The Osteonics® Extended Femoral Bearing Heads are indicated for single-use in patients requiring either a Total Hip Replacement or a Hemi- or Bipolar Hip Replacement. The Osteonics® Extended Femoral Bearing Heads are intended to be used with a specific listing of Osteonics® femoral stems identified in Attachment 1 and are designed to accommodate a patient population requiring the extended offset in order to restore ligamentous balance.
Device Story
Modular femoral bearing head; spherical design; 22, 26, 28, 32mm diameters; +12.5mm and +15mm neck extensions. Input: mechanical interface with compatible Osteonics femoral stems via Morse or C-Taper configurations. Operation: provides extended offset to restore ligamentous balance in hip arthroplasty. Output: physical bearing surface for hip joint articulation. Used in surgical settings by orthopedic surgeons. Benefits: restores joint stability and function in patients with hip disease or trauma requiring specific offset geometry.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Material: ASTM F-799 Cobalt chromium alloy. Configurations: Morse Taper and European (C-Taper) female tapers. Diameters: 22, 26, 28, 32mm. Extensions: +12.5mm, +15mm. Surface: Highly polished, optional nitrogen ion treatment. Mechanical device; no software or energy source.
Indications for Use
Indicated for patients requiring Total Hip, Hemi-, or Bipolar Hip replacement. Conditions include femoral head/neck fractures, non-unions, aseptic necrosis, osteo/rheumatoid/post-traumatic arthritis, failed previous arthroplasty, or conditions requiring conservative acetabular procedures. Designed for patients needing extended offset (+12.5mm or +15mm) to restore ligamentous balance.
Regulatory Classification
Identification
A hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis is a device intended to be implanted to replace a portion of the hip joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum. This generic type of device includes designs which are intended to be fixed to the bone with bone cement (§ 888.3027) as well as designs which have large window-like holes in the stem of the device and which are intended for use without bone cement. However, in these latter designs, fixation of the device is not achieved by means of bone ingrowth.
Predicate Devices
Osteonics® Morse Taper Heads
Osteonics® C-Taper Heads
DePuy Articule-Eze® Total Hip Balls
Submission Summary (Full Text)
{0}------------------------------------------------
OCT 21 1998
R982936
## 510(k) Premarket Notification Summary of Safety and Effectiveness for the Osteonics® Extended Femoral Bearing Head
#### Submission Information:
Name and Address of the Sponsor of the 510(k) Submission:
Contact Person:
Date of Summary Preparation:
Device Identification
Proprietary Name:
Common Name:
Classification Name/Reference:
Osteonics Corporation 59 Route 17 Allendale NJ 07401-1677
Marybeth Naughton Regulatory Affairs Team Member
August 20, 1998
Osteonics® Extended Femoral Bearing Head
Modular Femoral Bearing Head
Hip Joint, Metal/Polymer, Semi-constrained, Cemented Prosthesis 21 CFR §888.3350
## Predicate Device Identification:
The Osteonics® Extended Femoral Bearing Heads are substantially equivalent to the following predicate devices: The Osteonics® Morse Taper Heads, the Osteonics® C-Taper Heads, and the DePuy Articule-Eze® Total Hip Balls.
#### Description of Devices:
The Osteonics® Extended Femoral Bearing Heads are ASTM F-799 Cobalt chromium alloy devices available in two female taper configurations, a Morse Taper and a European Taper (Osteonics` trade name "C-Taper"). The devices are spherical in design and are available in 22, 26, 28, and 32mm diameters with +12.5mm and +15mm neck extensions. The outer surface of the Osteonics® Extended Femoral Bearing Heads are highly polished and are available both with and without nitrogen ion surface treatment.
{1}------------------------------------------------
## Intended Use:
The Osteonics® Extended Femoral Bearing Heads are indicated for single-use in patients requiring either a Total Hip Replacement or a Hemi- or Bipolar Hip Replacement. The Osteonics® Extended Femoral Bearing Heads are intended to be used with a specific listing of Osteonics® femoral stems identified in Attachment 1 and are designed to accommodate a patient population requiring the extended offset in order to restore ligamentous balance.
## Statement of Technological Comparison:
The design, materials, and intended use of the Osteonics® Extended Femoral Bearing Heads are substantially equivalent in design, materials and intended use to the predicate Osteonics® Morse Taper Heads, the Osteonics® C-Taper Heads and the DePuy Articule-Eze® Total Hip Balls.
## Performance Data:
The comparison of the Osteonics® Extended Femoral Bearing Head to predicate systems which have been determined to be substantially equivalent through 510(k) premarket notifications demonstrates that no additional performance data are necessary.
{2}------------------------------------------------
# ATTACHMENT 1 (Page 1)
# Osteonics® Femoral Stems to be used with Osteonics® Extended Femoral Bearing Heads
| Catalog Number | Description |
|-------------------|-------------------------------------------------------------------------------------|
| S-1162-HFXX | CoCr Alloy, Plasma Sprayed, Omnifit, Collared, Normalized, Morse Taper Trunnion |
| 1003-xxxx | CoCr Alloy, Satin Finish, Omnifit, Collarless, Normalized, Morse Taper Trunnion |
| 1004-xxxx | CoCr Alloy, Satin Finish, Omnifit, Collared, Normalized, Morse Taper Trunnion |
| 1005-xxxx | Ti6Al-4V Alloy, Satin Finish, Omnifit, Collarless, Normalized, Morse Taper Trunnion |
| 1006-xxxx | Ti6Al-4V Alloy, Satin Finish, Omnifit, Collared, Normalized, Morse Taper Trunnion |
| 1012-xxxx | Ti6Al-4V Alloy, Satin Finish, Omnifit, Collared, Normalized, Morse Taper Trunnion |
| 1017-xxxx | Ti6Al-4V Alloy, HA Coated, Omnifit, Collarless, Normalized, Morse Taper Trunnion |
| 1044-xxxx | CoCr Alloy, Satin Finish, Omnifit, Collared, Normalized, Morse Taper Trunnion |
| 1045-xxxx | Ti6Al-4V Alloy, Satin Finish, Omnifit, Collarless, Normalized, Morse Taper Trunnion |
| 1046-xxxx | Ti6Al-4V Alloy, Satin Finish, Omnifit, Collared, Normalized, Morse Taper Trunnion |
| 1055-xxxx | Ti6Al-4V Alloy, HA Coated, Omnifit, Collared, Normalized, Morse Taper Trunnion |
| 1092-xxxx | Ti6Al-4V Alloy, HA Coated, Omnifit, Collarless, Normalized, Morse Taper Trunnion |
| 1093-xxxx | Ti6Al-4V Alloy, HA Coated, Omnifit, Collared, Normalized, Morse Taper Trunnion |
| 1094-xxxx | Ti6Al-4V Alloy, HA Coated, Omnifit, Collarless, Normalized, Morse Taper Trunnion |
| 21-xxxx | CoCr Alloy, Satin Finish, Omnifit, Collared, Smooth, Morse Taper Trunnion |
| 6005-xxxx | Ti6Al-4V Alloy, Satin Finish, Omnifit, Collarless, Normalized, C-Taper Trunnion |
| 6006-xxxx | Ti6Al-4V Alloy, Satin Finish, Omnifit, Collared, Normalized, C-Taper Trunnion |
| 6010-xxxx | CoCr Alloy, Satin Finish, Omnifit, Collared, Normalized, C-Taper Trunnion |
| 6017-xxxx | Ti6Al-4V Alloy, HA Coated, Omnifit, Collarless, Normalized, C-Taper Trunnion |
| 6025-xxxx | Ti6Al-4V Alloy, Satin Finish, Omnifit. Collarless, Normalized, C-Taper Trunnion |
| 6026-xxxx | Ti6Al-4V Alloy, Satin Finish, Omnifit. Collared, Normalized, C-Taper Trunnion |
| 6033-xxxx | CoCr Alloy. Satin Finish. Omnifit. Collarless. Normalized, C-Taper Trunnion |
| Catalog<br>Number | Description |
| 6034-xxxx | CoCr Alloy, Satin Finish, Omnifit, Collared, Normalized, C-Taper Trunnion |
| 6070-xxxx | CoCr Alloy, Grit Blasted, Omnifit, Collared, Smooth, C-Taper Trunnion |
| 6072-xxxx | CoCr Alloy, Rough Finish, Omnifit, Collared, Smooth, C-Taper Trunnion |
| 6079-xxxx | CoCr Alloy, Satin Finish, Omnifit, Collared, Smooth, C-Taper Trunnion |
| 6088-xxxx | CoCr Alloy, Satin Finish, Omnifit, Collared, Normalized, C-Taper Trunnion |
| S0853-HF0x | CoCr Alloy, Grit Blasted, Omnifit, Collarless, Smooth C-Taper Trunnion |
| S1003-xxxx | CoCr Alloy, Plasma Sprayed, Omnifit, Collarless, Normalized, Morse Taper Trunnion |
| S1003-xxxx-A | CoCr Alloy, Satin Finish, Omnifit, Collarless, Normalized, Morse Taper Trunnion |
| S1003-xxxx-C | CoCr Alloy, Satin Finish, Omnifit, Collarless, Normalized, Morse Taper Trunnion |
| S1003-xxxx-D | CoCr Alloy, Satin Finish, Omnifit, Collarless, Normalized. Morse Taper Trunnion |
| S1012-xxxx-A | Ti6Al-4V Alloy, Satin Finish, Omnifit, Collarless, Normalized, Morse Taper Trunnion |
| S1149-xxxx | Ti6Al-4V Alloy, HA Coated, Omnifit, Collared, Normalized, Morse Taper Trunnion |
| S1296-HF0x | Ti6Al-4V Alloy, Satin Finish, Omnifit, Collared, Normalized, Morse Taper Trunnion |
| S1414-xxxx | CoCr Alloy, Grit Blasted, Omnifit, Collared, Smooth, C-Taper Trunnion |
| S1456-xxxx | Ti6Al-4V Alloy, Satin Finish, Omnifit, Collared, Normalized, Morse Taper Trunnion |
| S1462-HF0x | CoCr Alloy, Grit Blasted, Omnifit, Collared, Smooth, C-Taper Trunnion |
| S1966-xxxx | CoCr Alloy, Satin Finish, Omnifit, Collarless, Normalized, C-Taper Trunnion |
| S21-xxxx | CoCr Alloy, Satin Finish, Omnifit, Collared, Smooth, Morse Taper Trunnion |
| S21-xxxx-A | CoCr Alloy, Satin Finish, Omnifit, Collared, Smooth, Morse Taper Trunnion |
| S21-xxxx-B | CoCr Alloy, Satin Finish, Omnifit, Collared, Smooth, Morse Taper Trunnion |
| S21-xxxx-C | CoCr Alloy, Satin Finish, Omnifit, Collarless, Smooth, Morse Taper Trunnion |
| S2591-xxxx | CoCr Alloy, Satin Finish, Omnifit, Collared, Normalized, C-Taper Trunnion |
{3}------------------------------------------------
# ATTACHMENT 1 (PAGE 2)
## Osteonics® Femoral Stems to be used with Osteonics® Extended Femoral Bearing Heads
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure, composed of flowing lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 21 1998
Ms. Elizabeth A. Staub Director, Quality Assurance and Regulatory Affairs Osteonics Corporation 59 Route 17 07401-1677 Allendale, New Jersey
Re: K982936 Osteonics® Extended Femoral Bearing Head Trade Name: Regulatory Class: İI Product Code: LWJ Dated: August 20, 1998 Auqust 21, 1998 Received:
Dear Ms. Staub:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions The general controls provisions of the Act of the Act. include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the current Good Manufacturing Practice requirement, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic (QS) inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory In addition, FDA may publish further announcements action. concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
{5}------------------------------------------------
Page 2 - Ms. Elizabeth A. Staub
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4659. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
Celia M. Whitton, Ph.D.
~ Celia M. Witten, Ph.D., M.D. Director Division of General and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Page 1 of 1
# 510(k) Number (if Known): K 9829 36
Device Name: Osteonics® Extended Femoral Bearing Head
#### Indications For Use:
The Osteonics® Extended Femoral Bearing Heads are intended for single use in patients requiring either a Total Hip Replacement or Hemi- or Bipolar Hip Replacement. The Osteonics® Extended Femoral Bearing Heads are designed to accommodate a patient population requiring extended offset provided by a plus 12.5mm or a plus 15mm femoral head option when used in conjunction with the Osteonics® Femoral Stems specified in this submission.
#### Indications:
For use as a Bipolar or Hemi-Hip Replacement:
- · Femoral head/neck fractures or non-unions.
- Aseptic necrosis of the femoral head. ●
- Osteo-, rheumatoid, and post-traumatic arthritis of the hip with minimal . acetabular involvement or distortion.
Other considerations for use as a Bipolar or Hemi-Hip Replacement:
- · Pathological conditions or age considerations which indicate a more conservative acetabular procedure and an avoidance of the use of bone cement in the acetabulum.
- Salvage of failed total hip arthroplasty. .
For use as a Total Hip Replacement:
- · Painful, disabling joint disease of the hip resulting from: degenerative arthritis, rheumatoid arthritis, post-traumatic arthritis or late stage avascular necrosis.
- · Revision of previous unsuccessful femoral head replacement, cup arthroplasty or other procedure.
- Clinical management problems where arthrodesis or alternative reconstructive ● techniques are less likely to achieve satisfactory results.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(per 21 CFR 801.109)
OROver-The-Counter Use No
(Optional Format 1-2-96)
(Division Sign-Off)
Division of General Restorative Devices
510(k) Number K482936
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.