The HEARTLink II System is intended to provide cardiac monitoring from a low-risk patient's home a central monitoring facility pursuant to to physician prescription.
Device Story
HEARTLink II is a microprocessor-based ECG arrhythmia detection and alarm system. It receives radiofrequency-encoded ECG signals from a patient-worn unit; processes signals to detect arrhythmias; and transmits data via standard telephone lines to a remote Central Station. The Central Station, located in a healthcare facility, displays ECG data and sounds alarms for review by medical personnel. The system supports up to 30 patients per Central Station. It is intended for home use by low-risk patients. Healthcare providers use the output to evaluate cardiac rhythm, monitor post-operative recovery, and assess conditions like atrial fibrillation or syncope. The device benefits patients by enabling remote cardiac monitoring outside of a hospital setting.
Clinical Evidence
Two studies were conducted: a usability study at Saint Francis Medical Center and Washington Center Hospital, and a transmissibility study using healthy volunteers wearing the HEARTLink II and a predicate device simultaneously for four days. Results showed the system is as usable as the predicate and transmits ECG data with equal fidelity. Bench testing confirmed compliance with safety, emissions, and effectiveness standards including FCC Parts 15/68, AAMI EC-13, and EN 60601-1-2.
Technological Characteristics
Microprocessor-based ECG arrhythmia detection and alarm system. Operates via radiofrequency transmission of ECG signals to a monitoring unit, then via telephone lines to a Central Station. Supports 30-patient capacity per station. Complies with EN 60601-1-2 (EMC), FCC Parts 15 and 68, and AAMI EC-13 standards. Database management software used for non-physiological information.
Indications for Use
Indicated for low-risk patients requiring cardiac monitoring, including those with dizziness, lightheadedness, palpitations, syncope of unknown etiology, non-life-threatening arrhythmias (e.g., atrial fibrillation, supra-ventricular arrhythmias, bradyarhythmias, intermittent bundle branch block), post-CABG monitoring, co-morbid conditions (hyperthyroidism, chronic lung disease), obstructive sleep apnea for nocturnal arrhythmia evaluation, and stroke/transient cerebral ischemia evaluation.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".
December 6, 2023
Cardiac Telecom Corp. Jonathan S. Kahan Hogan & Hartson L.L.P. 555 Thirteenth Street. N.W. Washington, DC 20004-1109
Re: K982803
Trade/Device Name: Heartlink II (Arrhythmia Detector and Alarm System) Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector And Alarm (Including St-Segment Measurement And Alarm) Regulatory Class: Class II Product Code: QYX, DSI
Dear Jonathan S. Kahan:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated November 13, 1998. Specifically, FDA is updating this SE Letter because FDA has created a new product code to better categorize your device technology.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Jennifer Kozen, OHT2: Office of Cardiovascular Devices, 301-796-5813, Jennifer.Kozen@fda.hhs.gov.
Sincerely,
# Jennifer W. Shih -S
Jennifer Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three tail feathers, representing the department's mission to protect the health of all Americans and provide essential human services. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 1 3 1998
Cardiac Telecom Corporation c/o Mr. Jonathan S. Kahan Hogan & Hartson L.L.P. 555 Thirteenth Street, N.W. Washington, DC 20004-1109
K982803 Re : HeartLink II (Arrhythmia Detector and Alarm System) Requlatory Class: III (three) Product Code: 74 DSI Dated: Auqust 10, 1998 Received: August 11, 1998
Dear Mr. Kahan:
We have reviewed your Section 510(k) notification of intent to market we nave reviews for above and we have determined the device is che device forcemed above a the indications for use stated in the enclosure) to legally marketed predicate devices marketed in enclosure) to regally marketed problems the enactment date of the Interstate Commerce proble of to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and In actoriation with one may, therefore, market the device, subject to Cosmeral controls provisions of the Act. The general controls che general concreas provinements for annual registration, provisions of the not on manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special II your device ID Srabbing approval), it may be subject to such controls) of Crabb III- (isting major regulations affecting your device ( audicional Concervior - Enzor-Federal Regulations, Title 21, Parts 800 can be really equivalent determination assumes compliance co 093. If babbaneadad
with the Current Good Manufacturing Practice requirements, as set with the Ourlone oty System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS ocherar rogalation on Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in aboumperone. In addition, FDA may publish further announcements regaratory acoronvice in the Federal Register. Please note: this concerning your actroom submission submission does not affect any response to your premaire sections 531 through 542 of the Act obrigation you might well and the Radiation Control provisions, or other Federal laws or regulations.
-
{2}------------------------------------------------
#### Page 2 - Mr. Jonathan S. Kahan
This letter will allow you to begin marketing your device as described
ing the begineer tification - The FDA finding of substantial This letter will allow you to begin marketing your and of substantial
in your 510(k) premarket notification. The Finding device in your 510(k) premarket notification. The best breaticate device equivalence of your device to a regarly marroos.
results in a classification for your device and thus, permits your resures I proceed to the market.
If you desire specific advice for your device on our labeling
If you desire specific advice book a bodditionally 809 10 for in vi If you desire specific advice for your actlys 09 10 for in vitro
regulation (21 CFR Part 801 and additionally 809.10 for in vitro regulation (2) CFR Part our and actressff oof Compliance at diagnostic devices), please contact on the promotion and (301) 594–4648. Additionaly, for questions on effec promocoment.
advertising of your device, please contact the office of Compliance at
advertising of your device, please co advertising of your device, prouse the regulation entitled, (301) 594-4639. Also, please note the regarication"(21 CFR 807.97).
"Misbranding by reference to premarket notification"(21 CFR 807.97) "Misbranding by reference to premation on the Act may Other general information of Small Manufacturers Assistance at its
be obtained from the Division of Small Manufacturers Assistance at its internet be obtained from the Division of Silations of Subles
toll-free number (800) 638–2041 or (301) 443–6597, or at its internet
toll-free number (800) esse-stadeb (dama/ghtml") toll-free number (800) 036-2011 01:51:00 address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Thomas J. Callahan
Thomas J. Callahan, Ph.D. Director Division of Cardiovascular, Vision of Caralo Neurological Devices Office of Device Evaluation Center for Devices and Radiclogical Health
Enclosure
{3}------------------------------------------------
Page 1 of of 1 a
510(k) Number (If known): K982803
HEARTLink II _________________________________________________________________________________________________________________________________________________________________ Device Name:
Indications For Use:
- 1. Patients who have demonstrated a need for cardiac monitoring and are at low risk of developing primary ventricular fibrillation or sustained ventricular tachycardia.
- 2. Specific criteria for Indications for use are as follows :
- -Patients with dizziness or lightheadedness
- Patients with palpitations .
- Patients with syncope of unknown edglogy .
- Patients who require monitoring for non life-threstening arrhythmias, such as atrial fibrillation. . other supra-vestricular armythmias, evaluation of various bradyarhythmias and intermittent bundle branch block. This indudes post operative monitoring for these rhythims
- Patients recovering from coronary artery bypess graft (CABG) surgery who require monitoring . for arrivthmias
- Patlents requiring monitoring for antitythmia-inducing co-morbid conditions such as . hyperthyroidism or chronic lung disasse
- Patients with obstructive steep apnea to evaluate possible nocturnal arrhythmias ◀
- Patients requiring arrhythmia evaluation for ediology of stroke or translent cerebral ischemia. ● possibly secondery to atrial fibrillation
(PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation(ODE)
Vene
mar Prescription Use V (Per 21 CFR 801.109)
(Division Sign-Off) Division of Cardiovascular, Respiratory, and Neurological Devices Over-The Counter Use
510(k) Number ________________________________________________________________________________________________________________________________________________________________
(Optional Format 1-2-96)
{4}------------------------------------------------
NOV 1 3 1998
# Premarket Notification 510(k) Summary [807.92]
k982803
for
# CARDIAC TELECOM CORPORATION's
# "HEARTLink II " [ECG Arrhythmia Detector and Alarm System]
Prepared on August 1, 1998 [807.92(a)(1)]
**cardiactelecom**
corporation
We Don't Miss a Beat! 503 Braddock Ave. Turtle Creek, PA 15145 412-824.6600 Fax 412-824.4225
[A Division of Telemed Technologies, International]
{5}------------------------------------------------
# 10. 510(k) Summary:
### 10.1. Company Name and Contact Person:
# Cardiac Telecom Corporation, Inc.
503 Braddock Avenue Turtle Creek, PA 15145
412.824.6600 (Voice) 412.824.4225 (FAX)
LEE G. DENEAULT [Director of Operations]
#### 10.2. Name of the Device:
Proprietary Name: HEARTLink, Model II
Common or Usual Name: ECG Arrhythmia Detection and Alarm System
#### Classification Name:
Radio frequency physiological signal transmitter and receiver (21 C.F.R. § 870.2910)
Telephone electrocardiograph transmitter and receiver (21 C.F.R. § 870.2920)
Arrhythmia detector and alarm (21 C.F.R. & 870.1025)
{6}------------------------------------------------
# 10.3. Predicate or Legally Marketed Devices:
- 。 HEARTLink I (K934913) Cardiac Telecom Corporation (Turtle Creek, PA)
- King of Hearts (K880626) Instromedix, Inc (Hillsboro, OR)
- · Aegis (K843503) Medical Concepts, Inc (Gibbsboro, NJ)
# 10.4. Device Description:
The HEARTLink II System receives and processes radiofrequency encoded transmitted ECG signals on the Tele-Link monitoring unit. Abnormal ECGs and arrhythmias detected by the Tele-Link can be displayed on an optional video display monitor, while being simultaneously transmitted through standard telephone lines to a remote Central Station. The Central Station, located within health care institution or patient monitoring facility, is where an alarm is sounded and the ECG is again displayed for the review of medical personnel. The Central Station system has four dedicated phone lines and can provide remote monitoring of ECG data received from 30 individual Tele-Links. A total of 30 patients can be monitored at one time by a single one Central Station.
# 10.5. Intended Use:
The HEARTLink II System is intended to provide cardiac monitoring from a low-risk patient's home
{7}------------------------------------------------
a central monitoring facility pursuant to to physician prescription.
### 10.6. Technological Characteristics:
The HEARTLink II System is a microprocessorbased ECG arrhythmia detection and alarm system. The HEARTLink II is substantially equivalent to marketed ECG monitoring systems with arrhythmia detection capabilities, that have the same intended use and have received 510(k) premarket notification clearance from FDA (K880626 and K843503). The HEARTLink II is also technically equivalent to the hospital-based version of the HEARTLink I System (a.k.a. The HEARTrac System) that has already received clearance by the FDA as an arrhythmia detector and alarm system. The HEARTLink II System is intended for use in a low-risk patient's home environment and with the exception of a name change and a minimal change to the nonphysiological information in the database management software is identical to the HEARTLink I System. The HEARTLink I was approved for use by the FDA on June 5, 1995 (K934913).
### 10.7. Summary of Performance Data:
### Bench Testing:
The HEARTLink I System (in K934913) complied with all safety, emissions, and effectiveness standards under the FDA's MDS 201-0004, the
{8}------------------------------------------------
FCC's Parts 15 and 68 Rules, and the draft version of the AAMI EC-13-R-4/91 standards. The FDA's MDS 201-0004 standard has since been updated with the current version of the EMC Directive (EN 60601-1-2) and the draft of the AMMI standard has been adopted as a standard (with minimal changes). The FCC's Parts 15 and 68 have not changed since K934913. Since the HEARTLink II is technically equivalent to the HEARTLink I System, the compliance to the draft of the AAMI and the FCC's standards still apply. The HEARTLink II System, however, was safety and emissions re-tested and now complies with the current EMC Directive.
### Clinical and Healthy Volunteers Trials:
Since the HEARTLink II System's intended environment for use is in the home of low-risk patients, this new environment raised following two safety and effectiveness issues:
- · Would the patient be able to effectively use the HEARTLink II System in the home environment (Usability Issue).
- · Would the ECG data be able to be accurately transmitted to the Central Monitoring facility (Transmissability Issue).
To test these two safety and effectiveness issues, two studies were designed. To test the Usability Issue, clincal trials were done at Saint Francis Medical Center (Pittsburgh, PA) and at Washington
{9}------------------------------------------------
Center Hospital (Washington, DC). The results of this study determined that the HEARTLink II System is "as usable" as the predicate Event Recorder System. To test the "Transmissability Issue", healthy volunteers were to wear both the HEARTLink II System and the predicate for a total of four days. Data collected simultaneously on these two systems were transmitted to their appropriate Central Monitoring facilities and subsequently analyzed for transmission quality. The results of this study determined that the HEARTLink II System is capable of transmitting data with equal fidelity as its predicate Event Recorder System.
#### 10.8. Conclusions:
The HEARTLink II System is substantially equivalent to the HEARTLink I, King of Hearts, and the Aegis ECG monitoring systems. Performance data collected in laboratory bench-testing and in both clinical and healthy volunteer trails demonstrated that there are no new safety and effectiveness questions raised by the modifications made to the currently marketed HEARTLink I System.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.