BARD INLAY LUBRICIOUS DOUBLE PIGTAIL URETERAL STENT WITH SUTURE (HEREINAFTER REFERRED TO AS BARD LUBRICIOUS URETERAL STE
K983498 · C.R. Bard, Inc. · FAD · Dec 15, 1998 · Gastroenterology, Urology
Device Facts
Record ID
K983498
Device Name
BARD INLAY LUBRICIOUS DOUBLE PIGTAIL URETERAL STENT WITH SUTURE (HEREINAFTER REFERRED TO AS BARD LUBRICIOUS URETERAL STE
Applicant
C.R. Bard, Inc.
Product Code
FAD · Gastroenterology, Urology
Decision Date
Dec 15, 1998
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 876.4620
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Bard Lubricious Ureteral Stent is indicated to relieve obstruction in a variety of benign, malignant and post-traumatic conditions in the ureter such as presence of stones and/or stone fragments, or other ureteral obstructions such as those associated with ureteral stricture, carcinoma of abdominal organs, retroperitoneal fibrosis or ureteral trauma, or in association with Extracorporeal Shock Wave Lithotripsy (ESWL). The stent may be placed using endoscopic surgical techniques or percutaneously using standard radiographic technique.
Device Story
The Bard InLay Lubricious Double Pigtail Ureteral Stent is a polyurethane device with a hydrophilic coating designed to facilitate insertion and removal while enhancing patient comfort. It features a monofilament suture loop for retrieval. The stent is placed by physicians using endoscopic or percutaneous radiographic techniques to bypass ureteral obstructions. The device acts as a conduit to maintain ureteral patency, allowing urine flow from the kidney to the bladder. It is intended for use in clinical settings by trained medical professionals. The device is supplied sterile and disposable.
Clinical Evidence
No clinical data. Performance and functional testing were conducted in accordance with the FDA draft 'Guidance for the Content of Premarket Notifications for Ureteral Stents' (1993) and draft ASTM F.04.70.01 'Standard Test Method for Ureteral Stents' (1997).
Technological Characteristics
Polyurethane stent with hydrophilic polymer coating; monofilament nylon suture loop; radiopaque; 4.7, 6, 7, and 8 Fr. sizes. Push catheter made of High Density Polyethylene (HDPE). Sterilized, disposable. Mechanical design includes double pigtail configuration with 360° curvature and 45° overlap.
Indications for Use
Indicated for patients requiring relief of ureteral obstruction due to stones, stone fragments, ureteral stricture, carcinoma of abdominal organs, retroperitoneal fibrosis, ureteral trauma, or in association with Extracorporeal Shock Wave Lithotripsy (ESWL).
Regulatory Classification
Identification
A ureteral stent is a tube-like implanted device that is inserted into the ureter to provide ureteral rigidity and allow the passage of urine. The device may have finger-like protrusions or hooked ends to keep the tube in place. It is used in the treatment of ureteral injuries and ureteral obstruction.
Predicate Devices
Bard Lubricious-Coated Ureteral Stent with Suture (K903345)
Submission Summary (Full Text)
{0}------------------------------------------------
K9834498
Image /page/0/Picture/3 description: The image shows the word "BARD" in a bold, sans-serif font. The letters are all capitalized and appear to be filled with a solid black color. The font style is blocky and geometric, giving the word a strong and impactful appearance. The letters are evenly spaced and aligned horizontally.
### SECTION VI
# 510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.
A. Submitter's Information:
Submitter's Name: Address:
Contact Person: Contact Person's Phone: Contact Person's Fax: Date of Preparation:
C. R. Bard, Inc., Urological Division 8195 Industrial Blvd. Covington, Georgia 30014 Georgia C. Abernathy (770) 784-6454 (770) 784-6419 September 30, 1998
- B. Device Name:
| Trade Name: | Bard® InLay™ Lubricious Double Pigtail Ureteral Stent with<br>Suture |
|----------------------|----------------------------------------------------------------------|
| Common / Usual Name: | Bard Lubricious Ureteral Stent |
| Classification Name: | Ureteral Stent |
- C. Predicate Device Name:
Bard Lubricious-Coated Ureteral Stent with Suture Trade Name:
- D. Device Description:
The Bard Lubricious Ureteral Stent is a polyurethane stent with a hydrophilic (lubricious) coating which is intended to aid in stent insertion/removal and which has the potential to enhance patient comfort while indwelling; the stent also incorporates a monofilament suture loop which aids in stent removal.
- E. Intended Use:
The Bard Lubricious Ureteral Stent is indicated to relieve obstruction in a variety of benign, malignant and post-traumatic conditions in the ureter such as presence of stones and/or stone fragments, or other ureteral obstructions such as those associated with ureteral stricture, carcinoma of abdominal organs, retroperitoneal fibrosis or ureteral trauma, or in association with Extracorporeal Shock Wave Lithotripsy (ESWL). The stent may be placed using endoscopic surgical techniques or percutaneously using standard radiographic technique.
{1}------------------------------------------------
- F. Technological Characteristics Summary:
-
Table VI-1 provides a tabulated comparison summary of the technological characteristics of the Bard Lubricious Ureteral Stent versus the predicate device.
| Product Characteristic | Bard Lubricious Ureteral Stent<br>(this 510(k)) | Bard Lubricious-Coated Ureteral<br>Stent with Suture<br>(Predicate device) (#K903345) | Difference |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|
| | Stent | | |
| Indications or Intended<br>Use | The Bard Lubricious Ureteral Stent is<br>indicated to relieve obstruction in a<br>variety of benign, malignant and post-<br>traumatic conditions in the ureter such as<br>presence of stones and/or stone<br>fragments, or other ureteral obstructions<br>such as those associated with ureteral<br>stricture, carcinoma of abdominal organs,<br>retroperitoneal fibrosis or ureteral<br>trauma, or in association with<br>Extracorporeal Shock Wave Lithotripsy<br>(ESWL). The stent may be placed using<br>endoscopic surgical techniques or<br>percutaneously using standard<br>radiographic technique. | The Bard Lubricious-Coated Ureteral<br>Stent with Suture is indicated to relieve<br>obstruction in a variety of benign,<br>malignant and post-traumatic conditions<br>in the ureter such as presence of stones<br>and/or stone fragments, or other ureteral<br>obstructions such as those associated<br>with ureteral stricture, carcinoma of<br>abdominal organs, retroperitoneal<br>fibrosis or ureteral trauma, or in<br>association with Extracorporeal Shock<br>Wave Lithotripsy (ESWL). The stent<br>may be placed using endoscopic surgical<br>techniques or percutaneously using<br>standard radiographic technique. | None |
| Disposable | Yes | Yes | None |
| Sterile | Yes | Yes | None |
| Stent Base Material | Polyurethane | Polyurethane | None |
| X-Ray Opaque | Yes | Yes | None |
| Coating<br>Double Pigtail | Hydrophilic polymer (lubricious)* | Hydrophilic polymer (lubricious) | Different formula,<br>different supplier |
| Multilength | Hydrophilic polymer (lubricious) | None | Not previously<br>coated |
| Fr. Sizes Available | 4.7**, 6, 7 and 8 Fr. | 4.7, 6, 7 and 8 Fr. | 4.7 Fr. stent will<br>have a taper on the<br>kidney-end tip |
| Double Pigtail Lengths | 14, 20-30cm* | 14-30cm | 16 and 18cm<br>lengths deleted |
| Multilength Lengths | 22-32cm (one overall adjustable length) | 22-32cm (one overall adjustable length) | None |
| Pigtail Geometry<br>Double Pigtail | 360° curvature + 45° overlap (both ends) | "J" curl configuration (no overlap) | Additional curl<br>length |
| Multilength<br>Suture Loop | Yes-USP Medical Grade black nylon<br>monofilament; 3-0 | Yes-USP Medical Grade dark blue nylon<br>monofilament; 3-0 | Color change, both<br>biocompatible |
| Guidewire Interface | 4.7 Fr. = .035" diameter*<br>6, 7, 8 Fr. = .038" diameter* | 4.7 Fr. = .025" diameter<br>6, 7, 8 Fr. = .038" diameter* | 4.7 Fr. interfaces<br>with larger diameter<br>guidewire |
| Stent Clamp? | Yes | Yes | None |
| Table VI-1 |
|-----------------------------------------------------|
| Comparison Summary of Technological Characteristics |
{2}------------------------------------------------
| Product Characteristic | Bard Lubricious Ureteral Stent<br>(this 510(k)) | Bard Lubricious-Coated Ureteral<br>Stent with Suture<br>(Predicate device) (#K903345) | Difference |
|------------------------|-----------------------------------------------------------|---------------------------------------------------------------------------------------|----------------------------------------------------------------|
| Push Catheter | | | |
| Color | Orange* | Green | Color change |
| Material | High Density Polyethylene (HDPE) | High Density Polyethylene (HDPE) | None |
| Guidewire Interface | 4.7 Fr. = .035" diameter*<br>6, 7/8 Fr. = .038" diameter* | 4.7 Fr. = .025" diameter<br>6/7/8 Fr. = .038" diameter | 4.7 Fr. interfaces<br>with larger diameter<br>guidewire |
| Radiopacity | Radiopaque marker band near distal end* | N/A | Add radiopaque<br>band |
| Length | 4.7, 6, 7, 8 Fr. = 17.75"* | 4.7 Fr. = 22"<br>6/7/8 Fr. = 16" | Length changed,<br>OD on new push<br>catheter same as<br>stent |
New feature(s) or change this 510(k) *
- ** 4.7 Fr. stent will have a taper on the kidney-end tip
#### G. Performance Data Summary:
The Bard Lubricious Ureteral Stent referenced in this submission is held to the same design, manufacture, and performance specifications as those stents currently manufactured. Performance and functional testing standards are based on the FDA draft "Guidance for the Content of Premarket Notifications for Ureteral Stents" dated February 10, 1993 and the draft ASTM F.04.70.01, "Standard Test Method for Ureteral Stents" dated July 9, 1997.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the seal is an abstract symbol that resembles an eagle or bird-like figure with stylized wings and body. The symbol is rendered in a thick, black line.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 1 5 1998
Georgia C. Abernathy Regulatory Affairs Associate C.R. Bard, Inc. 8195 Industrial Blvd. Covington, GA 30209
Re: K983498
Bard® Inlay™ Lubricious Double Pigtail Ureteral Stent Dated: September 30, 1998 Received: October 5, 1998 Regulatory Class: II 21 CFR 876.4620/Procode: 78 FAD
Dear Ms. Georgia:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Eederal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compirance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Capt. Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### SECTION I - D
## INDICATIONS FOR USE STATEMENT
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: Bard Lubricious Ureteral Stent
Indications for Use:
The Bard Lubricious Ureteral Stent is indicated to relieve obstruction in a variety of benign, malignant and post-traumatic conditions in the ureter such as presence of stones and/or stone fragments, or other ureteral obstructions such as those associated with ureteral stricture, carcinoma of abdominal organs, retroperitoneal fibrosis or ureteral trauma, or in association with Extracorporeal Shock Wave Lithotripsy (ESWL). The stent may be placed using endoscopic surgical techniques or percutaneously using standard radiographic technique.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
CONCURRENCE OF CDRH, OFFICE OF DEVICE EVALUATION (ODE)
Prescription Use (Per 21 CFR 801.109)
،
Over The-Counter Use ____________________ OR
(Optional Format 1/2/96)
David L. Ingram
ivision Sign-Off) Division of Reproductive, Abdominal, ENT and Radiological Devices 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.