MEDRAD ANTERIOR NECK COIL, 1.5T & 1.0T MODEL NUMBERS M64ANC AND M42ANC
K982932 · Medrad, Inc. · LNH · Nov 2, 1998 · Radiology
Device Facts
Record ID
K982932
Device Name
MEDRAD ANTERIOR NECK COIL, 1.5T & 1.0T MODEL NUMBERS M64ANC AND M42ANC
Applicant
Medrad, Inc.
Product Code
LNH · Radiology
Decision Date
Nov 2, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
The Medrad Anterior Neck Coil is a receive only coil intended to be used in MRI Scanner Systems for imaging of the area superior to the oral pharynx and inferior to the brachial plexus. The Medrad Anterior Neck Coil is intended for use only under the supervision of a physician who is trained in the field of Diagnostic Magnetic Resonance Imaging.
Device Story
Medrad Anterior Neck Coil is a receive-only surface coil for MRI scanners; designed to enhance imaging of the neck region (pharynx, throat, tongue, larynx, thyroid). Device captures hydrogen nuclei signals during 2D/3D MRI procedures. Operated by trained MRI technicians/physicians in clinical settings. Output is raw MR signal data processed by the host MRI scanner to generate diagnostic images. Device facilitates anatomical visualization for clinical decision-making. Medrad manufactures the device, which is identical to the predicate GE Anterior Neck Coil except for branding/labeling.
Clinical Evidence
Bench testing included Signal-to-Noise Ratio (SNR) comparison, NEMA-standard image uniformity evaluation, and assessment of geometric distortion and RF field distortion. Clinical images were provided to demonstrate effectiveness for the 1.5T coil; 1.0T performance was inferred from SNR results. No clinical trials were required due to identical design to the predicate.
Technological Characteristics
Receive-only surface coil; hydrogen nuclei excitation; compatible with GE Superconducting MRI scanners. Materials are identical to the predicate GE device. Design includes blocking network to prevent induced currents. Static magnetic field distortion is mitigated by component positioning.
Indications for Use
Indicated for patients requiring MRI imaging of the anatomy superior to the oral pharynx and inferior to the brachial plexus, including the pharynx, throat, tongue, larynx, and thyroid regions. Intended for use under physician supervision.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
General Electric (GE) Anterior Neck Coil (PMA P830074)
Submission Summary (Full Text)
{0}------------------------------------------------
NOV 2 1998
### 510(k) SUMMARY MEDRAD 1.0T, 1.5T ANTERIOR NECK COILS
| OFFICIAL CONTACT: | Jim Ferguson, Jr.<br>Sr. Regulatory Affairs Associate<br>Medrad, Inc.<br>One Medrad Drive<br>Indianola, PA 15051<br>(412) 767-2400 Ext. 3326 |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| CLASSIFICATION NAME: | Magnetic Resonance Diagnostic Accessory<br>[21 CFR 892.1000} |
| COMMON/USUAL NAME: | MR Imaging Surface Coil |
| PROPRIETARY NAME: | Medrad Anterior Neck Coil |
| PREDICATE DEVICES: | General Electric (GE) Anterior Neck Coil |
## DEVICE DESCRIPTION:
The Medrad Anterior Neck Coil is a receive only coil designed to enhance the MR Imaging of the area superior to the oral pharynx and inferior to the brachial plexus. This area includes the pharynx, throat, tongue, larynx and thyroid regions.
### INTENDED USE:
The Medrad Anterior Neck Coil is a receive only coil intended to be used with the General Electric Superconducting MRI Scanners. This coil is intended to facilitate complete MR imaging of the area superior to the oral pharynx and inferior to the brachial plexus. This area includes the pharynx, throat, tongue, larynx and thyroid regions anatomy.
| Anatomical Region: | The area superior to the oral pharynx and<br>inferior to the brachial plexus. This area<br>includes the pharynx, throat, tongue, larynx and<br>thyroid regions anatomy. |
|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Nuclei Excited: | Hydrogen |
| Diagnostic Uses: | 2D and 3D Imaging |
{1}------------------------------------------------
510(k) Summary Medrad Anterior Neck Coil Page 2 of 3
# PROPOSED MEDRAD ANTERIOR NECK COIL TECHNICAL COMPARISON TO PREDICATE DEVICES:
The following table compares the predicate GE Anterior Neck Coil and the proposed Medrad Anterior Neck Coil.
| GE Anterior Neck Coil<br>(PMA P830074) | Medrad Anterior Neck Coil |
|----------------------------------------|---------------------------|
| GE labeling | Medrad labeling |
Medrad was responsible for the design and manufacturing of the predicate device, and will remain responsible for the design and manufacturing for the proposed device. The device will be labeled and marketed as "Medrad Product". This is the only change to the device.
# PROPOSED MEDRAD ANTERIOR NECK TECHNICAL COMPARISON TO PREDICATE DEVICES
| Patient contacting materials comparison information | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| GE Anterior Neck Coil | Medrad Anterior Neck Coil |
| Reference PMA P830074<br>for material information | All materials used are the<br>same as the GE Anterior Neck coil. (Predicate<br>device) |
| Certification: Medrad, Inc. certifies that the patient<br>contacting materials and formulations for<br>the proposed Medrad Anterior Neck Coils<br>are unchanged from currently marketed<br>devices. | |
{2}------------------------------------------------
#### PERFORMANCE TEST DATA:
#### SIGNAL TO NOISE RATIO (SNR)
A Signal to Noise Ratio(SNR) study was conducted to generate a Signal -To-Noise ratio comparison between the proposed Medrad Anterior Neck coil and the predicate GE Anterior Neck Coil.
IMAGE UNIFORMITY - The Medrad Anterior Neck Coil was evaluated using NEMA Standards to characterize the non-uniformity of the proposed coil. Contours of the images obtained with the coil were constructed for the axial image and sagittal image.
GEOMETRIC DISTORTION: None. Static magnetic field distortion results from the use of magnetic materials in the construction of a surface coil. This device contains slightly magnetic materials or components. However, such components have been positioned within the surface coil so that no observable distortion of the static magnetic field is present.
TRANSMIT RF FIELD IB11 DISTORTION - Analysis of the electrical design of the coil and its blocking network demonstrates that no significant currents are induced. No artifacts of any type were observed during imaging.
RESOLUTION, SLICE THICKNESS, AND CONTRAST - These performance parameters are not affected by the use of a surface coil and were not separately tested in the performance evaluation of the proposed Medrad Anterior Neck Coil.
CLINICAL EVALUATION - Clinical images for the proposed 1.5T Anterior Neck Coil have been provided with this submission to demonstrate the clinical effectiveness of the Anterior Neck coils. Based on these results, and the SNR results for the 1.0T Anterior Neck Coil, Medrad concludes that the proposed 1.0T Anterior Neck Coil will produce similar clinical image results.
CONCLUSION - Extensive safety, verification, durability and clinical testing was conducted on the predicate GE device (PMA P830074). This device is exactly the same as the proposed device except for the labeling. The proposed device is deemed by Medrad to be substantially equivalent.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol, with three curved lines forming the body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the bird symbol.
Re: ·
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 2 1998
Jim Ferguson, Jr. Sr. Regulatory Affairs Associate Medrad, Inc. One Medrad Drive Indianola, PA 15051
K982932 Medrad Anterior Neck Coil Dated: August 19, 1998 Received: August 20, 1998 Regulatory class: II 21 CFR 892.1000/Procode: 90 LNH
Dear Mr. Ferguson:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in ritto diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdri/dsmadsmamain.html".
Sincerely yours,
Lillian Yin, Ph.D.
Lillian Yin, Ph.D. Director, Division of Reproductive Abdominal, Ear, Nose and Throa and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) NUMBER (IF KNOWN): K982932
Medrad Anterior Neck Coil DEVICE NAME:
INDICATIONS FOR USE:
The Medrad Anterior Neck Coil is a receive only coil intended to be used in MRI Scanner Systems for imaging of the area superior to the oral pharynx and inferior to the brachial plexus.
The Medrad Anterior Neck Coil is intended for use only under the supervision of a physician who is trained in the field of Diagnostic Magnetic Resonance Imaging.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Prescription Use
(Per 21 CFR 801.109) √
OR
Over-The-Counter-Use (Optional Format 1-2-96)
Vivid G. Segura
---
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological D
510(k) Number K982432
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.