K983274 · Nellcor Puritan Bennett, Inc. · BZK · Dec 9, 1998 · Anesthesiology
Device Facts
Record ID
K983274
Device Name
STARTRACK INFANT GRAPHICS MONITOR
Applicant
Nellcor Puritan Bennett, Inc.
Product Code
BZK · Anesthesiology
Decision Date
Dec 9, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.1850
Device Class
Class 2
Attributes
Pediatric
Indications for Use
StarTrack is intended for prescription use only for the measurement and display of breathing flow, volume and pressure as delivered from a ventilator through the patient's endotracheal tube. StarTrack is not an Apnea Monitor. The intended patient population comprises neonatal and pediatric patients who require an endotracheal tube <5.0mm I.D. and/or do not exceed a peak breathing flow of 30 L/Min directly through the ET tube. The intended environments of use comprise Hospital type and Intra-Hospital Transport environments. Hospital use typically covers such areas as general care floors, operating rooms, special procedure areas, intensive and critical care areas, within the hospital plus hospital-type facilities such as Surgicenters, Sub-acute Centers and Special Nursing Facilities, outside of the hospital. Intra-hospital transport includes transport with a patient within the hospital or hospital-type facility.
Device Story
StarTrack Infant Graphics Monitor measures/displays breathing flow, volume, and pressure delivered via ventilator through endotracheal (ET) tube. Input: flow data via heated wire anemometer sensor; pressure data via internal or external transducer. Operation: heated wire anemometer maintains constant temperature; heating current changes proportional to flow. Output: LCD screen displays graphics (Flow/Volume/Pressure vs Time, loops) and numeric breath data (Tidal/Minute Volume, ET Leak, Breath Rate, Compliance). Used in hospital/transport settings by clinicians. Provides visual/audible alarms for patient-related (e.g., high volume/rate) and system-related (e.g., sensor disconnect) conditions. Assists clinicians in monitoring ventilation status to optimize patient care.
Clinical Evidence
Bench testing only. In-vitro testing conducted to confirm accuracy of flow, volume, and pressure measurements.
Technological Characteristics
Metal enclosure; heated wire anemometer sensor (constant temperature mode); internal pressure transducer; LCD display. Embedded software algorithm for flow calculation. Connectivity: supports external ventilator pressure transducer input.
Indications for Use
Indicated for neonatal and pediatric patients requiring endotracheal tubes <5.0mm I.D. or peak breathing flow ≤30 L/Min. Not for use as an apnea monitor.
Regulatory Classification
Identification
A monitoring spirometer is a device used to measure continuously a patient's tidal volume (volume of gas inhaled by the patient during each respiration cycle) or minute volume (the tidal volume multiplied by the rate of respiration for 1 minute) for the evaluation of the patient's ventilatory status.
Predicate Devices
Bear Medical Systems Neonatal Volume Monitor, model NVM-1 (K890724)
Submission Summary (Full Text)
{0}------------------------------------------------
3274
DEC 9 1998
Section 16, 510(k) Summary
### Section 16.a Date Summary Prepared
30 November 1998
## Section 16.b Company Information
Establishment:
Official Correspondent:
Nellcor Puritan Bennett Inc. 4280 Hacienda Drive Pleasanton, CA 94588
David A.C. Green Manager, Requiatory Affairs Nelicor Puritan Bennett Inc. 2200 Faraday Avenue Carlsbad, CA 92008 (760) 603-5978 (direct phone) (760) 603-5907 (fax)
### Section 16.c Name of Device
Proprietary:
Common/Usual:
StarTrack Infant Graphics Monitor
Volume Monitor
Classification:
Monitoring Spirometer(§868.1850/73BZK)
## Section 16.d Equivalent Devices
Substantial equivalence to the following legally marketed predicate device with the same or similar indications for use has been demonstrated by a comparison of product features as described in the labeling and promotional literature for the predicate device and the StarTrack Infant Graphics Monitor, StarTrack. Safety and environmental testing to accepted industry sfandards has been conducted as well as in-vitro testing to confirm the accuracy of StarTrack. The predicate device is as follows:
Bear Medical Systems Neonatal Volume Monitor, model NVM-1, K890724.
#### Section 16.e Device Description
The StarTrack Infant Graphics Monitor, StarTrack is contained in a rectangular metal enclosure measuring 9.25 inches wide by 6 inches high by 5.7 inches deep and weighing 7.2 Ibs.
A front panel LCD screen enables the presentation of graphics, numeric breath data and alarm settings. User-selectable graphics screens present Flow and Volume vs Time. Pressure and Volume vs Time, Flow and Pressure vs Time, Flow / Volume Loop and Pressure / Volume Loop. Numeric breath data are Expiratory Tidal Volume, Expiratory Minute Volume, ET Tube Leak, Breath Rate and Dynamic Compliance. Alarm Settings are High Minute Volume, Low Minute Volume, High ET Tube Leak, High Breath Rate and Breath Interval.
{1}------------------------------------------------
StarTrack also provides visual and audible indicators. Alarms are divided into two categories: patient-related alams and system-related alarms. A patient alarm causes a flashing bell to display next to the violated alarm setting. This is accompanied by an audible sound and the flashing Patient-Related Alarm LED. When the alarm condition is corrected, the audible sound ceases and the bell and LED stop flashing. These displays will turn off when the operator presses the Visual Reset button. Patient-related alarms can be silenced for sixty seconds by pressing the Alarm Silence button.
System-related alarms concem problems with the hardware, such as the sensor, sensor cable and the StarTrack electronics. System-related alarms have both visual and audible indicators. The audible indicators cannot be silenced. The System-Related alarms are Sensor Disconnect, Sensor Defect, Sensor Contaminated, AC Power Loss and System Failure.
#### Section 16.f Intended Use
StarTrack is intended for prescription use only for the measurement and display of breathing flow, volume and pressure as delivered from a ventilator through the patient's endotracheal tube. StarTrack is not an Apnea Monitor. The intended patient population comprises neonatal and pediatric patients who require an endotracheal tube <5.0mm I.D. and/or do not exceed a peak breathing flow of 30 L/Min directly through the ET tube. The intended environments of use comprise Hospital tvpe and Intra-Hospital Transport environments. Hospital use typically covers such areas as general care floors. operating rooms, special procedure areas, intensive and critical care areas, within the hospital plus hospital-type facilities such as Surgicenters, Sub-acute Centers and Special Nursing Facilities, outside of the hospital. Intra-hospital transport includes transport with a patient within the hospital or hospital-type facility.
## Section 16.g Technological Characteristics
StarTrack's function is to monitor the status of the patient's airway flow. volume and pressure as delivered from a ventilator through the endotracheal (ET) tube in infants. A heated wire anemometer sensor located in series with the patient airway measures flow in the constant temperature mode. Two heated wires within the airway flow sensor enable the determination of flow direction as well as flow magnitude. Current passing through the heated wires maintains a constant sensor resistance and, thus, a constant sensor temperature. As flow changes, the heat loss changes and thus the heating current changes. Knowledge of this current value enables the determination of a corresponding airway flow.
StarTrack also has the capability of displaying proximal pressure information. A pressure transducer resides within StarTrack. The operator can select either StarTrack's pressure transducer or the Infant Star Ventilator's pressure transducer or the Operator can disable pressure monitoring altogether.
The embedded software contains substantially the same software algorithm for determining airway flow values as used on the predicate device, the Bear Medical Systems Neonatal Volume Monitor, model NVM-1, K890724.
{2}------------------------------------------------
# Section 16.h Certification Statement
In accordance with the requirements of 21 CFR 807.87(j), the following certification is provided:
Nellcor Puritan Bennett Inc., believes that all data and information submitted in this Premarket Notification are truthful and accurate and no material fact has been omitted.
David A.e. Green.
David A. C. Green Manager, Regulatory Affairs for Nelicor Puritan Bennett Inc.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract design that resembles a stylized caduceus, a symbol often associated with medicine and healthcare. The design features a staff with a serpent coiled around it, representing healing and medicine.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 9 1998
Mr. David A. C. Green Nellcor Puritan Bennett Inc. 2200 Faraday Avenue Carlsbad, CA 92008
Re: K983274 StarTrack Infant Graphics Monitor Regulatory Class: II (two) Product Code: 73 BZK September 16, 1998 Dated: Received: September 17, 1998
Dear Mr. Green:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General requlation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such Failure to comply with the GMP regulation may result in assumptions. regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
Page 2 - Mr. David A. C. Green
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Thomas J. Callahan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page
| 510(k) Number (if known): | K983274 |
|---------------------------|-----------------------------------|
| Device Name: | StarTrack Infant Graphics Monitor |
# Indications For Use:
The StarTrack Infant Graphics Monitor's intended use is the measurement and display of breathing flow, volume and pressure as delivered trough the patient's endotracheal tube. The intended patient population comprises neonatal and pediatric patients requiring an endotracheal tube <5.0mm I.D. and/or who do not exceed a peak breathing flow of 30 L/Min directly through the ET tube. The intended environments of use comprise Hospital, Hospital-type and Intra-Hospital Transport environments. The device is for prescription use only. Hospital use typically covers such areas as general care floors, operating rooms, special procedure areas, intensive and critical care areas, within the hospital plus hospital-type facilities such as Surgicenters, Sub-acute Centers and Special Nursing Facilities, outside of the hospital transport includes transport with a patient within the hospital or hospital-type facility.
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109) \$\checkmark\$
OR
Over-The-Counter Use
(Optional Format 1-2-96)
(Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices
510(k) Number K963274
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.