ECHO-SCREEN

K982642 · Fischer-Zoth Audiologic Systems, Inc. · EWO · Oct 21, 1998 · Ear, Nose, Throat

Device Facts

Record IDK982642
Device NameECHO-SCREEN
ApplicantFischer-Zoth Audiologic Systems, Inc.
Product CodeEWO · Ear, Nose, Throat
Decision DateOct 21, 1998
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 874.1050
Device ClassClass 2
AttributesPediatric

Indications for Use

The Echo-Screen Otoacoustic Emissions Test Instrument is a hand-held instrument with an attached probe (that is placed in the outer ear canal) which emits sound that evokes otoacoustic emissions. A microphone in the probe picks up the emissions if any are present, and the device then analyzes the emissions and mathematically separates noise from cochlear emissions. The presence of these emissions indicates cochlear function in adults, infants and children. The instrument requires no active participation from the patient.

Device Story

Handheld, battery-powered screening system for detecting cochlear dysfunction via Otoacoustic Emissions (OAE). Device emits click-bursts (40-100Hz) into ear canal via miniature probe; microphone records cochlear response. Automated evaluation software processes signals to separate noise from emissions; provides 'PASS' indication based on signal statistics. Used by trained personnel in medical or school environments. Output includes evaluation results, stimulus stability, and artifact rate. Facilitates early identification of hearing disorders; does not prescribe or fit assistive devices.

Clinical Evidence

Bench testing only. No clinical data provided. Safety and effectiveness supported by compliance with electrical safety standards (IEC601-1, UL2601-1, VDE750) and verification of automated signal processing algorithms for OAE detection.

Technological Characteristics

Handheld device (21.5x10x5.4cm, 600g); 6V rechargeable battery. Probe contains one speaker and one microphone. Stimulus: 40-100Hz click-bursts; bandwidth 1.4kHz-4kHz. Max output 85dB SPL, software-limited. Connectivity: None during measurement. Software: C-based automated evaluation algorithm. Standards: ISO 9001, EN46001, IEC601-1, UL2601-1.

Indications for Use

Indicated for the detection of cochlear dysfunction in neonates, infants, children, and adults. Contraindications: None stated.

Regulatory Classification

Identification

An audiometer or automated audiometer is an electroacoustic device that produces controlled levels of test tones and signals intended for use in conducting diagnostic hearing evaluations and assisting in the diagnosis of possible otologic disorders.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K982642 OCT 21 1998 '510 (k) Summary' Summary and safety effectiveness Submitter's Name & Address: Fischer-Zoth Audiologic Systems 11309 South Brandon Park Dr. 3 Sandy, Utah, 84092 Contact person & Telephone: Mr. Michael Soerensen (801) 253 0704 May 5, 1998 Data summary prepared: Device name: Classification Name - Audiometer Common/Usual Name - Screening Audiometer Proprietary name - Echo-Screen Predicate Device: Grason Stadler Inc. model GS170 AudioPath Screener (ref. 510(k) #K974237) and Otodynamics LTD. model ILO88 (ref. 510(k) #K962995, Device description , intended Use & Effectiveness: The Fischer-Zoth Audiologic System model 'Echo-Screen' is based upon Otoacoustic Emissions and it is similar to the Grason Stadler Inc. GSI70 AudioPath Screener and the Otodynamics OAE Analyser, model ILO88. __ ____________________________________________________________________________________________________________________________________ The purpose of the device is the reliable and earliest possible detection of hearing disorders due to cochlear dysfunction. It can be used beginning from neonates to adults. The 'echo-screen' is a handheld, automated OAE-based screeming-system which is easy to use . The measurement flow is menu guided and the evaluation is based upon signal statistics. It is intended to be used by trained personnel in a medical or school environment. The echo-screen is not intended for use in prescribing or fitting assisstive listening devices such as hearing aids or cochlea implants. Technological Characterisitics: see also Attachment A - Probe: miniature probe easy to apply even for premature infants - Hardware: - **Hardware:** Hand held device : size 21,5x10x5,4cm , weight 600g. Battery powered : 6V/1000mAh, rechargeable Stimulus Rate: 40 ... 100Hz Signal Bandwidth : 1,4KHz to 4KHz - Software: echo-screen evaluation SW Automated stimulus adjustment and online control Signal statistical criterion for 'PASS' condition Additional hints to measurement quality (stimulus stability and artefact rate) Safety: Battery powered, no connection to mains required Sound Pressure level: 70-85 dB SPL maximum {1}------------------------------------------------ Safety: The 'Echo-Screen' Screening System is designed to provide electrical safety to the patient as well as the user. The system is designed to meet the following standards related electrical safety: IEC601-1, UL2601 and VDE750 (German MedGv). The echo-screen is a stand alone-system. It is battery powered with a 6V battery. During operation there are no connections to the mains. There are no conductive connections to the patient or to the user during measurement. To prevent excessive tone levels within the ear, the echo-screen system continiously controls the level of the output tone or click-burst. The max. output level is controlled via Software to 85dB SPL (sound pressure level) and the electronic hardware is build in such a way, that the speakers within the probe are incapable of producing enough decibels to permanently damage the ear. The system will also be certified according to the following standards: - ISO 9001 (EN29001) Quality Management System - EN46001 Class I or Class IIa device - Group 1 (medical equipment, conducted and radiated emissions) - EN 25011 - IEC801-3 Radiated Electromagnetic Field Susceptibility - Conforms the provisions of Eruopean council Directive 93/42EEC - CE Mark concerning medical devices. ## ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------Summary of Effectiveness: ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ The 'echo-screen' is a portable OAE based screening system. Its aim is to detect and track hearing loss and abnormal cochlear conditions as early as possible in the child's life. The automated test makes it easy to use by trained personnel in a medical or clinical environment. {2}------------------------------------------------ ## Attachement A i ------ ۰ ----- | Technological<br>Specification | GSI70 | echo-screen | |--------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Power source | Powered by rechargeable<br>battery | Powered by rechargeable<br>battery | | Battery Low<br>Indication | Battery low indicator | Battery low indicator | | Safety Compliance | IEC 601-1 and UL2601-1 | IEC601-1 and UL2601-1 | | Computer interface | IRDA communication | none, during measurement | | Audiometry | Tones are presented to a<br>miniature speaker within the<br>probe which perceived is<br>indicated by the patient using a<br>response switch. Tone quality<br>and frequency specified per<br>ANSIS3.6 | Tones (clicks) are presented to a<br>miniature speaker within the<br>probe and controlled via SW.<br>The response from the cochlea is<br>received and registrated.<br>Click tone quality and amplitude is<br>controlled via SW and HW and is<br>limited to a max. output level<br>85dB SPL. | | OAE probe design | Two speakers and a<br>microphone present and<br>record response from<br>cochlea of ear. | One speaker and a microphone<br>present and record response<br>from cochlea of ear. | | SW | GSI 60 SW | Echo-screen Automated Evaluation SW | | Supporting SW | Windows based SW that<br>shows result., allows export<br>of data, and updating of<br>patient ID | C-based SW, that shows evaluation results,<br>artefact rate, stimulus stability.<br>After measurement export of test-result<br>and patient data are possible | {3}------------------------------------------------ Image /page/3/Picture/1 description: The image is a circular seal for the Department of Health & Human Services - USA. The seal features the department's name encircling an emblem. The emblem is a stylized image of an eagle with three overlapping profiles, suggesting a sense of community or collaboration. The overall design is simple and professional, reflecting the department's role in public health and human services. Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 OCT 21 1998 David G. Wood, Esq. Fischer-Zoth Audiologic System, Inc. 11309 Brandon Park Drive Sandy, Utah 84092 Re: K982642 Echo-Screen Dated: July 27, 1998 Received: July 29, 1998 Regulatory class: II 21 CFR 874.1050/Procode: 77 EWO Dear Mr. Wood: We have reviewed your Section 510(k) notification of intent to market the device referenced above and we-have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). . You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Current Good Manufacturing Practice requirements, as set forth in the Quality System Regulation (QS) for Medical Devices: General regulation (21 CFR Part 820) and that, through periodic QS inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations. This letter will allow you to begin marketing your device as described in your 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your-device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4613. Additionally, for questions on the programs on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). Other general information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address "http://www.fda.gov/odff/dsmaldsmam.html". Sincerely yours, Lillian Yin, Ph.D. Lillian Yin, Ph.D. Director, Division of Reproductive Abdominal, Ear, Nose and Throa and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ | 510(k) Number (if known): | K982642 | |---------------------------|-------------| | Device Name: | ECHO-SCREEN | | Indications For Use: | | The Echo-Screen Otoacoustic Emissions Test Instrument is a hand-held instrument with an attached probe (that is placed in the outer ear canal) which emits sound that evokes otoacoustic emissions. A microphone in the probe picks up the emissions if any are present, and the device then analyzes the emissions and mathematically separates noise from cochlear emissions. The presence of these emissions indicates cochlear function in adults, infants and children. The instrument requires no active participation from the patient. Otoacoustic emissions (OAEs) are low-level audio-frequency sounds that are generated by the outer hair cells of the cochlea in response to sound input. The presence of OAEs is an indication of cochlear function. If the Echo-Screen signals a "pass" reading as it analyzes the presence of OAEs, the clinician can determine that the cochlea is operating normally. (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Reproductive, Abdominal, ENT, and Radiological Devices 510(k) Number K982642 | Prescription Use<br>(Per 21 CFR 801.109) | OR | Over-The-Counter Use | |------------------------------------------|----|----------------------| |------------------------------------------|----|----------------------| (Optional Format 1-2-96)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...