RT DESKTOP

K982713 · Elekta Oncology Systems, Ltd. · IYE · Nov 2, 1998 · Radiology

Device Facts

Record IDK982713
Device NameRT DESKTOP
ApplicantElekta Oncology Systems, Ltd.
Product CodeIYE · Radiology
Decision DateNov 2, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5050
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The EOS RT Desktop, as with the predicate SL/SLi Series user interface, is intended to be used as the user interface for the SL/SLi Series Linear Accelerators that are used for the radiation therapy treatments of malignant neoplastic diseases, as determined by a licensed medical practitioner.

Device Story

RT Desktop serves as user interface for SL/SLi Series Linear Accelerators; replaces legacy interface to address component obsolescence. Device utilizes industry-standard computer components to facilitate control and monitoring of radiation therapy delivery. Operated by clinical staff in radiation oncology settings; provides interface for managing treatment parameters. Enhances system reliability without altering fundamental radiation delivery function. Benefits patient by ensuring continued availability and support of therapy delivery systems.

Clinical Evidence

Bench testing only. No clinical data provided. Safety and effectiveness supported by hazard analysis, compliance with international safety standards (IEC 60601-1, IEC 60601-2-1, IEC 950/UL 1950), and established performance history of the predicate SL/SLi interface.

Technological Characteristics

User interface system utilizing industry-standard computer components. Complies with IEC 60601-1 (medical electrical equipment safety), IEC 60601-2-1 (linear accelerators), and IEC 950/UL 1950 (IT equipment). Designed for integration with SL/SLi Series Linear Accelerators. Quality system certified to ISO 9001, EN 46001, and MDD 93/42/EEC.

Indications for Use

Indicated for patients with malignant neoplastic diseases requiring radiation therapy treatment via SL/SLi Series Linear Accelerators.

Regulatory Classification

Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ NOV 2 1998 K982713 ## SUMMARY OF SAFETY & EFFECTIVENESS Elekta Oncology Systems Ltd hereby provides the following material summarising safety and effectiveness information for the Elekta Oncology Systems RT Desktop. This information is summarised as follows: - 1) The RT Desktop is an enhancement to the SL/SLi Interface (Javelin) which has previously been cleared for commercial distribution. This device has an established and proven track record for safety. The enhancement contains industry standard computer components to address the issue of component obsolescence currently associated with SL/SLi interface (Javelin). The RT Desktop does not raise additional types of safety or effectiveness considerations. - 2) accompanying documents provided for the user contain The comprehensive information to ensure safe and effective use. Past experience with substantially equivalent predicate devices has shown our device to be safe and effective when used as directed by the accompanying documents provided for the user. - 3) It is our opinion that the RT Desktop does not have technological characteristics that raise additional types of safety or effectiveness questions, and that we consider them an enhancement to the existing SL/SLi device. - 4) The RT Desktop is subject to compliance testing as defined in the internationally recognised safety standards IEC 60601-1 and IEC 60601-2-1. As appropriate, proprietary information technology equipment is procured to the internationally recognised standards IEC 950 and/or UL 1950. - The RT Desktop is designed to bear the CE mark affirming compliance ર) with all relevant European Directives in force, in particular the European Device Directive and the European Electromagnetic Medical Compatibility Directive. As a result of this, products may be sold freely without restriction throughout the entire European Union. - Elekta Oncology Systems is a registered medical device manufacturer of () assessed capability against the requirements of ISO 9001, EN 46001, and the Medical Device Directive 93/42/EEC Annex II. | REF: 804/<br>PC52MAO8452 | | |---------------------------------------------------------------------------------------------|----------| | MO8QA039 | | | Summary of Safety & Effectiveness Information<br>for the Elekta Oncology Systems RT Desktop | | | N.C. 4513 364 5769<br>Attachment No: 13 | | | Page 1 of 2 | 28/07/98 | © 1998 Elekta Oncology Systems Ltd. All rights reserved ELEKTA ONCOLOGY SYSTEMS LTD, CRAWLEY. UK. {1}------------------------------------------------ - Elekta Oncology Systems Quality System has been established to satisfy 7) the requirements of ISO 9001, EN 46001, the Medical Device Directive 93/42/EEC Annex II, and 21 CFR 820. Elekta Oncology Systems has developed the RT Desktop using an established and documented Quality Management System. - In accordance with the above requirements, all parts of the Quality 8) System are subject to periodic and systematic internal Quality Audits. These audits are performed by trained personnel not having direct responsibilities in the functions being audited. - The Quality System is subject to regular, planned and documented ਰੇ ) quality system audits conducted by external auditors from SGS Yarsley (UK Notified Body) and the FDA. - Elekta Oncology Systems has conducted hazard analysis on the RT 10) Desktop and has concluded that it does not introduce hazards that raise new types of safety or effectiveness considerations. After considering the Guidance for the Content of Pre-Market Notification Submissions of Medical Devices Containing Software EOS has concluded the level of concern appropriate to the device is "Higher". Signature P.A. Hart Vice President Research & Development Signature........................................ P.J. Gaccione Vice President Field Support Signature C.R. Everett Quality and Regulatory Affairs Manager | REF: 804/<br>PC52MAO8452<br>MO8QA039 | Summary of Safety & Effectiveness Information<br>for the Elekta Oncology Systems RT Desktop | N.C. 4513 364 5769<br>Attachment No: 13 | | |---------------------------------------------------------|---------------------------------------------------------------------------------------------|-----------------------------------------|----------| | | | Page 2 of 2 | 28/07/98 | | © 1998 Elekta Oncology Systems Ltd. All rights reserved | | | | | ELEKTA ONCOLOGY SYSTEMS LTD, CRAWLEY. UK. | | | | {2}------------------------------------------------ Image /page/2/Picture/0 description: The image is a black and white logo for the Department of Health & Human Services USA. The logo features a stylized eagle with its wings forming a human profile. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 2 1998 NOV Paul L. Sumner Director, Regulatory Affairs & Quality Assurance Elekta Instruments, Inc. 8 Executive Park West Atlanta, GA 30329 Re: K982713 EOS RT Desktop (RTD) System Dated: August 3, 1998 Received: August 4, 1998 Regulatory class: II 21 CFR 892.5050/Procode: 90 IYE Dear Mr. Sumner: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (OS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitn diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/cdr/dsmardsmamain.html". Sincerely yours, Lillian Yin, Ph.D. Lillian Yin, Ph.D. Director, Division of Reproductive Abdominal, Ear, Nose and Throa and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ | 510(k) Number (if known): | K982713 | |---------------------------|---------| |---------------------------|---------| EOS RT Desktop Device Name: Indication for Use: The EOS RT Desktop, as with the predicate SL/SLi Series user interface, is intended to be used as the user interface for the SL/SLi Series Linear Accelerators that are used for the radiation therapy treatments of malignant neoplastic diseases, as determined by a licensed medical practitioner. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use (Per 21 CFR 801.109) OR Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ Skirid le. Sypor (Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Dey 510(k) Number (Optional Format 1-2-96)
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...