K983110 · Toshiba America Mri, Inc. · LNH · Feb 25, 1999 · Radiology
Device Facts
Record ID
K983110
Device Name
FLEXART,FLEXART/HYPER, VISART, VISART/HYPER
Applicant
Toshiba America Mri, Inc.
Product Code
LNH · Radiology
Decision Date
Feb 25, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
Diagnostic imaging of the whole body (including head, abdomen, breast, heart, pelvis, joints, neck, TMJ, spine, blood vessels, limbs, and extremities), fluid visualization, 2D and 3D imaging, MR angiography/MR Vascular Imaging and MR fluoroscopy. [Application terms include MRCP (MR Cholangiopancreatography), MR Urography, MR Myelography, SAS (Surface Anatomy Scan), Dynamic Scan, Cine Imaging, and Cardiac Tagging.]
Device Story
Software upgrade for FLEXART and VISART MRI systems; enables new imaging sequences including cardiac tagging and BOLD imaging. Device processes magnetic resonance signals to produce diagnostic images of human anatomy. Operated by trained MRI technologists/radiologists in clinical settings. Output viewed by clinicians to assist in diagnostic decision-making for various pathologies. Benefits include enhanced diagnostic capabilities for cardiac and vascular assessment.
Clinical Evidence
Bench testing only. Acoustic noise levels measured per NEMA guidelines. Imaging performance parameters (specification volumes) provided. Sample clinical images presented for new sequences.
Technological Characteristics
MRI diagnostic accessory; software upgrade for existing systems. Operates at 0.5T (FLEXART) and 1.5T (VISART). SAR <0.4 W/kg (FLEXART) and <1.0 W/kg (VISART). Acoustic noise levels 98.5-105.3 dB. Supports 2D/3D imaging, MR angiography, MR fluoroscopy, and BOLD imaging.
Indications for Use
Indicated for diagnostic imaging of the whole body, including head, abdomen, breast, heart, pelvis, joints, neck, TMJ, spine, blood vessels, limbs, and extremities, for patients requiring MR imaging, fluid visualization, 2D/3D imaging, MR angiography, MR vascular imaging, MR fluoroscopy, and BOLD imaging.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
K98 3110
# 510(k) Premarket Notification
# SUMMARY OF SAFETY AND EFFECTIVENESS
| 1. | DEVICE NAME:<br>Model Number: | Magnetic Resonance Diagnostic Device Accessory<br>MRT-50GP /E2 (FLEXARTTM), /H2 (FLEXARTTM/Hyper)<br>MRT-150 /F1 (VISARTTM), /F2 (VISARTTM/Hyper) | | |
|----------------------------------------------------|------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|----------------------------|
| | Trade/Proprietary Name: | FLEXARTTM, FLEXARTTM/Hyper, VISARTTM,<br>VISARTTM/Hyper | | |
| 2. | ESTABLISHMENT REGISTRATION: | 2636923 | | |
| 3. | U.S. AGENT NAME AND ADDRESS: | Toshiba America MRI, Inc.<br>280 Utah Avenue<br>South San Francisco, CA 94080 | | |
| | CONTACT PERSON: | Ken Nehmer<br>(650)872-2722 ext. 6083 | | |
| 4. | MANUFACTURING SITE: | Toshiba Corporation<br>1385 Shimoisigami<br>Otawara-shi, Tochigi-Ken<br>Japan 324 | | |
| 5. | DATE OF SUBMISSION: | September 1, 1998 | | |
| 6. | DEVICE DESCRIPTION: | This submission consists of a software upgrade to the MRT-<br>50GP/E2 (FLEXARTTM), MRT-50GP/H2 (FLEXARTTM/Hyper),<br>MRT-150/F1 (VISARTTM), MRT-150/F2 (VISARTTM/Hyper) | | |
| 7. | SAFETY PARAMETERS:<br>Maximum static field strength: | FLEXARTTM &<br>FLEXARTTM/Hyper | V3.5 s/w<br>0.5 T | V4.0 s/w<br>0.5 T |
| | | VISARTTM &<br>VISARTTM/Hyper | 1.5T | 1.5T |
| | Rate of change of magnetic field: | FLEXARTTM<br>FLEXARTTM/Hyper | 11 T/sec.<br>13.3 T/sec. | 11 T/sec.<br>13.3 T/sec. |
| | | VISARTTM<br>VISARTTM/Hyper | 13.3 T/sec.<br>19.5 T/sec. | 13.3 T/sec.<br>19.5 T/sec. |
| Maximum radio frequency power<br>deposition (SAR): | | V3.5 s/w | V4.0 s/w | |
| FLEXART™ &<br>FLEXART™/Hyper | | <0.4 W/kg | <0.4 W/kg | |
| VISART™ &<br>VISART™/Hyper | | <1.0 W/kg | <1.0 W/kg | |
| Acoustic noise levels (maximum): | | | | |
| FLEXART™ | | 100.2 dB(A) | 100.2 dB(A) | |
| FLEXART™/Hyper | | 98.5 dB(A) | 98.5 dB(A) | |
| VISART™ | | 105.3 dB | 105.3 dB | |
| VISART™/Hyper | | 105.1 dB | 105.1 dB | |
{1}------------------------------------------------
Acoustic noise data was measured in accordance with NEMA guidelines. The user is cautioned to have the patient wear acoustic noise protection during scanning.
#### IMAGING PERFORMANCE PARAMETERS: 8.
| Specification volume: Head: | FLEXART™ &<br>VISART™ † | V3.5 s/w | V4.0 s/w |
|-----------------------------|-------------------------|----------|----------|
| | | 16cm dsv | 16cm dsv |
| Body: | FLEXART™ &<br>VISART™ † | 28cm dsv | 28cm dsv |
t = specifications for both standard and hyper versions
Sample clinical images are presented for new sequences.
### 9. INTENDED USE
| | Anatomical regions: Head, body, extremity, spine, neck, TMJ, and heart |
|-----------------|----------------------------------------------------------------------------------|
| Nuclei excited: | Hydrogen |
| Diagnostic use: | Diagnostic imaging of the whole body (including head, abdomen, breast, |
| | heart, pelvis, joints, neck, TMJ, spine, blood vessels, limbs, and extremities), |
| | fluid visualization, 2D and 3D imaging, MR angiography/MR Vascular Imaging |
| | and MR fluoroscopy. [Application terms include MRCP (MR |
| | Cholangiopancreatography), MR Urography, MR Myelography, SAS (Surface |
| | Anatomy Scan), Dynamic Scan, Cine Imaging, and Cardiac Tagging.] |
#### 10. EQUIVALENCY INFORMATION:
Toshiba America MRI, Inc. (TAMI), believes that the Version 4.0 software upgrade for the FLEXART™ and VISART™ systems is substantially equivalent to the software version 3.5 for FLEXART™ (K970573) which was cleared on 7/21/97 and VISART™ (K965068) which was cleared on 7/15/97. TAMI believes that the introduction of cardiac tagging in V4.0 is substantially equivalent to Cardiac Tagging Techniques (K973799) which was cleared on 1/2/98, and CINE Imaging which was cleared with version 3.1 for FLEXART™ (K962138) on 12/23/96. The modifications added to the Version 4.0 software do not raise new questions of safety or efficacy.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three abstract shapes that resemble birds in flight. The shapes are arranged in a diagonal line, with the largest shape at the top and the smallest shape at the bottom. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circle around the image.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 25 1999
Ken Nehmer Quality Engineer Toshiba America MRI, Inc. 280 Utah Avenue South San Francisco, CA 94080 Re:
K983110 Software Version 4.0 for Flexart and Visart Dated: December 17, 1998 . Received: December 21, 1998 Regulatory class: II 21 CFR 892.1000/Procode: 90 LNH
Dear Mr. Nehmer:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have vie nave reviewed your occain oroly nealisations for uses stated in the encosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the manical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls r cocial rood, Drag, and e comers controls provisions of the Act include requirements for annual registration, listing of provisions of the Act. "The general ce, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may if your device to such additional controls. Existing major regulations affecting your device can be found in the Code of be subject o battractional controlly equivalent determination assumes compliance with r occurrent Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for the Gallers Good management (21 CFR Part 820) and that, through periodic QS inspections, the Food and Mculan Donoral rogalerity such assumptions. Failure to comply with the GMP regulation may result in Drug Action. In addition, FDA may publish further announcements concerning your device in the Eederal regarding about. In addition - 27 has your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA I his reter will and you to bogin markeding your os a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in if you doon opeonices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Capt. Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Page _________________________________________________________________________________________________________________________________________________________________________
| "
510(k) Number (if known):
K983110 ______________________________________________________________________________________________________________________________________________________________________
Device Name:
Version 4,0 Software Upgrade (FLEXART™ & VISART™)
Indications for Use:
Imaging of:
-
- The Whole Body (including head, abdomen, breast, heart, pelvis, joints, neck, TMJ, The Whole Doug (inoluding noad, are ities). [Application terms include MRCP (MR spirie, blood vessels , linios and oxientition ("Myelography, MR Fluoroscopy, SAS Onolanglopanoroatography); manic Scan, Cine Imaging, and Cardiac tagging.]
- Fluid Visualization -
- 2D/3D Imaging •
- MR Angiography/MR Vascular Imaging "
- Blood Oxygenation Level Dependent (BOLD) Imaging .
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|--------------------------------------------------------------------|---------|
| Division of Reproductive, Abdominal, ENT, and Radiological Devices | |
| 510(k) Number | K983110 |
| Prescription Use (Per 21 CFR§801.109) | |
|---------------------------------------|--|
|---------------------------------------|--|
OR
| Over-The-Counter Use (Optional Format 1-2-96) | |
|-----------------------------------------------|--|
|-----------------------------------------------|--|
CONFIDENTIAL
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.