K982586 · Shiva Medicare , Ltd. · KGO · Aug 11, 1998 · General, Plastic Surgery
Device Facts
Record ID
K982586
Device Name
POWDER FREE SURGEONS GLOVE
Applicant
Shiva Medicare , Ltd.
Product Code
KGO · General, Plastic Surgery
Decision Date
Aug 11, 1998
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4460
Device Class
Class 1
Indications for Use
Powder free Surgeon's glove is a sterile powder free medical device intended to be worn by Operating room personnel to protect a surgical wound from contamination.
Device Story
Sterile, powder-free latex surgical glove; intended for use by operating room personnel to protect surgical wounds from contamination. Device acts as a physical barrier between the clinician and the patient. Manufactured to meet ASTM D3577 standards for dimensions, physical properties (tensile strength, elongation), and biocompatibility. Used in clinical/surgical settings. Provides protection against cross-contamination during surgical procedures.
Clinical Evidence
Bench testing only. Device performance evaluated against ASTM D3577 standards, including length, width, thickness, tensile strength, elongation at break, and modulus. Biocompatibility confirmed. Sterility verified per Indian Pharmacopea/USP standards. No clinical data provided.
Indicated for use by operating room personnel to protect surgical wounds from contamination. No specific age or gender restrictions.
Regulatory Classification
Identification
A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Class I Surgeon's Glove (Powder free) 79KGO (meets ASTM D3577)
Submission Summary (Full Text)
{0}------------------------------------------------
AUG | | | 1998
Image /page/0/Picture/1 description: The image shows a close-up of handwritten numbers and a letter. The numbers "98" and "2586" are visible, along with the letter "K". A horizontal line is drawn beneath the numbers, possibly indicating a total or a separation of values. The handwriting appears to be done with a thick marker or pen, giving the characters a bold and somewhat rough texture.
Image /page/0/Picture/2 description: The image shows a logo with a triangular shape at the top and the word "SHIV" written below it. The triangle is filled with a dark color, and there is a white letter "S" inside the triangle. The word "SHIV" is written in bold, uppercase letters.
:17
1. 510K SUMMARY as required by: 807.92( c )
#### 1.0 APPLICANT
| NAME | M/s. SHIVA MEDICARE LIMITED |
|---------|---------------------------------------------------------------------------------------------------------|
| ADDRESS | A 15, 16, 17(PART)<br>MADRAS EXPORT PROCESSINGZONE<br>KADAPERI, TAMBARAM,<br>CHENNAI 600 045.<br>INDIA. |
| PH.NO. | : 91-44-2368011/2368322/2362656/2362657. |
| FAX NO | : 91-44-2368327 |
| EMAIL | : shivam@md2.vsnl.net.in |
#### CONTACT PERSON: MR. SATISH JAIN EXECUTIVE DIRECTOR.
3. DEVICE TRADE NAME : NIL
.. Surgeon's Glove COMMON NAME
: Powder free Surgeon's Glove Classification Name
- 4. Legally marketed device to which the company claiming equivalence: Class I Surgeon's Glove ( Powder free) 79KGOthat meets all the requirements of ASTM D3577.
- 5. DESCRIPTION OF THE DEVICE :
Class I Powder free Surgical Glove 79KGO that meets all the requirements of ASTM D3577.
Intended use of the Device: 6.0
Powder free Surgeon's glove is a Powder free Medical Device intended tobe worn by Operating room personnel to protect a surgical wound from contamination.
Image /page/0/Picture/15 description: The image shows a logo for Shiva Medicare LTD. The logo features a black cross with the word "MEDICARE" written diagonally across it. Below the cross is a circular seal with the words "SHIVA MEDICARE LTD" around the edge and "Madras-600 045" in the center.
**Shiva Medicare Limited**
(A 100% Export Oriented Unit)
Registered Office & Works : A 15-17. M.E.P.Z., Kadaperi, Tambaram, Chenna1 - 600 045 (INDIA) Phones : 044/2368011/2368322 EMAIL: Shivani @md2.vsnl-net.in Telefax : 091-44-2368327. Grams : "SHIVMEDI" Corporate Office : 808, Pragaii Tower, 26, Rajendra Place, New Delhi - 110 008 (INDIA) ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{1}------------------------------------------------
# 7.0 Technological characteristics of the device compared to predicate device.
| Measured Parameters of Latex<br>Surgeon's gloves (Powder free)<br>manufactured by Shiva Medicare Ltd., | | | ASTM D3577<br>Requirement for<br>Latex Surgeon's glove<br>(Powder free) |
|--------------------------------------------------------------------------------------------------------|------|---------------|-------------------------------------------------------------------------|
| Characteristics | SIZE | Value | |
| 1. Length | 5 ½ | 270-272 mm | 245 mm minimum |
| | 6 | 270 – 272 mm | 265 mm minimum |
| | 6 ½ | 270 - 272 mm | 265 mm minimum |
| | 7 | 270 - 272 mm | 265mm minimum |
| | 7 ½ | 270 –272 mm | 265 mm minimum |
| | 8 | 270 -272 mm | 265 mm minimum |
| | 8 ½ | 270 -- 272 mm | 265 mm minimum |
| | 9 | 270 - 272 mm | 265 mm minimum |
| 2. Width | 5 ½ | 68 mm | 70 +/- 6 mm |
| | 6 | 73mm | 76 +/- 6mm |
| | 6 ½ | 79mm | 83 +/- 6 mm |
| | 7 | 87mm | 99 +/- 6 mm |
| | 7 ½ | 92mm | 95 +/- 6 mm |
| | 8 | 103mm | 105+/- 6 mm |
| | 8 ½ | 106mm | 108+/-6 mm |
| | 9 | 112mm | 114+/-6 mm |
- 3. Thickness at cuff , Palm and finger tip of all the size is 0 .12, 0 .16 and 0.19mm.
ASTM D3577 requirement for thickness at cuff, palm and finger tip is 0.1 mm minimum.
# PHYSICAL PROPERTIES:
| BEFORE AGEING | | AFTER AGEING | | |
|---------------------------------|-----------|------------------------------|-----------|------------------------------|
| Characteristics | SML Value | ASTM<br>D3577<br>Requirement | SML Value | ASTM<br>D3577<br>Requirement |
| Tensile Strength | 27 mpa | 24 mpa | 20 mpa | 18 mpa min |
| Elongation at break % | 850% | 750% | 750% | 560% min |
| Modulus at 500 %<br>elongation. | 3 mpa | 5.5 mpa<br>(max) | - | - |
Image /page/1/Picture/6 description: The image shows a circular stamp with the words "AAIHA MEDICARE LTD" around the outer edge. Inside the circle, the text "Madras-600 045" is visible. The stamp appears to be somewhat faded or worn, giving it a vintage or official document-like appearance.
{2}------------------------------------------------
## ERFORMANCE REQUIREMENT:
| Characteristics | Related<br>defects | Level followed<br>By<br>SML | Level As<br>per ASTM<br>D 3577 | AQL followed | AQL as per ASTM<br>D3577. |
|-----------------------|-------------------------------------------------------------------------------|-----------------------------|--------------------------------|--------------|---------------------------|
| Sterility | Fails sterility | As per IP* | As per<br>USP*. | NA | NA |
| Freedom from<br>Holes | Holes | S4 | S4 | 1.5 | 1.5 |
| Dimension | Width ,<br>Length<br>Thickness. | S2 | S2 | 4 | 4 |
| Physical<br>Property | Tensile<br>strength ,<br>Elongation at<br>break before<br>and after<br>ageing | S2 | S2 | 4 | 4 |
## IP – INDIAN PHARMACOPEA
#### OWDER CONTENT
| CIT 57<br>ﻟ<br>STATUTE OF COLUMN | ARFOURFMENT<br>A C'T'T |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|
| Nil P<br>Powder<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | mo/olove<br>max |
## PROTEIN CONTENT:
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>CONTINE<br>1 - 1 - 1 | DE CITIENTT |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------|
| 00<br>mmm<br>Actually College<br>A & Acres and assessment compares<br>of the property of the a second and in any of<br>Property of Canadian College of Children | alue not fixed.<br>STATUTE OF COLUMN OF CHEARTH AND THE PERSONAL PROPERTY |
#### MOISTURE CONTENT:
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>THE PHONOR A - A - A 16 - 415 - 411 - 411 - 411 - 411 - 411 - 411 - 411 - 411 - 411 - 411 - 411 - 411 - 411 - 411 - 411 - 411 - 411 - 411 - 411 - 411 - 411 - 411 - 411 - 411<br>A STATUS MARK VITE THE THE STORE THE<br>דיין<br>***<br>AN VALUE<br>A bear and and all different the later the later the later than | דיד גדוד לוח לה במוז<br>TY-13<br>TT<br>1<br>- |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 00<br>mau<br>11 1641<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>and and and the proposed for at the first can a first a program and | CMILD<br>TIVAC<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>www.bandrer and the transmit of the company of the company of the company of the company of the first to the country of the first to the company of the company of the first t |
#### BIOCOMPATABILITY:
| SML GLOVE | FDA REQUIREMENT |
|-------------------------|-------------------------|
| Biologically Compatible | Biologically Compatible |
Image /page/2/Picture/11 description: The image shows a circular stamp with the text "SHIVA MEDICARE LTD" around the border. In the center of the stamp, the text "Madras-600 045" is visible. The stamp appears to be of low quality, with some blurring and distortion.
{3}------------------------------------------------
## 8.0. Performance Data:
The performance test data of the powder free surgical gloves manufactured by Shiva Medicare Limited is given below.
| Measured Parameters of Latex | | |
|--------------------------------------|------|--------------|
| Surgeon's gloves (Powder free) | | |
| manufactured by Shiva Medicare Ltd., | | |
| Characteristics | SIZE | Value |
| 1. Length | 5 ½ | 270 – 272 mm |
| | 6 | 270 – 272 mm |
| | 6 ½ | 270 – 272 mm |
| | 7 | 270 – 272 mm |
| | 7 ½ | 270 – 272 mm |
| | 8 | 270 – 272 mm |
| | 8 ½ | 270 – 272 mm |
| | 9 | 270 – 272 mm |
| 2. Width | 5 ½ | 68 mm |
| | 6 | 73 mm |
| | 6 ½ | 79 mm |
| | 7 | 87 mm |
| | 7 ½ | 92 mm |
| | 8 | 103 mm |
| | 8 ½ | 106 mm |
| | 9 | 112 mm |
- 3. Thickness at cuff , Palm and finger tip of all the size is 0 .12, 0.16 and 0.19 mm.
PHYSICAL PROPERTIES:
| Characteristics | Before Ageing | After Ageing |
|---------------------------------|---------------|--------------|
| Tensile Strength | 27 mpa | 20 mpa |
| Elongation at break % | 850% | 750% |
| Modulus at 500 %<br>elongation. | 3 mpa | - |
Image /page/3/Picture/6 description: The image shows a circular stamp with the text "SHIVA MEDICARE LTD" around the outer edge. Inside the circle, the text "Madras-600 045" is printed. The text is in a simple, sans-serif font and is arranged in a way that it fills the circular space effectively. The stamp appears to be somewhat worn, with some parts of the text slightly faded or smudged.
{4}------------------------------------------------
#### PERFORMANCE REQUIREMENT:
| Characteristics | Related defects | LEVEL | AQL |
|-----------------------|--------------------------------------------------------------------------------|------------------------------|-----|
| Sterility | Fails sterility | As per Indian<br>Pharmacopea | |
| Freedom from<br>Holes | Holes | S4 | 1.5 |
| Dimension | Width , Length<br>Thickness. | S2 | 4 |
| Physical<br>Property | Tensile<br>strength ,<br>Elongation at<br>break before<br>and after<br>ageing. | S2 | 4 |
POWDER CONTENT : 1 + /- 1 mg per Glove
PROTEIN CONTENT: 80 +/- 20 ppm
MOISTURE CONTENT: 0.8 % max
BIOCOMPATABILITY: Biologically Compatible
NA 9. Clinical Data :
10. CONCLUSION OF PERFORMANCE TEST DATA:
The Powder free Surgeon's gloves manufactured by Shiva Medicare Limited
Meet or exceed the ASTM D3577 -
- Meet FDA Pin hole Requirement. -
- Meet labeling claim as shown by the data in 6 -
11. ANY OTHER INFORMATION:
Any other information required by FDA regarding product safety and effectiveness will be provided on request.
Image /page/4/Picture/14 description: The image shows a circular seal with the text "AHIA MEDICARE LT" around the outer edge. Inside the circle, the text "Madras-600 045" is printed. The text is in a simple, sans-serif font and is arranged to fit within the circular shape of the seal. The seal appears to be a stamp or an official mark, possibly from a medical or healthcare-related organization located in Madras (now Chennai), India.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the top half of the circle. Inside the circle is a stylized image of a human figure, represented by three overlapping profiles, with a wing-like shape extending from the head.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
AUG | | 1998
M/s. Shiva Medicare Ltd. C/O Mr. James F. Logan Medical Reports Exchange 314 N Huntet Street Baltimore, Maryland 21202
Re : K982586 Sterile Powder-Free Surgical Glove, Trade Name: Disposable Regulatory Class: I Product Code: KGO July 24, 1998 Dated: Received: July 24, 1998
Dear Mr. Logan:
We have reviewed your Section 510(k) notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the qeneral controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Requlations, Title 21, Parts 800 to 895. A substantially equivalent determination assumes compliance with the Good Manufacturing Practice for Medical Devices: General (GMP) regulation (21 CFR Part 820) and that, through periodic GMP inspections, the Food and Drug Administration (FDA) will verify such assumptions. Failure to comply with the GMP regulation may result in regulatory action. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please note: this response to your premarket notification submission does not affect any obligation you might have under sections 531 through 542 of
{6}------------------------------------------------
Page 2 - Mr. Logan
the Act for devices under the Electronic Product Radiation Control provisions, or other Federal laws or regulations.
This letter will allow you to begin marketing your device as described in your 510 (k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801 and additionally 809.10 for in vitro diagnostic devices), please contact the Office of Compliance at (301) 594-4692. Additionally, for questions on the promotion and advertising of your device, please contact the Office of Compliance at (301) 594-4639. Also, please note the regulation entitled, "Misbranding by reference to
premarket notification" (21 CFR 807.97). Other qeneral information on your responsibilities under the Act may be obtained from the Division of Small Manufacturers Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its internet address "http://www.fda.gov/cdrh/dsmamain.html".
Sincerely yours,
S. Dutram for
Timothy A. Ulatowski Director Division of Dental, Infection Control, and General Hospital Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### 3.0 INDICATIONS FOR USE:
| APPLICANT | SHIVA MEDICARE LIMITED |
|-------------|-----------------------------|
| 510(K) No. | K98 2586 |
| DEVICE NAME | POWDER FREE SURGEON'S GLOVE |
INDICATIONS FOR USE:
Powder free Surgeon's glove is a sterile powder free medical device intended to be worn I oncer free bargen ersonnel to protect a surgical wound from contamination.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Qim S. L
Division Sign. 280
Division Sign-( iff) Division of Der. Infection Control, ind General Hospital Devices "10(k) Number
Prescription Use (Per 21 CFR 801.109)
OR
Over-The-Counter Use X
(Optional Format 1-2-9
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.