← Product Code [HIR](/productcode/HIR) · K983052

# PERITRON PERINEOMETER MODEL 9300V OR 9300A (K983052)

_Cardio Design Pty. , Ltd. · HIR · Oct 19, 1998 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K983052

## Device Facts

- **Applicant:** Cardio Design Pty. , Ltd.
- **Product Code:** [HIR](/productcode/HIR.md)
- **Decision Date:** Oct 19, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.1425
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology
- **Attributes:** Therapeutic

## Indications for Use

Peritron is used in assessing the strength of pelvic floor muscles, teaching pelvic floor muscle exercises and for providing feedback during pelvic floor muscles exercises.

## Device Story

Peritron Perineometer is a biofeedback device for pelvic floor muscle assessment and exercise. It consists of a vaginal or anal sensor, a readout unit, and an interconnecting PVC tube. In operation, the user performs voluntary pelvic floor muscle contractions, which displace air within the sensor; this pressure change is transmitted through the tube to the readout unit. The device does not induce contractions. The readout unit uses a pressure transducer and microprocessor to interpret air pressure, displaying results numerically in cm water column or as an analogue bar graph. It is used in clinical settings by healthcare providers to assess muscle strength and teach exercises, or at home by patients for biofeedback. The device is battery-powered, handheld, and operated via a single push-button. It provides objective feedback on muscle contraction strength, assisting in the treatment of stress incontinence and pelvic floor muscle training.

## Clinical Evidence

No clinical data provided. Safety and effectiveness are supported by bench testing, including biocompatibility testing of the sensor sheath (Wacker Elastosil LR 3003/50A and LR/50B) and electrical emission/safety testing.

## Technological Characteristics

Device consists of a vaginal/anal sensor, PVC tubing, and a microprocessor-based readout unit. Sensor sheath is injection-molded liquid silicone rubber (Wacker Elastosil LR 3003/50A and LR/50B) meeting FDA 177.2600. Sensing principle: pneumatic pressure transducer. Power: 9V battery. Form factor: handheld plastic case. Connectivity: none. Sterilization: not specified.

## Regulatory Identification

A perineometer is a device consisting of a fluid-filled sack for intravaginal use that is attached to an external manometer. The devices measure the strength of the perineal muscles by offering resistence to a patient's voluntary contractions of these muscles and is used to diagnose and to correct, through exercise, uninary incontinence or sexual dysfunction.

## Predicate Devices

- Contimed II (Hollister, Illinois)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white graphic. The graphic contains a thick, black, semi-circle shape on the left side of the image. To the right of the semi-circle are several short, horizontal lines. The lines are also black and thick. 

K983052

# OUT 19 1998

## Safety and Effectiveness Summary

This summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.

### Background:

The Peritron Perineometer comprises a vaginal or anal Sensor, Readout Unit and interconnecting tube.

Perineometers are used in the diagnosis and treatment of stress incontinence in patients and for providing exercise feedback.

In the clinic Peritron is used to assess the condition of the pelvic floor muscles, (strength of contraction), teach pelvic floor exercises with the view to improving muscle strength and may be used at home as a biofeedback unit by patients performing pelvic floor exercises.

In operation, air displaced from the sensor caused by a voluntary contraction of the muscles of the pelvic floor travel to the readout units via the interconnecting tube. Peritron does not cause the contraction. These muscles are under the voluntary control of the user at all times.

#### 1 Safety

None of these products diminish the safety level set by the substantially equivalent Contimed II™ (Hollister, Illinois) of Appendix 4(a) due to the following

#### 1.1 Sensors

ﻣﻌﺘﺴﺴﺴﺴﺴﺴﺴﺴﺴﺴﺴﺴ

The key element concerning patient/user safety and the only contact component is the sensor.

As a principal manufacturer of catheters and tubing packs for Cardiac Surgery (Registered in Australia under the Therapeutic Goods Act as a GMP manufacturer but with no goods registered or offered for sale in the USA) we believe we have a good understanding of appropriate materials and manufacturing procedures necessary to produce a product of the highest biocompatibility and patient safety.

Accordingly we designed the sheath of the sensor (the only patient-contact surface) to be moulded in one piece without joins or crevices to potentially harbour bacteria. (In this respect it is safer than the vaginal sensor of the substantially equivalent Contimed unit, the several elements of which join together on the patient contact surface.) The sheath is injection moulded from liquid silicon rubber, Wacker Elastosil cat LR 3003/50A and LR/50B, that meets the requirements of FDA regulation 177.2600. Refer to documents in appendix 6(a)

As the sheath of the sensor contains 600ppm of blue pigment (Wacker cat RAL 5010 ref enclosure 6(c)) of unknown biocompatibility the FDA previously advised that we should conduct testing of the finished product. Refer to appendix 6 (a) for the test results.

{1}------------------------------------------------

### 1.2 Readout Unit

Peritron incorporates a pressure transducer and microprocessor to interpret the air pressure in the sensor and display it either numerically in cm water column pressure or as an analogue bar graph.

Hand-held Peritron weighs 220 gm and is powered by a standard 9 volt battery housed in an integral, user-accessible compartment separated physically from the electronic components.

All operating functions are controlled by one push-button.

The case of Peritron is of injection moulded plastic. There are no external parts that are of metal or are sharp and no electronic parts are accessible without the use of a tool.

At the time of the previous 510(k) application we understand electrical testing of the device was not required. However, since the previous approval we have had Peritron tested for Emission and Safety. The testing shows that Peritron is safe for the intended use. Refer to appendix 6(b) for test results.

#### 1.3 Summary of Safety Features of Peritron

- * Peritron is a passive device - it measures the result of a voluntary action of the user.
- * There are no external wires, electrodes or electrical items of any kind.
- * Internal user adjustments are neither required nor possible.
- * A PVC tube is the only connection between the vaginal probe of the patient/user and the Readout unit.
- * There are no liquids or dangerous items incorporated.
- * All tests show that Peritron is safe for the intended use and meets standards where they exist.

### Safety Experience

Since its inception in 1991, over 1200 units have been put into operation and at no time during the development or usage of the Peritron has there been an incident that in any way threatened the safety of any individual.

#### Effectiveness 2.

Peritron does not alter the effectiveness compared to substantially equivalent Contimed II™ device as both devices measure the increase in air pressure in a sensor caused by a voluntary contraction of pelvic floor muscles.

Like Peritron, Contimed 11 conveys air from its sensor to the readout unit by means of a plastic tube.

Peritron is tested during manufacture for compliance with the performance criteria set out in the Brochure and Handbook of Appendix five. We would be pleased to make test results available on request.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the caduceus symbol. The text is in all capital letters.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 19 1998

Mr. Neville Motley Manager cardio design pty ltd P.O. Box 5407 BHBC Baulkham Hills NSW 2153 AUSTRALIA

Re: K983052

Peritron, Model 9300A with Anal Sensor Model 9300V with Vaginal Sensor Dated: August 12, 1998 Received: September 1, 1998 Regulatory Class: II 21CFR 884.1425/Procode: 85 HIR

Dear Mr. Motley:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Ouality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note; this response to your premarket notification submission does not affect any obligation you might have under sections 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrh/dsmaldsmamain.html".

Sincerely yours

Lillian Yin, Ph.D.

Director, Division of Reproducti Abdominal, Ear, Nose and Thro and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

Page

K983052 510(k) Number (if known):_

PERITRON PERINEOMETER- Model 9300V or 9300A Device Name:_

Indications For Use:

ns For Use.
Peritron is used in assessing the strength of pelvic floor muscles, r entroll is about in associety
 teaching pelvic floor muscle exercises and for providing feedback during pelvic floor muscles exercises. **********

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Whind h. Segen

(Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number .

Prescription Jise_V (Per 21 CFR 801.109)

OR

Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________

(Optional Format 1-2-96)

---

**Source:** [https://fda-staging.innolitics.com/device/K983052](https://fda-staging.innolitics.com/device/K983052)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
