← Product Code [GCJ](/productcode/GCJ) · K983567

# IMAGE TECHNOLOGIES STERICAM, COUPLER-DRAPE AND TROVIEW IMAGING SYSTEM (K983567)

_Med Images, Inc. · GCJ · Jan 7, 1999 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K983567

## Device Facts

- **Applicant:** Med Images, Inc.
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Jan 7, 1999
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The Image Technologies Corporation's SteriCam, Coupler-Drape and TroView Imaging System consists of a CCD camera, coupler and a computer-based documentation system. The SteriCam is intended to attach to a standard commercially available endoscope for visualization of body cavities, hollow organs, and canals. The SteriCam Coupler is intended to couple the SteriCam to the standard commercially available endoscope. The TroView System is intended to provide documentation/storage of the image in the field of view of the SteriCam.

## Device Story

System comprises CCD camera, coupler-drape, and computer-based documentation unit. Camera attaches to standard endoscopes to capture visual data from body cavities/hollow organs; transmits video signal to TroView system. Coupler-drape maintains sterile field by covering non-sterile camera/cable. TroView unit displays real-time video on monitor; allows adjustment of brightness, hue, contrast, saturation; provides software-based zoom. Physician uses sterile remote controller within sterile field to operate camera/computer functions. System enables image storage to 3.5" floppy disk or printing to dedicated printer. Benefits include real-time visualization and documentation of endoscopic procedures.

## Clinical Evidence

Bench testing only. Image quality evaluated using USAF 1951 Test Target per MIL-STD-150. Results demonstrated resolution of 3.17 lines per mm.

## Technological Characteristics

CCD camera; computer-based documentation system; sterile disposable drape component. Connectivity via video signal cable. Image processing includes brightness, hue, contrast, saturation adjustment and software-based zoom. Storage via 3.5" floppy disk. Sterile remote controller for interface. Resolution 3.17 lines/mm.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

### 7 1999 .IAN

K983567

# 510(k) Summary Image Technologies Corporation SteriCam, Coupler-Drape and TroView System

#### 1. SPONSOR

Image Technologies Corporation 27 Wormwood Street Boston, MA 02210-1625 Telephone: 617-428-7595 Facsimile: 617-428-7599

Contact Person: Lunn Sawyer

Date Prepared: October 9, 1998

#### DEVICE NAME 2.

Classification Name: Endoscopes and accessories, 21 CFR 876.1500 TroCam/TroView Endoscopy System Proprietary Name:

#### 3. INTENDED USE

The Image Technologies Corporation's SteriCam, Coupler-Drape and TroView Imaging System consists of a CCD camera, coupler and a computer-based documentation system. The SteriCam is intended to attach to a standard commercially available endoscope for visualization of body cavities, hollow organs, and canals. The SteriCam Coupler is intended to couple the SteriCam to the standard commercially available endoscope. The TroView System is intended to provide documentation/storage of the image in the field of view of the SteriCam camera.

#### 4. DEVICE DESCRIPTION

The Image Technologies Corporation's SteriCam, SteriCam Coupler-Drape and TroView Imaging System consists of a CCD camera, coupler and a computer based documentation system. The SteriCam takes the image that normally would be seen by the naked eye, and displays it on a color monitor. The SteriCam provides 

{1}------------------------------------------------

imaging through standard commercially available endoscopes and is designed to be used with the sterile SteriCam Coupler-Drape. The Coupler-Drape includes a sterile drape component. The SteriCam is placed within the sterile drape of the Coupler-Drape to prevent contamination in the sterile field. The video signal cable connects the SteriCam to the TroView Documentation System.

The SteriCam Coupler-Drape's distal end attaches to the eyepiece of a conventional endoscope and uses the SteriCam CCD camera to display the image on the computer monitor. The SteriCam Coupler-Drape simply connects the conventional endoscope to the SteriCam. The SteriCam Coupler-Drape is sterile and includes the drape so that the non-sterile SteriCam Camera and cable are covered during the surgical procedures. Although the SteriCam Coupler-Drape attaches to various endoscopes, it is designed to be used only with the SteriCam camera.

The TroView System is a non-sterile reusable unit that is intended to provide real time image displays as well as storage of the image in the field of view of the SteriCam. As the TroView receives and displays the SteriCam digitized video output, the images can be adjusted for brightness, hue, contrast and saturation. The TroView System allows for printing to a dedicated printer or saving the image in to a 1.44 MB 3.5" floppy disk or other storage medium. A software based zoom is also permitted.

A sterile disposable Remote Controller is also offered with the TroView System which permits remote control access to key camera and computer functions. The Remote Controller looks and performs identically to the controller pad located on the faceplate of the computer. The physician is able to place the sterile Remote Controller in the sterile field whereas the faceplate of the computer is non-sterile and situated outside the sterile field.

#### 5. PERFORMANCE TESTING

Image Quality testing was performed on the SteriCam, Coupler-Drape, and TroView System using the USAF 1951 Test Target designed to MIL-STD-150. The results showed that the TroView System resulted in a resolution of 3.17 lines per mm, which is comparable to that of the predicate devices.

{2}------------------------------------------------

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

7 1999 JAN

Mary McNamara-Cullinane, RAC Staff Consultant Medical Device Consultants, Inc. 49 Plain Street North Attleboro, Massachusetts 02760

K983567 Re:

> Trade Name: Image Technologies SteriCam, Coupler-Drape and TroView Imaging System Regulatory Class: II Product Code: GCJ Dated: October 9, 1998 Received: October 13, 1998

Dear Ms. McNamara-Cullinane:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

## Page 2 - Ms. Mary McNamara-Cullinane, RAC

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours,

Celia M. Witten, Ph.D., M.D.
Director
Division of General and
Restorative Devices
Office of Device Evaluation

Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

0(k) Number (if known): 1983567

Image Technologies SteriCam, Coupler-Drape and TroView Imaging Device Name: System

Indications For Use:

The Image Technologies Corporation SteriCam and TroView Imaging System consists of a CCD camera, coupler and a computer-based documentation system. The SteriCam is intended to attach to a standard commercially available endoscope for visualization of body cavities, hollow organs, and canals. The System includes a Coupler-Drape Assembly which is intended to couple the SteriCam to a standard commercially available endoscope. The TroView System is intended to provide documentation/storage of the image in the field of view of the SteriCam.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use Over-The-Counter Use (Per 21 CFR 801.109) (Division Sign-Off) Division of General Restorative Devices 1983356 510(k) Number

Image Technologies Corporation 510(k) 10/9/98 SteriCam, Coupler-Drape, and TroView System

CONFIDENTIAL

---

**Source:** [https://fda-staging.innolitics.com/device/K983567](https://fda-staging.innolitics.com/device/K983567)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
