← Product Code [GCT](/productcode/GCT) · K983628

# XENON LIGHT SOURCE AUTO LIP 5123, MODEL 2123.011 (K983628)

_Richard Wolf Medical Instruments Corp. · GCT · Dec 11, 1998 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K983628

## Device Facts

- **Applicant:** Richard Wolf Medical Instruments Corp.
- **Product Code:** [GCT](/productcode/GCT.md)
- **Decision Date:** Dec 11, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

Light Projector AUTO LP 5123 provides light for examination, diagnostic and therapeutic applications, in particular, in endoscopy. The Xenon Light Source is classified as Cardiac Floating (CF) which permits its use for cardiac procedures when used in conjunction with the proper instrumentation for entry into the cardiac system.

## Device Story

Auto LP 5123 Xenon Light Projector provides illumination for endoscopes. Device utilizes 180W xenon arc lamp (approx. 5600K color temperature) to generate light, transmitted via light guide to endoscope. Illumination level is automatically controlled by video input. Used in clinical settings (e.g., OR, endoscopy suites) by physicians/clinicians. Device classified as Cardiac Floating (CF) for cardiac system entry. Output provides necessary visualization for diagnostic/therapeutic interventions, aiding clinical decision-making during endoscopic procedures.

## Clinical Evidence

No clinical tests performed. Bench testing conducted per IEC601-1 (amendment 1 and 2), IEC601-1-1, and IEC601-2-18 standards.

## Technological Characteristics

180 W xenon arc lamp; color temperature ~5600 K; service life ~500 hours. Type CF applied part per IEC601-1. Automatic illumination control via video input. Light output via light guide port.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Xenon Light Source ([K952696](/device/K952696.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

문

Pg 1 of 2

353 Corporate Woods Parkway Vernon Hills, Illinois 60061 Phone: 847.913.1113 Fax: 847.913.1488

RICHARD MEDICAL INSTRUMENTS C

# 510(k) Summary of Safety and Effectiveness

| Submitter                            |                                        |
|--------------------------------------|----------------------------------------|
| Company / Institution name:          | Richard Wolf Medical Instruments Corp. |
| Division name (if applicable):       | N.A.                                   |
| Street address:                      | 353 Corporate Woods Parkway            |
| City:                                | Vernon Hills                           |
| State/Province:                      | Illinois                               |
| Country:                             | USA                                    |
| ZIP/Postal Code:                     | 60061                                  |
| Contact name:                        | Mr. Robert L. Casarsa                  |
| Contact title:                       | Quality Assurance Manager              |
| Date of Preparation                  | October 13, 1998                       |
| FDA establishment regulation number: | 14 184 79                              |
| Phone number (include area code):    | (847) 913-1113                         |
| FAX number (include area code):      | (847) 913-0924                         |
| Product Information:                 |                                        |
| Trade name:                          | Auto LP 5123 Xenon Light Projector     |
| Model number:                        | 5123.011                               |
| Common name:                         | Xenon Light Source                     |
| Classification Name:                 | Xenon Light Source                     |

Information on devices to which substantial equivalence is claimed:

| 510(k) Number | Trade or proprietary or model name | Manufacturer          |
|---------------|------------------------------------|-----------------------|
| 1 K952696     | 1 Xenon Light Source               | 1 Richard Wolf M.I.C. |
| 2             | 2                                  | 2                     |
| 3             | 3                                  | 3                     |

### 1.0 Description

Light Projector AUTO LP 5123 uses a xenon arc lamp to provide illumination for an endoscope via a light guide which is connected to the projector's light port. The illumination level is automatically controlled by video.

8 - 1

{1}------------------------------------------------

K 983628

pg 2 of 2

#### 2.0 Intended Use

Light Projector AUTO LP 5123 provides light for examination, diagnostic and therapeutic applications, in particular, in endoscopy.

The Xenon Light Source is classified as Cardiac Floating (CF) which permits its use for cardiac procedures when used in conjunction with the proper instrumentation for entry into the cardiac system.

#### Technological Characteristics 3.0

The color temperature of the 180 W xenon arc lamp is about 5600 K. The service life is approximately 500 hours and will decline with time due to the number of starts. The light projector is a type CF applied part per IEC601-1.

#### 4.0 Substantial Equivalence

The submitted devices pose the same type of questions about safety and effectiveness as existing devices. The new technological characteristics have not diminished safety or effectiveness. The submitted devices are substantially equivalent to existing 510(k) devices sold by Richard Wolf.

#### 5.0 Performance Data

Independent laboratories tested the Xenon Light Source 5123 according to specified standards IEC601-1 amendment 1 and 2, IEC601-1-1, and IEC601-2-18. The certifications for UL and CE compliance are pending.

#### 6.0 Clinical Tests

No clinical tests performed.

#### 7.0 Conclusions Drawn

These devices are designed and tested to guarantee the safety and effectiveness when used according to the instruction manual.

By: Robert J. Casassa

Robert L. Casarsa Quality Assurance Manager

Date: Oct 12 98

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows a logo with a circular border containing the words "U.S. DEPARTMENT OF HEALTH & HUMAN SERVICES" in a sans-serif font. Inside the circle is a stylized symbol consisting of three overlapping human profiles or figures, arranged in a way that suggests movement or progression. The symbol is rendered in black, contrasting with the likely white or light background of the logo.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DEC 1 1 1998

Mr. Robert L. Casarsa Quality Assurance Manager Richard Wolf Medical Instruments Corp. 353 Corporate Woods Parkway Vernon Hills, Illinois 60061

Re: K983628 Auto LP 5123 Xenon Light Projector Dated: October 13, 1998 Received: October 15, 1998 Regulatory Class: II 21 CFR 876.1500/Procode: 78 GCT

Dear Mr. Casarsa:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrf/dsmaldsmamain.html".

Sincerely yours

Lillian Yin, Ph.D.
Director, Division of Reproductive

Director, Division of Reproductive, Abdominal, Ear, Nose and Throat and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

ndication

510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________

Device Name:__Light Projector AUTO LP 5123

## Intended Use:

Light Projector AUTO LP 5123 provides light for examination, diagnostic and therapeutic applications, in particular, in endoscopy.

The Xenon Light Source is classified as Cardiac Floating (CF) which permits its use for cardiac procedures when used in conjunction with the proper instrumentation for entry into the cardiac svstem.

(PI.E.ASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concursence of CDRH Office of Device Evaluation (ODE)

Voriel a. Seymon
(Division Sign-Off)

Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number _

---

**Source:** [https://fda-staging.innolitics.com/device/K983628](https://fda-staging.innolitics.com/device/K983628)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).