The ALDEN CLASSIC PROSTHETIC (polymacon) Tinted, Soft Contact Lens is indicated for daily wear to enhance or alter the apparent eye color, including ocular masking, either in sighted or nonsighted eyes that require a prosthetic contact lenses for cosmetic management of conditions such as corneal, iris or lens abnormalities. The lens may also be prescribed for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and not aphakic persons or for occlusive therapy for conditions such as diplopia, amblyopia or extreme photophobia. The lens may be disinfected with either a chemical or a heat disinfection system.
Device Story
The Alden Classic Prosthetic is a lathe-cut, hydrophilic soft contact lens fabricated from polymacon. It incorporates specific color additives (Vat Brown 1, Vat Blue 6, Vat Green 1) to alter or enhance eye appearance or provide ocular masking. Available in various designs including clear lenses with black pupils, black occluder lenses, and black annular designs. The lens is prescribed by eye care professionals for cosmetic management of ocular abnormalities, refractive correction, or occlusive therapy. In its hydrated state (38% water), the lens conforms to the corneal curvature. It is intended for daily wear and requires maintenance in compatible storage solutions. The device provides a color-altering optical surface to benefit patients with disfiguring eye conditions or those requiring light occlusion.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on bench testing, material characterization, and established safety profiles of similar previously cleared devices.
Technological Characteristics
Material: Polymacon (38% water content). Fabrication: Lathe-cut. Refractive Index: 1.43 (hydrated). Oxygen Permeability: 9 x 10^-11 Fatt units. Colorants: Vat Brown 1, Vat Blue 6, Vat Green 1. Form factor: Spherical back surface, various pupil/annular designs. Sterilization: Standard manufacturing procedures.
Indications for Use
Indicated for daily wear in sighted or non-sighted patients requiring prosthetic contact lenses for cosmetic management of corneal, iris, or lens abnormalities. Also indicated for correction of refractive ametropia (myopia, hyperopia, astigmatism) in aphakic and non-aphakic persons, or for occlusive therapy for diplopia, amblyopia, or extreme photophobia.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
Predicate Devices
WESLEY-JESSEN PROSTHETIC (phemfilcon A) Soft Contact Lens
{0}------------------------------------------------
# 510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is:
#### K983053
### Applicant information:
| Date Prepared: | August 28, 1998 |
|-----------------|-----------------------------------------|
| Name: | Alden Optical Laboratories |
| Address | 13295 Broadway<br>Alden, New York 14004 |
| Contact Person: | Charles H. Creighton |
| Phone Number: | 800.253.3669 |
| USA Consultant: | MedVice Consulting, Inc. |
| Phone Number: | Martin Dalsing<br>(970) 243-5490 |
#### Device Information:
| Device Classification: | Class II |
|------------------------|--------------------------------------------------------------------------------|
| Classification Number: | LPL |
| Trade Name: | ALDEN CLASSIC PROSTHETIC (polymacon) Tinted,<br>Soft Contact Lens (lathe-cut). |
| Classification Name: | Lenses, Soft Contact, Daily Wear |
{1}------------------------------------------------
#### Substantially Equivalent Devices:
The ALDEN CLASSIC PROSTHETIC (polymacon) Tipted, Soft Contact Lens is substantially equivalent to the "WESLEY-JESSEN PROSTHETIC" (phemfilcon A) Soft Contact Leas and the BENZ-38 (polymacon) Soft Contact Lens, the predicate devices.
#### Device Descriptive Characteristics:
The ALDEN CLASSIC PROSTHETIC (polymacon) Tinted, Soft Contact Leas is fabricated from polymacon, which in the dry (unhydrated) state may be machined and polished. The hydrophilic nature of this material allows the lens to become soft and pliable when immersed in an aqueous solution. A tunt mixture containing: Blue 7,16-Dichloro-6, 15-dihydro-5,9,14,18-anthrazinetetrone, Green 16,17-Dimethoxydinaphtho [1,2,3 -> 1' - Im] perylene-5,10-dione, and Brown 16,23 -Dibydrodinaphtho [2,3-a:2',3' -i] napth [2',3';.6,7] indolo [2,3-c] carbazole-5,10,15,17,22,24-hexone is added to the lens.
The tint mixture (BLACK) is processed into the contact lens to enhance or alter the apparent eye color, including ocular masking, either in sighted or non-sighted eyes that require a prosthetic contact lenses for cosmetic management of conditions such as corneal, iris or leas abnormalities. The lens may also be prescribed for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and not aphakic persons or for occhisive therapy for conditions such as diplopia, amblyopia or extreme photophobia.
Tim Patterns Available:
- 1. Clear lens with Black Pupil. Pupil sizes available in 2.0 mm to 12,5 mm.
- 2. Black Occluder Lens. A Central Black area that occludes light. Available to full lens diameter in 0.5 mm increments.
- 3. Black Annular with clear pupil. Black Annular diameter range 7.5 mm to full lens diameter in 0,5 mm increments. Clear Pupil diameter range 2.0 mm to 7,5 mm in 0.5 mm increments.
- 4. Tinted lens with Black Pupil. Uses the Alden Classic Tinted (polymacon 38%) contact lens with black pupil. Pupil sizes available in 2.0 mm to 12.5 mm diameter.
In the hydrated state, the leas conforms to the curvature of the eye covering the cornes and extending sightly beyond the limbus forming a color altering optical surface. The (polymacon) soft hydrophilic contact leas a spherical back surface. The hydrophilic properties of the lens require that it be maintained in a fully hydrated state in a solution compatible with the eye. If the leas dries out, it will become hard and appear somewhat warped; however, it will return to its proper configuration when completely rehydrated in the proper storage solution.
{2}------------------------------------------------
The hydrophilic characteristics allow aqueous solutions to enter the lens and in its fully hydrated state the lens is approximately 38% water by weight. The physical properties of the lens are:
| Refractive Index | 1.52 (dry) 1.43 (hydrated) |
|---------------------|-----------------------------------------------------------------------------|
| Light Transmission: | greater than 70% T |
| Water Content | 38 % |
| Specific Gravity | 1.28 (dry) 1.18 (hydrated) |
| Color additives | Vat Brown 1, Vat Blue 6, Vat Green 1 |
| Oxygen Permeability | 9 X $10^{-11}$ Fatt Units (cm²/sec)(ml O₂/ml x mm Hg @ 35°C), (revised Fatt |
| | method) |
#### Intended Use:
The ALDEN CLASSIC PROSTHETIC (polymacon) Tinted, Soft Contact Lens is indicated for daily wear to enhance or alter the apparent eye color, including ocular masking, either in sighted or nonsighted eyes that require a prosthetic contact lenses for cosmetic management of conditions such as corneal, iris or lens abnormalities. The lens may also be prescribed for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and not aphakic persons or for occlusive therapy for conditions such as diplopia, amblyopia or extreme photophobia. The lens may be disinfected with either a chemical or a heat disinfection system.
#### Substantial Equivalence:
The device will be manufactured according to specified process controls and a quality assurance program. The device will undergo manufacturing, packaging and sterilization procedures similar to devices currently marketed and distributed by Alden Optical Laboratories. The established safety profile (preclinical toxicology and manufacturing/chemistry data) of the device is equivalent to the Alden Classic, and Alden Classic Tinted (polymacon), 510(k)'s #K973967 and #K980554. Being similar with respect to materials, physical construction and safety & effectiveness to the predicate device, this meets the requirements per section 510(k) of the act regarding substantial equivalence and does not raise different questions of safety and effectiveness than the predicate devices identified above.
{3}------------------------------------------------
The following mattix illustrates the production methods, leas functions and material characteristics of the ALDEN CLASSIC PROSTHETIC (polymacon) Tinted, Soft Contact Leas as well as the predicate device.
#### Substantial Equivalence Matrix
| | Characteristic | ALDEN CLASSIC<br>PROSTHETIC | WESLEY-JESSEN<br>PROSTHETIC |
|-----|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1.) | PRODUCTION<br>METHOD | Lathe-Cut | Lathe-Cut |
| 2.) | LENS FUNCTION | Enhance or alter the apparent eye color,<br>including ocular masking, either in sighted<br>or non-sighted eyes that require a<br>prosthetic contact lenses for cosmetic<br>management of conditions such as corneal,<br>iris or lens abnormalities. The lens may<br>also be prescribed for the correction of<br>refractive ametropia (myopia, hyperopia<br>and astigmatism) in aphakic and not<br>aphakic persons or for occlusive therapy for<br>conditions such as diplopia, amblyopia or<br>extreme photophobia. | Enhance or alter the apparent eye color,<br>including ocular masking, either in sighted<br>or non-sighted eyes that require a<br>prosthetic contact lenses for cosmetic<br>management of conditions such as corneal,<br>iris or lens abnormalities. The lens may<br>also be prescribed for the correction of<br>refractive ametropia (myopia, hyperopia<br>and astigmatism) in aphakic and not<br>aphakic persons or for occlusive therapy for<br>conditions such as diplopia, amblyopia or<br>extreme photophobia. |
| 3.) | MATERIAL | Hydrophilic Polymer | Hydrophilic Polymer |
| a. | Polymer | Polymacon | phemfilcon A |
| b. | Water Content | 38% | 38% |
| c. | Polymer Content | 62% | 62% |
| d. | DK Value | 9 | 12.9 |
| e. | Refractive Index | 1.43 (hydrated) | 1.440 (hydrated) |
| f. | Specific Gravity | 1.18 (hydrated) | 1.17 (hydrated) |
| g. | Light Transmission | Varies depending on prosthetic lens<br>design | Varies depending on prosthetic lens<br>design |
| h. | Color Additives | Vat Green 1,<br>Vat Brown 1,<br>Vat Blue 6, | Iron oxides, chromium oxide greens,<br>titanium dioxide, [phthalocyaninato<br>(2)] copper, carbazole violet and<br>phthalocyanine green. |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines forming its body and head. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle. The text is in all capital letters and is evenly spaced around the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 21 1998
Alden Optical Laboratories, Inc. c/o Mr. Martin Dalsing 623 Glacier Drive Grand Junction, CO 81503
Re: K983053
Trade Name: ALDEN CLASSIC PROSTHETIC (polymacon) Tinted, Soft Daily Wear Contact Lens (lathe-cut) Regulatory Class: II Product Code: 86 LPL Dated: August 28, 1998 Received: September 1, 1998
Dear Mr. Dalsing:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations .
{5}------------------------------------------------
Page 2 - Mr. Martin Dalsing
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on. the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
A. Ralph Rosenthal
A. Ralph Rosenthal, M.D. Director Division of Ophthalmic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## INDICATIONS FOR USE STATEMENT
Device Name:
ALDEN CLASSIC PROSTHETIC (polymacon) Tinted, Soft Contact Lens (Lathe cut).
INDICATIONS FOR USE:
The ALDEN CLASSIC PROSTHETIC (polymacon) Tiated, Soft Contact Leas (Lathe-cut) is indicated for daily wear to enhance or alter the apparent eye color, including ocular masking, either in sighted or non-sighted eyes that require a prosthetic contact lenses for cosmetic management of conditions such as corneal, ins or lens abnormalities. The lens may also be prescribed for the correction of refractive ametropia (myopia, hyperopia and astigmatism) in aphakic and not aphakic persons or for occlusive therapy for conditions such as diplopia, amblyopia or extreme photophobia. The lens may be disinfected with either a chemical or a heat disinfection system.
(PLEASE DO OT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|--------------------------------------------------------|---------|
| (Division Sign-Off) | |
| Division of Ophthalmic Devices | |
| 510(k) Number | K983053 |
or
| Prescription Use | X | Over-The-Counter Use | |
|----------------------|--------------------------|----------------------|--|
| (Per 21 CFR 801.109) | (Optional Format 1-2-96) | | |
ALDEN CLASSIC PROSTHETIC (Polyanson) Trated, Soll Carned Lathousel)
… 15
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.