This device, Polymer OBL SB Series and Cinch Series Anchor/Suture Combination is intended for use only for the fixation of non-absorbable synthetic sutures for the following indications. The Polymer Cinch SB Anchor/Suture Combination is intended only for the fixation of surgical suture material to the pelvis for the purpose of bladder neck suspensions for female urinary incontinence due to urethral hypermobility of intrinsic sphincter deficiency. Shoulder: Bankhart Repair, SLAP Lesion Repair, Acromio-clavicular separation, Rotator Cuff Repair, Capsule and Capsulolabral Reconstruction, Biceps Tenodesis, Deltoid Repair. Hand, Wrist, Elbow: Scapholunate ligament reconstruction, Ulnar Collateral Ligament Reconstruction, Lateral Collateral Ligament Reconstruction, Biceps Tendon Reattachment, Elbow Medial/Lateral Repair of Tendons in the Elbow. Foot: Hallux Valgus Reconstruction, Mid and Forefoot Reconstruction. Knee: medial collateral ligament, lateral collateral ligament, posterior oblique ligament, Joint capsule closure, Illiotibial band tendonesis, VMO Advancement. Urinary: Bladder Neck Suspension
Device Story
Polymer OBL SB Anchor/Suture Combination is a sterile, single-use bone anchor system for soft tissue-to-bone fixation. Device consists of a polyacetal cylinder anchor preloaded with braided polyester suture and a manual insertion tool (polypropylene handle, stainless steel sleeve/piston). Surgeon creates bone hole using accessory drills/awls; anchor is pushed into site via inserter; rotation initiated by tapered ends secures anchor in bone. Pulling suture engages anchor against bone walls. Used in open or arthroscopic procedures by orthopedic surgeons. Shape distributes load to reduce bone breakage/pullout risk. Output is a secure anchor-suture construct for tissue attachment, facilitating ligament/tendon repair or bladder neck suspension.
Clinical Evidence
Bench testing only. Pullout strength tested in cancellous bone model: 2x8mm anchor achieved 9.6 lbs (vs 9 lbs for predicate); 6x15mm anchor achieved 14 lbs. Suture break strength tested: smallest version handled 19.4 lbs, largest 23.2 lbs, exceeding USP knot break strength for #2 synthetic suture (14 lbs).
Indicated for fixation of non-absorbable synthetic sutures in orthopedic soft tissue-to-bone repairs (shoulder, hand, wrist, elbow, foot, knee) and bladder neck suspension for female urinary incontinence due to urethral hypermobility or intrinsic sphincter deficiency.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Predicate Devices
2.8 and 3.5mm ROC Suture Bone Fastener by Innovasive Devices (K963402)
SB Bone Anchor by Orthopaedic Biosystems Ltd., Inc (K950592)
Submission Summary (Full Text)
{0}------------------------------------------------
982996
**510(k) SUMMARY**
Image /page/0/Picture/2 description: The image shows the logo for Orthopaedic Biosystems LTD., INC. The logo is made up of the letters OBL in a stylized font. The letters are black and are arranged in a horizontal line. The words "ORTHOPAEDIC BIOSYSTEMS LTD., INC." are written in a smaller font below the logo.
Date Summary was prepared: October 20, 1998
# SUBMITTER:
Contact: Jeffrey P. Baldwin Orthopaedic Biosystems Ltd., Inc. 15990 N. Greenway-Hayden Loop #100 Scottsdale, AZ 85260 Phone: 602-596-4066
# DEVICE:
Name of Device: Polymer OBL SB Anchor / Suture Combination Trade or Proprietary Name: Polymer OBL SB Anchor / Suture Combination Common or Usual Name: Polymer SB Anchor / Suture Combination Classification Name: FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Class: II
# PREDICATE DEVICE(S):
OBL claims that the Polymer OBL SB Anchor / Suture Combination has features substantially equivalent to two 510(k) approved devices. These include:
- "2.8 and 3.5mm ROC Suture Bone Fastener" by Innovasive Devices (K963402).
- "SB Bone Anchor" by Orthopaedic Biosystems Ltd., Inc (K950592)
# DEVICE DESCRIPTION:
The Polymer OBL SB Anchor is a simple modified polyacetal cylinder designed to be pushed through a hole in bone chosen by the surgeon as a site for anchoring soft tissue to bone. Each anchor is preloaded with suture and includes an insertion device.. These components are packaged together sterile in double mylar/tyvek pouches or formed trays. The device is single use. The device can be used for open and arthroscopic procedures.
The anchor is a modified cylinder made of polyacetal. From the end, the anchor is round with two grooves milled 180° apart. From the side, the anchor is a parallelogram with a hole through it. The anchors will be referred to by the diameter and the overall length, for example 2x8 means the diameter is 2.0 mm and the length is 8 mm. Refer to the drawing on page 12 for the dimensions "C" and "A".
The largest anticipated size of the anchor is 3 mm in diameter and 15 mm in length (3x15). The smallest anticipated size of the anchor is 2 mm in diameter and 8 mm in length (2x8). The purpose of the hole through the device is to hold suture. The smooth longitudinal grooves on each side of the device allow the suture to pass into the bone hole without pinching the suture. The ends of the device are designed to initiate rotation of the device as it is pushed and subsequently pulled into place. The ends are tapered at angles that range from 30° to 60°. The anchor is symmetric, allowing it to be inserted with regard to which end enters first. The shape
{1}------------------------------------------------
of the device distributes the load over a large bone surface reducing the risk of bone breakage or pullout.
The Insertion Device is a manual surgical instrument comprised of a handle, sleeve, and piston. The handle is constructed from polypropylene and serves the functions of holding all pieces together and to interface between the surgeon's hand and the device. The sleeve houses the piston and the Anchor. The sleeve keeps the Anchor and piston in-line to allow the piston to push the Anchor into the implantation site.
The Suture is Braided Polyester Suture from Surgical Specialties (See Appendix 3 for Approval Letters. N80950). The largest anticipated diameter of the suture USP #2. The smallest anticipated diameter of suture is USP #2-0. The suture is preloaded in the anchor eyelet.
### Materials
Anchor: Polyacetal Inserter: Piston and Sleeve: Stainless Steel, Handle: Polypropylene Suture: Braided Polyester from Surgical Specialties, See Appendix 3.
## Use
To implant the Anchor, the surgeon must first create an implantation site in bone. OBL supplies accessory drills and awls to prepare these sites.
Once the implantation site is prepared, the surgeon simply aligns the Anchor to the hole, pushes the handle toward the bone, retracts, and disconnects the Suture from the Inserter. The Suture will be retained by the Anchor. Pulling back on the Suture further engages the Anchor into the walls of the hole. The surgeon is then left with an Anchor in the bone and Suture to attach to the soft tissue.
## INTENDED USE:
This device. Polymer OBL SB Series and Cinch Series Anchor/Suture Combination is intended for use only for the fixation of non-absorbable synthetic sutures for the following indications.
The Polymer Cinch SB Anchor/Suture Combination is intended only for the fixation of surgical suture material to the pelvis for the purpose of bladder neck suspensions for female urinary incontinence due to urethral hypermobility of intrinsic sphincter deficiency.
Shoulder:
- Bankhart Repair 1.
- SLAP Lesion Repair
- Acromio-clavicular separation
- Rotator Cuff Repair 4.
- 5. Capsule and Capsulolabral Reconstruction
- Biceps Tenodesis 6.
- 7. Deltoid Repair
Hand, Wrist, Elbow:
- Scapholunate ligament reconstruction 1.
- Ulnar Collateral Ligament Reconstruction 2.
{2}------------------------------------------------
- Lateral Collateral Ligament Reconstruction 3.
- Biceps Tendon Reattachment 4.
- Elbow Medial/Lateral Repair of Tendons in the Elbow ഗ
#### Foot:
- Hallux Valgus Reconstruction 1.
- Mid and Forefoot Reconstruction 2.
#### Knee:
- medial collateral ligament 1.
- lateral collateral ligament 2.
- posterior oblique ligament 3.
- Joint capsule closure 4.
- న్న Illiotibial band tendonesis
- 6. VMO Advancement
Urinary: Bladder Neck Suspension
## COMPARISON OF CHARACTERISTICS:
## OBL Polymer OBL SB Anchor / Suture Combination vs. OBL SB Bone Anchor
The design of the OBL Polymer OBL SB Anchor is similar to the SB Bone Anchor. The only difference between the two implants is the material. The SB Bone Anchor is constructed of Titanium Alloy Ti6Al4VELI. The OBL Polymer OBL SB Anchor is constructed from Polyacetal. The indications for use for the two products is identical.
# OBL Polymer OBL SB Anchor / Suture Combination vs. Innovasive 2.8 mm and 3.5 mm ROC Suture Bone Fastener
Both devices utilize Polyacetal as a material in the construction of the anchor. Both devices utilize suture to join soft tissue to the anchor. Both devices can be EtO sterilized.
## PERFORMANCE DATA
The anchors were tested for two physical characteristics. Pullout Strength and Suture Break Strength.
Pullout Strength was determined in a cancellous bone model. The smallest version of the proposed devices measures 2 mm by 8 mm and provides 9.6 lbs of pullout strength. The largest version of the proposed devices measures 6 mm by 15 mm and provides 14 lbs of pullout strength. By comparison, the predicate Innovasive product showed 9 lbs of pullout strength in this model.
The knot is considered the weakest part of a suture because of the stress riser it creates. The USP listed knot break strength for #2 synthetic suture is 14 lbs. The smallest version of the proposed devices handles 19.4 lbs of strength before the suture breaks. The largest version of the proposed devices handles 23.2 lbs of strength before the suture breaks. This means that as force is applied to the suture at the knot will fail before the suture at the anchor.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked one behind the other, forming a visual representation of people.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR - 4 1999
Mr. Jeffrey B. Baldwin Vice President of Engineering Orthopaedic Biosystems Limited, Inc. 15990 North Greenway-Hayden Loop, Suite 100 Scottsdale, Arizona 85260
K982996 Re: Polymer OBL SB Suture/Anchor Combination Requlatory Class: II Product Codes: JDR, MBI, and GAT Dated: December 2, 1998 December 8, 1998 Received:
Dear Mr. Baldwin:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. ਜੋ substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{4}------------------------------------------------
## Page 2 - Mr. Jeffrey B. Baldwin
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Nunk U.S. Sayre
n Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# STATEMENT OF INDICATIONS FOR USE
510(k) Number: K982996
Device Name: Polymer OBL SB Anchor / Suture Combination
This device, Polymer OBL SB Series and Cinch Series Anchor/Suture Combination is intended I ms device, I oryther ODD of non-absorbable synthetic sutures for the following indications.
The Polymer Cinch SB Anchor/Suture Combination is intended only for the fixation of surgical r no receive and to the pelvis for the purpose of bladder neck suspensions for female urinary incontinence due to urethral hypermobility of intrinsic sphincter deficiency.
#### Shoulder:
- : Bankhart Repair 】.
- SLAP Lesion Repair 2.
- Acromio-clavicular separation 3.
- Rotator Cuff Repair 4.
- Capsule and Capsulolabral Reconstruction న్.
- Biceps Tenodesis 6.
- Deltoid Repair 7.
#### Hand, Wrist, Elbow:
- Scapholunate ligament reconstruction 1.
- Ulnar Collateral Ligament Reconstruction 2.
- Lateral Collateral Ligament Reconstruction 3.
- Biceps Tendon Reattachment 4.
- Elbow Medial/Lateral Repair of Tendons in the Elbow રું.
#### Foot:
- Hallux Valgus Reconstruction 1.
- Mid and Forefoot Reconstruction 2.
#### Knee:
- medial collateral ligament 1.
- lateral collateral ligament 2.
- posterior oblique ligament 3.
- Joint capsule closure 4.
- న్ Illiotibial band tendonesis
- 6. VMO Advancement
Urinary: Bladder Neck Suspension
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| | <i>(Sign-Off)</i> |
|------------------|------------------------------------------|
| | <i>in of General Restorative Devices</i> |
| <i>K) Number</i> | K982996 |
| 20 | Prescription Use |
|----|----------------------|
| | (Per 21 CFR 801.109) |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.