K983701 · Roche Diagnostic Systems, Inc. · LDJ · Dec 11, 1998 · Clinical Toxicology
Device Facts
Record ID
K983701
Device Name
MODIFICATION TO ABUSCREEN ONLINE FOR CANNABINOIDS
Applicant
Roche Diagnostic Systems, Inc.
Product Code
LDJ · Clinical Toxicology
Decision Date
Dec 11, 1998
Decision
SESE
Submission Type
Special
Regulation
21 CFR 862.3870
Device Class
Class 2
Indications for Use
Abuscreen ONLINE for Cannabinoids is an in vitro diagnostic test for the qualitative and semiquantitative detection of cannabinoids in human urine on automated clinical chemistry analyzers at cutoff concentrations of 20 ng/mL, 50 ng/mL, and 100 ng/mL. Semiquantitative test results may be obtained that permit laboratories to assess assay performance as part of a quality control program. Measurements obtained by this device are used in the diagnosis of cannabinoid use or abuse.
Device Story
Abuscreen ONLINE Cannabinoids is an in vitro diagnostic assay for detecting cannabinoids in human urine. It utilizes kinetic interaction of microparticles in a solution, measured by changes in light transmission. The device is designed for use on automated clinical chemistry analyzers, specifically the Hitachi 917. The assay reagents include a monoclonal antibody (mouse) reagent, a conjugated cannabinoid derivative microparticle reagent, and a buffer diluent. The system automates sample processing, photometric measurement, and result transmission. Healthcare providers use the qualitative or semiquantitative results to assist in the diagnosis of cannabinoid use or abuse. The device provides laboratories with the ability to assess assay performance via semiquantitative results as part of a quality control program.
Clinical Evidence
Bench testing only. Precision studies (within-run and day-to-day) were conducted at 20, 50, and 100 ng/mL cutoffs, reporting mean OD and CV% values. Accuracy was evaluated by comparing results against confirmed positive samples (N=69 for proposed device vs N=50 for predicate). Limit of detection was reported as 7 ng/mL.
Technological Characteristics
Methodology: Kinetic interaction of microparticles in solution measured by light transmission. Reagents: Mouse monoclonal antibody, conjugated cannabinoid derivative microparticles, and buffer diluent. Platform: Automated clinical chemistry analyzer (Hitachi 917).
Indications for Use
Indicated for qualitative and semiquantitative detection of cannabinoids and metabolites in human urine for diagnosis of cannabinoid use or abuse. Intended for use on automated clinical chemistry analyzers (specifically Hitachi 917) at 20, 50, and 100 ng/mL cutoffs.
Regulatory Classification
Identification
A cannabinoid test system is a device intended to measure any of the cannabinoids, hallucinogenic compounds endogenous to marihuana, in serum, plasma, saliva, and urine. Cannabinoid compounds include delta-9-tetrahydrocannabinol, cannabidiol, cannabinol, and cannabichromene. Measurements obtained by this device are used in the diagnosis and treatment of cannabinoid use or abuse and in monitoring levels of cannabinoids during clinical investigational use.
Special Controls
*Classification.* Class II (special controls). A cannabinoid test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
{0}------------------------------------------------
# Roche
# Diagnos
# 510(k) Summary
# Abuscreen ONLINE® Cannabinoids
In accordance with the Safe Medical Devices Act of 1990, a 510(k) summary is provided as outlined in 21 CFR 807.92.
The assigned 510(k) number is:
### Identification of 510(k) Sponsor: I.
Roche Diagnostic Systems, Inc. a subsidiary of Hoffmann-La Roche, Inc. Branchburg Township 1080 U.S. Highway 202 Somerville, New Jersey 08876-3771
510(k) Submission dated October 20, 1998
Rita Smith Contact: Senior Regulatory Affairs Associate Phone: (908) 253-7545 (908) 253-7547 Fax:
Branchburg Township 1080 U.S. Highway 202 Somerville, New Jersey 08876-3771
{1}------------------------------------------------
### II. Device Name:
The device name, including both the trade/proprietary name and the classification name are provided in the table below.
| C<br>1<br>abl<br>1<br>le<br>1 |
|-------------------------------|
|-------------------------------|
| Product Name | Classification<br>Name | Product<br>Code | CFR<br>Number and<br>Regulatory Class |
|--------------------------------------|-------------------------------------|-----------------|---------------------------------------|
| Abuscreen ONLINE for<br>Cannabinoids | Enzyme Immunoassay,<br>Cannabinoids | LDJ | 862.3870<br>Class II |
### Identification of the legally marketed device to which the 510(k) sponsor claims III. equivalence:
The following table identifies the legally marketed devices to which Roche Diagnostic Systems, Inc. claims equivalence.
# Table 2
| Product Name | Predicate Product Name | 510(k) Number and Date<br>Predicate Cleared |
|--------------------------------------|--------------------------------------|---------------------------------------------|
| Abuscreen ONLINE for<br>Cannabinoids | Abuscreen ONLINE for<br>Cannabinoids | K981585 5/28/98 |
{2}------------------------------------------------
### IV. Description of the Device/Statement of Intended Use:
Abuscreen ONLINE Cannabinoids is an in vitro diagnostic test for the qualitative and semiquantitative detection of cannabinoids and its metabolites in human urine on automated clinical chemistry analyzers at cutoff concentrations of 20 ng/mL, 50 ng/mL, and 100 ng/mL. Semiquantitative test results may be obtained that permit laboratories to assess assay performance as part of a quality control program. Measurements obtained by this device are used in the diagnosis of cannabinoid use or abuse.
The proposed Abuscreen ONLINE Cannabinoids test kit is specifically intended for use on the Hitachi 917 Analyzer and future similar analyzer models. It was adapted from the currently marketed Abuscreen ONLINE Cannabinoids test kit. The labeling and packaging have been changed for use on the Hitachi 917 Analyzer as well as an addition of a surfactant to the diluent. This modified test kit is not a replacement to the currently marketed kit.
The Hitachi 917 Analyzer System is a fully automatic, computer-controlled system for clinical chemistry. It was conceived for both quantitative and qualitative in vitro determination using a large variety of tests for analysis, e.g. in serum and urine. Integrated in the system is an ion-selective unit for determination of electrolytes. The throughout per hour is 800 tests for clinical chemistry (1200 with electrolytes). The system consists of the analyzer which performs all functions required for fully automatic sample and test processing. Beginning with the automatic recording of patient samples - provided that they are supplied in barcode-labeled vessels - up to the photometric measurement and results transmission to the computer unit. Additional detailed information about the Hitachi 917 Analyzer is contained in volume II of the premarket notification (K953239) cleared on September 25, 1995.
### V. Summary of the technological characteristics of the new device in comparison to those of the predicate.
Tables 3 outlines the technological characteristics (methodologies) of the Abuscreen ONLINE Cannabinoids in comparison to that of the legally marketed product.
{3}------------------------------------------------
### Brief discussion of the clinical and nonclinical tests relied on for a determination of VI. substantial equivalence:
Tables 3 demonstrates the results of clinical and nonclinical studies performed using the Abuscreen ONLINE Cannabinoids test kit. The significant performance characteristics relied upon for a determination of substantial equivalence are summarized in this chart. This information concludes that the performance of this device is essentially equivalent to the legally marketed predicate device.
{4}------------------------------------------------
# Abuscreen ONLINE Cannabinoids for Hitachi 917 Table 1
| | Proposed: | | Previously Cleared: (K981585) |
|-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Abuscreen ONLINE | | Abuscreen ONLINE |
| | | Cannabinoids for Hitachi 917 | Cannabinoids (1000 Test Kit) |
| Methodology | | Kinetic interaction of microparticles<br>in a solution as measured by<br>changes in light transmission | Kinetic interaction of microparticles<br>in a solution as measured by<br>changes in light transmission |
| Sample type | urine | | urine |
| Intended Use | qualitative and semiquantitative<br>detection of cannabinoids and its<br>metabolites | | qualitative and semiquantitative<br>detection of cannabinoids |
| Calibrator | | Abuscreen ONLINE Cannabinoids<br>Calibration Pack | Abuscreen ONLINE Cannabinoids<br>Calibration Pack or Abuscreen<br>Cannabinoids Calibration Pack 20 |
| Cutoff(s) | 20, 50, 100 ng/mL | | 20, 50, 100 ng/mL |
| Reagent (active<br>ingredients) | 1. Ab reagent: cannabinoids<br>monclonal (mouse) antibody in<br>buffer<br>2. Microparticle reagent:<br>Conjugated cannabinoid derivative<br>microparticles in buffer<br>3. Diluent: Buffer | | 1. Ab reagent: cannabinoids<br>monclonal (mouse) antibody in<br>buffer<br>2. Microparticle reagent:<br>Conjugated cannabinoid derivative<br>microparticles in buffer<br>3. Diluent: Buffer |
| Performance Characteristics: | | | |
| Precision Qualitative ( 20 ng/mL Cutoff): | | | |
| | >95% negative at 15 ng/mL | | >95% negative at 15 ng/mL |
| | >95% positive at 25 ng/mL | | >95% positive at 25 ng/mL |
| Within Run | Mean (OD) | CV% | |
| 15 ng/mL | 5607 | 0.9 | |
| 20 ng/mL | 4990 | 1.0 | |
| 25 ng/mL | 4614 | 1.3 | |
| 50 ng/mL | 2660 | 1.1 | |
| 60 ng/mL | 2153 | 2.0 | |
| Day-to-Day | Mean (OD) | CV% | |
| 15 ng/mL | 5651 | 1.3 | |
| 20 ng/mL | 5054 | 1.4 | |
| 25 ng/mL | 4670 | 1.6 | |
| 50 ng/mL | 2729 | 2.2 | |
| 60 ng/mL | 2237 | 2.9 | |
{5}------------------------------------------------
______________________________________________________________________________________________________________________________________________________________________________
# Abuscreen ONLINE Cannabinoids for Hitachi 917 Table 1 (Continued)
| Proposed: | | | Previously Cleared: (K981585) | | |
|---------------------------------------------------------------|--------------------------------------------------------|-----|--------------------------------------------------------------------------------|-----------------------------|-----------|
| Abuscreen ONLINE | | | Abuscreen ONLINE Cannabinoids | | |
| Cannabinoids for Hitachi 917 | | | (1000 Test Kit) | | |
| | 50 ng/mLCutoff ):<br>Precision Qualitative | | | | |
| | >95% negative at 40 ng/mL<br>>95% positive at 60 ng/mL | | >95% negative at 40 ng/mL<br>>95% positive at 60 ng/mL | | |
| | | | | | |
| Within Run | Mean (OD) | CV% | | | |
| 25 ng/mL | 2403 | 2.9 | | | |
| 40 ng/mL | 1732 | 1.1 | | | |
| 50 ng/mL | 1340 | 1.5 | | | |
| 60 ng/mL | 1169 | 2.2 | | | |
| 75 ng/mL | 823 | 1.3 | | | |
| Day-to-Day | Mean (OD) | CV% | | | |
| 25 ng/mL | 2404 | 1.1 | | | |
| 40 ng/mL | 1749 | 1.4 | | | |
| 50 ng/mL | 1354 | 2.2 | | | |
| 60 ng/mL | 1177 | 2.5 | | | |
| 75 ng/mL | 824 | 4.2 | | | |
| Precision Qualitative | 100 ng/mL Cutoff ): | | | | |
| | >95% negative at 80 ng/mL | | >95% negative at 75 ng/mL | | |
| | >95% positive at 120 ng/mL | | | >95% positive at 125 ng/mLL | |
| Within Run | Mean (OD) | CV% | | | |
| 50 ng/mL | 3180 | 1.4 | | | |
| 80 ng/mL | 2469 | 1.3 | | | |
| 100 ng/mL | 1755 | 1.2 | | | |
| 120 ng/mL | 1547 | 2.5 | | | |
| 150 ng/mL | 1120 | 2.9 | | | |
| Day-to-Day | Mean (OD) | CV% | | | |
| 50 ng/mL | 3194 | 1.6 | | | |
| 80 ng/mL | 2483 | 1.8 | | | |
| 100 ng/mL | 1764 | 2.7 | | | |
| 120 ng/mL | 1536 | 2.7 | | | |
| 150 ng/mL | 1119 | 3.2 | | | |
| Precision Quantitative | ( 20 ng/mL Cutoff ): | | | | |
| Within Run | Mean (ng/mL) | CV% | Conc. (ng/mL) | Mean (ng/mL) | CV% |
| 15 ng/mL | 13 | 5.0 | 15 | 13 | 6 |
| 20 ng/mL | 20 | 3.2 | 20 | 19 | 7 |
| 25 ng/mL | 23 | 2.9 | 25 | 23 | 4 |
| 50 ng/mL | 50 | 1.5 | 40 | 40 | 2 |
| 60 ng/mL | 94 | 1.3 | | | |
| Proposed:<br>Abuscreen ONLINE<br>Cannabinoids for Hitachi 917 | | | Previously Cleared: (K981585)<br>Abuscreen ONLINE Cannabinoids (1000 Test Kit) | | |
| Precision Quantitative (20 ng/mL Cutoff ): | | | | | |
| Day-to-Day | Mean (ng/mL) | CV% | Conc. (ng/mL) | Mean (ng/mL) | CV% |
| 15 ng/mL | 14 | 9.1 | 15 | 14 | 7 |
| 20 ng/mL | 18 | 8.5 | 20 | 20 | 6 |
| 25 ng/mL | 24 | 4.6 | 25 | 25 | 5 |
| 50 ng/mL | 48 | 3.3 | 40 | 41 | 2 |
| 60 ng/mL | 97 | 7.6 | | | |
| Precision Quantitative ( 50 ng/mL Cutoff ): | | | | | |
| Within Run | Mean (ng/mL) | CV% | Conc. (ng/mL) | Mean (ng/mL) | CV% |
| 25 ng/mL | 23 | 2.9 | 25 | 23 | 4 |
| 40 ng/mL | 39 | 1.1 | 40 | 40 | 2 |
| 50 ng/mL | 50 | 1.5 | 50 | 49 | 1 |
| 60 ng/mL | 62 | 2.2 | 60 | 66 | 3 |
| 75 ng/mL | 94 | 1.3 | 100 | 101 | 1 |
| Day-to-Day | Mean (ng/mL) | CV% | Conc. (ng/mL) | Mean (ng/mL) | CV% |
| 25 ng/mL | 24 | 4.6 | 25 | 25 | 5 |
| 40 ng/mL | 39 | 4.0 | 40 | 41 | 2 |
| 50 ng/mL | 48 | 3.3 | 50 | 50 | 3 |
| 60 ng/mL | 60 | 5.8 | 60 | 68 | 3 |
| 75 ng/mL | 97 | 7.6 | 100 | 100 | 1 |
| Precision Quantitative ( 100 ng/mL Cutoff ): | | | | | |
| Within Run | Mean (ng/mL) | CV% | Conc. (ng/mL) | Mean (ng/mL) | CV% |
| 50 ng/mL | 54 | 2.3 | 50 | 50 | 4 |
| 80 ng/mL | 74 | 1.7 | 80 | 75 | 4 |
| 100 ng/mL | 103 | 2.0 | 100 | 95 | 2 |
| 120 ng/mL | 118 | 1.9 | 120 | 123 | 2 |
| 150 ng/mL | 154 | 1.5 | 150 | 148 | 1 |
| Day-to-Day | Mean (ng/mL) | CV% | Conc. (ng/mL) | Mean (ng/mL) | CV% |
| 50 ng/mL | 53 | 2.7 | 50 | 52 | 6 |
| 80 ng/mL | 75 | 2.6 | 80 | 76 | 3 |
| 100 ng/mL | 106 | 3.8 | 100 | 96 | 2 |
| 120 ng/mL | 120 | 3.2 | 120 | 125 | 2 |
| 150 ng/mL | 154 | 3.3 | 150 | 149 | 1 |
| Accuracy | | | | | |
| 20 ng/mL Cutoff | N= 69 Confirmed Pos.<br>69 Pos. 0 Neg. | | N= 50 Confirmed Pos.<br>50 Pos. 0 Neg. | | |
| 50 ng/mL Cutoff | N= 69 Confirmed Pos.<br>53 Pos. 16 Neg. | | N = 50 Confirmed Pos.<br>38 Pos. 12 Neg. | | |
| 100 ng/mL Cutoff | N= 69 Confirmed Pos.<br>37 Pos. 32 Neg. | | N= 50 Confirmed Pos.<br>21 Pos. 29 Neg. | | |
| Limit of Detection | 7 ng/mL | | | | < 5 ng/mL |
: 上一
{6}------------------------------------------------
# Abuscreen ONLINE Cannabinoids for Hitachi 917 Table 1 (Continued)
{7}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/7/Picture/2 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, depicted in black.
DEC 1 1 1998
Ms. Rita Smith Senior Regulatory Affairs Associate Roche Diagnostic Systems, Inc. A Subsidiary of Hoffmann-La Roche, Inc. Branchburg Township 1080 U.S. Highway 202 Somerville, New Jersey 08876-3771
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Re : K983701 Abuscreen ONLINE® Cannabinoids Assay Trade Name: Regulatory Class: II Product Code: LDJ Dated: October 20, 1998 October 21, 1998
Dear Ms. Smith:
Received:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to A substantially equivalent determination assumes compliance with 895. the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements concerning action. your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or requlations.
{8}------------------------------------------------
### Page 2
This letter will allow you to beqin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a leqally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other qeneral information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Steven Autman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radioloqical Health
Enclosure
{9}------------------------------------------------
Page of
510(k) Number (if known)
Device Name: Abuscreen ONLINE® Cannabinoids
Indications for Use:
Abuscreen ONLINE for Cannabinoids is an in vitro diagnostic test for the qualitative and semiquantitative detection of cannabinoids in human urine on automated clinical chemistry analyzers at cutoff concentrations of 20 ng/mL, 50 ng/mL, and 100 ng/mL. Semiquantitative test results may be obtained that permit laboratories to assess assay performance as part of a quality control program. Measurements obtained by this device are used in the diagnosis of cannabinoid use or abuse.
## (PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)

scription Use
1
Prescription Use (Per 21 CFR 801.109) OR
Over-The-Counter Use (Optional Format 1-2-96)
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K983701
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.