted hatching, intracytoplasmic sperm injection (ICSI), or other assisted reproduction methods.
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, and clinical testing). The device, when the assisted reproduction microtools (pipettes) are manufactured from glass, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.
Submission Summary (Full Text)
{0}------------------------------------------------
#### 8 1999 JUN
K983713
### 510(K) Summary
February 23, 1999
1) PRODIMED Laboratories c/o Washington Regulatory Services Randolph L. Cooke 23 Welisewitz Road Ringoes, NJ 08551 609-466-0510 Fax 609-466-4443
- 2) DEVICE NAME:
Proprietary Names: Prodimed Microinjection Pipettes
Common Name: Assisted Reproduction Microtools
Classification Name: Assisted Reproduction catheters are classified as Class II at 21 CFR 884.6130. The product code is 85 MQH.
3) IDENTIFICATION: The Assisted Reproduction microtool devices have been down classified FR Vol. 63, No. 175 September 10, 1998. Effective date: October 13, 1998.
## 4) DESCRIPTION OF DEVICE:
As found in the labeling/promotional material:
Quality and Security:
- Permanence and reproducibility of production .
- Manufactured according to ISO 9001 and EN 46001 standards .
- . Routine performance of tests with revelation of biological indicators, search for endotoxins and embryotoxicity tests
Performance: (Function)
- Clear view and full control of the sperm for precision manipulation .
- . Bevel equipped with a short, sharp spike for easy, non-traumatising penetration of the oocyte
- . Effective prehension by the holding pipettes to avoid rotation of the oocyte during the micro-injection
Ergonomics:
- Totally safe transport of pipettes thanks to a patented protective stand .
- . Easy, safe handling by simply pressing the base of the instrument between the thumb and index finger
00:
{1}------------------------------------------------
Wide Range:
- Standard availability of pipettes with spike and 8um opening, straight or curved at . 20° and 30°
- Availability on order of spiked or non-spiked pipettes, opening of 6um, or 8um, . straight or curved at 20° and 30°
- Wide choice suitable for every type of installation on all microscopes and . instrument stands
### UNILAM:
- Ready-to-use and easy to handle device ●
- Provides rapid location of the oocyte .
- . Guaranteed sterility of pipettes and perfect visibility
- Adapted to heating plates in order to keep the most convenient ternperature for ● gamete survival
Function of Device/Significant Physical Characteristics of Device:
The holding and injection pipettes were made from 30µl borosilicate glass capillary tubes, 78mm in length and with inner and outer diameters of 0.69 and 0.97mm respectively. The glass pipettes were obtained by drawing thin-walled glass capillary tubes using a horizontal microelectrode puller. The holding pipette was cut and firepolished on a microforge to obtain an outer diamter of 50um and an inner diameter of 20um. To prepare the injection pipette, the pulled capillary was opened on a microgriner to an outer diameter of 7um and an inner diameter of 5um; the bevel angle was 45°. This grinding step required ~3 minutes. The whetstone of the grinder was humidified by a slow water drip during the procedure. The microforge was then used to make a sharp spike on the injection piepette and to bend the edge of the holding and injection pipettes to an angle of ~35° to facilitate the injection procedure in the Petri dish.
Device Design/ Material Used/ Physical Properties:
Micro-Injection Pipettes: N51-A Device Part Borasilicate Glass
Material Used Borone Oxide Sodium Oxide
## 5) STATEMENT OF INTENDED USED
These devices are used to introduce or remove gametres, zygotes(s), preembryo)s) and/or embryo(s) into or from the body.
## MICRO-INJECTION PIPETTES - Box of 10 units
13810008 Opening 8um with Spike Straight
{2}------------------------------------------------
| 13812008 | Opening 8μm with Spike Angle 20° |
|----------|----------------------------------|
|----------|----------------------------------|
- Opening 8um with Spike Angle 30° 13813008
## HOLDING PIPETTES - Box of 10 units
| 13900015 | Opening 15μm Straight |
|----------|------------------------|
| 13902015 | Opening 15μm Angle 20° |
| 13903015 | Opening 15μm Angle 30° |
| 1303200 | UNILAM |
# SPECIAL MANUFACTURING
## MICOR-INJECTION PIPETTES - Box of 10 units Delivery delay of 8 weeks after confirmation order
| 13800008 | Opening 8µm without Spike Straight |
|----------|-------------------------------------|
| 13802008 | Opening 8µm without Spike Angle 20° |
| 13803008 | Opening 8µm without Spike Angle 30° |
| 13810006 | Opening 6µm with Spike Straight |
| 13812006 | Opening 6µm with Spike Angle 20° |
| 13813006 | Opening 6µm with Spike Angle 30° |
| 13800006 | Opening 6µm without Spike Straight |
| 13802006 | Opening 6µm without Spike Angle 20° |
| 13803006 | Opening 6µm without Spike Angle 30° |
## 6) TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE:
Design Material:
の --------------------------------------------------------------------------------------------------------------------------------------------------------------------------
ICSI Injection: Intracytoplasmic injection micropipettes are prepared from borosilicate glass tubing (OD: 1mm ID: 0,75m)
ICSI Holding: Intracytoplasmic injection micropipettes are prepared from borosilicate glass tubing (OD: 1mm ID: 0,75m)
| Chemical Composition | | |
|----------------------|--------------|---------------------------|
| Borasilicate Glass: | Boron Oxide | ACGIH TLV 10mg/m3 |
| | | OSHA PEL 15mg/m3 |
| | Sodium Oxide | ACGIH TLV Not Recommended |
| | | OSHA PEL Not Regulated |
{3}------------------------------------------------
Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized image of an eagle with three overlapping wing shapes, suggesting movement or flight.
JUN 8 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
PRODIMED Laboratories c/o Washington Regulatory Services, Inc. Mr. Randolph L. Cooke 23 Welisewitz Road Ringoes, N.J. 08551
Re: K983713
Prodimed Microinjection Pipettes Dated: March 11, 1999 Received: March 15, 1999 Regulatory Class: II 21 CFR 884.6130/Procode: 85 MQH
Dear Mr. Cooke:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls r outs an ook of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the requiration entitled. "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
CART Daniel C. Schultz, M.D.
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### 510(k) NUMBER (IF KNOWN)____________________________________________________________________________________
#### DEVICE NAME: Micro-Injection Pipettes
#### INDICATIONS FOR USE:
These devices are used for I.C.S.I. (Intra cytoplasmic sperm injection)
## MICRO-INJECTION PIPETTES - Box of 10 units
| 13810008 | Opening 8µm with Spike Straight |
|----------|----------------------------------|
| 13812008 | Opening 8µm with Spike Angle 20° |
| 13813008 | Opening 8µm with Spike Angle 30° |
### HOLDING PIPETTES - Box of 10 units
| 13900015 | Opening 15µm Straight |
|----------------|------------------------|
| 13902015 | Opening 15µm Angle 20° |
| 13903015 | Opening 15µm Angle 30° |
| 1303200 UNILAM | |
#### SPECIAL MANUFACTURING MICOR JINIT CITION PIPETTES - Box of 10 units Delivery delay of 8 weeks after confirmation order
| 13800008 | Opening 8µm without Spike Straight |
|----------|-------------------------------------|
| 13802008 | Opening 8µm without Spike Angle 20° |
| 13803008 | Opening 8µm without Spike Angle 30° |
| 13810006 | Opening 6µm with Spike Straight |
| 13812006 | Opening 6µm with Spike Angle 20° |
| 13813006 | Opening 6µm with Spike Angle 30° |
| 13800006 | Opening 6µm without Spike Straight |
| 13802006 | Opening 6µm without Spike Angle 20° |
| 13803006 | Opening 6µm without Spike Angle 30° |
#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use Per 21 CFE 801.109 OR
Over-The Counter-Use _________________________________________________________________________________________________________________________________________________________ (Options Format 1-2-96)
Ra-Pales
(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological De 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.