In-vitro semi-quantitative determination of urine analytes. This is the same intended use as previously cleared for the Chemstrip Mini Urine Analyzer (K943592).
Device Story
Chemstrip 101 Urine Analyzer is a reflectance photometer for semi-quantitative urinalysis. Input: Chemstrip 10 UA urine test strips. Operation: Motor-driven automated tray moves test strip into instrument; rotating reader head with LEDs performs reflectance photometry to measure analytes; software processes reflectance data to determine analyte levels. Output: Semi-quantitative results for 10 analytes; results stored in memory (up to 100 samples) with reprint capability. Used in clinical settings by healthcare professionals. Features include a calibration fail-safe mechanism preventing testing if weekly calibration is not performed. Benefits: Automated positioning and calibration control improve consistency and reliability compared to manual systems.
Clinical Evidence
No clinical data provided; substantial equivalence based on bench testing and comparison of technological characteristics to the predicate device.
Technological Characteristics
Reflectance photometer; uses Light Emitting Diodes (LEDs) as light source; motor-driven automated test strip tray; rotating reader head; memory capacity for 100 results; calibration fail-safe mechanism. Designed for use with Chemstrip 10 UA urine test strips.
Indications for Use
Indicated for in-vitro semi-quantitative determination of urine analytes including specific gravity, pH, leukocytes, nitrite, protein, glucose, ketones, urobilinogen, bilirubin, and blood.
Regulatory Classification
Identification
An automated urinalysis system is a device intended to measure certain of the physical properties and chemical constituents of urine by procedures that duplicate manual urinalysis systems. This device is used in conjunction with certain materials to measure a variety of urinary analytes.
{0}------------------------------------------------
OCT 2 7 1998
<983510
# 510(k) Summary
| Introduction | According to the requirements established in the Food and Drug<br>Administration's guidance document entitled "The New 510(k) Paradigm:<br>Alternate Approaches to Demonstrating Substantial Equivalence in Premarket<br>Notifications", the following information provides sufficient detail to<br>understand the basis for a determination of substantial equivalence. |
|-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1) Submitter<br>name, address,<br>contact | Boehringer Mannheim Corporation<br>“doing business as Roche Diagnostics”<br>9115 Hague Rd.<br>Indianapolis, IN 46250<br>(800) 428-5074 ext. 3742<br><br>Contact Person: Jennifer Tribbett<br><br>Date Prepared: October 5, 1998 |
| 2) Device name | Proprietary name: Chemstrip 101 Urine Analyzer<br><br>Common name: Automated Urinalysis System<br><br>Classification name: System, Automated Urinalysis, 75KQO<br>Device Class I |
| 3) Predicate<br>device | We claim substantial equivalence to the currently marketed Roche<br>Diagnostics, Boehringer Mannheim Chemstrip Mini UA Urine Analyzer. |
Continued on next page
{1}------------------------------------------------
## 510(k) Summary, Continued
| 4) Device<br>Description | The Chemstrip 101 Urine Analyzer is a reflectance photometer intended for<br>in-vitro semi-quantitative reading of Chemstrip 10 UA urine test strips for the<br>following analytes: specific gravity, pH, leukocytes, nitrite, protein, glucose,<br>ketones, urobilinogen, bilirubin and blood. |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 5) Intended use | In-vitro semi-quantitative determination of urine analytes. This is the same<br>intended use as previously cleared for the Chemstrip Mini Urine Analyzer<br>(K943592). |
| 6) Comparison | The table below indicates the similarities between the Chemstrip 101 Urine |
Analyzer and the predicate device, the Chemstrip Mini UA Urine Analyzer. to predicate device:
| Topic | Chemstrip 101 Urine Analyzer | Predicate: Chemstrip Mini<br>UA Urine Analyzer |
|-----------------------|----------------------------------------------------------------------------------------------------------------|------------------------------------------------|
| Intended Use | In-vitro semi-quantitative<br>determination of urine analytes | Same |
| Scientific Technology | Reflectance Photometer | Same |
| Urine Test Strips | For use with Chemstrip 10 UA test<br>strips | Same |
| Test Parameters | Specific gravity, pH, leukocytes,<br>nitrite, protein, glucose, ketones,<br>urobilinogen, bilirubin and blood. | Same |
Continued on next page
CONFIDENTIAL
י
{2}------------------------------------------------
# 510(k) Summary, Continued
| 6. Comparison<br>to Predicate<br>Device | The Chemstrip 101 Urine Analyzer is a reflectance photometer intended for<br>in-vitro semi-quantitative reading of Chemstrip 10 UA urine test strips for the<br>following analytes: specific gravity, pH, leukocytes, nitrite, protein, glucose,<br>ketones, urobilinogen, bilirubin, and blood. |
|-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | The modifications which have been incorporated into the Chemstrip 101<br>Urine Analyzer are: |
| | 1.) an additional calibration fail-safe mechanism;<br>2.) an automated test strip holder;<br>3.) a rotating reader head, and<br>4.) increased results memory capacity |
| | Calibration Fail-Safe: |
| | The Chemstrip 101 contains a fail-safe mechanism which prevents a test from<br>being performed after the weekly instrument calibration limit. The device<br>software will prompt the user, via a message display, to "REPEAT<br>CALIBRATION" every 7 days after the last valid calibration. This fail-safe<br>will not allow the user to run a test strip until calibration is performed. |
| | In the initial (unmodified) analyzer, the "Repeat Calibration" message was<br>also displayed after the 7 day calibration limit, but a lock-out mechanism was<br>not incorporated which prevented the user from performing an actual test. |
| | This modification is an enhanced fail-safe feature. |
| | Automated Test Strip Tray: |
| | The Chemstrip 101 has eliminated the manual test strip tray by incorporating<br>a motor driven (automated) test strip tray. The user will place the strip on the<br>motorized test tray and the device will automatically move the tray into the<br>instrument for measurement. This modification allows for the automatic<br>positioning of the test strip versus the manual positioning required by the<br>initial (unmodified) analyzer. |
| | Continued on next page |
| | |
{3}------------------------------------------------
### 510(k) Summary, Continued
#### F. Device description and Reader Head: comparison
Due to the test strip holder being manual and not automated, the original analyzer had 11 reader channels with fixed positioned Light Emitting Diodes (LEDs). The Chemstrip 101 utilizes the same light source (Light Emitting Diodes) and technology (reflectance photometry) as the original analyzer, but the modified device incorporates one reader head since the strip holder is capable of moving the test strip into position under the LEDs for analyte measurement.
### Results Memory Capacity:
The initial (unmodified) analyzer did not have memory capability. The Chemstrip 101 incorporates the ability to store up to 100 sample results with the option of reprinting the results from memory.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three heads, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is surrounded by a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it.
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
OCT 2 7 1998
Jennifer Tribbett Regulatory Affairs Consultant Boehringer Mannheim 9115 Haque Road P.O. Box 50457 Indianapolis, Indiana 46250-0457
Re: K983510 Chemstrip 101 Urine Analyzer Requlatory Class: II Product Code: KQO, CDM, CEN, JIL, JIN, JIO, JIR, JJB, JMT, JRE, LJX Dated: October 5, 1998 October 7, 1998 Received:
Dear Ms. Tribbet:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions aqainst misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Druq Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Reqister. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}------------------------------------------------
Page 2
Under the Clinical Laboratory Improvement Amendments of 1988 (CLIA-88), this device may require a CLIA complexity categorization. To determine if it does, you should contact. the Centers for Disease Control and Prevention (CDC) at (770) 488-7655.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4588. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Steven Gutman
Steven I. Gutman, M.D., M.B.A. Director Division of Clinical Laboratory Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
510(k) Number (if known): Device Name: Roche Diagnostics, Boehringer Mannheim Chemstrip 101 Urine Analyzer
Indications for Use:
The Roche Diagnostics, Boehringer Mannheim Chemstrip 101 Urine Analyzer is a reflectance photometer for in-vitro semi-quantitative determination of the following urine analytes: specific gravity, pH, leukocytes, nitrite, glucose, ketones, urobilinogen, bilirubin and blood.
According to the Code of Federal Regulations, Title 21 (Food and Drugs), Part 862.2900, an automated urinalysis system is a device intended to measure certain of the physical properties and chemical constituents of urine by procedures that duplicate manual urinalysis systems. This device is used in conjunction with certain materials to measure a variety of urinary analytes.
### (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109)
OR
Over-The-Counter Use
(Optional Format 1-2-96)
u
(Division Sign-Off)
Division of Clinical Laboratory Devices
510(k) Number K983510
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.