← Product Code [EWO](/productcode/EWO) · K982878

# EARSCAN 3 (K982878)

_Micro Audiometrics Corp. · EWO · Nov 3, 1998 · Ear, Nose, Throat · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K982878

## Device Facts

- **Applicant:** Micro Audiometrics Corp.
- **Product Code:** [EWO](/productcode/EWO.md)
- **Decision Date:** Nov 3, 1998
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 874.1050
- **Device Class:** Class 2
- **Review Panel:** Ear, Nose, Throat

## Indications for Use

Earscan 3 is a peripheral used with a computer that creates an air conduction, bone conduction and speech audiometer, as defined by the standards shown in the labeling of the About Screen and the Splash Screen of the Earscan 3 software.

## Device Story

Earscan 3 is a computer-interfaced audiometer; functions as a peripheral to test auditory pathways. Inputs include manual and automatic air conduction, bone conduction, and speech signals. Device transforms inputs into audiometric data; operates via software-controlled tone generation. Used in industrial, school, and medical settings by trained personnel. Firmware limits continuous tone duration to prevent excessive sound exposure; software enforces hearing level limits per ANSI S3.6-1996 and IEC 60645 standards. Output displayed on computer screen for healthcare provider review; assists in clinical assessment of hearing condition. Benefits include integrated testing capabilities and automated safety controls.

## Clinical Evidence

No clinical data provided. Effectiveness is established through bench testing, compliance with ANSI S3.6-1996 and IEC 60645 standards, and demonstration of substantial equivalence to legally marketed predicate devices.

## Technological Characteristics

Computer-interfaced audiometer; air conduction headband and bone conduction vibrator compliant with ANSI S3.6. Electrical safety per IEC 601-1.1992 and UL2601. Connectivity via USB 1.0. Firmware-based safety controls for tone duration and intensity limits. Operates as a Type 3C/Type 3 audiometer.

## Regulatory Identification

An audiometer or automated audiometer is an electroacoustic device that produces controlled levels of test tones and signals intended for use in conducting diagnostic hearing evaluations and assisting in the diagnosis of possible otologic disorders.

## Predicate Devices

- Micro Audiometrics Corp. Earscan ([K812529](/device/K812529.md))
- Micro Audiometrics Corp. Earscan II ([K812529](/device/K812529.md))
- Decibel Instruments, Inc. ProDigit 2000 ([K940999](/device/K940999.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## 510 (K) Summary of Safety and EFFECTIVENESS

| Submitter's Name:<br>Address:     | Micro Audiometrics Corp.<br>2200 So. Ridgewood Ave.<br>South Daytona, FL 32119                                                                    |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:<br>Voice:<br>Fax: | Patricia L. Travis<br>(904) 788-9331<br>(904) 788-9332                                                                                            |
| Date Prepared:                    | August 11, 1998                                                                                                                                   |
| Device Name:                      | Classification Name - Audiometer<br>Common/Usual Name - Air Conduction, bone conduction,<br>and speech audiometer<br>Proprietary name - Earscan 3 |
| Predicate Device:                 | Micro Audiometrics Corp. Earscan (K812529)<br>Micro Audiometrics Corp. Earscan II (K812529)<br>Decibel Instruments, Inc. ProDigit 2000 (K940999)  |

## Device Description, Intended Use & Effectiveness:

The Earscan 3 is similar to and predicated upon the Micro Audiometrics Corp. Earscan and Earscan II, and Decibel Instruments, Inc. ProDigit 2000. Earscan 3 is a computer peripheral whose purpose is to test the condition of the auditory pathway through manual and automatic air conduction audiometry, bone conduction audiometry and speech audiometry. Earscan 3 is a lower level of concern device and is intended to be used by trained personnel in the industrial, school and medical environment.

## Safety:

The design of Earscan 3 provides electrical safety to the patient and the user. Earscan 3 intends to meet electrical standards IEC 601-1.1992 and UL2601. The wall cube will comply with UL2601-1. To prevent the unlikely possibility of excessive exposure to high-level sounds, a firmware check was implemented which limits the maximum duration of continuous tones. The software also prevents the generation of a hearing level greater than specified by ANSI S3.6-1996 Type 3C audiometer and IEC 60645 Type 3 audiometer. In complying with USB Spec. 1.0, Earscan 3 will not interfere with other devices in its computer environment. With respect to mechanical safety, the air conduction headband comply with ANSI S3.6, and the bone conduction vibrator and headband comply with ANSI S3.6.

## Summary of Effectiveness:

Micro Audiometrics Corporation's Earscan 3 is a combination of three predicate audiometer devices into one computer interfaced audiometer equivalent or superior in effectiveness to its predicates in testing the auditory pathway.

{1}------------------------------------------------

Re:

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Image /page/1/Picture/15 description: The image shows a black and white logo or emblem. On the left side, there's a stylized, abstract design that could represent a figure or symbol. To the right of this design, there is a horizontal line. The image appears to be a cropped or partial view of a larger document or design.

NOV . 3 1998

Jason R. Keller President Micro Audiometrics Corp. 2200 So. Ridgewood Ave., US#1 South Dayton, FL 32119-3018

K982878 Earscan 3 Dated: August 11, 1998 Received: August 14, 1998 Regulatory class: II 21 CFR 874.1050/Procode: 77 EWO

Dear Mr. Keller:

We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.

This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdrl/dsmardsmamain.html".

Sincerely yours,

Lillian Yin, Ph.D.

Lillian Yin, Ph.D. Director, Division of Reproductiv Abdominal, Ear, Nose and Thros and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

510 (k) NUMBER (IF KNOWN) : K982878

Earscan 3 DEVICE NAME:

INDICATIONS FOR USE:

Earscan 3 is a peripheral used with a computer that creates an air conduction, bone conduction and speech audiometer, as defined by the standards shown in the labeling of the About Screen and the Splash Screen of the Earscan 3 software.

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED.)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use
(Per 21 CFR 801.109) √

OR

Over-The-Counter-Use (Optional Format 1-2-96)

Schmit G. Sezer

(Division Sign-Off) Division of Reproductive, Abdominal, ENT and Radiological I 510(k) Number

---

**Source:** [https://fda-staging.innolitics.com/device/K982878](https://fda-staging.innolitics.com/device/K982878)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
