Gynex Electrodes are sterile, single-use electrosurgical accessories designed for gynecologic procedures. The device consists of an insulated metallic shaft (2.5mm diameter) terminating in a tungsten wire loop (curved or square) for excisional procedures or a brass ball tip for fulguration and coagulation. Operated by a physician in a clinical setting, the electrode connects to an electrosurgical generator. High-frequency current passes through the tungsten wire or ball tip to destroy tissue via high temperatures under direct visual observation. The device facilitates tissue excision, biopsy, or treatment of chronic cervicitis. Benefits include precise tissue removal and coagulation, reducing the need for traditional scalpel surgery in specific gynecologic applications.
Clinical Evidence
No clinical data. Substantial equivalence is based on design, material, and intended use comparisons to the predicate device.
Technological Characteristics
Materials: Tungsten wire (0.18mm), stainless steel, brass, Teflon. Design: Insulated 2.5mm diameter shaft with curved/square wire loops or ball tips. Energy: Electrosurgical (high-frequency current). Sterilization: EtO gas (ANSI/AAMI/ISO 11135-1994). Standards: ISO 10993-7.2, ISO 9000 series, Quality Systems Regulations.
Indications for Use
Indicated for soft tissue cutting and coagulation in gynecologic procedures, including treatment of intraepithelial neoplasia, condylomata acuminata, and miscellaneous lesions such as polyps, molluscum contagiosum, nevi, seborrheic keratosis, and achricordon. Prescription use only.
Regulatory Classification
Identification
A gynecologic electrocautery is a device designed to destroy tissue with high temperatures by tissue contact with an electrically heated probe. It is used to excise cervical lesions, perform biopsies, or treat chronic cervicitis under direct visual observation. This generic type of device may include the following accessories: an electrical generator, a probe, and electrical cables.
Predicate Devices
Cooper Surgical; Single-Use Tungsten Wire LEEP Electrodes, Sterile (K952483)
Submission Summary (Full Text)
{0}------------------------------------------------
K982739
3/12/99
# Summary of Safety and
Effectiveness Information Premarket Notification, Section 510(k) Gynex Electrodes
Regulatory Authority: Safe Medical Devices Act of 1990, 21 CFR 807.92
#### Device Name: 1.
Trade Name: Gynex Electrodes Common Name: cutting & coagulation electrodes Classification Name: electrocautery, Gynecologic and accessories
#### Establishment Name & Registration Number: 2.
GYNEX Corporation Name: Number: 3032109
#### 3. Classification:
& 878.4120 Electrocautery, Gynecologic and accessories. (a) Identification. A gynecologic electrocautery is a device designed to destroy tissue with high temperatures by tissue contact with an electrically heated probe. It is used to excise cervical lesions, perform biopsies, or treat chronic cervicitis under direct visual observation. This generic type of device may include the following accessories: an electrical generator, a probe, and electrical cables. (b) Classification. Class II (performance standards).
Device Class: Class II
Classification Panel: OB/GYN
Product Code: HGI
#### 4. Preamendments Device:
1. Cooper Surgical; Single-Use Tungsten Wire LEEP Electrodes, Sterile, K952483
## Gynex Corporation Device
Curved electrode: 12cm x 1cm x 0.7cm Curved electrode: 12cm x 1cm x 1cm Curved electrode: 12cm x 1.5cm x 0.5cm Curved electrode: 12cm x 1.5cm x 0.7cm Curved electrode: 12cm x 2cm x 0.8cm Curved electrode: 12cm x 2cm x 1cm Square electrode: 12cm x 1cm x 1cm Square electrode: 5.5cm x 1cm x .64cm Ball electrode: 12cm x 5mm Ball electrode: 5.5cm x 5mm
## Equivalent Cooper Surgical Device
ltem #R1007 ZL: 12cm x 1cm x 0.7cm ltem #R1010 ZL: 12cm x 1cm x 1cm ltem #R1505 ZL: 12cm x 1.5cm x 0.5cm ltem #R1507 ZL: 12cm x 1.5cm x 0.7cm Item #R2008 ZL: 12cm x 2cm x 0.8cm item #R2010 ZL: 12cm x 2cm x 1cm ltem #S1010 ZL: 12cm x 1cm x 1cm ltem #S1007 ZL: 5.5cm x 1cm x 7.0mm ltem #B0512 ZL: 12cm x 5mm Item #B0555 ZL: 5.5cm x 5mm
#### 5. Device Description:
The Gynex Electrodes are made of thin 0.18mm hard tungsten wire with a high melting temperature and excellent conductiveness of electric current. The electrodes are available in three different styles intended for use in two different types of electrosurgical procedure. The three styles are curved wire loop electrodes, square wire loop electrodes and ball-type
{1}------------------------------------------------
electrodes. The two intended uses are excisional procedures (wire loop) and fulguration or coagulation procedures (ball electrode).
All electrode styles consist of an insulated conductive metallic shaft of varing length terminating in a bare metalic "ball" a "square wire loop" or a "curved wire loop". The wire loop electrodes vary in both width and depth of the loop. Electrode shaft diameter is constant at 2,5mm.
#### 6. Applicant Name & Address:
GYNEX Corporation 16700 NE 79th St., Suite 204 Redmond, WA 98052
#### 7. Company Contact:
Mr. Steve Angelo, President GYNEX Corporation 16700 NE 79th St., Suite 204 Redmond, WA 98052 425.882.1179 * 425.895.0115 - fax
#### 8. Submission Correspondent:
Mr. David W. Schlerf Buckman Company, Inc. 200 Gregory Lane, Suite C-100 Pleasant Hill, CA 94523-3389 925.356.2640 - 925.356.2654 - fax
#### 9. Voluntary Standards:
United States Food and Drug Administration mandated performance standards for this device do not exist. Various voluntary performance standards are utilized. Voluntary standards utilized include ASTM, GYNEX Corporation. Standard Operating Procedures (SOP), vendor certification and qualification procedures, Quality Systems Regulations, ISO materials standards and ISO 9000 series quality requiations.
#### 10. Storage, Packaging & Sterilization Information:
Gynex brand Electrodes are supplied "STERILE". The sterilization method is EtO gas sterilization and the exposure cycle has been validated to achieve a sterility assurance level of 10°. The sterilization process, the validation procedure and the nature of the gas residuals are conducted and characterized in accordance with ANSI/AAMI/ISO 11135 -1994; EN550: 1994; AAMI method designation EOR; ISO/DIS 10993-7.2; FDA Proposed "Maximum Residue Limits and Maximum Exposure" and the Pharmacopeia of the United States.
Packaging materials are typical medical grade peel-type pouches of the generic mylar/nonwoven sandwich variety. All packaging should be inspected on arrival for evidence of shipping damage. Damaged packaging may indicate the presence of unsafe product and it should not be used until carefully inspected. Any shipping damaged product should be returned promptly. Sterile product must be handled, opened and placed into use following accepted sterile technique.
{2}------------------------------------------------
### Comparison Table: 11.
| FEATURE | Gynex Disposable Electrodes | Cooper Surgical Single-Use Wire<br>Electrodes | SE? |
|---------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----|
| Intended Use: | Soft tissue cutting & coagulation<br>Intraepithelial neoplasia<br>Condylomata acuminata<br>Misc. lesions; polyps, molluscum<br>contagiosum, nevi, seborrheic keratosis,<br>achricordon | Same | YES |
| Design: | Tugnsten wire<br>Curved loop<br>Square loop<br>Brass ball electrode | Tugnsten wire<br>Curved loop<br>Square loop<br>Brass ball electrode | YES |
| Sterile: | Yes | Yes | YES |
| Sizes: | Curved electrode: 12cm x 1cm x 0.7cm<br>Curved electrode: 12cm x 1cm x 1cm<br>Curved electrode: 12cm x 1.5cm x 0.5cm<br>Curved electrode: 12cm x 1.5cm x 0.7cm<br>Curved electrode: 12cm x 2cm x 0.8cm<br>Curved electrode: 12cm x 2cm x 1cm<br>Square electrode: 12cm x 1cm x 1cm<br>Square electrode: 5.5cm x 1cm x 6.4mm<br>Ball electrode: 12cm x 5mm<br>Ball electrode: 5.5cm x 5mm<br>Shaft diameter 2.5mm | Curved electrode: 12cm x 1cm x 0.7cm<br>Curved electrode: 12cm x 1cm x 1cm<br>Curved electrode: 12cm x 1.5cm x 0.5cm<br>Curved electrode: 12cm x 1.5cm x 0.7cm<br>Curved electrode: 12cm x 2cm x 0.8cm<br>Curved electrode: 12cm x 2cm x 1cm<br>Square electrode: 12cm x 1cm x 1cm<br>Square electrode: 5.5cm x 1cm x .7cm<br>Ball electrode: 12cm x 5mm<br>Ball electrode: 5.5cm x 5mm<br>plus additional size square, ball and needle-<br>type electrodes<br>Shaft diameter 2.5mm | YES |
| Material: | Tungsten wire<br>Stainless steel<br>Brass<br>Teflon | Tungsten wire<br>Stainless steel<br>Brass<br>Teflon | YES |
| Origin: | Europe | Europe | YES |
| Manufacturer: | Gynex Corporation | Cooper Surgical | YES |
| Product Code: | HGI | HGI | YES |
| K - Number: | Pending | K952483 | YES |
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a black and white seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is an abstract design featuring three stylized human profiles facing to the right, with three parallel lines above them, possibly representing wings or movement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 1 2 1999
Gynex Corporation c/o Mr. David W. Schlerf Official Correspondent BUCKMAN Company, Inc. 200 Gregory Lane, Suite C-100 Pleasant Hill, CA 94523-3389 Re: K982739
GYNEX Electrodes Dated: November 20, 1998 Received: February 24, 1999 Regulatory Class: II 21 CFR 878.4120/Procode: 85 HGI
Dear Mr. Schlerf:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"(21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597, or at its internet address "http://www.fda.gov/cdrh/dsma/dsmamain.html".
Sincerely yours,
CAPT Daniel G. Schultz, M.D. Acting Director, Division of Reproductive, Abdominal, Ear, Nose and Throat, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Page _1 __ of __ 1
**510(k) Number**: K982739
Device Name(s):
GYNEX Electrodes
# Intended Use of the Device:
- 1. Soft tissue cutting & coagulation
- 2. Intraepithelial neoplasia
- 3. Condylomata acuminata
- 4. Misc. Iesions; polyps, molluscum contagiosum, nevi, seborrheic keratosis, achricordon, etc.
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY
Concurrence of CDRH, Office of Device Evaluation (ODE)
David G. Segerson
(Division Sign-Off)
Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number_
Prescription Use X
(Per 21 CFR 801.109)
OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (Optional format 1-2-96)
GYNEX-ELECTRODES.doc
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.