K983648 · Perry Baromedical Services · CBF · Feb 24, 1999 · Anesthesiology
Device Facts
Record ID
K983648
Device Name
SIGMA PLUS/ II MULTIPLACE HYPERBARIC SYSTEM
Applicant
Perry Baromedical Services
Product Code
CBF · Anesthesiology
Decision Date
Feb 24, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5470
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The SIGMA II Dualplace Hyperbaric Chamber is used to provide Hyperbaric Oxygen Therapy, prescribed by a licensed physician, where medical grade oxygen is administered to patients while under pressure, by the patient breathing oxygen from a breathing hood or breathing mask with the chamber pressurized with breathing air. The breathing of oxygen at elevated pressure promotes the movement of oxygen into the patient's tissues.
Device Story
Multiplace hyperbaric chamber for two patients; provides hyperbaric oxygen therapy (HBOT). Input: compressed breathing quality air for chamber pressurization; medical grade oxygen for patient inhalation via hood or mask. Operation: ASME-certified pressure vessel; pneumatic control system for pressurization/depressurization; intrinsically safe communication system; fire suppression/extinguishing systems. Used in clinical settings; operated by trained personnel. Physician prescribes HBOT; patient breathes oxygen at elevated pressure to promote tissue oxygenation. Enhances ease of operation via moveable control arm and pedestal-mounted breathing system controls compared to predicate. Benefits: facilitates treatment of conditions like decompression sickness, wounds, and infections.
Clinical Evidence
No clinical data. Bench testing only; device relies on established ASME pressure vessel standards and substantial equivalence to a previously cleared predicate device.
Technological Characteristics
ASME pressure vessel (Section VIII, Div 1; PVHO-1); acrylic cylinder (40" ID, 65" length, 20:1 safety factor); pneumatic control system (no electrical/electronic components); intrinsically safe communication; fire suppression system; fiberglass cosmetic covers. Pressurization via compressed air.
Indications for Use
Indicated for patients requiring hyperbaric oxygen therapy for: Air or Gas Embolism, Carbon Monoxide/Smoke Inhalation, Compromised Skin Grafts/Flaps, Crush Injuries/Acute Traumatic Ischemias, Decompression Sickness, Enhanced Healing in Selected Problem Wounds, Exceptional Blood Loss (Anemia), Gas Gangrene (Clostridial Myonecrosis), Intracranial Abscess, Necrotizing Soft Tissue Infections, Radiation Tissue Damage (Osteoradionecrosis), Refractory Osteomyelitis, and Thermal Burns.
Regulatory Classification
Identification
A hyperbaric chamber is a device that is intended to increase the environmental oxygen pressure to promote the movement of oxygen from the environment to a patient's tissue by means of pressurization that is greater than atmospheric pressure. This device does not include topical oxygen chambers for extremities (§ 878.5650).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Perry Baromedical. The logo consists of a black circle with several black dots clustered on the left side, resembling a stylized atom or molecule. To the right of the circle, the words "PERRY" and "BAROMEDICAL" are stacked vertically, with "PERRY" on top and "BAROMEDICAL" below it. At the top of the image, there is a handwritten date that appears to be "2-24-99".
January 15, 1999
K 885648
7555 GARDEN ROAD Riviera Beach FLORIDA 33404-3411 USA
800 . 7 4 1 . 4 3 7 6 う61・840・0395 Fxx: 561 .840 . 0398 e-mail: baromed@aol.com web site: www.perrybaromedical.com
## Amended 510(k) K983648 SUMMARY Page 1 of 3
Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Blvd. Rockville, Maryland 20850
Document Control Clerk Attention:
## SIGMA Plus / II Multiplace Hyperbaric Chamber Summary of Safety & Subject: Effectiveness Information - Amended as of January 15, 1999
Gentlemen:
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
Substantial Equivalence: The Perry Baromedical Corporation SIGNA Plus / II Multiplace Substantial Equivalence. "The Perry Baromedical Corporation Corporation SIGMA Hyperbaric Chamber is substantially oquili K862198/A]. The SIGMA Plus / Il has the same If Dualplace Tryperbarte Chamoer, [220]. Investeristics as the SIGMA II predicate device.
Intended use and the same technological characteristics as the SIGMA II in in The scope and operation of the SIGMA Plus/ II is identical to that of the SIGMA II in the following areas:
- The critical component of both systems consists of an ASME pressure vessel that is l I he critical component of tested in accordance with the requirements of the ASME designed, labricated, and testode, Section VIII, Division I; and ASME PVHO-1, Safety Standard for Pressure Vessels for Human Occupancy.
- The pressure vessel consists of certified metal components constructed of materials in 2 The pressure vesser consists of occession Pressure Vessel Code, Section II - Materials. Contornatice with the rolling boundary consists of an actylic of an actylic 1 in addition to the mean parts, the requirements of ASME PVHO-1, Safety Standard for Pressure Vessels for Human Occupancy.
- The pressure vessel is protected from accidental over pressurization by an ASME "UV" 3 stamped pressure relief valve.
Material Production
{1}------------------------------------------------
SIGMA Plus 510(k) K983648 Amended Summary Page 2 of 3
- The medium used for pressurization of the chamber is compressed breathing quality air. 4
- Treatment capacity of the chamber is two (2) patients. 5
- Communication between the patients / attendants is through an intrinsically safe б communications system.
- The SIGMA II and SIGMA Plus/ II chambers are used to provide Hyperbaric Oxygen 7 Therapy, prescribed by a licensed physician, where medical grade oxygen is administered to patients while under pressure, by the patient breathing oxygen from a breathing hood or breathing mask with the chamber pressurized with breathing air. The breathing of oxygen at elevated pressure promotes the movement of oxygen into the patient's tissues.
- The chamber pressurization and depressurization control system is completely 8 pneumatic, with no electrical or electronic components.
- Both chambers are Class "A" multiplace hyperbaric systems, with Fire Suppression 9 Systems and independent hand line fire extinguishing systems.
Differences do exist between the SIGMA Plus/ II and the SIGMA II Dualplace chamber; they include:
- The PVHO-1 acrylic cylinder in the SIGMA Plus / II chamber is 40" inside diameter l and 65" long, and the PVHO-1 acrylic cylinder in the SIGMA II chamber is 40" inside diameter by 85" long. Both acrylic cylinders are designed with a 20 to 1 safety factor.
- The pneumatic controls for the SIGMA Plus / II chamber are mounted in a control 2 box that is suspended from a moveable arm, allowing use from either side of the chamber; while the pneumatic controls for the SIGMA II chamber are mounted in a console panel mounted on one side of the chamber.
{2}------------------------------------------------
SIGMA Plus 510(k) K983648 Amended Summary Page 3 of 3
- The SIGMA Plus/ II Multilace Hyperbaric Chamber has painted fiberglass cosmetic 3 covers on the sides and ends of the chamber; the SIGMA II Dualplace Hyperbaric Chamber has aluminum side cosmetic covers.
- The Built-In Breathing System controls for the SIGMA Plus / II are mounted in a 4 pedestal mounted moveable control panel, allowing positioning to suit the operator location. The SIGMA II Built-In Breathing System controls are mounted in the control console positioned at the side of the chamber.
- The SIGMA II Dualplace chamber is available with an Entry Lock and a Medical Lock 5 feature, which are not offered with the SIGMA Plus / II.
The changes to the SIGMA II system resulting in the SIGMA Plus/ II Dualplace Hyperbaric Chamber enhance its ease of operation, and do not adversely effect the safety and effectiveness of the device. The basic treatment method is identical, only the chamber acrylic window length size has been changed
None of the above information is confidential and all may be made available to the public upon written request.
Perry Baromedical Corporation
Aames Dodes 1-15.99
es P. Dodson Quality Assurance Manager
Date
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the edge. Inside the circle is an abstract symbol that resembles an eagle or bird in flight, with stylized wings and tail feathers.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 2 4 1999
Mr. James P. Dodson Perry Baromedical Services 7555 Garden Road Riviera Beach, FL 33404-3411
K983648 Re: SIGMA Plus/II Multiplace Hyperbaric Chamber Requlatory Class: II (two) Product Code: 73 CBF January 26, 1999 Dated: Received: February 2, 1999
Dear Mr. Dodson:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2 - Mr. James P. Dodson
This letter will allow you to begin marketing your device as described inis recei will arrow you co icention. The FDA finding of substantial in your Srown, promativice to a legally marketed predicate device equivalence or your dontion for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling II you actire bpoorted 801 and additionally 809.10 for in vitro reguration (Er Crs), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, (21 CFR 807. Hisbianaling by Formation on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its be obtained trom (300) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Thomas J. Callahan
Thomas J. Callahan, Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Perry Baromedical. The logo consists of a black graphic on the left, resembling a partial circle with several dots inside and around it. To the right of the graphic, the words "PERRY" and "BAROMEDICAL" are stacked vertically in a serif font. The text is in a lighter color, contrasting with the black graphic.
7555 GARDEN ROAD RIVIERA BEACH FLORIDA 33404-3411 USA
800 • 7 4 1 • 4 3 7 6 561 •840 •0395 Fxx: 561.840.0398 e-mail: baromed@aol.com web site: www.pcrrybaromedical.com
## INDICATIONS FOR USE
The SIGMA Plus / II Multiplace hyperbaric System is indicated for use for the following The SIGNA I has 7 21 Muniputo hyperomith guidelines established by the Undersea and Hyperbaric Medical Society, as follows:
- Air or Gas Embolism 1.
- Carbon Monoxide / Smoke Inhalation 2.
- Compromised Skin Grafts and Flaps 3.
- Crush Injuries / Acute Traumatic Ischemias 4.
- Decompression Sickness રે.
- Enhanced Healing in Selected Problem Wounds ર.
- Exceptional Blood Loss (Anemia) 7.
- Gas Gangrene (Clostridial Myonecrosis) 8.
- Intracranial Abscess ರು.
- Necrotizing Soft Tissue Infections 10.
- Radiation Tissue Damage (Osteoradionecrosis) 11.
- Refractory Osteomyelitis 12.
- Thermal Burns 13.
Prescription Device
mKramer
(Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number K453446
:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.