K983150 · Allergan, Inc. · LPN · Nov 25, 1998 · Ophthalmic
Device Facts
Record ID
K983150
Device Name
COMPLETE BRAND LUBRICATING AND REWETTING DROPS
Applicant
Allergan, Inc.
Product Code
LPN · Ophthalmic
Decision Date
Nov 25, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.5928
Device Class
Class 2
Indications for Use
COMPLETE® brand Lubricating and Rewetting Drops are indicated to lubricate and rewet soft (hydrophilic) contact lenses during lens wear.
Device Story
COMPLETE® brand Lubricating and Rewetting Drops are a sterile, isotonic, buffered solution for soft (hydrophilic) contact lens care. The solution contains hydroxypropyl methylcellulose (lubricant), polyhexamethylene biguanide 0.0001% (preservative), Poloxamer 237 (surfactant), edetate disodium (chelating agent), phosphate buffer, sodium chloride, potassium chloride, and purified water. The device is packaged in plastic bottles with controlled dropper tips for patient self-use. Patients apply the drops directly to the eye while wearing lenses to provide lubrication and rewetting, enhancing comfort. The formulation is free of chlorhexidine, thimerosal, and mercury-containing ingredients.
Clinical Evidence
No direct clinical study was performed for the lubricating and rewetting drop indication. Safety was supported by a 28-day ocular safety study in 12 rabbits, showing no clinically significant ocular irritation, discomfort, or histopathological changes. Clinical data from a separate study on the investigational multi-purpose solution (MPS) formulation (n=124) was provided as supporting evidence, showing no adverse device effects and substantial equivalence in safety and performance to the approved COMPLETE® MPS.
Technological Characteristics
Sterile, isotonic, buffered aqueous solution. Ingredients: hydroxypropyl methylcellulose, polyhexamethylene biguanide 0.0001%, Poloxamer 237, edetate disodium, phosphate buffer, sodium chloride, potassium chloride. Form factor: plastic bottle with controlled dropper tip. No chlorhexidine or thimerosal. Meets USP sterility and preservative effectiveness criteria.
Indications for Use
Indicated for lubrication and rewetting of soft (hydrophilic) contact lenses before application and during lens wear.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens care product is a device intended for use in the cleaning, rinsing, disinfecting, lubricating/rewetting, or storing of a soft (hydrophilic) contact lens. This includes all solutions and tablets used together with soft (hydrophilic) contact lenses and heat disinfecting units intended to disinfect a soft (hydrophilic) contact lens by means of heat.
Special Controls
*Classification.* Class II (Special Controls) Guidance Document: “Guidance for Industry Premarket Notification (510(k)) Guidance Document for Contact Lens Care Products.”
Predicate Devices
COMPLETE® brand Lubricating and Rewetting Drops (marketed formulation)
{0}------------------------------------------------
## 510(k) SUMMARY
K983150
# COMPLETE® brand Lubricating and Rewetting Drops
This summary uses the format provided in 21 CFR 807.92:
| (a)(1) | Submitter: | Paul J. Nowacki<br>Manager<br>Regulatory Affairs<br>Allergan<br>2525 Dupont Drive<br>Irvine CA 92612<br>Phone: (714) 246-6761<br>Fax: (714) 246-5457 |
|--------|--------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Summary Prepared: | November, 1998 |
| (a)(2) | Device Trade Name: | COMPLETE® brand Lubricating and Rewetting<br>Drops |
Device Common Name: Soft (Hydrophilic) Contact Lens Solution
Device Classification Names: Accessories to Contact Lens Solution (86LPN)
- Identification of Prodicate Device: COMPLETE® brand Lubricating and (a)(3) Rewetting Drops are substantially equivalent to the formulation of this product marketed now and to other contact lens multi-purpose, lubricating and rewetting solutions.
- Device Description: COMPLETE® brand Lubricating and Rewetting Drops (a)(4) are a sterile, isotonic, buffered, solution containing hydroxypropyl methylcellulose as a lubricant, preserved with polyhexamethylene biguanide 0.0001%, a phosphate buffer, Poloxamer 237 as a surfactant, edetate disodium as a chelating agent, sodium chloride, potassium chloride, and purified water. COMPLETE® brand Lubricating and Rewetting Drops contain no chlorhexidine or thimerosal and no other mercury containing ingredients.
Both current and reformulated products are clear, colorless solutions packaged in plastic bottles with controlled dropper tips.
- Intended Use: COMPLETE® brand Lubricating and Rewetting Drops are (a)(5) indicated to lubricate and rewet soft (hydrophilic) contact lenses during lens wear.
These uses are identical to the predicate, currently marketed product.
- (a)(6) Comparison of Technological Characteristics: The technological characteristics of the new formulation remain essentially the same as the current product.
{1}------------------------------------------------
510(k) SUMMARY COMPLETE® brand Lubricating and Rewetting Drops November, 1998
#### Discussion of Nonclinical: (b)(1)
With one exception, new COMPLETE® brand Lubricating and Rewetting Drops were evaluated for safety and efficacy using recommendations outlined in FDA's Premarket Notification [510(k)] Guidance Document for Contact Lens Care Products, issued May 1, 1997. A 28-day, ocular safety study in rabbits with repeated (4X/day) instillations, was completed to support product safety in lieu of a clinical study in humans. This product meets the suggested microbiological and toxicological criteria.
Microbiological Studies These studies are referenced from K981168 for COMPLETE® brand Multi-Purpose Solution.
- The product meets USP Modified criteria for Preservative Effectiveness.
- The product meets USP Sterility test requirements.
Toxicology Product safety was evaluated using the following tests:
- Cytotoxicity: .
Neutral Red Retention Assay: In an evaluation of ReNu MultiPlus vs. the old and new COMPLETE® solutions, using benzalkonium chloride (BAK) as a control, both new COMPLETE® and ReNu MultiPlus exhibited better neutral red retention than BAK. Old COMPLETE was better until the 2 hour time point, but dropped to approximately the same level as BAK after 3 hours of direct exposure. This indicates that new COMPLETE® may be less damaging to cells than old COMPLETE®.
CHO Clonal Growth Assay: In this comparison of ReNu MultiPlus vs. old COMPLETE® and new COMPLETE® Lubricating and Rewetting Drops. Chinese hamster ovary cells showed a significantly better survival rate with new COMPLETE® than the other two formulations. Relative survival was higher than the negative control at almost all concentration levels. Overall, new COMPLETE® showed the best relative survival in this assay. This indicates that new COMPLETE® solution may be less damaging to eye cells than the other products tested.
- . Sensitization: The potential contact sensitization (Maximization Test) of new COMPLETE® brand Lubricating and Rewetting Drops was evaluated in 25 female Hartley guinea pigs using a maximization test conducted following methods adapted from the procedure described by Magnusson and Kligman, 1970. The marketed COMPLETE® formulation was used as a control.
No dermal reactions were observed in either test or control groups. indicating that new COMPLETE® is comparable to the marketed formulation.
This study is referenced from K981168 for COMPLETE® brand Multi-Purpose Solution.
{2}------------------------------------------------
#### 510(k) SUMMARY COMPLETE® brand Lubricating and Rewetting Drops November, 1998
### Toxicology (Continued)
- Acute Oral Toxicity: The acute oral toxicity of new COMPLETE® . Lubricating and Rewetting Drops was evaluated in forty (20 male and 20 female) CD albino rats following a single oral administration and 14 days of observation. Test animals received new COMPLETE® while control animals received sterile water for injection, both at a dose of 20 mL/kg. All rats were sacrificed at the end of the 14-day observation period.
There were no treatment-related effects or deaths. New COMPLETE® caused no adverse effects when administered to rats at a single oral dose of 20 mLJkg.
This study is referenced from K981168 for COMPLETE® brand Multi-Purpose Solution.
- 28-Day Ocular Safety Study: The ocular safety of new COMPLETE® . Lubricating and Rewetting Drops in conjunction with COMPLETE® Weekly Enzymatic Cleaner, using SOFLENS® soft contact lenses was evaluated in twelve female New Zealand white rabbits.
Lenses were soaked overnight in approximately 4 mL of either the new or old formulation. Once a week, one COMPLETE® enzyme tablet was added to the overnight soak (1 tablet/4 mL). The test also involved topical instillations of new or old COMPLETE® solution at approximate 2 hour intervals (4X/day) for the entire study period. Lenses were worn for approximately 8 hours a day for 28 consecutive days.
Rabbits were observed daily at lens application and at each instillation for ocular discomfort. Observations for gross ocular irritation were made at each topical instillation and at lens removal. Pachometry, slit lamp examinations, and body weights were performed weekly. Ophthalmoscopy was performed prior to the start and at the end of the study. Histopathological evaluation was performed on both eyes of all rabbits at the conclusion of the study.
All animals remained healthy and growth was normal throughout the study. No clinically significant ocular discomfort or conjunctival irritation was observed at lens application or removal. No clinically significant changes in corneal thickness were detected by pachometry examinations. No clinically significant conjunctival, uveal or corneal abnormalities were noted at any slit lamp examination. No apparent ocular changes were observed by direct ophthalmoscopy. Histopathological examinations revealed no treatment related changes in ocular tissue.
The results of this study indicate that frequent eye contact with new COMPLETE® brand Lubricating and Rewetting Drops, in conjunction with COMPLETE® Weekly Enzymatic Cleaner, is safe for use with hydrophilic contact lenses.
Stability Testing using the protocol approved in K981168, to establish an expiration date, will be completed prior to marketing this product.
{3}------------------------------------------------
### 510(k) SUMMARY COMPLETE® brand Lubricating and Rewetting Drops November, 1998
#### Discussion of Clinical Data: (b)(2)
This clinical investigation involved use of the new COMPLETE® formulation as a multi-purpose solution. Although the product was not used as a lubricating and rewetting drop, its use as a multi-purpose solution involved indirect instillation. Subjects were instructed to either put the lens into the eye directly from soaking in the lens case, or to rinse the lens with multi-purpose solution prior to wearing. The 28-Day toxicology study summarized on the previous page provides the principal safety support for use as a lubricating and rewetting drop.
Six clinical investigators enrolled a total of 124 subjects, of whom 62 were assigned to the Investigational multi-purpose solution (MPS) group and 62 to the approved (COMPLETE®) MPS group. Two subjects in the Investigational MPS group were disqualified, leaving 60 evaluable subjects in the Investigational MPS group and 62 evaluable subjects in the COMPLETE® MPS group. Of these, 90.3% in the Investigational MPS group and 90.3% in the COMPLETE® MPS group completed the study.
#### Safety:
No adverse device effects were reported during this study and there were no statistically significant differences in the number of examinations with clinically significant slit lamp findings. There were statistically significant differences between the Investigational MPS and COMPLETE® MPS groups in the maximum severity grades for injection and tarsal anomaly, with more severe findings in the COMPLETE® MPS group than in the Investigational MPS group. Other safety variables were similar for both groups, with no findings directly attributed to the study regimens.
### Acceptability:
There was a statistically significant difference between the Investigational MPS and COMPLETE® MPS groups with higher maximum severity score for symptoms of discomfort in the Investigational MPS group. However, the overall incidence of these selected symptoms were low and not regimen related and therefore not clinically relevant to the performance of the investigational regimen. There was no statistically significant difference between the Investigational MPS and COMPLETE® MPS groups in the number of examinations with clinically significant ocular symptoms of discomfort.
There was no statistically significant difference between the Investigational MPS and COMPLETE® MPS for other findings which were unremarkable.
- Conclusions Drawn from Data Supporting Equivalence Determination: (b)(3) It is concluded that the safety, efficacy and performance of the Investigational product is substantially equivalent to the approved formulation currently on the market.
This study is referenced from K981168 for COMPLETE® brand Multi-Purpose Solution.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 25 1998
Mr. Paul J. Nowacki Manager, Regulatory Affairs Allergan 2525 Dupont Drive Irvine, CA 92623-9534
Re: K983150 Trade Name: Complete ® brand Lubricating and Rewetting Drops Regulatory Class: II Product Code: 86 LPN Dated: September 8, 1998 Received: September 9, 1998
Dear Mr. Nowacki:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations .
{5}------------------------------------------------
Page 2 -Mr. Paul J. Nowacki
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on.the.promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
A. Ralph Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
510(k) NUMBER: (IF KNOWN):
DEVICE NAME:
COMPLETE® brand Lubricating and Rewetting Drops
INDICATIONS FOR USE:
COMPLETE® brand Lubricating and Rewetting Drops are used to lubricate and rewet soft (hydrophilic) contact lenses before application and during lens wear.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use (Per 21 CFR 801.109
OR
Over-The-Counter-Use (Optional Format 1-2-96)
Mgra Smith
Jivision Sign-Off) . wision of Ophthalmic Devices 510(k) Number _
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.