← Product Code [GWG](/productcode/GWG) · K983365

# AESCULAP MINOP SYSTEM (K983365)

_Aesculap, Inc. · GWG · Dec 18, 1998 · Neurology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K983365

## Device Facts

- **Applicant:** Aesculap, Inc.
- **Product Code:** [GWG](/productcode/GWG.md)
- **Decision Date:** Dec 18, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.1480
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

Aesculab's MINOP System is intended for use in endoscope-assisted microneurosurgery and pure neuroendoscopy (i.e. ventriculoscopy) for direct visualization, diagnostic and/or therapeutic procedures such as ventriculostomies, biopsies and removal of cysts, tumors and other obstructions.

## Device Story

The MINOP System is a neuroendoscopic surgical toolset used in clinical neurosurgery settings by neurosurgeons. It consists of rigid, rod-lens neuroendoscopes (0° or 30° view), operating cannulas (1, 3, or 4 working channels), instruments, and electrodes. The system provides direct visualization of the surgical site; the eyepiece is angled at 90° to facilitate simultaneous use with a surgical microscope. An optional flexible titanium support arm allows for hands-free positioning of the endoscope. By providing clear visualization and access channels, the device enables surgeons to perform minimally invasive diagnostic and therapeutic procedures, such as biopsies and tumor removal, potentially reducing surgical trauma and improving patient outcomes.

## Clinical Evidence

No clinical data. Bench testing only. The system underwent evaluation for electrical, thermal, and irrigation safety, and complies with IEC 601-2-18 requirements for endoscopic equipment.

## Technological Characteristics

Rigid rod-lens neuroendoscopes (0°/30° view, 90° eyepiece); stainless steel construction; operating cannulas with 1, 3, or 4 working channels; optional flexible titanium support arm. Complies with IEC 601-2-18. Non-sterile, steam sterilizable.

## Regulatory Identification

A neurological endoscope is an instrument with a light source used to view the inside of the ventricles of the brain.

## Predicate Devices

- Aesculap Ventriculopscope
- Aesculap Angled Neuroendoscopes
- Codman Neuroendoscopes
- Neuro Navigational Neuroview
- Richard Wolf Neuro-Endoscope

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the text "DEC 18 1998" in a light gray font at the top. Below this text is the word "AESCULAP" in a large, bold, black font. The word is followed by a period.

K983365

Aesculan, Inc 000 Gateway Blvd. San Francisco. CA 94080-7028 (650) 876-7000 650) 876-7028

.aesculap.com

### 510(k) Summary of Safety and Effectiveness in Accordance with SMDA of 1990

#### Aesculap MINOP System

December 15, 1998

Aesculap®, Inc. Submitted by: 1000 Gateway Blvd. So. San Francisco, CA 94080 Contact: Mary Ellen Holden (650) 876-7000 x5072 Phone: Direct line: (650) 624-5072 (650) 589-3007 FAX:

> Product: Aesculap MINOP System Common Name: Neuroendoscopes

#### Intended Use

Aesculab's MINOP System is intended for use in endoscope-assisted microneurosurgery and pure neuroendoscopy (i.e. ventriculoscopy) for direct visualization, diagnostic and/or therapeutic procedures such as ventriculostomies, biopsies and removal of cysts, tumors and other obstructions.

#### Technological Characteristics

The MINOP System consists of trocars (cannulas with blunt obturators). neuroendoscopes, instruments and electrodes. There are three different styles of operating cannulas which function as the base for the device system. The MINOP System operating cannulas have either 1, 3, or 4 working channels.

The Neuroendoscopes for the MINOP System are rigid, rod-lens neuroendoscopes with either 0° or 30° direction of view. The lens evepiece is angled at 90° to allow an unobstructed view when used in conjunction with a surgical microscope. The MINOP System consists of reusable devices composed primarily of stainless steel.

An optional holding device, Aesculap's Flexible Titanium Support Arm, can be used to position and hold the neuroendoscopes. With use of the flexible support arm, the surgeon can freely operate with both hands.

## B BRAUN

{1}------------------------------------------------

## Performance Standards

No applicable performance standards have been promulgated under Section 514 of the Food, Drug and Cosmetic Act for these devices. However, Aesculap's MINOP System complies with the requirements of IEC 601-2-18 (Medical electrical equipment, Part 2: Particular requirements for the safety of endoscopic equipment.) The MINOP System has undergone evaluation for electrical, thermal and irrigation safety to ensure the system is safe for the intended use.

## Sterilization

The MINOP System is provided non-sterile and must be sterilized prior to use. The devices may be sterilized by steam sterilization.

## Substantial Equivalence

Aesculap's MINOP System shares similar features and function with corresponding devices distributed by:

- Aesculap (Ventriculopscope, Angled Neuroendoscopes) の
- Codman (Neuroendoscopes) 0
- Neuro Navigational (Neuroview) 0
- o Richard Wolf (Neuro-Endoscope)

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the caduceus. The logo is black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DEC 1 8 1998

Ms. Mary Ellen Holden Senior Regulatory Associate Aesculap, Inc. 1000 Gateway Boulevard South San Francisco, California 94080-7030

Re: K983365 Trade Name: Aesculap MINOP System Regulatory Class: II Product Code: GWG Dated: September 23, 1998 Received: September 24, 1998

Dear Ms. Holden:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (OS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

{3}------------------------------------------------

#### Page 2 -- Ms. Mary Ellen Holden

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4595. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".

Sincerely yours.

pcolle

Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# INDICATION FOR USE STATEMENT

983365 510(k) Number (if known):

Device Name:

Aesculap MINOP System

Indication for Use:

Aesculap's MINOP System is indicated for use in endoscope-assisted microneurosurgery and pure neuroendoscopy (i.e. ventriculoscopy) for direct visualization, diagnostic and/or therapeutic procedures such as ventriculostomies, biopsies and removal of cysts, tumors and other obstructions.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

t.coelloto

(Division Sign-Off)
Division of General Restorative Devices
510(k) Number 12983365

**Prescription Use**
(Per 21 CFR 801.109)

---

**Source:** [https://fda-staging.innolitics.com/device/K983365](https://fda-staging.innolitics.com/device/K983365)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
