Brachytherapy planning with PLATO Brachytherapy is intended for use with brachytherapy procedures, i.e. intercavitary, interstitial, intraluminal (including bronchial) and surface applicator treatments, involving manual or remote afterloading radioactive sources. The software program provides the physician with anatomical and dosimetric information (i.e. radiographs, transverse slices, or 3D coordinates) to determine the positioning and loading of the radioactive sources prior to insertion. The software also provides the treatment time and dose distribution for the specified loading. From this information the patient can be treated with radioactive sources. With remote afterloading brachytherapy treatment planning the PLATO Brachytherapy program exports the treatment data, i.e. treatment times to the remote afterloader. With manual brachytherapy treatment planning the program exports treatment data regarding insertion/removal time, total dose in permanent implantation and dose distributions for patient records. With both types of software the dose grid is exported and can be evaluated in conjunction with the Nucletron PLATO External Beam treatment planning software. The PLATO Brachytherapy (BPS v.14.0) is intended to be used for medical procedures on patients to be prescribed and performed by a suitably trained and certified medical professional.
Device Story
PLATO Brachytherapy (BPS v.14.0) is a software-based treatment planning system for brachytherapy. Input data includes patient anatomical information via radiographs, transverse slices, or 3D coordinates. The system processes this data to define radioactive source locations, reference points, and prescription doses; it calculates dose distributions and treatment times. The software provides evaluation tools like dose-volume histograms, verification points, and dose profiles. Used in clinical settings by trained medical professionals, the system outputs treatment data (times, loading parameters) for export to remote afterloading units via floppy disk, program card, or network, or for manual record-keeping. The software does not control the treatment unit; clinical staff verify all exported data. The system assists physicians in optimizing dose distribution prior to source insertion, potentially improving treatment accuracy and patient outcomes.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Software-based treatment planning system; runs on a dedicated workstation. Features include 3D reconstruction of implants, dose calculation algorithms, and data export capabilities for remote afterloaders. Connectivity via network, floppy, or program card. No specific hardware materials or sterilization methods applicable as it is a software-only accessory.
Indications for Use
Indicated for patients undergoing brachytherapy procedures, including intercavitary, interstitial, intraluminal (e.g., bronchial), and surface applicator treatments, requiring manual or remote afterloading of radioactive sources.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
{0}------------------------------------------------
4/2/99
: :
Premarket Notification Nucletron PLATO Brachytherapy Date : 20 September 1998
Image /page/0/Picture/2 description: The image shows the word "Nucletron" in a bold, sans-serif font. To the left of the word is a square logo. The logo contains a black circle with two white circles on either side. The logo and the word "Nucletron" are aligned horizontally.
# CONFIDENTIAL
K983343
NUCLETRON B.V. Waardgelder 1 3905 TH Veenendaal P.O.Box 930 3900 AX Veenendaal The Netherlands Phone +31 318 557133 +31 318 550485 Fax
Department of Health and Human Services Center of Devices and Radiological Health Office of Device Evaluation Pre-Market Notification section
## 510(K) SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION
as required by section 807.92(c)
#### Submitter of 510(k) a.
| Company name: | Nucletron Corporation |
|-----------------|--------------------------------------------------------|
| Registration # | 1121753 |
| Address: | 7080 Columbia Gateway Drive<br>Columbia, MD 21046-2133 |
| Contact Person: | Ralph E. Shuping<br>Regulatory Affairs Manager |
| Phone:. | 410-312-4100 |
| Fax: | 410-312-4197 |
#### b. Device Name:
| Trade/Proprietary Name: | PLATO Brachytherapy (BPS v. 14.0) |
|-------------------------|--------------------------------------------------------------|
| Common/Usual Name: | PLATO Brachytherapy planning system |
| Classification Name: | Accessory to remote afterloader<br>21 CFR 892.5700 Class II. |
#### Legally Marketed Predicate Device(s) C.
Our device is substantially equivalent to the legally marketed predicate devices cited in the table below.
| Manufacturer | Device | 510(k) # |
|--------------|-------------------------------------|----------|
| Nucletron BV | PLATO Brachytherapy Planning System | K915562 |
| Nucletron BV | Nucletron Planning System | K911612 |
Appendix 5, page 2
{1}------------------------------------------------
Premarket Notification Nucletron PLATO Brachytherapy Date : 20 September 1998
# CONFIDENTIAL
#### Description d.
PLATO Brachytherapy as described in this submission is a software package designed for treatment planning of brachytherapy procedures. This software runs on a PLATO radiation therapy treatment planning system workstation
Nucletron PLATO Brachytherapy software is capable of reconstructing the brachytherapy implant (i.e. radiographs, transverse slices, or 3D coordinates), defining the location of the radioactive sources within the implant, identifying a reference point, i.e. dose points, anatomical points. Once this data is entered the prescription dose is defined and the software program calculates treatment data, including dose distributions and treatment/implantation time.
The brachytherapy treatment planning session allows the physician to evaluate the implant prior to insertion of the radioactive sources in order to determine the most optimal dose distribution within the treatment volume. Once the physician approves the treatment plan the implant is loaded with radioactive sources manually or via the Nucletron remote afterloading equipment. The PLATO Brachytherapy software exports the treatment data to the Nucletron remote afterloading system via a floppy, program card or network. The PLATO Brachytherapy software does not control the treatment unit; it strictly exports treatment times and related information. The Nucletron remote afterloading system and the clinical staff verify this data input prior to treatment.
Nucletrons' PLATO Brachytherapy software includes remote afterloading radioactive source brachytherapy treatment planning and manual loaded radioactive source brachytherapy treatment planning. The program provides a variety of plan evaluation tools to assist in generating the most optimal dose distribution. i.e. dose volume histograms, dose verification points, dose profiles, etc.
#### Intended use e.
Brachytherapy planning with PLATO Brachytherapy is intended for use with brachytherapy procedures involving manual or remote afterloading radioactive sources. The software program provides the physician with anatomical and dosimetric information to determine the positioning and loading of the radioactive sources prior to insertion. The software also provides the treatment time and dose distribution for the specified loading. From this information the patient can be treated with radioactive sources.
With remote afterloading brachytherapy treatment planning the PLATO Brachytherapy program exports the treatment data, i.e. treatment times to the remote afterloader. With manual brachytherapy treatment planning the program exports treatment data regarding insertion/removal time, total dose in permanent implantation and dose distributions for patient records. With both types of software the dose grid is exported and can be evaluated in conjunction with the Nucletron PLATO External Beam treatment planning software.
{2}------------------------------------------------
CONFIDENTIAL
Premarket Notification Nucletron PLATO Brachytherapy Date : 20 September 1998
#### Summary of technological considerations f.
The PLATO Brachytherapy software is substantially equivalent to the predicate devices. It combines the functionality of the defined predicate devices into one user interface for brachytherapy treatment planning.
Signature
18- Sept-98
Date
Name: L. A. Johnson Product Manager Title Nucletron BV Veenendaal Netherlands
ﯩﺖ.
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Re:
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, consisting of a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the caduceus.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 2 1999
Ralph E. Shuping, Sc.D. Regulatory Affairs Manager Nucletron Corporation 7080 Columbia Gateway Drive Columbia, MD 21046-2133
K983343 PLATO Brachytherapy (BPS v14.0) Dated: January 7, 1999 Received: January 11, 1999 Regulatory class: II 21 CFR 892.5700/Procode: 90 MUJ
Dear Dr. Shuping:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
APT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## CONFIDENTIAL
Premarket Notification Nucletron PLATO Brachytherapy Date : 20 September 1998
Image /page/4/Picture/2 description: The image shows the word "Nucletron" in a bold, sans-serif font. To the left of the word is a square logo. The logo contains a black circle with two white circles on either side of it.
### NUCLETRON B.V.
Waardgelder 1 3905 TH Veenendaal 3900 AX Veenendaal P.O.Box 930 The Netherlands Phone +31 318 557133 +31 318 550485 Fax
Department of Health and Human Services Center of Devices and Radiological Health Office of Device Evaluation Pre-Market Notification section
## Statement of intended use
Device Name: PLATO Brachytherapy
## Intended use
Brachytherapy planning with PLATO Brachytherapy is intended for use with brachytherapy procedures, i.e. intercavitary, interstitial, intraluminal (including bronchial) and surface applicator treatments, involving manual or remote afterloading radioactive sources. The software program provides the physician with anatomical and dosimetric information (i.e. radiographs, transverse slices, or 3D coordinates) to determine the positioning and loading of the radioactive sources prior to insertion. The software also provides the treatment time and dose distribution for the specified loading. From this information the patient can be treated with radioactive sources.
With remote afterloading brachytherapy treatment planning the PLATO Brachytherapy program exports the treatment data, i.e. treatment times to the remote afterloader. With manual brachytherapy treatment planning the program exports treatment data regarding insertion/removal time, total dose in permanent implantation and dose distributions for patient records. With both types of software the dose grid is exported and can be evaluated in conjunction with the Nucletron PLATO External Beam treatment planning software.
### Prescription use
The PLATO Brachytherapy (BPS v.14.0) is intended to be used for medical procedures on patients to be prescribed and performed by a suitably trained and certified medical
professional.
Name L. A. Johnson
Product Manager Title Nucletron BV Veenendaal Netherlands
18-Sept.-98
Date
Yhmil h. Slym
(Division Sign-Off)
Division of Reproductive, Abdominal, El
510(k) Number K983342
Appendix 5, page 5
Prescription Use . (Per 21 CFR 801.109)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.