K983604 · Washington Regulatory Services · MQF · May 27, 1999 · Obstetrics/Gynecology
Device Facts
Record ID
K983604
Device Name
INTRA-UTERINE TRANSFER, INTRA-FOLOPIAN TRANSFER
Applicant
Washington Regulatory Services
Product Code
MQF · Obstetrics/Gynecology
Decision Date
May 27, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.6110
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
These devices are used to introduce or remove gametres, zygotes(s), preembryo)s) and/or embryo(s) into or from the body.
Device Story
Assisted reproduction catheters used for intra-uterine and intra-fallopian transfer of gametes, zygotes, and embryos. Devices consist of various catheter sets (Frydman, TDT, BI SET, Oliver) featuring flexible tips, mandrels for stiffening, and positioning markings. Operated by clinicians in a clinical setting. Devices are malleable to allow shape adjustment; mandrels occlude the lumen to prevent bodily fluid inflow during cervical insertion. Ultrasound markers facilitate visualization. Output is the delivery or retrieval of reproductive cells/embryos. Benefits include improved access for difficult embryo transfers (e.g., narrow or curved cervix) and reduced risk of embryo loss in crypts or mucus.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Materials include polyethylene, polypropylene, polyurethane, polyamide, stainless steel, and butadiene styrene. Features include malleable shafts, mandrels for stiffening, echogenic ultrasound markers, and various lumen configurations (single/double). Dimensions vary by model (lengths 14-33 cm; diameters 0.6-2.6 mm).
Indications for Use
Indicated for the introduction or removal of gametes, zygotes, pre-embryos, and/or embryos into or from the body for assisted reproduction procedures, including intra-uterine and intra-fallopian transfer.
Regulatory Classification
Identification
Assisted reproduction catheters are devices used in in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures to introduce or remove gametes, zygote(s), preembryo(s), and/or embryo(s) into or from the body. This generic type of device may include catheters, cannulae, introducers, dilators, sheaths, stylets, and component parts.
Special Controls
*Classification.* Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, biocompatibility testing, and clinical testing).
Submission Summary (Full Text)
{0}------------------------------------------------
2 1999 EEB
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Ms. Elizabeth Platt Regulatory Affairs Supervisor Bio-Rad Laboratories Diagnostics Group 9500 Jeronimo Road Irvine, California 92618-2017
K984604 Re:
> Trade Name: Architect™ Estradiol MasterCheck Regulatory Class: I Product Code: JJX Dated: December 21, 1998 Received: December 28, 1998
න් ක්රිඩ් බැටි පිහිටි පිහිටි විශ්වය විය විසින් පිහිටි විසින් පිහිටි විසින් පිහිටි විසින් පිහිටි විසින් පිහිටි විසින් පිහිටි පිහිටි පිහිටි විසින් පිහිටි විසින් පිහිටි විසින්
Dear Ms. Platt:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Flectronic Product .
{1}------------------------------------------------
#### Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours.
Steven Sutman
Steven I. Gutman, M.D. M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
page 1 of 6
Kag3604 March 3, 1999
### 510(K) Summary
1) PRODIMED Laboratories c/o Washington Regulatory Services Randolph L. Cooke 23 Welisewitz Road Ringoes. NJ 08551 609-466-0510 Fax 609-466-4443
#### 2) DEVICE NAME:
Proprietary Names: Prodimed Intra-uterine, Intra-fallopian Transfer Assisted Reproduction Catheters
Intra-uterine transfer
- 1) Frydman Catheter 4.5, 5.5
- 2) Frydman Catheter 4.5 with 2 lumen and removable mandrel
- 3) Long Frydman Set
- 4) Frydman Set
- 5) Soft Frydman Set
- 6) Soft Frydman Set with madrel
- 7) TDT Set
Intra-fallopian transfer
- 1) Biset for GIFT
- 2) Gift Catheter
- 3) Oliver Set
- 4) Rinsing Valve for Oliver Set
Common Name: Assisted Reproduction Catheters
Classification Name: Assisted Reproduction catheters are classified as Class II at 21 CFR 884.6110. The product code is 85 MQF.
3) IDENTIFICATION: The Assisted Reproduction microtool devices have been down classified FR Vol. 63, No. 175 September 10, 1998. Effective date: October 13, 1998.
- 4) DESCRIPTION OF DEVICE: As found in the labeling/promotional material:
{3}------------------------------------------------
## Frydman™ Catheter
Polyethylene catheter with a flexible end, internal diameter: 1.1mm, outer diameter: 1.6mm, and a distal opening.
For the 4.5 model:
- . Length of flexible end: 4.5cm
- Total length: 17.5cm .
- Two hysterometric markings at 5.5 and 6.5cm from the end .
For the 5.5 model:
- Length of flexible end: 5.5cm .
- Total length: 18.5cm .
- Two hysterometric markings at 6.5 and 7.5cm from the end .
### Double Lumen Frydman™ Catheter 4.5
- Polyethylene catheter, 17.5cm long, consisting of a central tube 2.1mm, . stiffened in its proximal part on 13cm by a second polypropylene tube 2.6mm, with two hysterometric markings at 5.5 and 6.5 from the distal opening.
- The central tube is a double lumen tube: its lateral lumen is provided with a ● malleable mandrel, and its central lumen 1.1mm, is intended to contain embryo(s).
Soft Frydman™ Set (with or without mandrel)
The set includes 2 catheters:
- · Catheter Nº 1: Polypropylene introducer (with a blue adapter)
- · 14.5 cm long
- 2.2mm with positioning markings at 1, 2, 3, 4, 5, 6 and 7 cm from its distal . end
- with a movable sliding stopper .
- Catheter Nº2 Polyurethane reimplantation catheter ●
- 23 cm long
- 1.53mm, 0.7mm .
- with markings spaced 1 cm apart in its lower part .
- . with a white adapter locked by a polyethylene cap
- Mandrel, 17.5 cm long, depending on the model ●
Short Frydman™ Set (for difficult embryo transfer)
The set includes 2 catheters:
- Catheter No. 1 polypropylene introducer (with an blue adapter) .
- · 14 cm long
- . 2.2mm, with positioning markings at 1, 2, 3, 4, 5, 6 and 7 cm from its distal end
- with a green mandrel and a movable sliding stopper .
- Catheter No. 2 polyethylene reimplantation catheter (with an orange adapter) .
- 21 cm long .
- . 1.6mm, 1.1mm
{4}------------------------------------------------
- with a steel tube and 3 markings spaced 1 cm apart in its lower part .
Long Frydman™ Set (for difficult embryo transfer)
The set includes 2 catheters:
- Catheter No. 1 polypropylene introducer (with an blue adapter) .
- 20 cm long .
- 2.2mm, with positioning markings at 1, 2, 3, 4, 5, 6 and 7 cm from its distal . end
- with a green mandrel and a movable sliding stopper .
- Catheter No. 2 polyethylene reimplantation catheter (with an orange adapter) .
- 25.5 cm long ●
- 1.6mm, 1.1mm .
- . with 1 marking its lower part
#### T.D.T.TM Set (for difficult embryo transfer)
The set includes 2 catheters:
- Catheter No. 1 polypropylene introducer .
- 17.5 cm long, with a flexible tip of 4.5 cm long, and two hysterometric ● markings at 5.5 and 6.5 cm from this end
- It is provided with a plastic coated metal mandrel. Two models are . available: with transparent mandrel or green mandrel.
- . Catheter No. 2 ultra-thin polyethylene reimplantation catheter
- 23.5 cm long, on a steel microtube, 1.0mm, 0.6mm, with a positioning ● mark at 2 cm from its lower part.
# BI SET™ for GIFT
The set includes 2 catheters:
- Catheter No. 1 introducer (with an blue adapter) .
- . 20 cm long
- with positioning markings at 1, 2, 3, 4, 5, 6 and 7 cm from its distal end . . with a green mandrel and a movable sliding stopper
- Catheter No. 2 polyethylene reimplantation catheter (with an orange adapter) ●
- 25.5 cm long .
- 1.6mm, 1.1mm ●
- with a steel tube and 1 marking at 2 cm from its lower part, with a movable . stopper and a graduated protective sheath
- Stainless steel trocar (with green adapter), 20 cm long .
- Two models are available:
- · 1 transfer catheter
- 2 transfer catheters .
#### Oliver Set MK II TM
The set includes:
- 1 polyamide, echogenic, introducer, 24 cm long, 1.9mm, 1.1mm, pre-curved . at 45°, with an loive-shape tip fitted with a ground stainless steel ring
{5}------------------------------------------------
4 of 6
- 1 metal mandrel, 27 cm long, 1mm. The introduction of this mandrel in the . introducer allows to stiffen it and to insert it more easily.
- 1 polyethylene transfer catheter, 33 cm long, 1mm, 0.6mm. When this . catheter is inserted in the introducer, its end protudes on 6 cm from the inrtoducer tip.
It is provided with a positioning system (markings) that allows to transfer at a chosen distance from the olive.
Function of Device/Significant Physical Characteristics of Device:
All devices are malleable, to allow a change in shape for the device; the mandrel is accluded to prevent the in-flow of bodily fluids when inserting into the crevix. All devices contain echolocater ultrasound markers.
The Frydman™ Catheter is used for embryo transfer when the permeability of the cervical canal is normal.
The Double Lumen Frydman™ Catheter 4.5 is used for embryo transfer with any permeability of the cervical canal.
The Soft Frydman™ Set, with or without a mandrel, is used for embryo transfer. The model with the mandrel may be used in the presence of a narrow or curved cervix that makes it difficult to introduce the model without the mandrel.
The Short Frydman™ Set is used for difficult embryo transfer. The set may be used in the presence of a narrow or curved cervix that makes it difficult to introduce a standard Frydman™ catheter.
The Long Frydman™ Set is also for difficult embryo transfer. The set may be used in the presence of a narrow or curved cervix that makes it difficult to introduce a standard Frydman™ catheter.
The T.D.T.TM Set is used for embryo transfer. The T.D.T.TM Set allows the reduction of the risk of setting in the crypts or occlusion by mucus that would impede the expulsion of the embryo.
The BI SET™ for GIFT is used for intrafallopian transfer of gametes via transparietal route under celioscopy.
The Oliver Set MK II ™ is used for intratubal transfer of gametes or zygotes via a transcervical approach.
Device Design/ Material Used/ Physical Properties:
Frydman Catheter:
- Device Part Tube E212 Tube E321 Tube E320 Tube E342 Fitting
Material Used Polyethylene Polypropylene Polypropylene Polypropylene Polyethylene
{6}------------------------------------------------
Frydman Set: Material Used Device Part Tube E212 Tube X092 Fitting Tube Fitting Tube E212 Tube 063 Plug Frydman Soft: Device Part Tube E212 Fitting Tube Fitting Tube E212 Tube 063 Plug TDT Set: Device Part Tube E241 Tube INOX Fitting Tube E212 Tube E321 Tube 320 Tube 342 Fitting Tube E212 Tube 063 Plug Oliver Set: Device Part
Tube E241
Polyethylene Stainless steel Polyethylene Polypropylene Polypropylene Polyethylene Metal Polypropylene Material Used Polyurethane Polyurethane Polypropylene Polypropylene Polyethylene Metal Polyethylene Material Used Polyethylene Stainless steel Butadiene styrene Polyethylene Polypropylene Polypropylene Polypropylene Polypropylene Polyethylene Metal Polyethylene
Material Used Polyethylene
{7}------------------------------------------------
cofle
Tube INOX Fitting
Stainless steel Butadiene styrene
Tube Fitting Polyamide Polyamide
# 5) STATEMENT OF INTENDED USED
These devices are used to introduce or remove gametres, zygotes(s), preembryo)s) and/or embryo(s) into or from the body.
Intra-uterine transfer Frydman Catheter 4.5, 5.5 Frydman Catheter 4.5 with 2 lumen and removable mandrel Long Frydman Set Frydman Set Soft Frydman Set Soft Frydman Set with madrel TDT Set
Intra-fallopian transfer Biset for GIFT Gift Catheter Oliver Set Rinsing Valve for Oliver Set
1
{8}------------------------------------------------
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAY 2 7 1999
Mr. Randolph L. Cooke PRODIMED Laboratories c/o Washington Regulatory Services. Inc. 23 Welisewitz Road Ringoes, N.J. 08551
Re: K983604
Prodimed Intra-uterine, Intra-fallopian Transfer Assisted Reproduction Catheters Sets Dated: March 11, 1999 Received: March 15, 1999 Regulatory Class: II 21 CFR §884.6110/Procode: 85 MQF
Dear Mr. Cooke:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
#### 510(k) NUMBER (IF KNOWN) K983604
DEVICE NAME: Prodimed Intra-uterine, Intra-fallopian Transfer Assisted Reproduction Catheters
#### INDICATIONS FOR USE:
These devices are used to introduce or remove gametres, zygotes(s), preembryo)s) and/or embryo(s) into or from the body.
Intra-uterine transfer
- 1) Frydman Catheter 4.5, 5.5
- 2) Frydman Catheter 4.5 with 2 lumen and removable mandrel
- 3) Long Frydman Set
- 4) Frydman Set
- 5) Soft Frydman Set
- 6) Soft Frydman Set with madrel
- 7) TDT Set
Intra-fallopian transfer
- 1) Biset for GIFT
- 2) Gift Catheter
- 3) Oliver Set
- 4) Rinsing Valve for Oliver Set
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use<br>Per 21 CFE 801.109 | <div style="text-align:center;">OR</div> | Over-The Counter-Use<br>(Options Format 1-2-96) |
|----------------------------------------|------------------------------------------|-------------------------------------------------|
|----------------------------------------|------------------------------------------|-------------------------------------------------|
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT,
and Radiological Devices
| 510(k) Number | K983604 |
|---------------|---------|
|---------------|---------|
000
3
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.