MEDI-TRACE 1610C COMBINATION DEFIBRILLATION, MONITORING AND PACING ELECTRODE
Applicant
Graphic Controls Corp.
Product Code
MKJ · Cardiovascular
Decision Date
Nov 5, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5310
Device Class
Class 3
Attributes
Therapeutic
Indications for Use
Intended for use in defibrillation procedures, cardioversion and pacing. This device is intended for use on adults.
Device Story
Medi-Trace 1610C is a pre-gelled, self-adhesive electrode for defibrillation, cardioversion, pacing, and ECG monitoring. Device consists of pre-attached leadwire, vinyl substrate, crimped wire connector, metal foil layer, conductive adhesive hydrogel, pressure-sensitive adhesive (PSA) ring, and release liner. Packaged in pairs in heat-sealed pouches. Used by clinicians in clinical settings to deliver electrical therapy and monitor cardiac signals. Device functions as interface between defibrillator/pacer and patient skin; conductive hydrogel ensures electrical contact for therapy delivery and signal acquisition. Benefits include simplified application and multi-function capability for emergency cardiac care.
Clinical Evidence
Bench testing only. Device met AAMI DF-39 specifications for electrical performance, energy throughput, and simulated use (multiple defibrillation shocks, maximum current pacing, and ECG monitoring). Biocompatibility testing per ISO-10993 confirmed device is non-irritating, non-cytotoxic, and non-sensitizing. Accelerated shelf-life studies performed.
Technological Characteristics
Materials: vinyl substrate, metal foil, conductive adhesive hydrogel, pressure-sensitive adhesive. Form factor: self-adhesive electrode with pre-attached leadwire. Connectivity: wired interface to defibrillator/pacer. Testing standards: AAMI DF-39, ISO-10993.
Indications for Use
Indicated for use in defibrillation, cardioversion, pacing, and ECG monitoring in adult patients.
Regulatory Classification
Identification
An automated external defibrillator (AED) system consists of an AED and those accessories necessary for the AED to detect and interpret an electrocardiogram and deliver an electrical shock (e.g., battery, pad electrode, adapter, and hardware key for pediatric use). An AED system analyzes the patient's electrocardiogram, interprets the cardiac rhythm, and automatically delivers an electrical shock (fully automated AED), or advises the user to deliver the shock (semi-automated or shock advisory AED) to treat ventricular fibrillation or pulseless ventricular tachycardia.
Predicate Devices
Cardiotronics Systems, Inc., a division of Ballard Medical Products, R2® Pads™ Pacing/Defibrillation and ECG Electrode
Submission Summary (Full Text)
{0}------------------------------------------------
5 1998 NOV
K98328
## 510(k) SUMMARY
المشاركة المستوى المنتخب المنتخب المنتخب المنتخب المنتخب المنتخب المنتخب المنتخب المنتخب المنتخب المنتخب المنتخب المنتخب المنتخب المنتخب المنتخب المنتخب المنتخب المنتخب المن
| | Date: | Graphic Controls<br>September 16, 1998 |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Manufacturing Facility: | Graphic Controls Corporation<br>189 Van Rensselaer Street<br>P.O. Box 1271<br>Buffalo, NY 14240<br>Registration Number 1317188 |
| | Telephone: | (716) 853-7500 |
| | Contact Person: | Kathleen Selover<br>Regulatory Affairs Specialist<br>(716) 853-7500, extension 7630<br>Fax: (716) 847-7531 |
| | Device Trade Name: | Medi-Trace® 1610C Combination Defibrillation,<br>Monitoring and Pacing Electrode |
| | Device Common Name: | Self Adhesive Defibrillation and Monitoring<br>Electrode |
| | Classification Name: | Electrode, Electrocardiograph, Multifunction |
| | Regulatory Reference: | 74MLN |
| | Predicate Device: | Cardiotronics Systems, Inc., a division of Ballard<br>Medical Products,.R2® Pads™ Pacing/Defibrillation<br>and ECG Electrode |
| | Description: | A pre-gelled conductive electrode consisting of a<br>pre-attached leadwire, and insulating layer of a vinyl<br>substrate, a secondary vinyl piece covering the<br>crimped wire connector, a layer of metal foil, a<br>conductive adhesive hydrogel, a pressure sensitive<br>adhesive (PSA) ring and a release liner. Device is<br>packaged in pairs in a heat sealed pouch; 10 pouches<br>are packaged into one shelf box shipper. |
| Intended/Indications for Use: | Intended for use in defibrillation procedures,<br>cardioversion and pacing. This device is intended for<br>use on adults. | |
| Physical/Technical Comparison: | Medi-Trace® 1610C Electrode is replaceable with<br>the Cardiotronics, Inc, a division of Ballard Medical<br>Products, R2® Pads™ Pacing/Defibrillation/ECG<br>Electrodes. Physical and technical characteristics,<br>including materials used in construction, size,<br>intended use and conductive gel type of these<br>electrodes are comparable. Since the formulation of<br>conductive gel of the predicate is considered to be<br>proprietary, exact chemical comparison could not be<br>made. | |
| Performance Summary: | The device and the predicate were subject to AAMI<br>electrical tests as described in DF-39, 3.3.19 and<br>energy through put testing. Test results for both the<br>device and the predicate meet the specifications as<br>established in DF-39 for self adhesive electrodes for<br>monitoring and defibrillation and pacing. | |
| | In addition, the device was subjected to simulated use<br>testing consisting of multiple defibrillation shocks,<br>maximum current pacing and ECG monitoring. Test<br>results for the device met the specifications as<br>established in DF-39 for self adhesive electrodes for<br>monitoring, defibrillation and pacing. | |
| Biocompatibility Testing: | The device was subjected to biocompatibility testing<br>as recommended in ISO-10993, Biological<br>Evaluation of Medical Devices Part I: Evaluation<br>and Testing." The device was found to be<br>non-irritating, non-cytotoxic and non-sensitizing. | |
| Shelf Life: | Data obtained in accelerated shelf life studies were<br>reviewed and found to substantiate our claimed shelf<br>life. | |
{1}------------------------------------------------
510(k) Summary
Page 2
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle faces right and appears to be in flight. Encircling the eagle is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 5 1998
Susan A. Krasny, Ph.D. Graphic Controls Corporation 189 Van Rensselaer Street P.O. Box 1271 Buffalo, NY 14240-1271
K983287 Re: Medi-Trace® 1610C Combination Defibrillation, Monitoring and Pacing Electrode Requlatory Class: III (Three) Product Code: 74 MLN Dated: September 16, 1998 Received: September 18, 1998
Dear Dr. Krasny:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual reqistration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in requlatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{3}------------------------------------------------
Page 2 - Susan A. Krasny, Ph.D.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
sincerely yours,
Thomas J. Callahon
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page _ 1 _ of ________________________________________________________________________________________________________________________________________________________________
510(k) Number, if known: _____________________________________________________________________________________________________________________________________________________ K983287
______________________________________________________________________________________________________________________________________________________________________________
Medi-Trace® 1610C Combination Defibrillation, Device Name: Monitoring and Pacing Electrode
Indications for Use:
The Medi-Trace® 1610C Combination Defibrillation, Monitoring and Pacing Electrode is intended for use in defibrillation procedures, cardioversion, pacing and ECG monitoring on adults.
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
______________________________________________________________________________________________________________________________________________________________________________
______________________________________________________________________________________________________________________________________________________________________________
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mark Kramer
| Prescription Use | OR | Over-the-Counter Use (Per 21 CFR 801.109) |
|------------------|----|-------------------------------------------|
|------------------|----|-------------------------------------------|
(Division Sign-Off)
Division of Cardiovascular, Respiratory, and Neurological Devices
| 510(k) Number | |
|---------------|--|
|---------------|--|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.