K982771 · Gyrus Medical , Ltd. · HIH · Jan 21, 1999 · Obstetrics/Gynecology
Device Facts
Record ID
K982771
Device Name
GYRUS HYSTEROSCOPIC RESECTOSCOPE
Applicant
Gyrus Medical , Ltd.
Product Code
HIH · Obstetrics/Gynecology
Decision Date
Jan 21, 1999
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.1690
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Gyrus Hysteroscopic Resectoscope is intended to enable the viewing of the cervical canal and the uterine cavity for the purpose of performing diagnostic and surgical procedures. The surgical applications include tissue cutting, removal, and dissection as required or encountered in gynecologic hysteroscopic electrosurgical procedures for the treatment of intrauterine myomas, polyps, adhesions, and septa.
Device Story
Gyrus Hysteroscopic Resectoscope is an electrosurgical instrument for gynecologic procedures; enables visualization of cervical canal and uterine cavity. Device consists of electrode-specific working element, telescope, and standard sheaths; utilizes 'Quick Connection' for assembly. Surgeon operates device manually via finger controls to manipulate electrodes for cutting, coagulating, and transecting tissue; high-frequency electrical current delivered via cable. Used in clinical settings by physicians to treat intrauterine myomas, polyps, adhesions, and septa. Provides direct visualization and surgical access; facilitates tissue removal and dissection; improves patient outcomes by enabling minimally invasive intrauterine surgery.
Clinical Evidence
Bench testing only. Pre-clinical and bench-top testing performed to verify performance requirements and efficacy for hysteroscopic usage.
Technological Characteristics
Materials: stainless steel, chrome-plated stainless steel, plastic, brass. Principle: electrosurgical resection/coagulation via high-frequency current. Form factor: hysteroscopic resectoscope with telescope and electrode-specific working element. Connectivity: none. Sterilization: not specified.
Indications for Use
Indicated for patients requiring diagnostic or surgical hysteroscopy, including excision of intrauterine myomas, excision of intrauterine polyps, lysis of intrauterine adhesions, and incision of uterine septa.
Regulatory Classification
Identification
A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.
Predicate Devices
COMEG Endoscopy Resectoscope
Karl Storz Hysteroscopic Resectoscope Models 27040/27050
Submission Summary (Full Text)
{0}------------------------------------------------
۱ ﻣﺴ
SECTION 7
| 510(k) Summary of Safety and Effectiveness | |
|--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Statement | Information supporting claims of substantial equivalence, as defined under the Federal Food, Drug, and Cosmetic Act, respecting safety and effectiveness is summarized below. For the convenience of the Reviewer, this summary is formatted in accordance with the Agency's final rule "...510(k) Summaries and 510(k) Statements..." (21 CFR 807) and can be used to provide a substantial equivalence summary to anyone requesting it from the Agency. |
| | MODIFIED DEVICE NAME: Gyrus Hysteroscopic Resectoscope |
| | PREDICATE DEVICE NAME: COMEG Endoscopy Resectoscope and Karl Storz Hysteroscopic Resectoscope Models 27040/27050. |
| Device Description | The Gyrus Hysteroscopic Resectoscope that we intend to market employs an electrode specific resectoscope-working element. All other sheaths etc., are standard COMEG components. A "Quick Connection" feature is used to attach the working element to the sheath. |
| | Endoscopic electrosurgical instruments and accessories are described under "Endoscope and Accessories" in 21 CFR 876.1500, and Hysteroscopes are described in 21 CFR 884.1690. |
| | The Hysteroscopic Resectoscope is intended to enable the viewing of the cervical canal and the uterine cavity for the purpose of performing diagnostic and surgical procedures. |
| | The surgeon performs the examination through the cervical canal. The working elements are the devices that house and control various electrodes and lasers used to remove cut, coagulate, and transect tissue. The surgeon controls the back and forth movement of the electrode, using finger controls. The working elements also house a telescope for visualization. Current is transmitted to the electrode through a high frequency cable. |
| | These devices are composed of stainless steel, stainless steel chrome plated, plastic and brass chrome plated parts. |
| | |
Continued on next page
{1}------------------------------------------------
______________________________________________________________________________________________________________________________________________________________________________
## 510(k) Summary of Safety and Effectiveness, Continued
| Intended Use | The Gyrus Hysteroscopic Resectoscope is intended to enable the viewing of<br>the cervical canal and the uterine cavity for the purpose of performing<br>diagnostic and surgical procedures. The surgical applications include tissue<br>cutting, removal, and dissection as required or encountered in gynecologic<br>hysteroscopic electrosurgical procedures for the treatment of intrauterine<br>myomas, polyps, adhesions, and septa. | |
|----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Indications<br>Statement | The Gyrus Hysteroscopic Resectoscope is used to permit the direct viewing of<br>the cervical canal and the uterine cavity for the purpose of performing<br>diagnostic and surgical procedures. | |
| Technological<br>characteristics | The modified device has the same technological characteristics as the predicate<br>devices. The form, fit and function is similar. | |
| Performance<br>Data | Pre-clinical as well as bench top testing has been performed to verify that the<br>product meets the performance requirements described. This substantiates the<br>efficacy of the resectoscope for the hysteroscopic usage. | |
| Conclusion | Based upon the 510(k) summaries and 510(k) statements (21 CFR 807) and<br>the information provided herein, we conclude that the new device is<br>substantially equivalent to the predicate devices under the Federal Food, Drug,<br>Cosmetic Act. | |
| Contact | David Kay<br>Director of Regulatory Affairs and Quality Assurance<br>Gyrus Medical Ltd.<br>Fortran Road<br>St. Mellons<br>Cardiff, CF3 0LT<br>United Kingdom | |
| Date | August 4, 1998 | |
{2}------------------------------------------------
Rockville MD 20850
Food and Drug Administration 9200 Corporate Boulevard
Image /page/2/Picture/2 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which consists of a stylized human figure with three wavy lines representing the head, body, and legs. The emblem is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
JAN 2 1 1999
David Kay Director, Regulatory Affairs & Quality Assurance Gyrus Medical Fortran Road St. Mellons Cardiff CF3 OLT UNITED KINGDOM
Re: K982771
Gyrus Axipoler™ Hysteroscopic Resectoscope Dated: November 23, 1998 Received: November 25, 1998 Regulatory Class: II 21 CFR 884.1690/Procode: 85 HIH
Dear Mr. Kay:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major requlations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verily such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling requlation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
Capt. Daniel G. Schultz, M.D.
Capt. Daniel G. Schultz. M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## INDICATIONS FOR USE
| 510(k) Number (if known): | K982771 |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | Gyrus Hysteroscopic Resectoscope |
| Indications for Use: | Used to permit direct viewing and access to the cervical canal and<br>the uterine cavity for the purpose of performing diagnostic and<br>surgical procedures. As used with the Gynecare Versapoint<br>System, the indications for use include:<br>Excision of intrauterine myomas<br>Excision of intrauterine polyps<br>Lysis of intrauterine adhesions<br>Incision of uterine septa |
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
signature
Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devi CD I S10(k) Number
Prescription Use
OR
Over-The-Counter Use _
(Optional Format 1-2-96)
4
Gyrus Hysteroscopic Resectoscope Gyrus Medical, Ltd.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.