INVASIVE BLOOD PRESSURE (IBP) FUNCTION, MODEL 9500 MULTIGAS MONITOR

K983250 · Magnetic Resonance Equipment Corp. · DSK · Dec 12, 1999 · Cardiovascular

Device Facts

Record IDK983250
Device NameINVASIVE BLOOD PRESSURE (IBP) FUNCTION, MODEL 9500 MULTIGAS MONITOR
ApplicantMagnetic Resonance Equipment Corp.
Product CodeDSK · Cardiovascular
Decision DateDec 12, 1999
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1110
Device ClassClass 2

Indications for Use

Invasive blood pressure function is an option for the Model 9500 Multigas monitor. This function would be used when a clinician determined the need for continuous monitoring of arterial or venous blood pressure for adult patients undergoing MRI examination.

Device Story

Invasive Blood Pressure (IBP) module for Model 9500 Multigas Monitor; provides two isolated channels for invasive pressure measurement in MRI environment. Inputs: arterial/venous pressure signals. Processing: signal acquisition and display of systolic, diastolic, and mean pressures; waveform visualization; trend data collection. Output: real-time pressure values and waveforms on monitor display. Used in clinical MRI settings by healthcare professionals. Benefits: enables continuous hemodynamic monitoring during MRI procedures where standard monitors may be unsuitable.

Clinical Evidence

Bench testing only. The device was tested for compliance with IEC 601-1, IEC 601-1-2, IEC 801-2, -3, -4, -5, and CISPR 11 standards. Performance testing confirmed the device operates in the MRI environment substantially equivalent to the predicate device.

Technological Characteristics

Two-channel isolated invasive blood pressure measurement module. Designed for integration into Model 9500 Multigas Monitor. Operates in MRI environment. Compliance with IEC 601-1 (general safety), IEC 601-1-2 (EMC), and CISPR 11 (electromagnetic interference).

Indications for Use

Indicated for adult patients undergoing MRI examination requiring continuous arterial or venous blood pressure monitoring.

Regulatory Classification

Identification

A blood pressure computer is a device that accepts the electrical signal from a blood pressure transducer amplifier and indicates the systolic, diastolic, or mean pressure based on the input signal.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ DEC 2 9 1999 K9f325-0 510(k) Summary ## Summary of Safety and Effectiveness | Submitter: | MR Equipment Corporation | |-----------------------------|--------------------------------------------------------------------------| | Address: | 6 Washington Avenue<br>Bay Shore, NY, 11706 | | Telephone: | (516) 243-3500 | | Contact: | President | | Prepared: | December 15, 1997 | | Proprietary Name: | Invasive Blood Pressure function for Model 9500<br>Multigas Monitor | | Common/Classification Name: | Invasive Blood Pressure function | | Predicate Devices: | BCI Model 9100 Multigas Monitor with Invasive Blood<br>Pressure function | New Device Description: The Invasive Blood Pressure function is an addition to a device legally marketed by MR Equipment Corporation. The device provides an additional measurement function to the Model 9500 Multigas Monitor. The existing 9500 monitor has pulse oximetry, ECG, non-invasive blood pressure (NIBP), capnograph and, optionally, oxygen and agent gas measurement functions. The Model 9500 is based on the BCI Model 9100 monitor with the same measuring functions, but the Model 9500 is specifically intended for use in the MR environment. Intended Use The Invasive Blood Pressure function provides for two isolated channels of invasive blood pressure measurement in the MR imaging environment. The IBO function is an optional module for the Model 9500. When active, the Invasive Pressure display shows the label chosen, systolic pressure and (optionally) its high alarm limit, diastolic pressure and (optionally) its low alarm limit, and the mean pressure. Waveforms may be displayed, and trend data collected for long term monitoring. {1}------------------------------------------------ ## 510(k) Summary ## Performance Testing: The Model 9500 Multigas monitor meets IEC 601-1, IEC 601-1-2, IEC 801-2, -3, -4, -5, CISPR 11. In addition, tests were performed to establish that the Model 9500 operated in the MRI environment substantially equivalent to the predicate device. This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The logo is black and white. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DEC 2 9 1999 6. Ronald Morris, Ph.D. Magnetic Resonance Equipment Corporation 6 Washington Avenue P.O. Box 5489 Bay Shore, NY 11706 Re: K983250 Model 9500 Multigas Monitor Requlatory Class: II (two) Product Code: 74 DSK Dated: September 30, 1999 Received: October 8, 1999 Dear Dr. Morris: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to leqally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual reqistration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Requlation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. {3}------------------------------------------------ Page 2 - G. Ronald Morris, Ph.D. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4648. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html". Sincerely yours, Ser, Lee R. Campbell Celia M. Witten, Ph.D. Acting Director Division of Cardiovascular, Respiratory, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ K983250 510(k) NUMBER (IF KNOWN) : Invasive Blood Pressure Function, Model 9500 Multigas Monitor DEVICE NAME: INDICATIONS FOR USE: Invasive blood pressure function is an option for the Model 9500 Multigas monitor. This function would be used when a clinician determined the need for continuous monitoring of arterial or venous blood pressure for adult patients undergoing MRI examination. Bette L. Campbell (Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices 510(k) Number_K98 3250 (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.) Concurrence of CDRH, Office of Device Evaluation i (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter-Use (Optional Format 1-
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...